Class 2 Device Recall Medtronic Disposable Pressure Display Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medtronic Disposable Pressure Display Setsee related information
Date Initiated by FirmSeptember 09, 2011
Date PostedOctober 20, 2011
Recall Status1 Terminated 3 on October 11, 2012
Recall NumberZ-0079-2012
Recall Event ID 59948
510(K)NumberK852232 
Product Classification Gauge, pressure, coronary, cardiopulmonary bypass - Product Code DXS
ProductMedtronic Intersept Custom Tubing Pack, which are designed as specified by the user. The following models contain a Pressure Display Set. Sterilized by Ethylene Oxide. Non-pyrogenic. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604. Product Usage: This product is intended for use in monitoring catheter, cannula, or line pressures associated with cardiopulmonary bypass equipment and/or related products.
Code Information Models 0K57R2, 0K82R15, 0M63R7, 0M68R6, 0N41R14, 0T50R5, 0V67R27, 0X52R3, 0Y35R, 0Y95R30, 0Z30R11, 1D41R16, 1F66R12, 1G59R1, 1J74R4, 1M90R8, 1P09R2, 2031R7, 2C88R8, 2C89R7, 2C90R7, 2C91R5, 2D61R2, 2D64R16, 2H08R3, 2K85R1, 2K87R1, 2K88R1, 2N90R9, 2T28R1, 3B74R2, 3T73R1, 3W29R, 4B42R5, 4C03R, 4D82R3, 4M55R, 4N04R, 4R43R, 5788R4, 5A45R3, 5B51R9, 5B94R, 5C33R2, 5C45R3, 5C92R9, 5D83R2, 5F68R1, 5G28R4, 5H65R, 5J00R2, 5P62R, 5W13R6, 5Z18R, 6G05R, 6J55R1, 6J88R4, 6J95R8, 6N12R, 6Q49R1, 6Q49R1, 6R08R, 6R09R, 6R44R2, 6S66R, 6S75R1, 6V25R, 6Y04R5, 6Y07R6, 6Y59R3, 7A69R4, 7A70R1, 7B56R1, CB0L44R16, CB0M02R5, CB0M79R8, CB1Q18R11, CB1S33R9, CB2D32R5, CB2W45R7, CB2W86R4, CB3K45R9, CB3K91R1, CB3M40R17, CB3Y46R9, CB3Y47R10, CB4800R18, CB4953R16, CB4958R16, CB4B68R8, CB4E78R7, CB4F45R15, CB4K64R6, CB4L83R4, CB4M60R2, CB4S43R9, CB4W10R10, CB4X84R, CB4Y71R14, CB4Z87R6, CB5A30R10, CB5F84R12, CB5J33R6, CB5L15R13, CB5L79R, CB5L90R16, CB5Q71R2, CB5Q74R1, CB5R04R1, CB5R46R5, CB5U81R, CB6A80R6, CB6G02R4, CB6H12R, CB6H45R4, CB6H76R, CB6H77R, CB6H78R, CB6J66R2, CB6J68R3, CB6J75R3, CB6L12R3, CB6L13R, CB6L14R, CB6L60R2, CB6M11R1, CB6Q15R, CB6Q16R1, CB6Q62R, CB6Q78R2, CB6S17R, CB6U12R, CB6U76R1, CB6X94R2, CB6X98R3, CB6Y02R6, CB6Y34R2, CB7C31R, DLP2X17R, HY0F32R17, HY2280R16, HY2U14R2, HY3U59R12, HY4A96R9, HY4V52R4, HY4V82R3, HY5D07R2, HY6B26R2, HY6G55R6, HY6G56R6, HY6H03R5, HY6H03R5, HY6J33R3, HY6J99R, HY6R36R1, HY6R36R1, HY6U70R2, HY6V34R2, HY6Z04R1, M331105H, M331105K, M336902L, M392810D, M392810E, M392811C, M392811D, M392811E, M440115G, M440115H, M440115J, M440115K, M440115L, M440129A, M440130A, M441002E, M443824F, M443825E, M492609A, M499301B, M530710A, M530710C, M530710D, M960208A, M960454B, P390806A, S330115A, SS6X67R1, SS6Z09R1, SSHY6E81R2, TL2H89R7, TL3Y45R1, TL3Z67R4, TL4E64R, TL4E66R10, TL4E66R10, TL5398R21, TL5C09R4, TL5M01R7, TL5R15R4, TL6A91R2, TL6C91R2, TL6D87R2, TL6H39R5, TL6W08R3, TL6X29R4, TL6X36R2, TL6X62R5, TL6Y92R, TL7A15R2.
Recalling Firm/
Manufacturer
Medtronic Perfusion Systems
7611 Northland Dr N
Brooklyn Park MN 55428-1088
For Additional Information ContactMedtronic Cardiovascular Tech Support
877-526-7890
Manufacturer Reason
for Recall
Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the system resulting in a pressure reading that is not representative of the actual line pressure.
FDA Determined
Cause 2
Device Design
ActionMedtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification. For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.
Quantity in Commerce74,804
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela, and Yemen.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DXS
-
-