| Class 2 Device Recall EZ Steer ThermoCool Nav | |
Date Initiated by Firm | November 22, 2010 |
Date Posted | February 07, 2012 |
Recall Status1 |
Terminated 3 on February 08, 2012 |
Recall Number | Z-0979-2012 |
Recall Event ID |
60577 |
PMA Number | P030031 |
Product Classification |
Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation - Product Code OAE
|
Product | EZ Steer ThermoCool Nav Part Number D-1292-01-S Diagnostic/Ablation Catheter instructions for use.
The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter is a multi-electrode luminal catheter with a deflectable tip designed to facilitate electrophysiological mapping of the heart and to transmit radiofrequency (RF) current to the catheter tip electrode for ablation purposes. The catheter shaft measures 7.5 F with 8 F ring electrodes. For ablation, the catheter is used in conjunction with an RF generator and a dispersive pad (indifferent electrode) for the treatment of: - Type I atrial flutter in patients age 18 or older. - Recurrent drug/device refractory sustained monomorphic ventricular tachycardia (VT) due to prior myocardial infarction (Mil in adults. - Drug refractory recurrent symptomatic paroxysmal atrial fibrillation, when used with compatible three-dimensional electroanatomic mapping systems. The EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Catheter provides location information when used with CARTO~ EP Navigation Systems. |
Code Information |
Lot Numbers: 14050236, 14069164, 14085850, 14085852, 14089636, 14098511, 14102112, 14138785, 14142301, and 15091243 |
Recalling Firm/ Manufacturer |
Biosense Webster, Inc. 15715 Arrow Hwy Irwindale CA 91706-2006
|
For Additional Information Contact | 909-839-8604 |
Manufacturer Reason for Recall | Biosense Webster, Inc. has recently become aware of the omission of the following two Warning and Precaution statements on the Instructions For Use (IFU) for the above catheter:
1. If phrenic nerve location is a concern, precautionary measures are recommended to evaluate the proximity of the nerve to the ablation electrode, such as pacing maneuvers.
2. To minimize risk of atrio-esophageal fist |
FDA Determined Cause 2 | Labeling design |
Action | Biosense Webster, Inc. sent an Urgent Field Safety Notice letter dated Novembe 22, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were advised to continue use of the affected product and that no further action is required on their part. Consignees were advised that the updated IFU can be found at www.jnjgatewayifu.com. Consignees were instructed to contact their local Biosense Webster representative if they have any questions or require additional information.
For questions regarding this recall call 909-839-8604. |
Quantity in Commerce | 24,713 units in total |
Distribution | Nationwide Distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database | PMAs with Product Code = OAE
|
|
|
|