| | Class 2 Device Recall GEMINI TF 16 PET/CT Xray Diagnostic Imaging System |  |
| Date Initiated by Firm | April 25, 2011 |
| Date Posted | December 01, 2011 |
| Recall Status1 |
Terminated 3 on February 22, 2013 |
| Recall Number | Z-0343-2012 |
| Recall Event ID |
60302 |
| 510(K)Number | K081135 |
| Product Classification |
System, tomography, computed, emission - Product Code KPS
|
| Product | Philips GEMINI TF 16 Diagnostic CT X-ray/PET Imaging System, 510(k) #K052640. System units equipped with software version: 3.5.2 are subject to recall/field correction.
The GEMINI TF I6, and TF 64 are diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and other organ imaging). Attenuation correction is accomplished by CT AC. The device also provides for list mode, dynamic, and gated acquisitions. Both subsystems (PET and CT) can also be operated independently as fully functional, diagnostic imaging systems including application of the CT scanner as a radiation therapy simulation scanner. |
| Code Information |
The recalled GEMINI TF 16 system units are identified as follows: Model Number: 882470 and Model #882473, Serial Numbers: 7228, 7230, 7233, 7235, 7239, 7240, 7242, 7244, and 7237. |
| FEI Number |
1525965
|
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Road Cleveland OH 44143-2131
|
| For Additional Information Contact | 440-483-7000 |
Manufacturer Reason for Recall | In response to complaints received by Philips Medical concerning certain identified software errors reported by their customers, the firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm believes that if these software defects were to recur, it could impact the performance of reliability of the these PET/CT system units. |
FDA Determined Cause 2 | Software design |
| Action | Philips Healthcare sent a "CPE CUSTOMER LETTER" dated October 28, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Additionally, device operation information is provided to customers. A Philips Field Service Engineer will be visiting each customer site to install the necessary software upgrades. Contact the Customer Care Solutions Center at 1-800-722-9377 for questions concerning this notice. |
| Quantity in Commerce | 9 units |
| Distribution | Nationwide Distribution including the states of CO, LA, MD, MI, MN, PA, TX, VA, VT, and WI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|