| | Class 2 Device Recall Millennium and HD120 MultiLeaf Collimators (MLC), Clinac |  |
| Date Initiated by Firm | March 17, 2011 |
| Date Posted | September 14, 2011 |
| Recall Status1 |
Terminated 3 on June 20, 2012 |
| Recall Number | Z-3204-2011 |
| Recall Event ID |
58474 |
| 510(K)Number | K100890 |
| Product Classification |
radiotherapy delivery system - Product Code IYE
|
| Product | Varian Medical Systems Inc., Millennium and HD-120 MultiLeaf Collimators (MLC), a component of the Clinac, All Varian Clinac using the collimator cover P/N 1106900-xx, Model Numbers:
H14, H18, H27, H29, HCX; part of the Trilogy Radiotherapy Delivery System. |
| Code Information |
LIST OF SERIAL NUMBERS H140001 ,H140049 ,H140097 ,H140144 ,H140235 ,H140742 ,H140788 ,H140856 ,H180473 ,H180645 ,H180718 ,H140002 ,H140051 ,H140098 ,H140145 ,H140236 ,H140743 ,H140789 ,H140882 ,H180485 ,H180647 ,H180719 ,H140003 ,H140052 ,H140099 ,H140146 ,H140237 ,H140744 ,H140790 ,H140889 ,H180487 ,H180649 ,H180720 ,H140004 ,H140053 ,H140100 ,H140147 ,H140238 ,H140745 ,H140791 ,H180003 ,H180529 ,H180650 ,H180721 ,H140005 ,H140054 ,H140101 ,H140148 ,H140239 ,H140746 ,H140792 ,H180056 ,H180532 ,H180651 ,H180722 ,H140006 ,H140055 ,H140102 ,H140149 ,H140240 ,H140747 ,H140793 ,H180088 ,H180535 ,H180655 ,H180723 ,H140007 ,H140056 ,H140103 ,H140150 ,H140700 ,H140748 ,H140794 ,H180102 ,H180536 ,H180657 ,H180724 ,H140008 ,H140057 ,H140105 ,H140151 ,H140701 ,H140749 ,H140795 ,H180115 ,H180542 ,H180659 ,H180725 ,H140009 ,H140058 ,H140106 ,H140152 ,H140702 ,H140750 ,H140796 ,H180117 ,H180548 ,H180660 ,H180726 ,H140010 ,H140060 ,H140107 ,H140153 ,H140703 ,H140751 ,H140797 ,H180144 ,H180550 ,H180661 ,H180728 ,H140011 ,H140061 ,H140108 ,H140154 ,H140704 ,H140752 ,H140798 ,H180166 ,H180553 ,H180665 ,H180729 ,H140012 ,H140062 ,H140110 ,H140201 ,H140705 ,H140753 ,H140799 ,H180167 ,H180555 ,H180667 ,H180730 ,H140014 ,H140063 ,H140111 ,H140202 ,H140707 ,H140754 ,H140800 ,H180173 ,H180556 ,H180668 ,H180731 ,H140015 ,H140064 ,H140112 ,H140203 ,H140708 ,H140755 ,H140801 ,H180201 ,H180559 ,H180669 ,H180733 ,H140016 ,H140065 ,H140113 ,H140204 ,H140709 ,H140756 ,H140802 ,H180204 ,H180561 ,H180670 ,H180734 ,H140017 ,H140066 ,H140114 ,H140205 ,H140710 ,H140757 ,H140803 ,H180206 ,H180575 ,H180671 ,H180735 ,H140018 ,H140067 ,H140115 ,H140206 ,H140711 ,H140758 ,H140804 ,H180210 ,H180581 ,H180672 ,H180736 ,H140019 ,H140068 ,H140116 ,H140207 ,H140712 ,H140759 ,H140805 ,H180217 ,H180588 ,H180674 ,H180737 ,H140021 ,H140069 ,H140117 ,H140208 ,H140713 ,H140760 ,H140806 ,H180220 ,H180591 ,H180676 ,H180738 ,H140022 ,H140070 ,H140118 ,H140209 ,H140714 ,H140761 ,H140807 ,H180221 ,H180593 ,H180679 ,H180739 ,H140023 ,H140071 ,H140119 ,H140210 ,H140715 ,H140762 ,H140808 ,H180222 ,H180596 ,H180680 ,H180741 ,H140024 ,H140072 ,H140120 ,H140211 ,H140716 ,H140763 ,H140809 ,H180228 ,H180597 ,H180681 ,H180744 ,H140025 ,H140073 ,H140121 ,H140212 ,H140717 ,H140764 ,H140810 ,H180235 ,H180599 ,H180682 ,H180745 ,H140026 ,H140074 ,H140122 ,H140213 ,H140718 ,H140765 ,H140811 ,H180255 ,H180600 ,H180683 ,H180746 ,H140027 ,H140075 ,H140123 ,H140214 ,H140719 ,H140766 ,H140812 ,H180256 ,H180601 ,H180684 ,H180747 ,H140028 ,H140076 ,H140124 ,H140215 ,H140720 ,H140767 ,H140813 ,H180260 ,H180603 ,H180685 ,H180750 ,H140029 ,H140077 ,H140125 ,H140216 ,H140721 ,H140768 ,H140814 ,H180268 ,H180604 ,H180687 ,H180751 ,H140030 ,H140078 ,H140126 ,H140217 ,H140722 ,H140769 ,H140815 ,H180271 ,H180605 ,H180690 ,H180752 ,H140032 ,H140079 ,H140127 ,H140218 ,H140723 ,H140770 ,H140817 ,H180283 ,H180607 ,H180691 ,H180753 ,H140033 ,H140080 ,H140128 ,H140219 ,H140724 ,H140771 ,H140818 ,H180284 ,H180608 ,H180692 ,H180754 ,H140034 ,H140082 ,H140129 ,H140220 ,H140725 ,H140772 ,H140819 ,H180290 ,H180614 ,H180693 ,H180756 ,H140035 ,H140083 ,H140130 ,H140221 ,H140727 ,H140773 ,H140820 ,H180292 ,H180619 ,H180694 ,H180757 ,H140036 ,H140084 ,H140131 ,H140222 ,H140728 ,H140774 ,H140821 ,H180294 ,H180625 ,H180695 ,H180758 ,H140037 ,H140085 ,H140132 ,H140223 ,H140729 ,H140775 ,H140822 ,H180298 ,H180626 ,H180696 ,H180759 ,H140038 ,H140086 ,H140133 ,H140224 ,H140730 ,H140776 ,H140823 ,H180324 ,H180627 ,H180698 ,H180760 ,H140039 ,H140087 ,H140134 ,H140225 ,H140731 ,H140778 ,H140824 ,H180405 ,H180629 ,H180699 ,H180761 ,H140040 ,H140088 ,H140135 ,H140226 ,H140732 ,H140779 ,H140825 ,H180407 ,H180630 ,H180703 ,H180763 ,H140041 ,H140089 ,H140136 ,H140227 ,H140733 ,H140780 ,H140826 ,H180409 ,H180632 ,H180704 ,H180769 ,H140042 ,H140090 ,H140137 ,H140228 ,H140734 ,H140781 ,H140827 ,H180415 ,H180633 ,H180707 ,H180770 ,H140043 ,H140091 ,H140138 ,H140229 ,H140735 ,H140782 ,H140828 ,H180427 ,H180634 ,H180709 ,H180774 ,H140044 ,H140092 ,H140139 ,H140230 ,H140736 ,H140783 ,H140829 ,H180446 ,H180637 ,H180710 ,H180775 ,H140045 ,H140093 ,H140140 ,H140231 ,H140737 ,H140784 ,H140830 ,H180447 ,H180640 ,H180714 ,H180776 ,H140046 ,H140094 ,H140141 ,H140232 ,H140738 ,H140785 ,H140837 ,H180466 ,H180641 ,H180715 ,H180777 ,H140047 ,H140095 ,H140142 ,H140233 ,H140739 ,H140786 ,H140844 ,H180467 ,H180643 ,H180716 ,H180778 ,H140048 ,H140096 ,H140143 ,H140234 ,H140740 ,H140787 ,H140849 ,H180470 ,H180644 ,H180717 ,H180780 ,H180784 ,H180863 ,H181043 ,H181142 ,H270209 ,H270473 ,H270783 ,H271027 ,H271238 ,H271372 ,H271465 ,H180785 ,H180864 ,H181045 ,H181143 ,H270214 ,H270476 ,H270787 ,H271029 ,H271248 ,H271375 ,H271466 ,H180786 ,H180866 ,H181046 ,H181145 ,H270215 ,H270483 ,H270789 ,H271033 ,H271249 ,H271376 ,H271468 ,H180791 ,H180870 ,H181047 ,H181146 ,H270216 ,H270489 ,H270791 ,H271035 ,H271250 ,H271377 ,H271471 ,H180792 ,H180871 ,H181053 ,H181147 ,H270219 ,H270498 ,H270796 ,H271038 ,H271251 ,H271378 ,H271473 ,H180794 ,H180872 ,H181055 ,H181148 ,H270227 ,H270501 ,H270797 ,H271041 ,H271256 ,H271381 ,H271474 ,H180795 ,H180873 ,H181057 ,H181149 ,H270229 ,H270502 ,H270800 ,H271051 ,H271257 ,H271384 ,H271475 ,H180796 ,H180877 ,H181058 ,H181152 ,H270235 ,H270503 ,H270803 ,H271061 ,H271258 ,H271385 ,H271476 ,H180803 ,H180878 ,H181061 ,H181155 ,H270244 ,H270504 ,H270804 ,H271067 ,H271260 ,H271386 ,H271477 ,H180804 ,H180879 ,H181062 ,H181156 ,H270256 ,H270505 ,H270805 ,H271071 ,H271264 ,H271387 ,H271478 ,H180806 ,H180880 ,H181065 ,H181157 ,H270260 ,H270512 ,H270806 ,H271076 ,H271266 ,H271390 ,H271479 ,H180807 ,H180881 ,H181066 ,H181158 ,H270270 ,H270513 ,H270808 ,H271079 ,H271271 ,H271391 ,H271481 ,H180810 ,H180882 ,H181068 ,H181159 ,H270273 ,H270514 ,H270813 ,H271085 ,H271273 ,H271393 ,H271485 ,H180812 ,H180883 ,H181069 ,H181160 ,H270280 ,H270518 ,H270814 ,H271086 ,H271276 ,H271398 ,H271487 ,H180815 ,H180884 ,H181070 ,H181161 ,H270281 ,H270522 ,H270817 ,H271093 ,H271277 ,H271399 ,H271490 ,H180816 ,H180885 ,H181071 ,H181162 ,H270285 ,H270526 ,H270822 ,H271101 ,H271280 ,H271401 ,H271492 ,H180817 ,H180886 ,H181073 ,H181163 ,H270293 ,H270532 ,H270825 ,H271108 ,H271281 ,H271404 ,H271493 ,H180819 ,H180887 ,H181075 ,H181164 ,H270297 ,H270607 ,H270826 ,H271109 ,H271282 ,H271406 ,H271494 ,H180820 ,H180888 ,H181077 ,H181165 ,H270302 ,H270613 ,H270828 ,H271113 ,H271285 ,H271407 ,H271495 ,H180821 ,H180890 ,H181078 ,H181166 ,H270323 ,H270622 ,H270829 ,H271114 ,H271286 ,H271410 ,H271497 ,H180822 ,H180894 ,H181082 ,H181167 ,H270345 ,H270635 ,H270833 ,H271115 ,H271287 ,H271411 ,H271498 ,H180823 ,H180895 ,H181083 ,H181168 ,H270352 ,H270639 ,H270836 ,H271123 ,H271288 ,H271412 ,H271499 ,H180824 ,H180896 ,H181084 ,H181170 ,H270356 ,H270644 ,H270837 ,H271124 ,H271289 ,H271413 ,H271500 ,H180825 ,H180898 ,H181085 ,H181173 ,H270360 ,H270646 ,H270840 ,H271126 ,H271291 ,H271414 ,H271501 ,H180826 ,H181002 ,H181086 ,H181175 ,H270370 ,H270647 ,H270842 ,H271132 ,H271296 ,H271415 ,H271502 ,H180827 ,H181003 ,H181089 ,H181177 ,H270371 ,H270649 ,H270844 ,H271141 ,H271299 ,H271417 ,H271503 ,H180828 ,H181004 ,H181090 ,H181180 ,H270378 ,H270652 ,H270846 ,H271142 ,H271304 ,H271419 ,H271504 ,H180829 ,H181005 ,H181091 ,H181181 ,H270384 ,H270669 ,H270847 ,H271143 ,H271305 ,H271420 ,H271505 ,H180830 ,H181006 ,H181092 ,H181184 ,H270386 ,H270671 ,H270848 ,H271149 ,H271307 ,H271422 ,H271506 ,H180831 ,H181007 ,H181093 ,H181185 ,H270391 ,H270675 ,H270852 ,H271151 ,H271308 ,H271424 ,H271507 ,H180832 ,H181008 ,H181097 ,H260915 ,H270392 ,H270684 ,H270854 ,H271157 ,H271309 ,H271425 ,H271509 ,H180834 ,H181009 ,H181098 ,H260916 ,H270397 ,H270690 ,H270861 ,H271159 ,H271314 ,H271426 ,H271511 ,H180835 ,H181011 ,H181101 ,H260923 ,H270399 ,H270699 ,H270863 ,H271163 ,H271315 ,H271427 ,H271512 ,H180837 ,H181012 ,H181103 ,H260924 ,H270401 ,H270703 ,H270864 ,H271166 ,H271317 ,H271428 ,H271514 ,H180839 ,H181013 ,H181108 ,H260928 ,H270410 ,H270711 ,H270865 ,H271167 ,H271319 ,H271429 ,H271515 ,H180840 ,H181015 ,H181109 ,H260931 ,H270416 ,H270712 ,H270877 ,H271175 ,H271322 ,H271430 ,H271516 ,H180842 ,H181018 ,H181114 ,H260932 ,H270420 ,H270714 ,H270879 ,H271183 ,H271325 ,H271431 ,H271517 ,H180843 ,H181019 ,H181116 ,H260938 ,H270421 ,H270715 ,H270880 ,H271199 ,H271327 ,H271432 ,H271518 ,H180844 ,H181020 ,H181118 ,H260939 ,H270426 ,H270717 ,H270883 ,H271201 ,H271332 ,H271435 ,H271519 ,H180850 ,H181022 ,H181119 ,H260942 ,H270427 ,H270718 ,H270888 ,H271204 ,H271333 ,H271436 ,H271520 ,H180852 ,H181023 ,H181124 ,H260946 ,H270430 ,H270726 ,H270889 ,H271205 ,H271334 ,H271438 ,H271522 ,H180853 ,H181026 ,H181127 ,H260947 ,H270431 ,H270727 ,H270890 ,H271209 ,H271338 ,H271439 ,H271523 ,H180854 ,H181029 ,H181128 ,H260948 ,H270437 ,H270735 ,H270891 ,H271214 ,H271347 ,H271440 ,H271525 ,H180855 ,H181030 ,H181132 ,H260950 ,H270438 ,H270739 ,H270894 ,H271218 ,H271349 ,H271447 ,H271526 ,H180856 ,H181031 ,H181133 ,H260951 ,H270439 ,H270741 ,H271002 ,H271222 ,H271355 ,H271448 ,H271527 ,H180857 ,H181032 ,H181134 ,H260952 ,H270447 ,H270757 ,H271005 ,H271223 ,H271357 ,H271451 ,H271528 ,H180858 ,H181033 ,H181135 ,H260953 ,H270449 ,H270759 ,H271006 ,H271228 ,H271360 ,H271453 ,H271529 ,H180859 ,H181036 ,H181136 ,H270073 ,H270450 ,H270762 ,H271013 ,H271231 ,H271361 ,H271459 ,H271530 ,H180860 ,H181038 ,H181137 ,H270096 ,H270457 ,H270767 ,H271016 ,H271234 ,H271362 ,H271460 ,H271531 ,H180861 ,H181039 ,H181138 ,H270118 ,H270461 ,H270778 ,H271021 ,H271235 ,H271367 ,H271463 ,H271532 ,H180862 ,H181042 ,H181140 ,H270201 ,H270464 ,H270781 ,H271024 ,H271236 ,H271370 ,H271464 ,H271533 ,H271534 ,H271599 ,H271656 ,H271716 ,H271774 ,H271831 ,H271890 ,H271943 ,H271997 ,H272053 ,H272106 ,H271535 ,H271600 ,H271658 ,H271717 ,H271775 ,H271832 ,H271891 ,H271944 ,H271998 ,H272054 ,H272107 ,H271536 ,H271601 ,H271659 ,H271718 ,H271776 ,H271833 ,H271892 ,H271945 ,H271999 ,H272055 ,H272108 ,H271537 ,H271602 ,H271660 ,H271719 ,H271777 ,H271835 ,H271893 ,H271946 ,H272000 ,H272056 ,H272109 ,H271538 ,H271603 ,H271661 ,H271720 ,H271778 ,H271838 ,H271894 ,H271948 ,H272001 ,H272057 ,H272110 ,H271539 ,H271605 ,H271662 ,H271721 ,H271779 ,H271839 ,H271895 ,H271949 ,H272002 ,H272058 ,H272111 ,H271540 ,H271606 ,H271663 ,H271722 ,H271780 ,H271840 ,H271896 ,H271950 ,H272003 ,H272059 ,H272112 ,H271542 ,H271607 ,H271664 ,H271723 ,H271781 ,H271841 ,H271897 ,H271951 ,H272004 ,H272060 ,H272113 ,H271543 ,H271608 ,H271666 ,H271724 ,H271782 ,H271842 ,H271898 ,H271952 ,H272005 ,H272061 ,H272114 ,H271544 ,H271609 ,H271667 ,H271729 ,H271783 ,H271843 ,H271899 ,H271953 ,H272006 ,H272062 ,H272115 ,H271545 ,H271610 ,H271669 ,H271731 ,H271784 ,H271844 ,H271901 ,H271954 ,H272007 ,H272064 ,H272116 ,H271546 ,H271611 ,H271670 ,H271732 ,H271785 ,H271845 ,H271902 ,H271955 ,H272008 ,H272065 ,H272117 ,H271547 ,H271612 ,H271671 ,H271733 ,H271786 ,H271846 ,H271903 ,H271956 ,H272009 ,H272066 ,H272118 ,H271548 ,H271613 ,H271672 ,H271734 ,H271787 ,H271848 ,H271904 ,H271957 ,H272010 ,H272067 ,H272119 ,H271549 ,H271614 ,H271673 ,H271735 ,H271788 ,H271849 ,H271905 ,H271958 ,H272011 ,H272068 ,H272120 ,H271550 ,H271615 ,H271674 ,H271736 ,H271789 ,H271850 ,H271907 ,H271959 ,H272012 ,H272069 ,H272121 ,H271551 ,H271617 ,H271675 ,H271737 ,H271790 ,H271851 ,H271908 ,H271960 ,H272013 ,H272070 ,H272122 ,H271552 ,H271618 ,H271676 ,H271738 ,H271791 ,H271852 ,H271909 ,H271961 ,H272014 ,H272071 ,H272123 ,H271554 ,H271619 ,H271677 ,H271739 ,H271793 ,H271854 ,H271910 ,H271962 ,H272015 ,H272072 ,H272124 ,H271555 ,H271620 ,H271678 ,H271740 ,H271794 ,H271855 ,H271911 ,H271964 ,H272017 ,H272073 ,H272125 ,H271559 ,H271621 ,H271680 ,H271741 ,H271796 ,H271856 ,H271912 ,H271965 ,H272018 ,H272074 ,H272126 ,H271560 ,H271622 ,H271681 ,H271742 ,H271798 ,H271857 ,H271913 ,H271966 ,H272019 ,H272075 ,H272127 ,H271561 ,H271624 ,H271682 ,H271743 ,H271800 ,H271858 ,H271914 ,H271967 ,H272020 ,H272076 ,H272128 ,H271563 ,H271625 ,H271683 ,H271744 ,H271801 ,H271859 ,H271915 ,H271968 ,H272021 ,H272077 ,H272129 ,H271564 ,H271626 ,H271684 ,H271746 ,H271802 ,H271860 ,H271916 ,H271969 ,H272022 ,H272078 ,H272130 ,H271565 ,H271627 ,H271685 ,H271747 ,H271803 ,H271861 ,H271917 ,H271970 ,H272023 ,H272079 ,H272131 ,H271567 ,H271629 ,H271686 ,H271748 ,H271804 ,H271862 ,H271918 ,H271971 ,H272024 ,H272080 ,H272132 ,H271569 ,H271630 ,H271687 ,H271749 ,H271805 ,H271864 ,H271919 ,H271972 ,H272025 ,H272081 ,H272133 ,H271570 ,H271631 ,H271689 ,H271750 ,H271806 ,H271865 ,H271920 ,H271973 ,H272026 ,H272082 ,H272134 ,H271571 ,H271632 ,H271691 ,H271751 ,H271807 ,H271866 ,H271921 ,H271974 ,H272027 ,H272083 ,H272135 ,H271572 ,H271634 ,H271692 ,H271752 ,H271808 ,H271867 ,H271922 ,H271975 ,H272028 ,H272084 ,H272136 ,H271573 ,H271635 ,H271693 ,H271753 ,H271809 ,H271868 ,H271923 ,H271976 ,H272029 ,H272085 ,H272137 ,H271574 ,H271636 ,H271695 ,H271754 ,H271810 ,H271869 ,H271924 ,H271977 ,H272030 ,H272086 ,H272138 ,H271575 ,H271637 ,H271696 ,H271755 ,H271811 ,H271870 ,H271925 ,H271978 ,H272031 ,H272087 ,H272140 ,H271576 ,H271638 ,H271697 ,H271756 ,H271812 ,H271871 ,H271926 ,H271980 ,H272032 ,H272088 ,H272141 ,H271577 ,H271639 ,H271698 ,H271757 ,H271813 ,H271872 ,H271927 ,H271981 ,H272033 ,H272089 ,H272142 ,H271578 ,H271640 ,H271700 ,H271758 ,H271814 ,H271873 ,H271928 ,H271982 ,H272034 ,H272090 ,H272143 ,H271580 ,H271641 ,H271701 ,H271759 ,H271815 ,H271874 ,H271929 ,H271983 ,H272035 ,H272091 ,H272144 ,H271581 ,H271642 ,H271702 ,H271760 ,H271816 ,H271875 ,H271930 ,H271984 ,H272036 ,H272092 ,H272145 ,H271582 ,H271643 ,H271703 ,H271761 ,H271817 ,H271876 ,H271931 ,H271985 ,H272039 ,H272093 ,H272146 ,H271583 ,H271644 ,H271704 ,H271762 ,H271819 ,H271877 ,H271932 ,H271986 ,H272040 ,H272094 ,H272147 ,H271587 ,H271645 ,H271706 ,H271763 ,H271820 ,H271878 ,H271933 ,H271987 ,H272041 ,H272095 ,H272148 ,H271588 ,H271646 ,H271707 ,H271764 ,H271821 ,H271879 ,H271934 ,H271988 ,H272043 ,H272096 ,H272149 ,H271589 ,H271647 ,H271708 ,H271765 ,H271822 ,H271881 ,H271935 ,H271989 ,H272045 ,H272097 ,H272150 ,H271590 ,H271648 ,H271709 ,H271766 ,H271823 ,H271883 ,H271936 ,H271990 ,H272046 ,H272098 ,H272151 ,H271591 ,H271649 ,H271710 ,H271767 ,H271825 ,H271884 ,H271937 ,H271991 ,H272047 ,H272099 ,H272152 ,H271593 ,H271650 ,H271711 ,H271768 ,H271826 ,H271885 ,H271938 ,H271992 ,H272048 ,H272100 ,H272153 ,H271594 ,H271651 ,H271712 ,H271770 ,H271827 ,H271886 ,H271939 ,H271993 ,H272049 ,H272101 ,H272154 ,H271595 ,H271653 ,H271713 ,H271771 ,H271828 ,H271887 ,H271940 ,H271994 ,H272050 ,H272102 ,H272155 ,H271596 ,H271654 ,H271714 ,H271772 ,H271829 ,H271888 ,H271941 ,H271995 ,H272051 ,H272103 ,H272156 ,H271597 ,H271655 ,H271715 ,H271773 ,H271830 ,H271889 ,H271942 ,H271996 ,H272052 ,H272104 ,H272157 ,H272158 ,H272209 ,H272264 ,H272316 ,H272369 ,H272421 ,H272472 ,H272829 ,H272886 ,H272937 ,H272990 ,H272159 ,H272210 ,H272266 ,H272317 ,H272370 ,H272422 ,H272473 ,H272830 ,H272887 ,H272938 ,H272991 ,H272160 ,H272211 ,H272267 ,H272318 ,H272371 ,H272423 ,H272475 ,H272831 ,H272888 ,H272939 ,H272992 ,H272161 ,H272212 ,H272268 ,H272319 ,H272372 ,H272424 ,H272476 ,H272833 ,H272889 ,H272940 ,H272993 ,H272162 ,H272213 ,H272269 ,H272320 ,H272373 ,H272425 ,H272477 ,H272834 ,H272890 ,H272941 ,H272994 ,H272163 ,H272214 ,H272270 ,H272321 ,H272374 ,H272426 ,H272478 ,H272836 ,H272891 ,H272942 ,H272995 ,H272164 ,H272216 ,H272271 ,H272322 ,H272375 ,H272427 ,H272479 ,H272837 ,H272892 ,H272943 ,H272996 ,H272165 ,H272217 ,H272272 ,H272323 ,H272376 ,H272428 ,H272480 ,H272838 ,H272893 ,H272944 ,H272997 ,H272166 ,H272218 ,H272273 ,H272324 ,H272377 ,H272429 ,H272482 ,H272839 ,H272894 ,H272946 ,H272998 ,H272167 ,H272219 ,H272274 ,H272325 ,H272378 ,H272430 ,H272483 ,H272840 ,H272895 ,H272947 ,H272999 ,H272168 ,H272220 ,H272275 ,H272326 ,H272379 ,H272431 ,H272484 ,H272841 ,H272896 ,H272948 ,H273000 ,H272169 ,H272222 ,H272276 ,H272327 ,H272380 ,H272432 ,H272485 ,H272842 ,H272897 ,H272949 ,H273001 ,H272170 ,H272223 ,H272277 ,H272328 ,H272381 ,H272433 ,H272488 ,H272843 ,H272898 ,H272950 ,H273002 ,H272171 ,H272224 ,H272278 ,H272329 ,H272382 ,H272434 ,H272489 ,H272844 ,H272899 ,H272951 ,H273003 ,H272172 ,H272225 ,H272279 ,H272330 ,H272383 ,H272435 ,H272490 ,H272845 ,H272900 ,H272952 ,H273004 ,H272173 ,H272226 ,H272280 ,H272331 ,H272384 ,H272436 ,H272491 ,H272846 ,H272901 ,H272953 ,H273005 ,H272174 ,H272227 ,H272281 ,H272332 ,H272385 ,H272437 ,H272492 ,H272847 ,H272902 ,H272954 ,H273006 ,H272175 ,H272228 ,H272282 ,H272333 ,H272386 ,H272438 ,H272493 ,H272848 ,H272903 ,H272955 ,H273007 ,H272176 ,H272229 ,H272283 ,H272334 ,H272387 ,H272439 ,H272494 ,H272849 ,H272904 ,H272956 ,H273008 ,H272177 ,H272230 ,H272284 ,H272335 ,H272388 ,H272440 ,H272495 ,H272850 ,H272905 ,H272957 ,H273010 ,H272178 ,H272231 ,H272285 ,H272336 ,H272389 ,H272441 ,H272496 ,H272851 ,H272906 ,H272958 ,H273011 ,H272179 ,H272232 ,H272286 ,H272337 ,H272391 ,H272442 ,H272497 ,H272852 ,H272907 ,H272959 ,H273012 ,H272180 ,H272233 ,H272287 ,H272338 ,H272392 ,H272443 ,H272500 ,H272853 ,H272908 ,H272960 ,H273013 ,H272181 ,H272234 ,H272288 ,H272339 ,H272393 ,H272444 ,H272800 ,H272854 ,H272909 ,H272961 ,H273014 ,H272182 ,H272235 ,H272289 ,H272340 ,H272394 ,H272445 ,H272801 ,H272855 ,H272910 ,H272962 ,H273016 ,H272183 ,H272236 ,H272290 ,H272341 ,H272395 ,H272446 ,H272802 ,H272856 ,H272911 ,H272963 ,H273017 ,H272184 ,H272237 ,H272291 ,H272342 ,H272396 ,H272447 ,H272803 ,H272857 ,H272912 ,H272964 ,H273018 ,H272185 ,H272238 ,H272292 ,H272343 ,H272397 ,H272448 ,H272805 ,H272858 ,H272913 ,H272965 ,H273019 ,H272186 ,H272239 ,H272293 ,H272344 ,H272398 ,H272449 ,H272806 ,H272859 ,H272914 ,H272967 ,H273020 ,H272187 ,H272240 ,H272294 ,H272345 ,H272399 ,H272450 ,H272807 ,H272862 ,H272915 ,H272968 ,H273021 ,H272188 ,H272241 ,H272295 ,H272346 ,H272400 ,H272451 ,H272808 ,H272863 ,H272916 ,H272969 ,H273022 ,H272189 ,H272242 ,H272296 ,H272347 ,H272401 ,H272452 ,H272809 ,H272864 ,H272917 ,H272970 ,H273023 ,H272190 ,H272243 ,H272297 ,H272348 ,H272402 ,H272453 ,H272810 ,H272865 ,H272918 ,H272971 ,H273024 ,H272191 ,H272244 ,H272298 ,H272349 ,H272403 ,H272454 ,H272811 ,H272866 ,H272919 ,H272972 ,H273025 ,H272192 ,H272245 ,H272299 ,H272350 ,H272404 ,H272455 ,H272812 ,H272867 ,H272920 ,H272973 ,H273026 ,H272193 ,H272246 ,H272300 ,H272352 ,H272405 ,H272456 ,H272813 ,H272869 ,H272921 ,H272974 ,H273027 ,H272194 ,H272247 ,H272301 ,H272353 ,H272406 ,H272457 ,H272814 ,H272870 ,H272922 ,H272975 ,H273028 ,H272195 ,H272248 ,H272302 ,H272354 ,H272407 ,H272458 ,H272815 ,H272871 ,H272923 ,H272976 ,H273029 ,H272196 ,H272249 ,H272303 ,H272356 ,H272408 ,H272459 ,H272816 ,H272872 ,H272924 ,H272977 ,H273030 ,H272197 ,H272250 ,H272304 ,H272357 ,H272409 ,H272460 ,H272817 ,H272873 ,H272925 ,H272978 ,H273031 ,H272198 ,H272251 ,H272305 ,H272358 ,H272410 ,H272461 ,H272818 ,H272875 ,H272926 ,H272979 ,H273032 ,H272199 ,H272253 ,H272306 ,H272359 ,H272411 ,H272462 ,H272819 ,H272876 ,H272927 ,H272980 ,H273033 ,H272200 ,H272254 ,H272307 ,H272360 ,H272412 ,H272463 ,H272820 ,H272877 ,H272928 ,H272981 ,H273034 ,H272201 ,H272255 ,H272308 ,H272361 ,H272413 ,H272464 ,H272821 ,H272878 ,H272929 ,H272982 ,H273035 ,H272202 ,H272256 ,H272309 ,H272362 ,H272414 ,H272465 ,H272822 ,H272879 ,H272930 ,H272983 ,H273036 ,H272203 ,H272258 ,H272310 ,H272363 ,H272415 ,H272466 ,H272823 ,H272880 ,H272931 ,H272984 ,H273037 ,H272204 ,H272259 ,H272311 ,H272364 ,H272416 ,H272467 ,H272824 ,H272881 ,H272932 ,H272985 ,H273038 ,H272205 ,H272260 ,H272312 ,H272365 ,H272417 ,H272468 ,H272825 ,H272882 ,H272933 ,H272986 ,H273039 ,H272206 ,H272261 ,H272313 ,H272366 ,H272418 ,H272469 ,H272826 ,H272883 ,H272934 ,H272987 ,H273040 ,H272207 ,H272262 ,H272314 ,H272367 ,H272419 ,H272470 ,H272827 ,H272884 ,H272935 ,H272988 ,H273041 ,H272208 ,H272263 ,H272315 ,H272368 ,H272420 ,H272471 ,H272828 ,H272885 ,H272936 ,H272989 ,H273042 ,H273043 ,H273098 ,H273151 ,H273208 ,H290272 ,H290342 ,H290400 ,H290537 ,H291030 ,H291087 ,H291142 ,H273044 ,H273099 ,H273152 ,H273209 ,H290273 ,H290343 ,H290401 ,H290538 ,H291031 ,H291088 ,H291143 ,H273046 ,H273100 ,H273153 ,H273210 ,H290274 ,H290346 ,H290402 ,H290539 ,H291032 ,H291089 ,H291145 ,H273047 ,H273101 ,H273155 ,H273547 ,H290275 ,H290347 ,H290403 ,H290540 ,H291034 ,H291090 ,H291146 ,H273048 ,H273102 ,H273156 ,H273711 ,H290276 ,H290348 ,H290404 ,H290541 ,H291035 ,H291091 ,H291147 ,H273049 ,H273103 ,H273157 ,H290003 ,H290278 ,H290349 ,H290405 ,H290542 ,H291036 ,H291092 ,H291148 ,H273050 ,H273104 ,H273158 ,H290012 ,H290279 ,H290350 ,H290406 ,H290544 ,H291037 ,H291093 ,H291149 ,H273051 ,H273105 ,H273159 ,H290021 ,H290281 ,H290351 ,H290408 ,H290545 ,H291038 ,H291094 ,H291150 ,H273052 ,H273106 ,H273160 ,H290030 ,H290282 ,H290352 ,H290409 ,H290546 ,H291039 ,H291096 ,H291151 ,H273053 ,H273107 ,H273162 ,H290031 ,H290284 ,H290353 ,H290410 ,H290547 ,H291040 ,H291097 ,H291152 ,H273054 ,H273108 ,H273163 ,H290033 ,H290285 ,H290354 ,H290411 ,H290548 ,H291041 ,H291098 ,H291153 ,H273055 ,H273109 ,H273164 ,H290104 ,H290286 ,H290355 ,H290413 ,H290549 ,H291042 ,H291099 ,H291154 ,H273056 ,H273110 ,H273165 ,H290109 ,H290287 ,H290356 ,H290414 ,H290550 ,H291043 ,H291101 ,H291155 ,H273057 ,H273111 ,H273166 ,H290110 ,H290289 ,H290357 ,H290415 ,H290551 ,H291044 ,H291102 ,H291156 ,H273058 ,H273112 ,H273167 ,H290122 ,H290290 ,H290358 ,H290422 ,H290552 ,H291045 ,H291103 ,H291157 ,H273059 ,H273113 ,H273168 ,H290123 ,H290291 ,H290360 ,H290423 ,H290553 ,H291046 ,H291104 ,H291158 ,H273060 ,H273115 ,H273169 ,H290151 ,H290292 ,H290362 ,H290425 ,H290554 ,H291047 ,H291105 ,H291159 ,H273061 ,H273116 ,H273170 ,H290154 ,H290293 ,H290363 ,H290426 ,H290555 ,H291048 ,H291106 ,H291160 ,H273062 ,H273117 ,H273171 ,H290155 ,H290295 ,H290364 ,H290500 ,H290556 ,H291049 ,H291107 ,H291161 ,H273063 ,H273118 ,H273172 ,H290165 ,H290296 ,H290365 ,H290501 ,H290558 ,H291050 ,H291108 ,H291162 ,H273064 ,H273119 ,H273173 ,H290166 ,H290297 ,H290366 ,H290502 ,H290559 ,H291051 ,H291109 ,H291163 ,H273065 ,H273120 ,H273174 ,H290192 ,H290301 ,H290367 ,H290503 ,H290560 ,H291052 ,H291111 ,H291164 ,H273066 ,H273121 ,H273175 ,H290193 ,H290302 ,H290368 ,H290504 ,H290561 ,H291053 ,H291112 ,H291165 ,H273067 ,H273122 ,H273177 ,H290198 ,H290304 ,H290369 ,H290505 ,H290600 ,H291054 ,H291113 ,H291166 ,H273068 ,H273123 ,H273178 ,H290206 ,H290305 ,H290370 ,H290506 ,H291001 ,H291055 ,H291114 ,H291167 ,H273069 ,H273124 ,H273179 ,H290208 ,H290306 ,H290371 ,H290508 ,H291003 ,H291056 ,H291115 ,H291168 ,H273070 ,H273125 ,H273180 ,H290210 ,H290307 ,H290372 ,H290510 ,H291004 ,H291057 ,H291116 ,H291169 ,H273071 ,H273126 ,H273181 ,H290211 ,H290308 ,H290373 ,H290511 ,H291005 ,H291059 ,H291117 ,H291170 ,H273072 ,H273127 ,H273182 ,H290216 ,H290312 ,H290374 ,H290512 ,H291006 ,H291060 ,H291118 ,H291171 ,H273074 ,H273128 ,H273183 ,H290223 ,H290313 ,H290375 ,H290513 ,H291007 ,H291061 ,H291119 ,H291172 ,H273075 ,H273130 ,H273184 ,H290224 ,H290314 ,H290376 ,H290514 ,H291008 ,H291062 ,H291120 ,H291173 ,H273077 ,H273131 ,H273186 ,H290225 ,H290315 ,H290377 ,H290515 ,H291009 ,H291063 ,H291121 ,H291174 ,H273078 ,H273132 ,H273187 ,H290228 ,H290317 ,H290378 ,H290516 ,H291010 ,H291064 ,H291122 ,H291175 ,H273079 ,H273133 ,H273188 ,H290233 ,H290319 ,H290379 ,H290517 ,H291011 ,H291065 ,H291123 ,H291176 ,H273080 ,H273134 ,H273189 ,H290235 ,H290320 ,H290380 ,H290518 ,H291013 ,H291067 ,H291124 ,H291177 ,H273081 ,H273135 ,H273190 ,H290236 ,H290321 ,H290381 ,H290519 ,H291014 ,H291068 ,H291125 ,H291178 ,H273082 ,H273136 ,H273191 ,H290240 ,H290322 ,H290382 ,H290520 ,H291015 ,H291069 ,H291126 ,H291179 ,H273083 ,H273137 ,H273192 ,H290241 ,H290323 ,H290383 ,H290521 ,H291016 ,H291070 ,H291127 ,H291180 ,H273084 ,H273138 ,H273193 ,H290243 ,H290326 ,H290384 ,H290522 ,H291017 ,H291071 ,H291128 ,H291181 ,H273085 ,H273139 ,H273195 ,H290246 ,H290327 ,H290386 ,H290523 ,H291018 ,H291072 ,H291129 ,H291183 ,H273086 ,H273140 ,H273196 ,H290249 ,H290328 ,H290387 ,H290524 ,H291019 ,H291073 ,H291130 ,H291184 ,H273087 ,H273141 ,H273197 ,H290255 ,H290329 ,H290388 ,H290525 ,H291020 ,H291074 ,H291131 ,H291185 ,H273088 ,H273142 ,H273198 ,H290256 ,H290331 ,H290389 ,H290526 ,H291021 ,H291075 ,H291132 ,H291186 ,H273089 ,H273143 ,H273199 ,H290259 ,H290332 ,H290390 ,H290527 ,H291022 ,H291079 ,H291133 ,H291187 ,H273090 ,H273144 ,H273200 ,H290260 ,H290334 ,H290392 ,H290529 ,H291023 ,H291080 ,H291134 ,H291188 ,H273091 ,H273145 ,H273201 ,H290261 ,H290335 ,H290393 ,H290530 ,H291024 ,H291081 ,H291135 ,H291189 ,H273092 ,H273146 ,H273202 ,H290262 ,H290336 ,H290394 ,H290531 ,H291025 ,H291082 ,H291136 ,H291190 ,H273093 ,H273147 ,H273203 ,H290267 ,H290337 ,H290395 ,H290532 ,H291026 ,H291083 ,H291137 ,H291191 ,H273094 ,H273148 ,H273204 ,H290268 ,H290338 ,H290396 ,H290533 ,H291027 ,H291084 ,H291138 ,H291192 ,H273095 ,H273149 ,H273205 ,H290269 ,H290340 ,H290398 ,H290535 ,H291028 ,H291085 ,H291140 ,H291193 ,H273097 ,H273150 ,H273206 ,H290271 ,H290341 ,H290399 ,H290536 ,H291029 ,H291086 ,H291141 ,H291194 ,H291195 ,H291248 ,H293324 ,H293377 ,H293432 ,H293484 ,H293540 ,H293594 ,H293649 ,H293703 ,H293760 ,H291196 ,H291249 ,H293325 ,H293378 ,H293433 ,H293485 ,H293541 ,H293595 ,H293650 ,H293705 ,H293762 ,H291197 ,H291250 ,H293326 ,H293379 ,H293434 ,H293486 ,H293542 ,H293596 ,H293651 ,H293706 ,H293763 ,H291198 ,H291251 ,H293327 ,H293380 ,H293435 ,H293487 ,H293543 ,H293597 ,H293652 ,H293707 ,H293764 ,H291199 ,H291252 ,H293328 ,H293381 ,H293436 ,H293488 ,H293544 ,H293598 ,H293653 ,H293708 ,H293765 ,H291200 ,H291253 ,H293329 ,H293382 ,H293437 ,H293489 ,H293545 ,H293599 ,H293654 ,H293710 ,H293766 ,H291201 ,H291254 ,H293330 ,H293383 ,H293438 ,H293490 ,H293546 ,H293600 ,H293655 ,H293712 ,H293767 ,H291202 ,H291255 ,H293331 ,H293384 ,H293439 ,H293491 ,H293548 ,H293601 ,H293656 ,H293713 ,H293769 ,H291203 ,H291256 ,H293332 ,H293385 ,H293440 ,H293492 ,H293549 ,H293602 ,H293657 ,H293714 ,H293770 ,H291204 ,H291257 ,H293333 ,H293386 ,H293441 ,H293493 ,H293550 ,H293603 ,H293658 ,H293715 ,H293771 ,H291205 ,H291258 ,H293334 ,H293387 ,H293442 ,H293494 ,H293551 ,H293604 ,H293659 ,H293716 ,H293772 ,H291206 ,H291259 ,H293335 ,H293388 ,H293443 ,H293495 ,H293552 ,H293605 ,H293660 ,H293717 ,H293773 ,H291207 ,H291260 ,H293336 ,H293389 ,H293444 ,H293496 ,H293553 ,H293606 ,H293661 ,H293718 ,H293774 ,H291208 ,H291261 ,H293338 ,H293390 ,H293445 ,H293497 ,H293554 ,H293609 ,H293662 ,H293719 ,H293775 ,H291209 ,H291262 ,H293339 ,H293391 ,H293446 ,H293498 ,H293555 ,H293610 ,H293663 ,H293720 ,H293776 ,H291211 ,H291263 ,H293340 ,H293393 ,H293447 ,H293499 ,H293556 ,H293611 ,H293664 ,H293721 ,H293777 ,H291212 ,H291264 ,H293341 ,H293394 ,H293448 ,H293500 ,H293557 ,H293612 ,H293665 ,H293723 ,H293778 ,H291213 ,H291265 ,H293342 ,H293395 ,H293449 ,H293501 ,H293558 ,H293613 ,H293666 ,H293724 ,H293779 ,H291214 ,H291267 ,H293343 ,H293396 ,H293450 ,H293502 ,H293559 ,H293614 ,H293667 ,H293725 ,H293780 ,H291215 ,H291268 ,H293344 ,H293397 ,H293451 ,H293503 ,H293560 ,H293616 ,H293668 ,H293726 ,H293781 ,H291216 ,H291269 ,H293345 ,H293399 ,H293452 ,H293504 ,H293561 ,H293617 ,H293669 ,H293727 ,H293783 ,H291217 ,H291270 ,H293346 ,H293401 ,H293453 ,H293505 ,H293562 ,H293618 ,H293670 ,H293728 ,H293784 ,H291218 ,H291271 ,H293347 ,H293402 ,H293454 ,H293506 ,H293563 ,H293619 ,H293671 ,H293729 ,H293785 ,H291219 ,H291272 ,H293348 ,H293403 ,H293455 ,H293508 ,H293564 ,H293620 ,H293672 ,H293730 ,H293786 ,H291220 ,H291273 ,H293349 ,H293404 ,H293456 ,H293509 ,H293565 ,H293621 ,H293673 ,H293731 ,H293787 ,H291221 ,H292832 ,H293350 ,H293405 ,H293457 ,H293510 ,H293566 ,H293622 ,H293674 ,H293732 ,H293788 ,H291222 ,H293207 ,H293351 ,H293406 ,H293458 ,H293512 ,H293567 ,H293623 ,H293675 ,H293733 ,H293789 ,H291223 ,H293300 ,H293352 ,H293407 ,H293459 ,H293513 ,H293568 ,H293624 ,H293676 ,H293734 ,H293790 ,H291224 ,H293301 ,H293353 ,H293408 ,H293460 ,H293514 ,H293569 ,H293625 ,H293677 ,H293735 ,H293791 ,H291225 ,H293302 ,H293354 ,H293409 ,H293461 ,H293515 ,H293570 ,H293626 ,H293678 ,H293736 ,H293792 ,H291226 ,H293303 ,H293355 ,H293410 ,H293462 ,H293516 ,H293571 ,H293627 ,H293679 ,H293737 ,H293793 ,H291227 ,H293304 ,H293356 ,H293411 ,H293463 ,H293517 ,H293572 ,H293628 ,H293681 ,H293738 ,H293794 ,H291228 ,H293305 ,H293357 ,H293412 ,H293464 ,H293518 ,H293573 ,H293629 ,H293682 ,H293739 ,H293795 ,H291229 ,H293306 ,H293359 ,H293413 ,H293465 ,H293519 ,H293574 ,H293630 ,H293683 ,H293740 ,H293796 ,H291230 ,H293307 ,H293360 ,H293414 ,H293466 ,H293521 ,H293575 ,H293631 ,H293684 ,H293741 ,H293797 ,H291231 ,H293308 ,H293361 ,H293415 ,H293468 ,H293522 ,H293576 ,H293632 ,H293685 ,H293742 ,H293798 ,H291232 ,H293309 ,H293362 ,H293416 ,H293469 ,H293523 ,H293577 ,H293633 ,H293686 ,H293743 ,H293799 ,H291233 ,H293310 ,H293363 ,H293417 ,H293470 ,H293524 ,H293578 ,H293634 ,H293687 ,H293744 ,H293800 ,H291234 ,H293311 ,H293364 ,H293418 ,H293471 ,H293525 ,H293579 ,H293635 ,H293688 ,H293745 ,H293801 ,H291235 ,H293312 ,H293365 ,H293419 ,H293472 ,H293527 ,H293581 ,H293637 ,H293689 ,H293746 ,H293802 ,H291236 ,H293313 ,H293366 ,H293420 ,H293473 ,H293529 ,H293583 ,H293638 ,H293692 ,H293747 ,H293803 ,H291237 ,H293314 ,H293367 ,H293421 ,H293474 ,H293530 ,H293584 ,H293639 ,H293693 ,H293748 ,H293804 ,H291238 ,H293315 ,H293368 ,H293422 ,H293475 ,H293531 ,H293585 ,H293640 ,H293694 ,H293749 ,H293805 ,H291239 ,H293316 ,H293369 ,H293423 ,H293476 ,H293532 ,H293586 ,H293641 ,H293695 ,H293750 ,H293806 ,H291240 ,H293317 ,H293370 ,H293424 ,H293477 ,H293533 ,H293587 ,H293642 ,H293696 ,H293751 ,H293807 ,H291241 ,H293318 ,H293371 ,H293425 ,H293478 ,H293534 ,H293588 ,H293643 ,H293697 ,H293752 ,H293808 ,H291242 ,H293319 ,H293372 ,H293426 ,H293479 ,H293535 ,H293589 ,H293644 ,H293698 ,H293753 ,H293809 ,H291243 ,H293320 ,H293373 ,H293428 ,H293480 ,H293536 ,H293590 ,H293645 ,H293699 ,H293754 ,H293812 ,H291244 ,H293321 ,H293374 ,H293429 ,H293481 ,H293537 ,H293591 ,H293646 ,H293700 ,H293755 ,H293813 ,H291245 ,H293322 ,H293375 ,H293430 ,H293482 ,H293538 ,H293592 ,H293647 ,H293701 ,H293756 ,H293814 ,H291247 ,H293323 ,H293376 ,H293431 ,H293483 ,H293539 ,H293593 ,H293648 ,H293702 ,H293758 ,H293815 ,H293816 ,H293869 ,H293920 ,H293976 ,H294031 ,H294085 ,H294147 ,H294203 ,H294257 ,H294484 ,H293817 ,H293870 ,H293921 ,H293977 ,H294034 ,H294086 ,H294148 ,H294204 ,H294258 ,H294500 ,H293818 ,H293871 ,H293922 ,H293978 ,H294035 ,H294087 ,H294149 ,H294205 ,H294260 ,H294519 ,H293819 ,H293872 ,H293924 ,H293979 ,H294036 ,H294088 ,H294150 ,H294206 ,H294261 ,H294542 ,H293820 ,H293873 ,H293925 ,H293980 ,H294037 ,H294089 ,H294151 ,H294207 ,H294262 ,H294550 ,H293821 ,H293874 ,H293926 ,H293981 ,H294038 ,H294090 ,H294152 ,H294208 ,H294263 ,H294565 ,H293822 ,H293875 ,H293928 ,H293982 ,H294039 ,H294091 ,H294153 ,H294209 ,H294264 ,H294572 ,H293823 ,H293876 ,H293929 ,H293983 ,H294040 ,H294092 ,H294154 ,H294210 ,H294265 ,H294573 ,H293824 ,H293877 ,H293930 ,H293984 ,H294041 ,H294093 ,H294155 ,H294211 ,H294266 ,H294592 ,H293825 ,H293878 ,H293931 ,H293985 ,H294042 ,H294094 ,H294156 ,H294212 ,H294267 ,H294595 ,H293826 ,H293879 ,H293932 ,H293986 ,H294043 ,H294095 ,H294157 ,H294213 ,H294268 ,H294603 ,H293827 ,H293880 ,H293933 ,H293987 ,H294044 ,H294097 ,H294158 ,H294214 ,H294269 ,H294618 ,H293828 ,H293881 ,H293934 ,H293988 ,H294045 ,H294098 ,H294159 ,H294215 ,H294270 ,H294637 ,H293829 ,H293882 ,H293935 ,H293989 ,H294046 ,H294099 ,H294160 ,H294216 ,H294271 ,H294653 ,H293830 ,H293883 ,H293936 ,H293990 ,H294047 ,H294100 ,H294161 ,H294217 ,H294272 ,H294656 ,H293831 ,H293884 ,H293937 ,H293991 ,H294048 ,H294101 ,H294162 ,H294218 ,H294273 ,H294658 ,H293832 ,H293885 ,H293938 ,H293992 ,H294049 ,H294102 ,H294163 ,H294219 ,H294274 ,H294660 ,H293834 ,H293886 ,H293940 ,H293993 ,H294050 ,H294103 ,H294164 ,H294220 ,H294275 ,H294674 ,H293835 ,H293887 ,H293941 ,H293994 ,H294051 ,H294107 ,H294165 ,H294221 ,H294276 ,H294679 ,H293836 ,H293888 ,H293942 ,H293995 ,H294052 ,H294108 ,H294166 ,H294222 ,H294277 ,H294681 ,H293837 ,H293889 ,H293943 ,H293996 ,H294053 ,H294109 ,H294167 ,H294224 ,H294278 ,H294698 ,H293838 ,H293890 ,H293944 ,H293997 ,H294054 ,H294110 ,H294168 ,H294225 ,H294279 ,H294713 ,H293839 ,H293891 ,H293946 ,H293998 ,H294055 ,H294111 ,H294169 ,H294226 ,H294280 ,H294714 ,H293840 ,H293892 ,H293947 ,H293999 ,H294056 ,H294112 ,H294170 ,H294227 ,H294281 ,H294715 ,H293841 ,H293893 ,H293948 ,H294000 ,H294057 ,H294113 ,H294171 ,H294228 ,H294283 ,H294719 ,H293842 ,H293894 ,H293949 ,H294001 ,H294058 ,H294114 ,H294172 ,H294229 ,H294284 ,H294723 ,H293843 ,H293895 ,H293950 ,H294002 ,H294059 ,H294116 ,H294173 ,H294230 ,H294285 ,H294726 ,H293844 ,H293896 ,H293951 ,H294003 ,H294060 ,H294117 ,H294174 ,H294231 ,H294286 ,H294735 ,H293845 ,H293897 ,H293952 ,H294004 ,H294061 ,H294118 ,H294175 ,H294232 ,H294287 ,H294737 ,H293846 ,H293898 ,H293953 ,H294005 ,H294062 ,H294119 ,H294176 ,H294233 ,H294288 ,H294752 ,H293847 ,H293899 ,H293954 ,H294006 ,H294063 ,H294120 ,H294177 ,H294234 ,H294301 ,H294820 ,H293848 ,H293900 ,H293955 ,H294007 ,H294064 ,H294122 ,H294179 ,H294235 ,H294302 ,HCX4138 ,H293849 ,H293901 ,H293956 ,H294008 ,H294065 ,H294123 ,H294180 ,H294236 ,H294308 ,H293850 ,H293902 ,H293957 ,H294009 ,H294066 ,H294124 ,H294181 ,H294237 ,H294311 ,H293851 ,H293903 ,H293958 ,H294010 ,H294067 ,H294125 ,H294182 ,H294238 ,H294341 ,H293852 ,H293904 ,H293959 ,H294011 ,H294068 ,H294126 ,H294183 ,H294239 ,H294351 ,H293854 ,H293905 ,H293960 ,H294012 ,H294069 ,H294127 ,H294184 ,H294240 ,H294355 ,H293855 ,H293906 ,H293961 ,H294013 ,H294070 ,H294128 ,H294185 ,H294241 ,H294358 ,H293856 ,H293907 ,H293962 ,H294015 ,H294071 ,H294129 ,H294186 ,H294242 ,H294362 ,H293857 ,H293908 ,H293963 ,H294016 ,H294072 ,H294130 ,H294187 ,H294243 ,H294380 ,H293858 ,H293909 ,H293964 ,H294017 ,H294073 ,H294131 ,H294188 ,H294244 ,H294395 ,H293859 ,H293910 ,H293965 ,H294021 ,H294074 ,H294132 ,H294190 ,H294245 ,H294408 ,H293860 ,H293911 ,H293966 ,H294022 ,H294075 ,H294133 ,H294191 ,H294246 ,H294413 ,H293861 ,H293912 ,H293967 ,H294023 ,H294076 ,H294134 ,H294194 ,H294247 ,H294421 ,H293862 ,H293913 ,H293968 ,H294024 ,H294077 ,H294135 ,H294195 ,H294248 ,H294427 ,H293863 ,H293914 ,H293969 ,H294025 ,H294078 ,H294136 ,H294197 ,H294250 ,H294447 ,H293864 ,H293915 ,H293970 ,H294026 ,H294079 ,H294140 ,H294198 ,H294252 ,H294448 ,H293865 ,H293916 ,H293971 ,H294027 ,H294080 ,H294141 ,H294199 ,H294253 ,H294450 ,H293866 ,H293917 ,H293972 ,H294028 ,H294081 ,H294142 ,H294200 ,H294254 ,H294458 ,H293867 ,H293918 ,H293974 ,H294029 ,H294083 ,H294143 ,H294201 ,H294255 ,H294465 ,H293868 ,H293919 ,H293975 ,H294030 ,H294084 ,H294146 ,H294202 ,H294256 ,H294482 |
| FEI Number |
2916710
|
Recalling Firm/ Manufacturer |
Varian Medical Systems, Inc.
911 Hansen Way
Palo Alto
CA
94304-1028
|
| For Additional Information Contact | Michael A. Pignataro 650-424-5731 |
Manufacturer Reason for Recall | An improperly installed collimator cover may detach and fall from the machine, potentially striking the patient or nearby personnel. |
FDA Determined Cause 2 | Component design/selection |
| Action | On 3/17/2011, Varian Medical Systems sent an Urgent Medical Device Correction Letter to all affected users, with a description of the problem and user corrective action steps. |
| Quantity in Commerce | 3791 units |
| Distribution | Worldwide distribution including USA, Australia, Argentina, Italy, Morocco, South Korea, Belgium, Canada, Netherlands, Denmark, Japan, Hong Kong, Malaysia, Hungary, New Zealand, Sweden, Taiwan, Switzerland, Romania, Brazil, Slovakia, India, Colombia, UK, Ireland, Spain, Singapore, Martinique, Reunion, Venezuela, Iceland, Portugal, Turkey, Israel, Bahrain, Macau, Chile, Saudi Arabia, Latvia, China, Finland, Kazakhstan, Thailand, France, Poland, and Germany. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = IYE
|
|
|
|