U.S. Department of Health and Human Services

Class 2 Device Recall Animas 2020 Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Animas 2020 Insulin Infusion Pumpsee related information
Date Initiated by FirmJuly 29, 2011
Date PostedAugust 12, 2011
Recall Status1 Terminated 3 on December 07, 2011
Recall NumberZ-3017-2011
Recall Event ID 59517
Product Classification Insulin Infusion Pump - Product Code LZG
ProductAnimas 2020 Insulin Infusion Pump. Animas Corp. An insulin infusion.
Code Information Animas 2020 Insulin Pump  Serial numbers 00-89133-12, 00-89207-12, 14-88703-12, 14-88710-12, 15-88704-12, 15-89190-12, 16-89096-12, 16-89191-12, 17-88720-12, 17-89097-12, 17-89185-12, 17-89611-12, 18-88721-12, 18-89612-12, 19-88715, 18-89099-12, 19-89187-12, 19-89194-12, 19-89620-12, 20-88716-12, 20-89188-12, 20-89621-12, 21-88717-12, 21-88731-12, 21-89196-12, 21-89615-12, 22-88732-12, 22-89616-12, 23-88733-12, 24-88734-12, 25-88728-12, 25-88735-12, 27-88737-12, 28-88738-12, 33-89824-12, 35-89819-12, 36-89827-12, 88-89100-12, 91-89110-12, 92-89111-12, 93-89112-12, 94-89106-12, 94-89120-12, 95-89107-12, 95-89114-12, 95-89121-12, 96-89115-12, 21-88386-12, 21-88393-12, 22-88387-12, 96-89122-12, 96-89210-12, 97-89116-12, 97-89123-12, 97-89130-12, 01-89215-12, 02-89216-12, 02-89223-12, 02-89230-12, 03-89217-12, 03-89224-12, 03-89400-12, 04-89218-12, 04-89225-12, 04-89232-12, 04-89401-12, 05-89219-12, 05-89226-12, 06-89227-12, 06-89234-12, 06-89241-12, 06-89410-12, 07-89235-12, 07-89242-12, 07-89404-12, 07-89411-12, 08-89236-12, 08-89412-12, 09-89237-12, 09-89406-12, 09-89413-12, 10-89238-12, 10-89414-12, 11-89239-12, 11-89408-12, 11-89415-12, 12-89409-12, 13-89417-12, 14-89418-12, 14-89520-12, 15-89521-12, 17-89509-12, 17-89516-12, 18-89517-12, 18-98524-12, 18-89531-12, 19-89525-12, 29-88672-12, 32-88668-12, 32-88682-12, 20-89533-12, 20-89540-12, 21-88643-12, 21-89541-12, 22-89528-12, 22-89535-12, 22-89542-12, 23-88638-12, 23-88740-12, 24-88741-12, 24-89544-12, 25-88647-12, 26-88655-12, 26-89553-12, 27-88751-12, 27-89385-12, 27-89554-12, 27-89561-12, 28-88745-12, 28-89555-12, 29-88760-12, 29-89556-12, 29-89563-12, 30-88747-12, 30-88754-12, 30-88761-12, 30-89395-12, 31-88755-12, 31-88762-12, 31-89558-12, 32-88756-12, 32-89397-12, 33-88757-12, 33-88771-12, 34-88758-12, 34-88772-12, 35-88759-12, 35-88766-12, 36-88774-12, 37-88768-12, 39-88777-12, 39-88939-12, 41-88948-12, 42-88787-12, 51-88979-12, 03-88414-12, 10-89245-12, 33-88690-12, 40-88697-12, 98-89131-12, 14-89256-12, 15-89257-12, 16-89258-12, 17-89273-12, 18-89267-12, 19-89444-12, 20-89269-12, 21-89460-12, 22-88901-12, 22-89461-12, 24-89463-12, 26-88905-12, 27-89466-12, 33-88852-12, 34-88853-12, 35-88854-12, 36-88855-12, 37-88856-12, 39-88872-12, 41-88874-12, 43-88869-12, 47-88887-12, 99-88403-12, 00-88404-12, 00-88411-12, 01-88412-12, 02-88406-12, 02-88413-12, 03-88407-12, 04-88408-12, 04-88415-12, 12-88370-12, 13-88371-12, 14-88372-12, 15-88373-12, 16-88367-12, 16-88381-12, 17-88375-12, 18-88376-12, 18-88383-12, 18-88390-12, 19-88377-12, 19-88384-12, 19-88391-12, 20-88378-12, 20-88385-12, 20-88392-12, 44-88789-12, 00-89140-12, 01-89141-12, 64-83146-12, 00-89221-12, 05-89240-12, 17-89266-12, 22-88394-12, 23-88388-12, 23-88395-12, 24-88389-12, 24-88396-12, 25-88397-12, 26-88398-12, 27-88399-12, 96-88400-12, 99-88410-12, 38-88695-12, 41-88698-12, 09-88603-12, 10-88604-12, 12-88606-12, 12-88613-12, 13-88614-12, 15-88616-12, 22-88637-12, 20-88642-12, 22-88644-12, 23-88645-12, 24-88646-12, 12-89180-12, 08-88602-12, 11-88605-12, 11-88612-12, 19-88627-12, 21-88629-12, 33-88588-12, 51-88993-12, 54-88989-12, 54-88996-12, 86-89003-12, 89-89013-12, 90-89007-12, 92-89016-12, 97-89028-12, 98-89029-12, 19-88641-12, 27-88656-12, 27-88663-12, 27-88670-12, 29-88658-12, 35-88678-12, 35-88685-12, 36-88679-12, 36-88686-12, 36-88950-12, 37-88951-12, 38-88952-12, 39-88953-12, 40-88947-12, 40-88954-12, 41-88955-12, 42-88949-12, 42-88963-12, 27-88920-12, 30-88923-12, 32-88932-12, 33-88933-12, 34-88927-12, 35-88928-12, 35-88935-12, 18-88802-12, 21-88805-12, 23-88807-12, 33-88838-12, 02-89135-12, 03-89136-12, 04-89144-12, 06-89160-12, 08-89148-12, 11-89077-12, 11-89158-12, 12-89159-12, 12-89173-12, 13-89167-12, 18-89179-12, 23-89293-12, 24-89294-12, 25-89295-12, 00-89126-12, 02-89128-12, 03-89129-12, 13-88702-12, 16-88705-12, 17-88706-12, 99-89125-12, 25-88742-12, 26-88743-12, 27-88744-12, 28-88752-12, 29-88739-12, 29-88746-12, 31-88748-12, 32-88749-12, 32-88763-12, 32-88770-12, 34-88765-12, 35-88773-12, 35-88780-12, 38-88769-12, 38-88776-12, 40-88778-12, 15-88380-12, 16-88374-12, 17-88368-12, 17-88382-12, 18-88369-12, 21-88379-12, 32-88851-12, 42-88868-12, 07-89330-12, 11-89334-12, 19-88803-12, 19-88810-12, 21-88812-12, 22-88820-12, 24-88808-12, 25-88809-12, 25-88816-12, 25-88830-12, 26-88831-12, 27-88818-12, 27-88832-12, 29-88827-12, 29-88834-12, 30-88828-12, 31-88829-12, 31-88836-12, 32-88837-12, 01-89303-12, 02-89142-12, 03-89150-12, 04-89151-12, 05-89138-12, 05-89145-12, 06-89139-12, 07-89161-12, 08-89155-12, 09-89149-12, 09-89156-12, 10-89083-12, 10-89090-12, 10-89157-12, 10-89171-12, 11-89165-12, 13-89174-12, 13-89181-12, 14-89094-12, 14-89168-12, 14-89175-12, 15-89176-12, 20-89276-12, 20-89290-12, 21-89277-12, 21-89291-12, 22-89278-12, 22-89292-12, 25-89288-12, 26-89289-12, 26-89296-12, 27-89297-12, 29-89299-12, 30-88930-12, 12-88701-12, 05-89233-12, 26-88750-12, 37-88775-12, 10-89164-12, 36-88693-12, 39-88696-12, 32-88675-12, 33-88676-12, and 30-88835-12.
FEI Number 3002920072
Recalling Firm/
Manufacturer
Animas Corporation
200 Lawrence Dr
West Chester PA 19380-3428
For Additional Information ContactJon Mulberg
877-937-7867 Ext. 1257
Manufacturer Reason
for Recall
Incorrect component (1nF capacitor) was used to produce 191 printed circuit boards used in six pump lots of which 380 pumps were distributed. This issue may result in the pump displaying a "Replace Battery" alarm not related to the condition of the battery.
FDA Determined
Cause 2
Mixed-up of materials/components
ActionAnimas Corporation sent an "Urgent: Insulin Pump Recall" letter dated July, 29 20011 to all affected customers. The letter included affected product and problem. They requested anyone who received the pump to send it back to Animas and a replacement pump will be sent at no charge. For additional information contact the Customer Support Department at (877) 937-7867.
Quantity in Commerce380
DistributionWorldwide Distribution: Nationwide (USA) distribution including the states of CA, CO, MA, and MD; and the countries of Australia, Finland, France, Israel, Italy, Sweden and the U.K.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.