• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Optical Guidance Platform, Version 2.6 and 2.6.1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Optical Guidance Platform, Version 2.6 and 2.6.1see related information
Date Initiated by FirmMarch 30, 2011
Date PostedMay 16, 2011
Recall Status1 Terminated 3 on June 19, 2012
Recall NumberZ-2200-2011
Recall Event ID 58505
510(K)NumberK071360 
Product Classification Accelerator, linear, medical - Product Code IYE
ProductOptical Guidance Platform, Version 2.6 and 2.6.1, Model Number: HZl, Manufactured and Distributed by: Varian Medical Systems Inc., Palo Alto, CA. For use with a charged particle accelerator to perform precise positioning of treatment target for stereotactic radiosurgery or radiotherapy treatments on cranial or extracranial lesions.
Code Information HZlOOO3 ,HZlO153 ,HZI0313 ,HZI0513 ,HZI0679 ,HZl1020 ,HZ12046 ,HZ12097 ,HZ13059 ,HZ17058 ,HZ18118  ,HZlOOO8 ,HZI0154 ,HZ10316 ,HZI0528 ,HZlO681 ,HZl1021 ,HZ12047 ,HZ12098 ,HZ13060 ,HZ17059 ,HZ18120  ,HZIOO13 ,HZI0155 ,HZ10326 ,HZI0534 ,HZI0684 ,HZl1022 ,HZ12048 ,HZ12099 ,HZ13062 ,HZ17061 ,HZ18126  ,HZIOO14 ,HZI0157 ,HZ10330 ,HZI0537 ,HZ10685 ,HZl1023 ,HZ12049 ,HZl2100 ,HZ13063 ,HZ17062 ,HZ18131  ,HZlOO15 ,HZI0159 ,HZI0333 ,HZ10540 ,HZ10689 ,HZl1024 ,HZ12050 ,HZl2101 ,HZ13064 ,HZ17065 ,HZ18132  ,HZIOO17 ,HZI0165 ,HZI0335 ,HZI0541 ,HZ10693 ,HZl1025 ,HZ12051 ,HZl2102 ,HZ15006 ,HZ17067 ,HZ18134   ,HZIOO19 ,HZI0170 ,HZI0337 ,HZ10542 ,HZI0698 ,HZ12001 ,HZ12052 ,HZ12103 ,HZ15070 ,HZ17068 ,HZ18142  ,HZIOO23 ,HZlOl72 ,HZI0343 ,HZI0544 ,HZ10700 ,HZ12002 ,HZ12054 ,HZ12104 ,HZ15118 ,HZ17069 ,HZ18143  ,HZlOO24 ,HZI0179 ,HZ10347 ,HZ10545 ,HZI0703 ,HZ12003 ,HZ12056 ,HZ12107 ,HZ15480 ,HZ17070 ,HZ18155  ,HZlOO25 ,HZI0181 ,HZI0352 ,HZ10547 ,HZ10706 ,HZ12004 ,HZ12057 ,HZ12108 ,HZ16006 ,HZ17071 ,HZ18175  ,HZIOO26 ,HZI0182 ,HZI0357 ,HZ10552 ,HZI0712 ,HZ12005 ,HZ12058 ,HZ12125 ,HZ16069 ,HZ17072 ,HZ18188  ,HZIOO27 ,HZI0183 ,HZI0368 ,HZ10553 ,HZlO715 ,HZ12006 ,HZ12059 ,HZ12126 ,HZ16070 ,HZ17074 ,HZ18193  ,HZIOO29 ,HZI0185 ,HZI0372 ,HZI0557 ,HZI0717 ,HZ12007 ,HZl2060 ,HZ12154 ,HZ17001 ,HZ17078 ,HZ19001  ,HZIOO44 ,HZI0187 ,HZ10376 ,HZI0560 ,HZI0722 ,HZ12008 ,HZ12061 ,HZ12188 ,HZ17002 ,HZ17079 ,HZ19003  ,HZIOO46 ,HZI0188 ,HZ10378 ,HZI0562 ,HZI0724 ,HZ12009 ,HZ12062 ,HZ12190 ,HZ17003 ,HZ17081 ,HZ19004  ,HZlOO48 ,HZI0190 ,HZ10379 ,HZ10568 ,HZlO728 ,HZ12010 ,HZ12063 ,HZ13006 ,HZ17006 ,HZ17106 ,HZ19cio5  ,HZIOO51 ,HZI0194 ,HZI0387 ,HZ10580 ,HZ10730 ,HZ12011 ,HZ12064 ,HZ13009 ,HZ17007 ,HZ17118 ,HZ19006  ,HZlOO52 ,HZI0196 ,HZI0392 ,HZI0584 ,HZ10731 ,HZ12012 ,HZ12065 ,HZ13013 ,HZ17010 ,HZl7126 ,HZ19007  ,HZlOO54 ,HZ10201 ,HZI0394 ,HZI0591 ,HZ10733 ,HZ12013 ,HZ12066 ,HZ13019 ,HZ17011 ,HZ17137 ,HZ19010  ,HZIOO55 ,HZ10203 ,HZ10398 ,HZI0592 ,HZI0737 ,HZ12015 ,HZ12067 ,HZ13020 ,HZ17012 ,HZl7142 ,HZ19012  ,HZIOO58 ,HZ10205 ,HZI0401 ,HZI0594 ,HZI0740 ,HZ12016 ,HZ12068 ,HZ13021 ,HZ17013 ,HZl7175 ,HZ19013  ,HZIOO61 ,HZI0209 ,HZI0403 ,HZI0598 ,HZI0743 ,HZ12018 ,HZ12069 ,HZ13022 ,HZ17016 ,HZ18001 ,HZ19014  ,HZlOO63 ,HZ10210 ,HZ10404 ,HZ10601 ,HZ10744 ,HZ12019 ,HZ12070 ,HZ13023 ,HZ17017 ,HZ18002 ,HZ19015  ,HZlOO65 ,HZ10220 ,HZ10408 ,HZI0607 ,HZI0759 ,HZ12020 ,HZ12071 ,HZ13024 ,HZ17018 ,HZ18004 ,HZ19017  ,HZIOO68 ,HZI0228 ,HZI0409 ,HZI0609 ,HZI0762 ,HZ12021 ,HZ12072 ,HZ13026 ,HZ17019 ,HZ18006 ,HZ19018  ,HZIOO70 ,HZI0237 ,HZI0411 ,HZI0611 ,HZI0764 ,HZ12022 ,HZ12073 ,HZ13029 ,HZ17020 ,HZ18013 ,HZ19019  ,HZIOO73 ,HZI0239 ,HZ10412 ,HZ10612 ,HZ10766 ,HZ12023 ,HZ12074 ,HZ13030 ,HZ17021 ,HZ18018 ,HZ19020  ,HZlOO76 ,HZI0240 ,HZI0425 ,HZI0613 ,HZI0771 ,HZ12024 ,HZ12075 ,HZ13031 ,HZ17023 ,HZ18020 ,HZ19023  ,HZlOO83 ,HZ10242 ,HZ10426 ,HZ10617 ,HZI0774 ,HZ12026 ,HZ12076 ,HZ13032 ,HZ17024 ,HZ18022 ,HZ19025  ,HZIOO84 ,HZ10245 ,HZI0430 ,HZI0620 ,HZI0776 ,HZ12027 ,HZ12077 ,HZ13038 ,HZ17027 ,HZ18024 ,HZ19026  ,HZlOO86 ,HZ10247 ,HZI0434 ,HZI0627 ,HZI0778 ,HZ12028 ,HZ12078 ,HZ13039 ,HZ17028 ,HZ18026 ,HZ19028  ,HZIOO88 ,HZI0249 ,HZI0438 ,HZI0629 ,HZll001 ,HZ12029 ,HZ12080 ,HZ13041 ,HZ17030 ,HZ18035 ,HZ19029  ,HZlOO90 ,HZI0254 ,HZI0442 ,HZI0632 ,HZll002 ,HZ12030 ,HZ12081 ,HZ13042 ,HZ17032 ,HZ18040 ,HZ19033  ,HZIOO92 ,HZI0257 ,HZI0450 ,HZI0640 ,HZll003 ,HZ12031 ,HZ12082 ,HZ13043 ,HZ17034 ,HZ18041 ,HZ19034  ,HZlOO95 ,HZI0258 ,HZI0453 ,HZI0644 ,HZll004 ,HZ12032 ,HZ12083 ,HZ13044 ,HZ17035 ,HZ18042 ,HZ19035  ,HZIOO98 ,HZI0260 ,HZI0460 ,HZI0646 ,HZll005 ,HZ12033 ,HZ12084 ,HZ13045 ,HZ17036 ,HZ18043 ,HZ19036  ,HZIOI02 ,HZ10271 ,HZI0467 ,HZI0647 ,HZll006 ,HZ12034 ,HZ12085 ,HZ13046 ,HZ17037 ,HZ18046 ,HZ19040  ,HZIOI08 ,HZI0273 ,HZI0479 ,HZI0649 ,HZll007 ,HZ12035 ,HZ12086 ,HZ13047 ,HZ17039 ,HZ18048 ,HZ19041  ,HZIOll0 ,HZ10274 ,HZI0489 ,HZI0650 ,HZll008 ,HZ12036 ,HZ12087 ,HZ13048 ,HZ17040 ,HZ18062 ,HZ19043  ,HZI0116 ,HZI0283 ,HZ10493 ,HZ10652 ,HZll010 ,HZ12037 ,HZ12088 ,HZ13049 ,HZ17041 ,HZ18066 ,HZ19044  ,HZI0117 ,HZI0293 ,HZI0495 ,HZ10656 ,HZll0ll ,HZ12038 ,HZ12089 ,HZ13050 ,HZ17042 ,HZ18068 ,HZ19045  ,HZI0121 ,HZlO294 ,HZI0496 ,HZI0659 ,HZll013 ,HZ12039 ,HZ12090 ,HZ13052 ,HZ17043 ,HZ18069 ,HZ19046  ,HZI0126 ,HZI0295 ,HZ10497 ,HZI0663 ,HZll014 ,HZ12040 ,HZ12091 ,HZ13053 ,HZ17044 ,HZ18070 ,HZ19051  ,HZI0138 ,HZI0297 ,HZI0500 ,HZ10665 ,HZll015 ,HZ12041 ,HZ12092 ,HZ13054 ,HZ17046 ,HZ18076 ,HZ19052  ,HZI0144 ,HZ10299 ,HZ10501 ,HZ10670 ,HZll016 ,HZ12042 ,HZ12093 ,HZ13055 ,HZ17048 ,HZ18077 ,HZ19053  ,HZlO146 ,HZlO303 ,HZI0507 ,HZI0674 ,HZll017 ,HZ12043 ,HZ12094 ,HZ13056 ,HZ17049 ,HZ18079 ,HZ19055  ,HZlO148 ,HZI0305 ,HZI0508 ,HZI0675 ,HZll018 ,HZ12044 ,HZ12095 ,HZ13057 ,HZ17051 ,HZ18085 ,HZ19056  ,HZlO151 ,HZI0311 ,HZI0509 ,HZI0676 ,HZll019 ,HZ12045 ,HZ12096 ,HZ13058 ,HZ17057 ,HZ18088 ,HZ19059   ,HZ19060 ,HZ19086 ,HZl9109 ,HZ19129 ,HZ19158 ,HZ19193 ,HZ19926   ,HZ19062 ,HZ19087 ,HZ19113 ,HZ19130 ,HZ19159 ,HZ19194 ,HZ19996   ,HZ19065 ,HZ19088 ,HZ19114 ,HZ19131 ,HZ19162 ,HZ19195 ,HZ19997   ,HZ19066 ,HZ19089 ,HZ19115 ,HZ19133 ,HZ19167 ,HZ19201 ,HZ19998   ,HZl9068 ,HZ19091 ,HZl9117 ,HZ19134 ,HZ19170 ,HZ19226   ,HZ19069 ,HZ19095 ,HZ19118 ,HZ19135 ,HZ19171 ,HZ19227   ,HZ19071 ,HZ19096 ,HZ19119 ,HZ19140 ,HZ19172 ,HZ19228   ,HZ19072 ,HZ19098 ,HZ19121 ,HZ19141 ,HZ19179 ,HZ19229   ,HZ19076 ,HZ19099 ,HZ19125 ,HZ19142 ,HZ19186 ,HZ19235   ,HZ19078 ,HZ19104 ,HZ19126 ,HZ19143 ,HZl9189 ,HZ19236   ,HZ19079 ,HZ19106 ,HZ19128 ,HZ19155 ,HZ19190 , and HZ19611.  
FEI Number 2916710
Recalling Firm/
Manufacturer
Varian Medical Systems, Inc. Oncology Systems
911 Hansen Way
Palo Alto CA 94304-1028
For Additional Information ContactMichael A. Pignataro
650-424-5731
Manufacturer Reason
for Recall
A software anomaly has been identified with the Optical Guidance Platform (OGP) Software v2.6 and v2.6.1 where the transfer of datasets from treatment planning systems other than the FastPlan system result in a lateral offset error.
FDA Determined
Cause 2
Software design
ActionOn 3/30/2011, Varian distributed an Urgent Medical Device Correction Urgent Field Safety Notice to all affected users, with a description of the problem and identification of the affected product. The letter also informed customers of how they can determine if the problem affects them. The letter states that there is no work around for the issue unless they possess a FastPlan system. Varian is preparing a software patch to correct the issue and it will be applied to all OGP v2.6 and v2.6.1 systems. Customers are to advise the appropriate personnel of the content of the letter. If further clarification is needed, customers are to contact their local Varian Customer Support District or Regional Manager.
Quantity in Commerce598 units
DistributionWorldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = IYE
-
-