Class 2 Device Recall Innovate Esmark Bandages

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Innovate Esmark Bandagessee related information
Date Initiated by FirmJanuary 28, 2011
Date PostedMarch 16, 2011
Recall Status1 Terminated 3 on September 04, 2012
Recall NumberZ-1667-2011
Recall Event ID 57880
Product Classification General surgery tray - Product Code LRO
ProductInnovate Esmark Bandages, a component of various Presource Custom Sterile Packs/Presource PBDS Modules; Packaged by Cardinal Health Medical Products and services, McGaw Park, IL 60085 U.S.A.; There is no label on the Esmark bandage in the Presource pack/module, Some are placed in Tyvek pouches and some are just loose in the pack, but neither have any labeling on them. The Esmark bandage is used to support and compress a part of the patient's body. Presource Packs/Modules are manufactured, according to customer requirements, for use during surgical procedures.
Code Information The Esmark bandages were a component of the following Presource Convenience Packs/Modules: Cat #PO12AUGUT, lot 9097; Cat #PO12HNGUQ, lot 9275; Cat #PO33FAPSF, lot 634802; Cat #PO53TKDMF, lot 633697; Cat #PO56TK48S, lot 634508; Cat #POMHTKWDI, lot 635911; Cat #POOCUEENK, lot 10301516; Cat #SBA11PHSLG, lot 10863; Cat #SBA32VCPAQ, lot 637321; Cat #SBA32VEMEG, lots 638023 and 634732; Cat #SBACGAAAUC, lot 986893; Cat #SBACZEXBRB, lot 633928; Cat #SCV15ACFSF, lot 634357; Cat #SOP08EPDMC, lot 985889; Cat #SOP11APCPZ, lot 8199; Cat #SOP11ARNHT, lot 10004; Cat #SOP11ATMTI, lot 634538; Cat #SOP11BKRJG, lots 9615 and 6734; Cat #SOP11ETCHA, lot 10627; Cat #SOP11EXMML, lot 9036; Cat #SOP11EXNYL, lot 10778; Cat #SOP11FASLD, lot 634159; Cat #SOP11FMLHN, lot 10708; Cat #SOP11FTJHK, lot 10593; Cat #SOP11FTSHO, lot 10244; Cat #SOP11HASWB, lot 9609; Cat #SOP11HDNSF, lot 9062; Cat #SOP11HNFHK, lot 634172; Cat #SOP11HNLDL, lot 10467; Cat #SOP11HNMOM, lot 8839; Cat #SOP11HNMUF, lot 9767; Cat #SOP11HNNBO, lot 9632; Cat #SOP11KAJEG, lot 9707; Cat #SOP11LXRJK, lot 9918; Cat #SOP11MAVUD, lot 9703; Cat #SOP11OASLE, lot 9452; Cat #SOP11OMICF, lot 8411; Cat #SOP11PDBPB, lot 635731; Cat #SOP11UPENH, lot 9510; Cat #SOP12ARKGN, lot 8802; Cat #SOP12ARMOD, lot 9744; Cat #SOP12HDCHL, lot 637167; Cat #SOP12MOVAF, lot 9751; Cat #SOP12ORKGJ, lot 8998; Cat #SOP13AROAN, lot 8160; Cat #SOP13ARXEH, lot 9704; Cat #SOP13FABOE, lot 637206; Cat #SOP13HABOD, lot 637491; Cat #SOP13HABSF, lot 9496; Cat #SOP13HAMGA, lot 10268; Cat #SOP13HNBIN, lot 9729; Cat #SOP13HNMGB, lot 9558; Cat #SOP13KNAOI, lot 9145; Cat #SOP13KNSHG, lot 8929; Cat #SOP13LEGLC, lot 11133; Cat #SOP13UEXEG, lot 636690; Cat #SOP14EXMSQ, lot 10002; Cat #SOP15AKFSG, lot 634345; Cat #SOP18ARCCE, lot 9100; Cat #SOP18ARMLC, lots 9996 and 7752; Cat #SOP18EXGHF, lot 635818; Cat #SOP18EXSLG, lot 637282; Cat #SOP18HNMFE, lot 10588; Cat #SOP18HNSFR, lot 9309; Cat #SOP18OMWBJ, lot 9985; Cat #SOP19EXFMM, lot 986794; Cat #SOP19FPSMB, lot 985369; Cat #SOP19KAEMD, lot 986851; Cat #SOP21KNSTB, lot 8140; Cat #SOP21TKHOC, lot 10023; Cat #SOP22FTSMP, lot 635580; Cat #SOP22KAMCI, lot 9023; Cat #SOP23ACBSA, lot 9966; Cat #SOP23ACUCO, lot 8999; Cat #SOP23APUCS, lot 10003; Cat #SOP23ARMZD, lot 635735; Cat #SOP23ATGMC, lot 9317; Cat #SOP23EXDUY, lot 9119; Cat #SOP23FTGME, lot 8493; Cat #SOP23HDVAI, lot 9045; Cat #SOP23HNRCQ, lot 636310; Cat #SOP23KTSMA, lot 9455; Cat #SOP23L1DUS, lot 9724; Cat #SOP23LMUCR, lot 10086; Cat #SOP23UPPCE, lot 9456; Cat #SOP23UXHWE, lot 8795; Cat #SOP24KAPSE, lot 633735; Cat #SOP29EXCFJ, lot 635695; Cat #SOP29HDUMA, lot 635594; Cat #SOP29HNMEC, lot 634278; Cat #SOP31ACOSI, lot 987042; Cat #SOP31AHSAE, lot 633267; Cat #SOP31ATMDD, lot 986850; Cat #SOP31EXLSD, lot 636577; Cat #SOP31EXNWL, lot 986251; Cat #SOP31EXRSG, lot 986895; Cat #SOP31EXSFT, lot 986268; Cat #SOP31EXSSF, lot 986831; Cat #SOP31HDCDI, lot 635667; Cat #SOP31HNEVU, lot 986962; Cat #SOP31HNSFU, lot 986275; Cat #SOP31KAMNH, lot 986943; Cat #SOP31KAOSO, lot 985700; Cat #SOP31KPEZB, lot 986891; Cat #SOP31PDJUA, lot 986837; Cat #SOP31RADGF, lot 984976; Cat #SOP31XTDSL, lot 985606; Cat #SOP32ATFSN, lot 635190; Cat #SOP32EXFCD, lot 986432; Cat #SOP32EXPAO, lot 635212; Cat #SOP32HARMI, lot 637186; Cat #SOP32HNMNN, lot 987081; Cat #SOP32HNWBO, lot 634987; Cat #SOP32KECYF, lot 986845; Cat #SOP32PDPCB, lot 635003; Cat #SOP32TKFSN, lot 637532; Cat #SOP32TKSAF, lot 633920; Cat #SOP33EXSFE, lot 985937; Cat #SOP33HDDWH, lot 635158; Cat #SOP34ACSTH, lot 636002; Cat #SOP34ARSKA, lots 985378 and 986541; Cat #SOP34TJHCD, lot 986839; Cat #SOP35ARCRJ, lot 985010; Cat #SOP35HDSTI, lots 986853 and 987204; Cat #SOP36ARAGQ, lot 9791; Cat #SOP36HACOD, lot 10586; Cat #SOP36HAFBQ, lot 9091; Cat #SOP36HNALR, lot 10257; Cat #SOP36KASRC, lot 9930; Cat #SOP36TKFBO, lot 10663; Cat #SOP36TNCAE, lot 9112; Cat #SOP3IKPIPD, lot 633880; Cat #SOP40PDDOG, lot 9098; Cat #SOP41HAMBC, lot 635589; Cat #SOP41HANSC, lot 635498; Cat #SOP41HFSUF, lot 635157; Cat #SOP41HNHMK, lot 9518; Cat #SOP41KNSOC, lot 634860; Cat #SOP41LEMBD, lot 636245; Cat #SOP41UEHMJ, lot 634162; Cat #SOP42HPKRD, lot 637131; Cat #SOP42KNKOD, lot 632967; Cat #SOP42MNKOM, lot 634092; Cat #SOP42PHOMK, lot 635492; Cat #SOP43ELLRB, lot 636119; Cat #SOP43KSLGF, lot 634293; Cat #SOP44HNSME, lot 634202; Cat #SOP44TKFSE, lot 633395; Cat #SOP44TKFSE, lot 635134; Cat #SOP45EXMKO, lot 634771; Cat #SOP45HNOCG, lot 634160; Cat #SOP45HPMVA, lot 635373; Cat #SOP48ACKAD, lot 633536; Cat #SOP48HPFAF, lot 634125; Cat #SOP48THLAA, lot 633072; Cat #SOP4AARNWC, lot 634726; Cat #SOP4AHASI1, lot 7545; Cat #SOP51ATSSC, lot 986838; Cat #SOP51EXDVG, lot 636019; Cat #SOP51LEMOG, lot 637145; Cat #SOP51XRKKB, lot 635228; Cat #SOP52FT73C, lot 634208; Cat #SOP53ARLEK, lot 634333; Cat #SOP53ETFOL, lot 635677; Cat #SOP53EXDCC, lot 635387; Cat #SOP53KNCCE, lots 633311 and 634813; Cat #SOP53KNSLC, lot 634179; Cat #SOP53MHPRI, lot 635312;  Cat #SOP53PPFRC, lot 637803; Cat #SOP54HDTCD, lot 634122; Cat #SOP54KAPRA, lot 987014; Cat #SOP55BCOKA, lot 633825; Cat #SOP56FTFOA, lot 985222; Cat #SOP56KN19R, lot 635155; Cat #SOP56KNSAQ, lot 636191; Cat #SOP56TKDCY, lot 634930; Cat #SOP57KNWGI, lot 633763; Cat #SOP59LXCPM, lot 634192; Cat #SOP59MNCHX, lot 637045; Cat #SOP5CAPGSC, lot 635171; Cat #SOP63KNSCL, lot 634454; Cat #SOP69AKPSB, lot 634161; Cat #SOP69FPEBB, lot 635340; Cat #SOP69OLFCC, lot 635445; Cat #SOP6NHDOIC, lot 986830; Cat #SOP73EXFSC, lot 636681; Cat #SOP73LOBDA, lot 635147; Cat #SOP73LXTCK, lot 634327;  Cat #SOP73TKSJB, lot 634055; Cat #SOP7CEMBRD, lot 8822; Cat #SOP92ONSID, lot 634481; Cat #SOPBOKANFE, lot 636064; Cat #SOPCGCTAUA, lot 987152; Cat #SOPCGHNWMC, lot 985228; Cat #SOPCGKAAUA, lot 987068; Cat #SOPCGTKAUB, lot 987207; Cat #SOPHBKNBIC, lot 9719; Cat #SOPHDEXRVB, lot 636398; Cat #SOPOCACPAB, lot 9760; Cat #SOPOCBOMSA, lot 10558; Cat #SPP99HD1AA, lot 636774 and 633896; Cat #SPP99KA2AA, lot 636315; Cat #SPP99PD1AA, lot 635104; Cat #SPP99TN1AA, lot 636786; Cat #SPP99TN3AA, lot 636027; Cat #SRH32STFGA, lot 635101; Cat #SRH73ACLSB, lot 635369; Cat #SRH73EXCCA, lot 986309; Cat #SRH73EXLSB, lot 635099; Cat #SRIOR0715Z, lot 9883; Cat #SUT69FBCPO, lot 633481;  
FEI Number 3001236905
Recalling Firm/
Manufacturer
Cardinal Health
1430 Waukegan Rd
Mc Gaw Park IL 60085-6726
For Additional Information ContactMichele B. Donatich R.N.
800-292-9332
Manufacturer Reason
for Recall
Esmark bandages placed in some Presource kits/modules may not have been adequately sterilized.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe firm, Cardinal Health, sent an "URGENT: PRODUCT RECALL" letter dated January 28, 2011 to all affected consignees/customers via UPS next day air. The letter described the product, problem and action to be taken. The customers were instructed to quarantine and return the packs/modules to Cardinal Health for destruction; contact the appropriate Customer Service group listed in the letter for return instructions, and complete and return the enclosed Acknowledgment Form via fax to 847-689-9101, regardless of whether or not they have inventory. Should you have any questions, relating to this recall, please feel free to contact Cardinal Health Quality Systems, 800-292-9332.
Quantity in Commerce4.395 modules
DistributionNationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-