| | Class 3 Device Recall LUCAS 2 Chest Compression System |  |
| Date Initiated by Firm | February 08, 2011 |
| Date Posted | March 16, 2011 |
| Recall Status1 |
Terminated 3 on April 24, 2012 |
| Recall Number | Z-1664-2011 |
| Recall Event ID |
57967 |
| 510(K)Number | K090422 |
| Product Classification |
External Cardiac Compressor - Product Code DRM
|
| Product | LUCAS 2 Chest Compression System.
The device's Instruction for Use is labeled in part: "LUCAS 2 Chest Compression System***jolife***www.lucas-cpr.com***a product by JOLIFE***MANUFACTURER, MAIN OFFICE JOLIFE AB Scheelevagen 17 SE-223 70 LUND Sweden***www.jolife.com***".
Designed to uninterrupted chest compressions at a consistent rate and depth. |
| Code Information |
SERIAL NUMBERS for 546 UNITS DISTRIBUTED BY PHYSIO-CONTROL IN THE US: 30090182, 30090188, 30090216, 30090232, 30090242, 30090273, 30090275, 30090276, 30090277, 30090278, 30090279, 30090280, 30090281, 30090282, 30090284, 30090285, 30090286, 30090305, 30090314, 30090318, 30090325, 30090327, 30090328, 30090331, 30090333, 30090335, 30090426, 30090428, 30090429, 30090430, 30090432, 30090433, 30090434, 30090436, 30090437, 30090438, 30090439, 30090440, 30090442, 30090443, 30090445, 30090448, 30090449, 30090450, 30090451, 30090452, 30090453, 30090454, 30090457, 30090458, 30090459, 30090460, 30090461, 30090471, 30090499, 30090500, 30090501, 30090502, 30090503, 30090504, 30090505, 30090506, 30090507, 30090508, 30090509, 30090510, 30090511, 30090512, 30090513, 30090514, 30090515, 30090516, 30090517, 30090550, 30090564, 30090565, 30090566, 30090567, 30090568, 30090569, 30090575, 30090577, 30090580, 30090581, 30090583, 30090584, 30090629, 30090633, 30090653, 30090661, 30090665, 30090666, 30090670, 30090671, 30090672, 30090673, 30090674, 30090675, 30090676, 30090677, 30090678, 30090679, 30090680, 30090681, 30090682, 30100752, 30100753, 30100754, 30100755, 30100756, 30100757, 30100758, 30100759, 30100760, 30100761, 30100762, 30100763, 30100764, 30100765, 30100816, 30100817, 30100818, 30100819, 30100820, 30100821, 30100822, 30100823, 30100824, 30100825, 30100826, 30100827, 30100828, 30100829, 30100830, 30100831, 30100832, 30100834, 30100835, 30100838, 30100839, 30100840, 30100841, 30100842, 30100843, 30100844, 30100845, 30100846, 30100847, 30100848, 30100849, 30100850, 30100851, 30100852, 30100853, 30100854, 30100856, 30100858, 30100859, 30100860, 30100861, 30100863, 30100867, 30100868, 30100873, 30100898, 30100909, 30100944, 30100945, 30100946, 30100947, 30100948, 30100949, 30100950, 30100951, 30100952, 30100953, 30100954, 30100955, 30100956, 30100957, 30100958, 30100959, 30100960, 30100961, 30100962, 30100963, 30100964, 30100965, 30100966, 30100967, 30100968, 30100969, 30100970, 30100971, 30100972, 30100973, 30100974, 30100975, 30100976, 30100977, 30100978, 30100979, 30100980, 30100982, 30100983, 30100984, 30100986, 30101060, 30101062, 30101063, 30101064, 30101065, 30101066, 30101067, 30101074, 30101075, 30101077, 30101084, 30101085, 30101086, 30101087, 30101088, 30101124, 30101125, 30101160, 30101170, 30101171, 30101172, 30101173, 30101174, 30101175, 30101176, 30101177, 30101178, 30101179, 30101180, 30101181, 30101182, 30101183, 30101184, 30101185, 30101186, 30101187, 30101188, 30101189, 30101190, 30101191, 30101192, 30101193, 30101194, 30101195, 30101196, 30101197, 30101198, 30101199, 30101200, 30101201, 30101202, 30101203, 30101204, 30101205, 30101206, 30101207, 30101208, 30101209, 30101210, 30101211, 30101212, 30101213, 30101214, 30101215, 30101216, 30101217, 30101218, 30101219, 30101220, 30101221, 30101222, 30101223, 30101224, 30101225, 30101226, 30101227, 30101228, 30101229, 30101230, 30101231, 30101232, 30101233, 30101234, 30101235, 30101236, 30101237, 30101252, 30101254, 30101265, 30101266, 30101267, 30101268, 30101276, 30101277, 30101278, 30101279, 30101280, 30101281, 30101282, 30101283, 30101284, 30101285, 30101286, 30101287, 30101288, 30101290, 30101292, 30101293, 30101294, 30101295, 30101333, 30101334, 30101337, 30101338, 30101343, 30101344, 30101345, 30101346, 30101347, 30101349, 30101350, 30101351, 30101353, 30101354, 30101355, 30101356, 30101380, 30101381, 30101382, 30101386, 30101388, 30101389, 30101391, 30101392, 30101393, 30101394, 30101395, 30101396, 30101397, 30101398, 30101399, 30101400, 30101401, 30101402, 30101403, 30101404, 30101405, 30101406, 30101407, 30101408, 30101409, 30101411, 30101412, 30101413, 30101414, 30101452, 30101453, 30101455, 30101456, 30101457, 30101459, 30101460, 30101463, 30101464, 30101465, 30101473, 30101475, 30101476, 30101481, 30101495, 30101496, 30101497, 30101503, 30101504, 30101505, 30101506, 30101507, 30101508, 30101509, 30101510, 30101511, 30101512, 30101513, 30101514, 30101515, 30101516, 30101517, 30101518, 30101519, 30101520, 30101521, 30101522, 30101523, 30101524, 30101525, 30101526, 30101527, 30101528, 30101529, 30101530, 30101531, 30101532, 30101533, 30101534, 30101535, 30101536, 30101537, 30101538, 30101539, 30101540, 30101541, 30101542, 30101544, 30101545, 30101547, 30101548, 30101549, 30101551, 30101553, 30101554, 30101555, 30101556, 30101557, 30101559, 30101560, 30101573, 30101576, 30101577, 30101578, 30101579, 30101584, 30101607, 30101609, 30101610, 30101611, 30101612, 30101614, 30101615, 30101616, 30101617, 30101618, 30101620, 30101621, 30101622, 30101623, 30101624, 30101661, 30101662, 30101663, 30101665, 30101666, 30101667, 30101668, 30101670, 30101673, 30101674, 30101675, 30101676, 30101677, 30101678, 30101679, 30101681, 30101682, 30101683, 30101690, 30101696, 30101698, 30101701, 30101704, 30101705, 30101706, 30101707, 30101709, 30101724, 30101725, 30101726, 30101727, 30101728, 30101729, 30101730, 30101731, 30101732, 30101733, 30101734, 30101735, 30101736, 30101737, 30101751, 30101752, 30101762, 30101788, 30101789, 30101790, 30101791, 30101792, 30101793, 30101794, 30101795, 30101796, 30101797, 30101798, 30101799, 30101800, 30101805, 30101806, 30101808, 30101811, 30101814, 30101815, 30101816, 30101817, 30101818, 30101820, 30101821, 30101865, 30101866, 30101867, 30101868, 30101869, 30101870, 30101871, 30101872, 30101873, 30101874, 30101875, 30101876, 30101878, 30101879, 30101880, 30101881, 30101882, 30101883, 30101885, 30101887, 30101919, 30101921, 30101923, 30101927, and 30101928. SERIAL NUMBERS FOR 892 UNITS DISTRIBUTED BY PHYSIO-CONTROL TO THEIR FOREIGN CONSIGNEES: 30090054, 30090055, 30090059, 30090060, 30090061, 30090062, 30090063, 30090064, 30090065, 30090067, 30090068, 30090069, 30090070, 30090071, 30090072, 30090073, 30090074, 30090075, 30090076, 30090077, 30090078, 30090079, 30090080, 30090081, 30090082, 30090083, 30090084, 30090085, 30090087, 30090088, 30090111, 30090112, 30090113, 30090114, 30090115, 30090116, 30090117, 30090141, 30090142, 30090164, 30090165, 30090166, 30090168, 30090169, 30090171, 30090172, 30090173, 30090174, 30090175, 30090178, 30090180, 30090181, 30090183, 30090184, 30090185, 30090186, 30090187, 30090189, 30090190, 30090191, 30090192, 30090193, 30090194, 30090195, 30090196, 30090197, 30090198, 30090199, 30090206, 30090208, 30090209, 30090213, 30090221, 30090223, 30090224, 30090225, 30090226, 30090228, 30090233, 30090235, 30090237, 30090238, 30090239, 30090240, 30090241, 30090243, 30090244, 30090245, 30090246, 30090247, 30090248, 30090249, 30090250, 30090251, 30090252, 30090253, 30090254, 30090255, 30090256, 30090257, 30090258, 30090259, 30090260, 30090261, 30090262, 30090263, 30090264, 30090265, 30090266, 30090267, 30090268, 30090269, 30090270, 30090271, 30090272, 30090274, 30090283, 30090287, 30090288, 30090289, 30090290, 30090291, 30090292, 30090293, 30090294, 30090295, 30090296, 30090297, 30090298, 30090299, 30090300, 30090301, 30090302, 30090303, 30090304, 30090306, 30090307, 30090308, 30090309, 30090310, 30090311, 30090312, 30090313, 30090317, 30090319, 30090320, 30090321, 30090323, 30090324, 30090329, 30090330, 30090332, 30090334, 30090336, 30090337, 30090338, 30090339, 30090340, 30090341, 30090342, 30090343, 30090344, 30090345, 30090346, 30090347, 30090348, 30090349, 30090350, 30090351, 30090352, 30090353, 30090354, 30090355, 30090356, 30090357, 30090358, 30090359, 30090360, 30090361, 30090362, 30090363, 30090364, 30090365, 30090366, 30090367, 30090368, 30090369, 30090370, 30090371, 30090372, 30090373, 30090374, 30090375, 30090376, 30090377, 30090378, 30090379, 30090380, 30090381, 30090382, 30090383, 30090384, 30090385, 30090386, 30090387, 30090388, 30090389, 30090390, 30090391, 30090392, 30090393, 30090394, 30090395, 30090396, 30090397, 30090398, 30090399, 30090400, 30090401, 30090402, 30090403, 30090404, 30090405, 30090406, 30090407, 30090408, 30090409, 30090410, 30090411, 30090412, 30090413, 30090414, 30090415, 30090416, 30090417, 30090418, 30090419, 30090420, 30090421, 30090422, 30090423, 30090424, 30090425, 30090427, 30090455, 30090456, 30090462, 30090463, 30090464, 30090524, 30090525, 30090526, 30090527, 30090528, 30090536, 30090544, 30090547, 30090548, 30090551, 30090552, 30090553, 30090554, 30090556, 30090559, 30090560, 30090561, 30090562, 30090563, 30090585, 30090586, 30090587, 30090589, 30090590, 30090591, 30090592, 30090593, 30090594, 30090595, 30090596, 30090597, 30090598, 30090599, 30090600, 30090601, 30090602, 30090603, 30090604, 30090605, 30090606, 30090607, 30090608, 30090609, 30090610, 30090611, 30090612, 30090613, 30090614, 30090615, 30090634, 30090637, 30090640, 30090641, 30090642, 30090643, 30090644, 30090646, 30090647, 30090650, 30090651, 30090652, 30090654, 30090655, 30090656, 30090657, 30090658, 30090659, 30090660, 30090662, 30090663, 30090664, 30090694, 30090695, 30090696, 30090697, 30090698, 30090699, 30090700, 30090701, 30090702, 30090703, 30090704, 30090705, 30090706, 30090707, 30090708, 30090710, 30090711, 30090712, 30090713, 30090714, 30100766, 30100767, 30100768, 30100769, 30100770, 30100771, 30100772, 30100773, 30100774, 30100775, 30100776, 30100777, 30100778, 30100779, 30100780, 30100781, 30100782, 30100783, 30100784, 30100785, 30100786, 30100787, 30100788, 30100789, 30100790, 30100791, 30100792, 30100793, 30100794, 30100795, 30100796, 30100797, 30100798, 30100799, 30100800, 30100801, 30100802, 30100803, 30100804, 30100805, 30100806, 30100807, 30100808, 30100809, 30100810, 30100811, 30100812, 30100813, 30100814, 30100815, 30100862, 30100870, 30100871, 30100872, 30100874, 30100875, 30100882, 30100883, 30100884, 30100885, 30100886, 30100887, 30100888, 30100889, 30100890, 30100891, 30100892, 30100893, 30100894, 30100895, 30100896, 30100897, 30100899, 30100900, 30100901, 30100902, 30100903, 30100904, 30100905, 30100906, 30100907, 30100908, 30100910, 30100911, 30100912, 30100913, 30100914, 30100915, 30100916, 30100917, 30100918, 30100919, 30100920, 30100921, 30100922, 30100923, 30100924, 30100925, 30100926, 30100927, 30100928, 30100929, 30100930, 30100931, 30100932, 30100933, 30100934, 30100935, 30100936, 30100937, 30100938, 30100939, 30100940, 30100941, 30100942, 30100943, 30100987, 30100988, 30100989, 30100990, 30100991, 30100992, 30100993, 30100994, 30100995, 30100996, 30100997, 30101000, 30101001, 30101002, 30101003, 30101004, 30101005, 30101006, 30101007, 30101008, 30101009, 30101029, 30101036, 30101046, 30101055, 30101056, 30101057, 30101058, 30101059, 30101061, 30101068, 30101069, 30101070, 30101071, 30101072, 30101073, 30101076, 30101078, 30101079, 30101080, 30101081, 30101082, 30101083, 30101089, 30101092, 30101093, 30101094, 30101095, 30101096, 30101097, 30101098, 30101099, 30101100, 30101101, 30101102, 30101103, 30101104, 30101105, 30101106, 30101107, 30101108, 30101109, 30101110, 30101111, 30101112, 30101113, 30101114, 30101115, 30101116, 30101117, 30101118, 30101119, 30101120, 30101122, 30101123, 30101126, 30101128, 30101129, 30101130, 30101131, 30101132, 30101133, 30101136, 30101138, 30101143, 30101144, 30101147, 30101149, 30101150, 30101151, 30101152, 30101153, 30101154, 30101159, 30101239, 30101244, 30101245, 30101246, 30101248, 30101249, 30101251, 30101255, 30101256, 30101257, 30101258, 30101259, 30101260, 30101261, 30101262, 30101263, 30101264, 30101269, 30101270, 30101271, 30101272, 30101273, 30101274, 30101275, 30101291, 30101296, 30101297, 30101298, 30101299, 30101300, 30101301, 30101302, 30101304, 30101305, 30101306, 30101307, 30101308, 30101309, 30101310, 30101311, 30101312, 30101313, 30101314, 30101316, 30101317, 30101318, 30101320, 30101321, 30101322, 30101323, 30101324, 30101325, 30101326, 30101327, 30101328, 30101329, 30101330, 30101331, 30101332, 30101337, 30101340, 30101341, 30101348, 30101358, 30101359, 30101360, 30101361, 30101362, 30101363, 30101364, 30101365, 30101366, 30101367, 30101368, 30101369, 30101370, 30101371, 30101372, 30101373, 30101374, 30101375, 30101376, 30101377, 30101378, 30101379, 30101415, 30101416, 30101417, 30101418, 30101419, 30101420, 30101421, 30101422, 30101423, 30101424, 30101425, 30101426, 30101427, 30101428, 30101429, 30101430, 30101431, 30101434, 30101435, 30101436, 30101437, 30101438, 30101439, 30101440, 30101441, 30101442, 30101443, 30101444, 30101445, 30101446, 30101447, 30101448, 30101449, 30101450, 30101451, 30101454, 30101458, 30101461, 30101462, 30101466, 30101467, 30101468, 30101469, 30101470, 30101471, 30101472, 30101474, 30101477, 30101478, 30101479, 30101480, 30101482, 30101483, 30101484, 30101485, 30101486, 30101487, 30101488, 30101490, 30101491, 30101492, 30101493, 30101494, 30101498, 30101499, 30101500, 30101501, 30101502, 30101552, 30101561, 30101562, 30101563, 30101564, 30101565, 30101566, 30101567, 30101568, 30101569, 30101570, 30101571, 30101572, 30101574, 30101575, 30101580, 30101581, 30101582, 30101583, 30101585, 30101586, 30101587, 30101588, 30101589, 30101590, 30101591, 30101592, 30101593, 30101594, 30101595, 30101596, 30101597, 30101598, 30101599, 30101600, 30101602, 30101603, 30101604, 30101605, 30101606, 30101608, 30101613, 30101625, 30101626, 30101627, 30101628, 30101629, 30101631, 30101632, 30101633, 30101634, 30101635, 30101636, 30101637, 30101638, 30101639, 30101640, 30101641, 30101642, 30101643, 30101644, 30101645, 30101646, 30101647, 30101648, 30101649, 30101650, 30101651, 30101652, 30101653, 30101654, 30101655, 30101656, 30101657, 30101658, 30101659, 30101660, 30101684, 30101685, 30101686, 30101687, 30101688, 30101689, 30101691, 30101692, 30101693, 30101694, 30101695, 30101697, 30101699, 30101700, 30101702, 30101703, 30101708, 30101710, 30101711, 30101712, 30101713, 30101714, 30101715, 30101716, 30101717, 30101718, 30101719, 30101720, 30101721, 30101722, 30101723, 30101738, 30101739, 30101740, 30101741, 30101742, 30101743, 30101745, 30101746, 30101747, 30101748, 30101749, 30101750, 30101753, 30101754, 30101755, 30101756, 30101757, 30101758, 30101759, 30101760, 30101761, 30101764, 30101768, 30101771, 30101773, 30101775, 30101776, 30101778, 30101779, 30101781, 30101783, 30101786, 30101787, 30101801, 30101802, 30101803, 30101804, 30101813, 30101819, 30101822, 30101823, 30101824, 30101825, 30101826, 30101827, 30101828, 30101829, 30101830, 30101832, 30101833, 30101834, 30101835, 30101836, 30101837, 30101838, 30101839, 30101840, 30101841, 30101846, 30101848, 30101850, 30101851, 30101852, 30101853, 30101854, 30101855, 30101856, 30101857, 30101858, 30101859, 30101860, 30101893, 30101909, 30101910, 30101911, 30101912, 30101913, 30101914, 30101916, 30101930, and 30101932. SERIAL NUMBERS FOR 224 DEMO UNITS OWNED by PHYSIO-CONTROL: 30090026, 30090027, 30090028, 30090029, 30090031, 30090032, 30090033, 30090034, 30090035, 30090086, 30090089, 30090090, 30090091, 30090092, 30090093, 30090094, 30090095, 30090096, 30090097, 30090098, 30090099, 30090100, 30090102, 30090103, 30090104, 30090105, 30090106, 30090108, 30090109, 30090110, 30090118, 30090119, 30090120, 30090121, 30090122, 30090123, 30090124, 30090125, 30090126, 30090127, 30090128, 30090129, 30090130, 30090131, 30090132, 30090133, 30090134, 30090135, 30090136, 30090137, 30090138, 30090139, 30090140, 30090143, 30090144, 30090145, 30090146, 30090147, 30090148, 30090149, 30090150, 30090151, 30090152, 30090153, 30090154, 30090155, 30090156, 30090157, 30090158, 30090160, 30090161, 30090162, 30090163, 30090167, 30090170, 30090200, 30090201, 30090202, 30090203, 30090204, 30090205, 30090207, 30090210, 30090211, 30090212, 30090214, 30090215, 30090217, 30090218, 30090219, 30090220, 30090222, 30090227, 30090229, 30090230, 30090231, 30090234, 30090236, 30090326, 30090441, 30090444, 30090446, 30090447, 30090466, 30090467, 30090468, 30090469, 30090470, 30090472, 30090473, 30090474, 30090475, 30090476, 30090477, 30090478, 30090479, 30090480, 30090481, 30090482, 30090483, 30090484, 30090485, 30090486, 30090487, 30090488, 30090489, 30090490, 30090491, 30090492, 30090493, 30090494, 30090495, 30090496, 30090497, 30090498, 30090518, 30090519, 30090520, 30090521, 30090522, 30090523, 30090529, 30090530, 30090531, 30090532, 30090533, 30090534, 30090535, 30090537, 30090538, 30090539, 30090540, 30090541, 30090542, 30090543, 30090545, 30090546, 30090549, 30090555, 30090557, 30090558, 30090570, 30090571, 30090572, 30090573, 30090574, 30090576, 30090578, 30090579, 30090582, 30090588, 30090616, 30090617, 30090618, 30090619, 30090620, 30090621, 30090622, 30090623, 30090624, 30090625, 30090626, 30090627, 30090628, 30090630, 30090631, 30090632, 30090635, 30090636, 30090638, 30090639, 30090645, 30090683, 30090684, 30090685, 30090686, 30090687, 30090688, 30090689, 30090690, 30090691, 30090692, 30090693, 30100880, 30100881, 30100985, 30101090, 30101091, 30101289, 30101335, 30101336, 30101339, 30101342, 30101352, 30101383, 30101384, 30101385, 30101387, 30101390, 30101410, 30101619, 30101669, 30101672, and 30101680. SERIAL NUMBERS FOR 181 UNITS DISTRIBUTED BY JOLIFE, SWEDEN: 30090016, 30090018, 30090019, 30090020, 30090021, 30090022, 30090023, 30090024, 30090025, 30090030, 30090036, 30090037, 30090038, 30090039, 30090040, 30090041, 30090042, 30090043, 30090044, 30090045, 30090046, 30090047, 30090048, 30090049, 30090050, 30090051, 30090052, 30090053, 30090056, 30090057, 30090058, 30090101, 30090107, 30090159, 30090315, 30090316, 30090322, 30090431, 30090431, 30090435, 30090465, 30090648, 30090649, 30090667, 30090668, 30090669, 30090709, 30090715, 30090716, 30090717, 30090718, 30090719, 30090720, 30090721, 30090722, 30090723, 30090724, 30090725, 30090726, 30090727, 30090728, 30090729, 30090730, 30090731, 30090732, 30090733, 30090734, 30090735, 30090736, 30090737, 30090738, 30090739, 30090740, 30100741, 30100742, 30100743, 30100744, 30100745, 30100746, 30100747, 30100748, 30100749, 30100750, 30100751, 30100833, 30100836, 30100837, 30100855, 30100857, 30100864, 30100865, 30100866, 30100869, 30100876, 30100877, 30100878, 30100879, 30100998, 30100999, 30101010, 30101011, 30101012, 30101013, 30101014, 30101015, 30101016, 30101017, 30101018, 30101019, 30101020, 30101021, 30101022, 30101023, 30101024, 30101025, 30101026, 30101027, 30101028, 30101030, 30101031, 30101032, 30101033, 30101034, 30101035, 30101037, 30101038, 30101039, 30101040, 30101041, 30101042, 30101043, 30101044, 30101045, 30101047, 30101048, 30101049, 30101050, 30101051, 30101052, 30101053, 30101054, 30101121, 30101127, 30101134, 30101135, 30101137, 30101139, 30101140, 30101141, 30101142, 30101145, 30101146, 30101148, 30101155, 30101156, 30101157, 30101158, 30101161, 30101162, 30101163, 30101164, 30101165, 30101166, 30101167, 30101168, 30101169, 30101238, 30101240, 30101241, 30101242, 30101243, 30101247, 30101250, 30101253, 30101303, 30101315, 30101319, 30101432, 30101433, 30101766, and 30101886. |
| FEI Number |
3005445717
|
Recalling Firm/ Manufacturer |
Jolife AB
Ideon, Scheelev, 17
Lund
Sweden
|
| For Additional Information Contact | Physio-Control Technical Support 800-442-1142 Ext. 5 |
Manufacturer Reason for Recall | LUCAS 2 Chest Compression System may have malfunctions resulting from the failure of a hood probe on the device that connects to the battery. The device may stop in fail safe mode and the LEDs will start flickering or the device will switch off. |
FDA Determined Cause 2 | Other |
| Action | Medical Device Correction letter, dated February 2011, was sent to customers and states that Jolife AB with the servicing support assistant of Physio-Control is conducting a voluntary upgrade of the LUCAS 2 Chest Compression System with the Serial number range of 30090016 through 30101932.
Consignees were advised to keep LUCAS 2 in service and continue to use it in accordance with the Instruction for Use, Section 3.8-Safety Precautions. WARNING-MALFUNCTION. If there are interruptions, or the compressions are not sufficient or something unusual occurs during operation: Push ON/OFF for 1 second to stop LUCAS and remove the device. Start Manual chest compression.
Physio-Control service representative will contact the consignees to schedule a service call to replace the affected battery connector board in the hood of the device. The new connector board will include a new probe design in which current cannot affect the spring.
Consignees can contact Physio-Control Technical Support at 1-800-442-1142, #5 between 6 am-4 pm (Pacific Time), Monday-Friday or visit the website www.physio-control-notices.com/LUCAS2. If customers no longer have the affected device or the device location has changed they should contact Physio-Control Technical Support. |
| Quantity in Commerce | 1843 units |
| Distribution | Worldwide Distribution -- United States, Australia, Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirates, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = DRM
|
|
|
|