• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall SYSTEM 1E liquid chemical sterilant processing system

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall SYSTEM 1E liquid chemical sterilant processing systemsee related information
Date Initiated by FirmMarch 14, 2011
Date PostedApril 12, 2011
Recall Status1 Terminated 3 on March 13, 2013
Recall NumberZ-1960-2011
Recall Event ID 58290
510(K)NumberK090036 K101409 K102462 
Product Classification Sterilant, medical devices - Product Code MED
ProductSTERIS SYSTEM 1E Liquid Chemical Sterilant Processing System The SYSTEM 1 E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned impressible, reusable, critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories. Devices processed in a SYSTEM IE have been chemically sterilized using S40 Sterilant Concentrate, a peracetic acid liquid chemical sterilant and rinsed with extensively treated, potable water.
Code Information All codes manufactured from 5/5/2010 through 3/8/2011, are subject to recall/correction.
FEI Number 3000251274
Recalling Firm/
Manufacturer
Steris Corporation Hopkins Facility
6515 Hopkins Rd
Mentor OH 44060-4307
For Additional Information ContactLindsey McGowan
440-392-7601
Manufacturer Reason
for Recall
The printout, display and Operator Manual for the STERIS System 1E Liquid Chemical Sterilant Processing System are subject to recall/correction by the firm because in the event of an incomplete or aborted cycle, the language of the warning messages used in the printout, display, and Operator Manual omits the qualifying term: *liquid chemically* (not **sterilized), when alerting the user to inco
FDA Determined
Cause 2
Other
ActionThe firm, STERIS, Issued an 'Urgent Field Correction Notice' dated March 14, 2011 addressed to their Hospital Administrator, or Manager, and Risk Manager customers on March 18, 2011. The notification letter describes the product subject to field correction and the problem that exists with the language omission present in the printout display and Operator Manual messages which warn of an incomplete or aborted processing cycle. The notification goes on to state that there is no health hazard involved in this error, as the users are still made aware that the processing cycle has not been successfully completed. The notification goes on to remind the user/customers that in case of any incomplete, cancelled or aborted cycle, the users are instructed to process devices in the cycle following the directions contained in the Operator Manual. Lastly, the notification letter informs the user/customers that the company will be installing new software on the recalled systems in order to revise the language on the cycle printout and the display screen warning of a disrupted cycle. It also states that a STERIS Service Representative will be contacting each facility in order to schedule installation of the new software. The notification also provides the customers with a 1-800 telephone number for STERIS Field Service Dispatch, to be used if the customers have questions or want to request additional information. It also provides a local telephone no. 1-440-392-7455, for obtaining additional information from the firm's Director, Div. of Low Temperature Sterilization at: 1-440-392-7455.
Quantity in Commerce106 units
DistributionNationwide distribution: USA including states of :CA, DE, FL, GA, IA, ID, IL, KS, MA, MO, MT, NE, NJ, NM, NY, OH, OR, SD, TX, VA, WA, WI, and WV.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = MED
-
-