Class 2 Device Recall Varian Unique Single Energy Linear Accelerator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Varian Unique Single Energy Linear Accelerator see related information
Date Initiated by Firm November 28, 2011
Date Posted January 11, 2012
Recall Status1 Terminated 3 on June 02, 2015
Recall Number Z-0483-2012
Recall Event ID 60659
510(K)Number K101751  
Product Classification Accelerator, linear, medical - Product Code IYE
Product Varian Unique Single Energy Linear Accelerator, C-Series Clinac or Trilogy, versions 7.x and 8.x, Reference/FSCA Identifier: CP-0661; Model Numbers: H14, H18, H27, H29; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA

Varian Unique Single Energy Linear Accelerator: The UNIQUE is a Single Energy Linear Accelerator intended to be used for conventional radiotherapy and includes modifications to the previously cleared Varian Trilogy. The UNIQUE provides additional features, safety improvements, and usability improvements. The UNIQUE is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Code Information H140001 ,H140145 ,H140711 ,H140772 ,H140832 ,H140893 ,H140958 ,H271265 ,H271742 ,H272014 ,H272235 ,H272416 ,H272858  ,H140011 ,H140146 ,H140712 ,H140773 ,H140833 ,H140895 ,H140959 ,H271277 ,H271755 ,H272016 ,H272236 ,H272418 ,H272859  ,H140018 ,H140147 ,H140713 ,H140774 ,H140834 ,H140896 ,H140962 ,H271293 ,H271760 ,H272023 ,H272237 ,H272421 ,H272860  ,H140025 ,H140148 ,H140714 ,H140775 ,H140835 ,H140897 ,H140963 ,H271308 ,H271762 ,H272029 ,H272239 ,H272422 ,H272862  ,H140026 ,H140150 ,H140715 ,H140776 ,H140836 ,H140898 ,H180455 ,H271334 ,H271763 ,H272031 ,H272240 ,H272423 ,H272864  ,H140027 ,H140151 ,H140716 ,H140777 ,H140837 ,H140899 ,H182000 ,H271362 ,H271773 ,H272033 ,H272243 ,H272425 ,H272866  ,H140029 ,H140152 ,H140717 ,H140778 ,H140838 ,H140900 ,H182001 ,H271370 ,H271780 ,H272035 ,H272250 ,H272426 ,H272868  ,H140031 ,H140153 ,H140718 ,H140779 ,H140839 ,H140901 ,H182002 ,H271378 ,H271783 ,H272043 ,H272259 ,H272427 ,H272872  ,H140038 ,H140154 ,H140719 ,H140780 ,H140840 ,H140902 ,H182003 ,H271385 ,H271785 ,H272045 ,H272260 ,H272428 ,H272873  ,H140039 ,H140201 ,H140720 ,H140781 ,H140841 ,H140903 ,H182005 ,H271393 ,H271797 ,H272049 ,H272271 ,H272432 ,H272876  ,H140043 ,H140202 ,H140721 ,H140782 ,H140842 ,H140904 ,H182006 ,H271407 ,H271801 ,H272050 ,H272277 ,H272434 ,H272885  ,H140045 ,H140203 ,H140722 ,H140783 ,H140843 ,H140905 ,H182007 ,H271413 ,H271802 ,H272052 ,H272283 ,H272438 ,H272889  ,H140046 ,H140204 ,H140723 ,H140784 ,H140844 ,H140906 ,H182008 ,H271414 ,H271811 ,H272062 ,H272290 ,H272440 ,H272890  ,H140047 ,H140205 ,H140724 ,H140785 ,H140845 ,H140907 ,H182009 ,H271420 ,H271825 ,H272064 ,H272291 ,H272444 ,H272893  ,H140048 ,H140206 ,H140725 ,H140786 ,H140846 ,H140908 ,H182010 ,H271425 ,H271831 ,H272069 ,H272293 ,H272445 ,H272899  ,H140058 ,H140207 ,H140726 ,H140787 ,H140847 ,H140909 ,H182011 ,H271429 ,H271833 ,H272078 ,H272299 ,H272446 ,H272905  ,H140062 ,H140208 ,H140727 ,H140788 ,H140848 ,H140910 ,H182012 ,H271438 ,H271835 ,H272081 ,H272301 ,H272454 ,H272908  ,H140067 ,H140209 ,H140728 ,H140789 ,H140849 ,H140911 ,H182013 ,H271439 ,H271838 ,H272085 ,H272302 ,H272456 ,H272909  ,H140070 ,H140210 ,H140729 ,H140790 ,H140850 ,H140912 ,H182014 ,H271440 ,H271845 ,H272086 ,H272306 ,H272458 ,H272910  ,H140071 ,H140211 ,H140730 ,H140791 ,H140851 ,H140913 ,H182016 ,H271443 ,H271853 ,H272089 ,H272307 ,H272460 ,H272911  ,H140073 ,H140212 ,H140731 ,H140792 ,H140852 ,H140914 ,H182017 ,H271452 ,H271856 ,H272111 ,H272309 ,H272462 ,H272912  ,H140076 ,H140213 ,H140732 ,H140793 ,H140853 ,H140915 ,H182018 ,H271490 ,H271858 ,H272115 ,H272310 ,H272464 ,H272914  ,H140077 ,H140214 ,H140733 ,H140794 ,H140854 ,H140916 ,H182019 ,H271497 ,H271860 ,H272118 ,H272311 ,H272466 ,H272922  ,H140078 ,H140215 ,H140734 ,H140795 ,H140855 ,H140917 ,H182020 ,H271502 ,H271862 ,H272129 ,H272313 ,H272469 ,H272926  ,H140079 ,H140216 ,H140735 ,H140796 ,H140856 ,H140918 ,H182021 ,H271507 ,H271866 ,H272134 ,H272318 ,H272470 ,H272932  ,H140088 ,H140217 ,H140736 ,H140797 ,H140857 ,H140919 ,H182022 ,H271515 ,H271870 ,H272136 ,H272321 ,H272473 ,H272933  ,H140090 ,H140218 ,H140737 ,H140798 ,H140858 ,H140920 ,H182023 ,H271519 ,H271872 ,H272137 ,H272324 ,H272476 ,H272936  ,H140092 ,H140219 ,H140738 ,H140799 ,H140859 ,H140921 ,H182030 ,H271523 ,H271875 ,H272138 ,H272325 ,H272479 ,H272938  ,H140096 ,H140220 ,H140739 ,H140800 ,H140860 ,H140922 ,H182032 ,H271526 ,H271877 ,H272143 ,H272326 ,H272480 ,H272941  ,H140097 ,H140221 ,H140740 ,H140801 ,H140861 ,H140923 ,H260905 ,H271531 ,H271879 ,H272153 ,H272327 ,H272488 ,H272946  ,H140098 ,H140222 ,H140742 ,H140802 ,H140862 ,H140924 ,H260909 ,H271534 ,H271881 ,H272155 ,H272328 ,H272491 ,H272951  ,H140101 ,H140223 ,H140743 ,H140803 ,H140863 ,H140925 ,H260922 ,H271559 ,H271890 ,H272156 ,H272329 ,H272492 ,H272953  ,H140107 ,H140224 ,H140744 ,H140804 ,H140864 ,H140926 ,H260924 ,H271582 ,H271893 ,H272157 ,H272333 ,H272494 ,H272956  ,H140109 ,H140225 ,H140745 ,H140805 ,H140865 ,H140927 ,H260925 ,H271587 ,H271895 ,H272158 ,H272336 ,H272497 ,H272960  ,H140111 ,H140226 ,H140746 ,H140806 ,H140866 ,H140929 ,H260928 ,H271588 ,H271896 ,H272159 ,H272339 ,H272500 ,H272961  ,H140113 ,H140227 ,H140747 ,H140807 ,H140867 ,H140930 ,H260931 ,H271593 ,H271903 ,H272162 ,H272341 ,H272801 ,H272963  ,H140114 ,H140228 ,H140748 ,H140808 ,H140868 ,H140931 ,H260936 ,H271594 ,H271905 ,H272163 ,H272343 ,H272803 ,H272964     ,H140115 ,H140229 ,H140749 ,H140809 ,H140869 ,H140932 ,H260940 ,H271602 ,H271906 ,H272168 ,H272345 ,H272804 ,H272968  ,H140116 ,H140230 ,H140750 ,H140810 ,H140870 ,H140933 ,H260942 ,H271607 ,H271918 ,H272169 ,H272346 ,H272805 ,H272971  ,H140117 ,H140231 ,H140751 ,H140811 ,H140871 ,H140934 ,H260946 ,H271618 ,H271921 ,H272172 ,H272349 ,H272806 ,H272973  ,H140118 ,H140232 ,H140752 ,H140812 ,H140872 ,H140935 ,H260949 ,H271624 ,H271926 ,H272173 ,H272354 ,H272811 ,H272978  ,H140119 ,H140233 ,H140753 ,H140813 ,H140873 ,H140936 ,H260950 ,H271626 ,H271931 ,H272174 ,H272355 ,H272816 ,H272981  ,H140121 ,H140234 ,H140754 ,H140814 ,H140874 ,H140937 ,H260951 ,H271628 ,H271935 ,H272178 ,H272359 ,H272820 ,H272982  ,H140122 ,H140235 ,H140755 ,H140815 ,H140875 ,H140938 ,H260952 ,H271635 ,H271941 ,H272179 ,H272360 ,H272822 ,H272985  ,H140124 ,H140236 ,H140756 ,H140816 ,H140876 ,H140939 ,H260953 ,H271640 ,H271942 ,H272180 ,H272362 ,H272824 ,H272986  ,H140125 ,H140237 ,H140757 ,H140817 ,H140877 ,H140940 ,H270767 ,H271662 ,H271951 ,H272184 ,H272365 ,H272825 ,H272987  ,H140129 ,H140238 ,H140758 ,H140818 ,H140878 ,H140941 ,H270788 ,H271663 ,H271954 ,H272187 ,H272370 ,H272826 ,H272990  ,H140132 ,H140239 ,H140759 ,H140819 ,H140879 ,H140942 ,H270794 ,H271664 ,H271961 ,H272188 ,H272372 ,H272827 ,H272993  ,H140133 ,H140240 ,H140760 ,H140820 ,H140881 ,H140943 ,H271041 ,H271669 ,H271964 ,H272195 ,H272377 ,H272828 ,H272996  ,H140134 ,H140700 ,H140761 ,H140821 ,H140882 ,H140944 ,H271066 ,H271678 ,H271972 ,H272197 ,H272380 ,H272834 ,H272999  ,H140135 ,H140701 ,H140762 ,H140822 ,H140883 ,H140945 ,H271105 ,H271681 ,H271973 ,H272202 ,H272381 ,H272837 ,H273003  ,H140136 ,H140702 ,H140763 ,H140823 ,H140884 ,H140947 ,H271157 ,H271691 ,H271974 ,H272203 ,H272388 ,H272841 ,H273004  ,H140137 ,H140703 ,H140764 ,H140824 ,H140885 ,H140948 ,H271158 ,H271693 ,H271975 ,H272204 ,H272394 ,H272842 ,H273006  ,H140138 ,H140704 ,H140765 ,H140825 ,H140886 ,H140949 ,H271159 ,H271711 ,H271976 ,H272209 ,H272397 ,H272846 ,H273011  ,H140139 ,H140705 ,H140766 ,H140826 ,H140887 ,H140950 ,H271167 ,H271712 ,H271993 ,H272216 ,H272401 ,H272847 ,H273016  ,H140140 ,H140706 ,H140767 ,H140827 ,H140888 ,H140951 ,H271194 ,H271713 ,H271994 ,H272219 ,H272405 ,H272848 ,H273026  ,H140141 ,H140707 ,H140768 ,H140828 ,H140889 ,H140952 ,H271214 ,H271714 ,H272000 ,H272220 ,H272407 ,H272849 ,H273027  ,H140142 ,H140708 ,H140769 ,H140829 ,H140890 ,H140953 ,H271227 ,H271720 ,H272001 ,H272223 ,H272410 ,H272854 ,H273029  ,H140143 ,H140709 ,H140770 ,H140830 ,H140891 ,H140956 ,H271242 ,H271724 ,H272003 ,H272226 ,H272412 ,H272856 ,H273030  ,H140144 ,H140710 ,H140771 ,H140831 ,H140892 ,H140957 ,H271258 ,H271729 ,H272009 ,H272231 ,H272413 ,H272857 ,H273031  ,H273032 ,H273133 ,H273197 ,H290501 ,H291030 ,H291094 ,H291158 ,H291220 ,H293306 ,H293369 ,H293432 ,H293494 ,H293558  ,H273035 ,H273134 ,H273198 ,H290503 ,H291031 ,H291095 ,H291159 ,H291221 ,H293307 ,H293370 ,H293433 ,H293495 ,H293559  ,H273036 ,H273135 ,H273199 ,H290504 ,H291032 ,H291096 ,H291160 ,H291222 ,H293308 ,H293371 ,H293434 ,H293496 ,H293560  ,H273039 ,H273136 ,H273200 ,H290505 ,H291034 ,H291097 ,H291161 ,H291223 ,H293309 ,H293372 ,H293435 ,H293497 ,H293561  ,H273040 ,H273137 ,H273201 ,H290508 ,H291035 ,H291098 ,H291162 ,H291224 ,H293310 ,H293373 ,H293436 ,H293498 ,H293562  ,H273041 ,H273138 ,H273202 ,H290509 ,H291036 ,H291099 ,H291163 ,H291225 ,H293311 ,H293374 ,H293437 ,H293499 ,H293563  ,H273042 ,H273139 ,H273203 ,H290510 ,H291037 ,H291101 ,H291164 ,H291226 ,H293312 ,H293375 ,H293438 ,H293500 ,H293564  ,H273048 ,H273140 ,H273204 ,H290511 ,H291038 ,H291102 ,H291165 ,H291227 ,H293313 ,H293376 ,H293439 ,H293501 ,H293565  ,H273049 ,H273141 ,H273205 ,H290512 ,H291039 ,H291103 ,H291166 ,H291228 ,H293314 ,H293377 ,H293440 ,H293502 ,H293566  ,H273052 ,H273142 ,H273206 ,H290517 ,H291040 ,H291104 ,H291167 ,H291229 ,H293315 ,H293378 ,H293441 ,H293503 ,H293567  ,H273053 ,H273143 ,H273208 ,H290519 ,H291041 ,H291105 ,H291168 ,H291230 ,H293316 ,H293379 ,H293442 ,H293504 ,H293568  ,H273055 ,H273144 ,H273209 ,H290520 ,H291042 ,H291106 ,H291169 ,H291231 ,H293317 ,H293380 ,H293443 ,H293505 ,H293569  ,H273056 ,H273145 ,H273210 ,H290530 ,H291043 ,H291107 ,H291170 ,H291232 ,H293318 ,H293381 ,H293444 ,H293506 ,H293570  ,H273057 ,H273146 ,H273547 ,H290533 ,H291044 ,H291108 ,H291171 ,H291233 ,H293319 ,H293382 ,H293445 ,H293507 ,H293571  ,H273063 ,H273147 ,H273711 ,H290538 ,H291045 ,H291109 ,H291172 ,H291234 ,H293320 ,H293383 ,H293446 ,H293508 ,H293572  ,H273064 ,H273148 ,H273929 ,H290540 ,H291046 ,H291111 ,H291173 ,H291235 ,H293321 ,H293384 ,H293447 ,H293509 ,H293573  ,H273065 ,H273149 ,H290186 ,H290541 ,H291047 ,H291112 ,H291174 ,H291236 ,H293323 ,H293385 ,H293448 ,H293510 ,H293574  ,H273066 ,H273150 ,H290192 ,H290547 ,H291048 ,H291113 ,H291175 ,H291237 ,H293324 ,H293386 ,H293449 ,H293511 ,H293575  ,H273072 ,H273151 ,H290201 ,H290548 ,H291049 ,H291114 ,H291176 ,H291238 ,H293325 ,H293387 ,H293450 ,H293512 ,H293576  ,H273075 ,H273152 ,H290208 ,H290549 ,H291050 ,H291115 ,H291177 ,H291239 ,H293326 ,H293388 ,H293451 ,H293513 ,H293577  ,H273077 ,H273153 ,H290222 ,H290550 ,H291051 ,H291116 ,H291178 ,H291240 ,H293327 ,H293389 ,H293452 ,H293514 ,H293578  ,H273080 ,H273155 ,H290224 ,H290551 ,H291052 ,H291117 ,H291179 ,H291241 ,H293328 ,H293390 ,H293453 ,H293515 ,H293579  ,H273082 ,H273156 ,H290233 ,H290552 ,H291053 ,H291118 ,H291180 ,H291242 ,H293329 ,H293391 ,H293454 ,H293516 ,H293581  ,H273083 ,H273157 ,H290241 ,H290553 ,H291054 ,H291119 ,H291181 ,H291243 ,H293330 ,H293392 ,H293455 ,H293517 ,H293583  ,H273084 ,H273158 ,H290243 ,H290554 ,H291055 ,H291120 ,H291182 ,H291244 ,H293331 ,H293393 ,H293456 ,H293518 ,H293584  ,H273086 ,H273159 ,H290248 ,H290555 ,H291056 ,H291121 ,H291183 ,H291245 ,H293332 ,H293394 ,H293457 ,H293519 ,H293585  ,H273088 ,H273160 ,H290249 ,H290556 ,H291057 ,H291122 ,H291184 ,H291246 ,H293333 ,H293395 ,H293458 ,H293520 ,H293586  ,H273089 ,H273162 ,H290256 ,H290557 ,H291059 ,H291123 ,H291185 ,H291247 ,H293334 ,H293396 ,H293459 ,H293521 ,H293587  ,H273092 ,H273163 ,H290262 ,H290558 ,H291060 ,H291124 ,H291186 ,H291248 ,H293335 ,H293397 ,H293460 ,H293522 ,H293588  ,H273093 ,H273164 ,H290272 ,H290559 ,H291061 ,H291125 ,H291187 ,H291249 ,H293336 ,H293398 ,H293461 ,H293523 ,H293589  ,H273100 ,H273165 ,H290274 ,H290560 ,H291062 ,H291126 ,H291188 ,H291250 ,H293337 ,H293399 ,H293462 ,H293524 ,H293590  ,H273101 ,H273166 ,H290278 ,H290561 ,H291063 ,H291127 ,H291189 ,H291251 ,H293338 ,H293401 ,H293463 ,H293525 ,H293591  ,H273102 ,H273167 ,H290282 ,H290600 ,H291064 ,H291128 ,H291190 ,H291252 ,H293339 ,H293402 ,H293464 ,H293527 ,H293592  ,H273103 ,H273168 ,H290287 ,H291001 ,H291065 ,H291129 ,H291191 ,H291253 ,H293340 ,H293403 ,H293465 ,H293528 ,H293593  ,H273104 ,H273169 ,H290295 ,H291002 ,H291066 ,H291130 ,H291192 ,H291254 ,H293341 ,H293404 ,H293466 ,H293529 ,H293594  ,H273105 ,H273170 ,H290304 ,H291003 ,H291067 ,H291131 ,H291193 ,H291255 ,H293342 ,H293405 ,H293467 ,H293530 ,H293595  ,H273106 ,H273171 ,H290308 ,H291004 ,H291068 ,H291132 ,H291194 ,H291256 ,H293343 ,H293406 ,H293468 ,H293531 ,H293596  ,H273107 ,H273172 ,H290313 ,H291005 ,H291069 ,H291133 ,H291195 ,H291257 ,H293344 ,H293407 ,H293469 ,H293532 ,H293597  ,H273108 ,H273173 ,H290317 ,H291006 ,H291070 ,H291134 ,H291196 ,H291258 ,H293345 ,H293408 ,H293470 ,H293533 ,H293598  ,H273109 ,H273174 ,H290321 ,H291007 ,H291071 ,H291135 ,H291197 ,H291259 ,H293346 ,H293409 ,H293471 ,H293534 ,H293599  ,H273110 ,H273175 ,H290326 ,H291008 ,H291072 ,H291136 ,H291198 ,H291260 ,H293347 ,H293410 ,H293472 ,H293535 ,H293600  ,H273111 ,H273176 ,H290341 ,H291009 ,H291073 ,H291137 ,H291199 ,H291261 ,H293348 ,H293411 ,H293473 ,H293536 ,H293601  ,H273112 ,H273177 ,H290352 ,H291010 ,H291074 ,H291138 ,H291200 ,H291262 ,H293349 ,H293412 ,H293474 ,H293537 ,H293602  ,H273113 ,H273178 ,H290358 ,H291011 ,H291075 ,H291139 ,H291201 ,H291263 ,H293350 ,H293413 ,H293475 ,H293538 ,H293603  ,H273114 ,H273179 ,H290362 ,H291012 ,H291076 ,H291140 ,H291202 ,H291264 ,H293351 ,H293414 ,H293476 ,H293539 ,H293604     ,H273115 ,H273180 ,H290371 ,H291013 ,H291077 ,H291141 ,H291203 ,H291265 ,H293352 ,H293415 ,H293477 ,H293540 ,H293605  ,H273116 ,H273181 ,H290372 ,H291014 ,H291078 ,H291142 ,H291204 ,H291266 ,H293353 ,H293416 ,H293478 ,H293541 ,H293606  ,H273117 ,H273182 ,H290376 ,H291015 ,H291079 ,H291143 ,H291205 ,H291267 ,H293354 ,H293417 ,H293479 ,H293542 ,H293607  ,H273118 ,H273183 ,H290379 ,H291016 ,H291080 ,H291144 ,H291206 ,H291268 ,H293355 ,H293418 ,H293480 ,H293543 ,H293608  ,H273119 ,H273184 ,H290381 ,H291017 ,H291081 ,H291145 ,H291207 ,H291269 ,H293356 ,H293419 ,H293481 ,H293544 ,H293609  ,H273120 ,H273185 ,H290386 ,H291018 ,H291082 ,H291146 ,H291208 ,H291270 ,H293357 ,H293420 ,H293482 ,H293545 ,H293610  ,H273121 ,H273186 ,H290392 ,H291019 ,H291083 ,H291147 ,H291209 ,H291271 ,H293358 ,H293421 ,H293483 ,H293546 ,H293611  ,H273122 ,H273187 ,H290397 ,H291020 ,H291084 ,H291148 ,H291210 ,H291272 ,H293359 ,H293422 ,H293484 ,H293548 ,H293612  ,H273123 ,H273188 ,H290399 ,H291021 ,H291085 ,H291149 ,H291211 ,H291273 ,H293360 ,H293423 ,H293485 ,H293549 ,H293613  ,H273124 ,H273189 ,H290400 ,H291022 ,H291086 ,H291150 ,H291212 ,H293009 ,H293361 ,H293424 ,H293486 ,H293550 ,H293614  ,H273125 ,H273190 ,H290401 ,H291023 ,H291087 ,H291151 ,H291213 ,H293207 ,H293362 ,H293425 ,H293487 ,H293551 ,H293615  ,H273126 ,H273191 ,H290402 ,H291024 ,H291088 ,H291152 ,H291214 ,H293300 ,H293363 ,H293426 ,H293488 ,H293552 ,H293616  ,H273127 ,H273192 ,H290403 ,H291025 ,H291089 ,H291153 ,H291215 ,H293301 ,H293364 ,H293427 ,H293489 ,H293553 ,H293617  ,H273128 ,H273193 ,H290405 ,H291026 ,H291090 ,H291154 ,H291216 ,H293302 ,H293365 ,H293428 ,H293490 ,H293554 ,H293618  ,H273130 ,H273194 ,H290414 ,H291027 ,H291091 ,H291155 ,H291217 ,H293303 ,H293366 ,H293429 ,H293491 ,H293555 ,H293619  ,H273131 ,H273195 ,H290422 ,H291028 ,H291092 ,H291156 ,H291218 ,H293304 ,H293367 ,H293430 ,H293492 ,H293556 ,H293620  ,H273132 ,H273196 ,H290423 ,H291029 ,H291093 ,H291157 ,H291219 ,H293305 ,H293368 ,H293431 ,H293493 ,H293557 ,H293621  ,H293622 ,H293686 ,H293749 ,H293812 ,H293875 ,H293938 ,H294002 ,H294065 ,H294129 ,H294192 ,H294254 ,H294317 ,H294381  ,H293623 ,H293687 ,H293750 ,H293813 ,H293876 ,H293940 ,H294003 ,H294066 ,H294130 ,H294193 ,H294255 ,H294318 ,H294382  ,H293624 ,H293688 ,H293751 ,H293814 ,H293877 ,H293941 ,H294004 ,H294067 ,H294131 ,H294194 ,H294256 ,H294319 ,H294383  ,H293625 ,H293689 ,H293752 ,H293815 ,H293878 ,H293942 ,H294005 ,H294068 ,H294132 ,H294195 ,H294257 ,H294320 ,H294384  ,H293626 ,H293690 ,H293753 ,H293816 ,H293879 ,H293943 ,H294006 ,H294069 ,H294133 ,H294196 ,H294258 ,H294321 ,H294385  ,H293627 ,H293691 ,H293754 ,H293817 ,H293880 ,H293944 ,H294007 ,H294070 ,H294134 ,H294197 ,H294259 ,H294322 ,H294386  ,H293628 ,H293692 ,H293755 ,H293818 ,H293881 ,H293945 ,H294008 ,H294071 ,H294135 ,H294198 ,H294260 ,H294323 ,H294387  ,H293629 ,H293693 ,H293756 ,H293819 ,H293882 ,H293946 ,H294009 ,H294072 ,H294136 ,H294199 ,H294261 ,H294324 ,H294388  ,H293630 ,H293694 ,H293758 ,H293820 ,H293883 ,H293947 ,H294010 ,H294073 ,H294137 ,H294200 ,H294262 ,H294325 ,H294389  ,H293631 ,H293695 ,H293759 ,H293821 ,H293884 ,H293948 ,H294011 ,H294074 ,H294139 ,H294201 ,H294263 ,H294326 ,H294390  ,H293632 ,H293696 ,H293760 ,H293822 ,H293885 ,H293949 ,H294012 ,H294075 ,H294140 ,H294202 ,H294264 ,H294328 ,H294391  ,H293633 ,H293697 ,H293761 ,H293823 ,H293886 ,H293950 ,H294013 ,H294076 ,H294141 ,H294203 ,H294265 ,H294329 ,H294392  ,H293634 ,H293698 ,H293762 ,H293824 ,H293887 ,H293951 ,H294014 ,H294077 ,H294142 ,H294204 ,H294266 ,H294330 ,H294393  ,H293635 ,H293699 ,H293763 ,H293825 ,H293888 ,H293952 ,H294015 ,H294078 ,H294143 ,H294205 ,H294267 ,H294331 ,H294394  ,H293636 ,H293700 ,H293764 ,H293826 ,H293889 ,H293953 ,H294016 ,H294079 ,H294144 ,H294206 ,H294268 ,H294332 ,H294395  ,H293637 ,H293701 ,H293765 ,H293827 ,H293890 ,H293954 ,H294017 ,H294080 ,H294145 ,H294207 ,H294269 ,H294333 ,H294396  ,H293638 ,H293702 ,H293766 ,H293828 ,H293891 ,H293955 ,H294018 ,H294081 ,H294146 ,H294208 ,H294270 ,H294334 ,H294397  ,H293639 ,H293703 ,H293767 ,H293829 ,H293892 ,H293956 ,H294020 ,H294083 ,H294147 ,H294209 ,H294271 ,H294335 ,H294398  ,H293640 ,H293704 ,H293768 ,H293830 ,H293893 ,H293957 ,H294021 ,H294084 ,H294148 ,H294210 ,H294272 ,H294336 ,H294399  ,H293641 ,H293705 ,H293769 ,H293831 ,H293894 ,H293958 ,H294022 ,H294085 ,H294149 ,H294211 ,H294273 ,H294337 ,H294400  ,H293642 ,H293706 ,H293770 ,H293832 ,H293895 ,H293959 ,H294023 ,H294086 ,H294150 ,H294212 ,H294274 ,H294338 ,H294401  ,H293643 ,H293707 ,H293771 ,H293833 ,H293896 ,H293960 ,H294024 ,H294087 ,H294151 ,H294213 ,H294275 ,H294339 ,H294402  ,H293644 ,H293708 ,H293772 ,H293834 ,H293897 ,H293961 ,H294025 ,H294088 ,H294152 ,H294214 ,H294276 ,H294340 ,H294403  ,H293645 ,H293709 ,H293773 ,H293835 ,H293898 ,H293962 ,H294026 ,H294089 ,H294153 ,H294215 ,H294277 ,H294341 ,H294404  ,H293646 ,H293710 ,H293774 ,H293836 ,H293899 ,H293963 ,H294027 ,H294090 ,H294154 ,H294216 ,H294278 ,H294342 ,H294405  ,H293648 ,H293712 ,H293775 ,H293837 ,H293900 ,H293964 ,H294028 ,H294091 ,H294155 ,H294217 ,H294279 ,H294343 ,H294406  ,H293649 ,H293713 ,H293776 ,H293838 ,H293901 ,H293965 ,H294029 ,H294092 ,H294156 ,H294218 ,H294280 ,H294345 ,H294407  ,H293650 ,H293714 ,H293777 ,H293839 ,H293902 ,H293966 ,H294030 ,H294093 ,H294157 ,H294219 ,H294281 ,H294346 ,H294408  ,H293651 ,H293715 ,H293778 ,H293840 ,H293903 ,H293967 ,H294031 ,H294094 ,H294158 ,H294220 ,H294283 ,H294347 ,H294409  ,H293652 ,H293716 ,H293779 ,H293841 ,H293904 ,H293968 ,H294032 ,H294095 ,H294159 ,H294221 ,H294284 ,H294348 ,H294410  ,H293653 ,H293717 ,H293780 ,H293842 ,H293905 ,H293969 ,H294033 ,H294097 ,H294160 ,H294222 ,H294285 ,H294349 ,H294411  ,H293654 ,H293718 ,H293781 ,H293843 ,H293906 ,H293970 ,H294034 ,H294098 ,H294161 ,H294223 ,H294286 ,H294350 ,H294412  ,H293655 ,H293719 ,H293782 ,H293844 ,H293907 ,H293971 ,H294035 ,H294099 ,H294162 ,H294224 ,H294287 ,H294351 ,H294413  ,H293656 ,H293720 ,H293783 ,H293845 ,H293908 ,H293972 ,H294036 ,H294100 ,H294163 ,H294225 ,H294288 ,H294352 ,H294414  ,H293657 ,H293721 ,H293784 ,H293846 ,H293909 ,H293974 ,H294037 ,H294101 ,H294164 ,H294226 ,H294289 ,H294353 ,H294415  ,H293658 ,H293722 ,H293785 ,H293847 ,H293910 ,H293975 ,H294038 ,H294102 ,H294165 ,H294227 ,H294290 ,H294354 ,H294416  ,H293659 ,H293723 ,H293786 ,H293848 ,H293911 ,H293976 ,H294039 ,H294103 ,H294166 ,H294228 ,H294291 ,H294355 ,H294417  ,H293660 ,H293724 ,H293787 ,H293849 ,H293912 ,H293977 ,H294040 ,H294104 ,H294167 ,H294229 ,H294292 ,H294356 ,H294418  ,H293661 ,H293725 ,H293788 ,H293850 ,H293913 ,H293978 ,H294041 ,H294105 ,H294168 ,H294230 ,H294293 ,H294357 ,H294419  ,H293662 ,H293726 ,H293789 ,H293851 ,H293914 ,H293979 ,H294042 ,H294106 ,H294169 ,H294231 ,H294294 ,H294358 ,H294420  ,H293663 ,H293727 ,H293790 ,H293852 ,H293915 ,H293980 ,H294043 ,H294107 ,H294170 ,H294232 ,H294295 ,H294359 ,H294421  ,H293664 ,H293728 ,H293791 ,H293854 ,H293916 ,H293981 ,H294044 ,H294108 ,H294171 ,H294233 ,H294296 ,H294360 ,H294422  ,H293665 ,H293729 ,H293792 ,H293855 ,H293917 ,H293982 ,H294045 ,H294109 ,H294172 ,H294234 ,H294297 ,H294361 ,H294423  ,H293666 ,H293730 ,H293793 ,H293856 ,H293918 ,H293983 ,H294046 ,H294110 ,H294173 ,H294235 ,H294298 ,H294362 ,H294424  ,H293667 ,H293731 ,H293794 ,H293857 ,H293919 ,H293984 ,H294047 ,H294111 ,H294174 ,H294236 ,H294299 ,H294363 ,H294425  ,H293668 ,H293732 ,H293795 ,H293858 ,H293920 ,H293985 ,H294048 ,H294112 ,H294175 ,H294237 ,H294300 ,H294364 ,H294426  ,H293669 ,H293733 ,H293796 ,H293859 ,H293921 ,H293986 ,H294049 ,H294113 ,H294176 ,H294238 ,H294301 ,H294365 ,H294427  ,H293670 ,H293734 ,H293797 ,H293860 ,H293922 ,H293987 ,H294050 ,H294114 ,H294177 ,H294239 ,H294302 ,H294366 ,H294428  ,H293671 ,H293735 ,H293798 ,H293861 ,H293923 ,H293988 ,H294051 ,H294115 ,H294178 ,H294240 ,H294303 ,H294367 ,H294429  ,H293672 ,H293736 ,H293799 ,H293862 ,H293924 ,H293989 ,H294052 ,H294116 ,H294179 ,H294241 ,H294304 ,H294368 ,H294430  ,H293673 ,H293737 ,H293800 ,H293863 ,H293925 ,H293990 ,H294053 ,H294117 ,H294180 ,H294242 ,H294305 ,H294369 ,H294431     ,H293674 ,H293738 ,H293801 ,H293864 ,H293926 ,H293991 ,H294054 ,H294118 ,H294181 ,H294243 ,H294306 ,H294370 ,H294432  ,H293675 ,H293739 ,H293802 ,H293865 ,H293927 ,H293992 ,H294055 ,H294119 ,H294182 ,H294244 ,H294307 ,H294371 ,H294433  ,H293676 ,H293740 ,H293803 ,H293866 ,H293928 ,H293993 ,H294056 ,H294120 ,H294183 ,H294245 ,H294308 ,H294372 ,H294434  ,H293677 ,H293741 ,H293804 ,H293867 ,H293929 ,H293994 ,H294057 ,H294121 ,H294184 ,H294246 ,H294309 ,H294373 ,H294435  ,H293678 ,H293742 ,H293805 ,H293868 ,H293930 ,H293995 ,H294058 ,H294122 ,H294185 ,H294247 ,H294310 ,H294374 ,H294436  ,H293679 ,H293743 ,H293806 ,H293869 ,H293931 ,H293996 ,H294059 ,H294123 ,H294186 ,H294248 ,H294311 ,H294375 ,H294437  ,H293681 ,H293744 ,H293807 ,H293870 ,H293933 ,H293997 ,H294060 ,H294124 ,H294187 ,H294249 ,H294312 ,H294376 ,H294438  ,H293682 ,H293745 ,H293808 ,H293871 ,H293934 ,H293998 ,H294061 ,H294125 ,H294188 ,H294250 ,H294313 ,H294377 ,H294439  ,H293683 ,H293746 ,H293809 ,H293872 ,H293935 ,H293999 ,H294062 ,H294126 ,H294189 ,H294251 ,H294314 ,H294378 ,H294440  ,H293684 ,H293747 ,H293810 ,H293873 ,H293936 ,H294000 ,H294063 ,H294127 ,H294190 ,H294252 ,H294315 ,H294379 ,H294442  ,H293685 ,H293748 ,H293811 ,H293874 ,H293937 ,H294001 ,H294064 ,H294128 ,H294191 ,H294253 ,H294316 ,H294380 ,H294443  ,H294444 ,H294508 ,H294577 ,H294640 ,H294702 ,H294768 ,H294832 ,H294897 ,H294963 ,H295030 ,H295097 ,H295170 ,H295275  ,H294445 ,H294509 ,H294578 ,H294641 ,H294704 ,H294769 ,H294833 ,H294898 ,H294965 ,H295031 ,H295099 ,H295171 ,H295276  ,H294446 ,H294510 ,H294579 ,H294642 ,H294705 ,H294770 ,H294834 ,H294899 ,H294966 ,H295032 ,H295100 ,H295172 ,H295277  ,H294447 ,H294511 ,H294580 ,H294643 ,H294706 ,H294771 ,H294835 ,H294900 ,H294967 ,H295033 ,H295101 ,H295173 ,H295279  ,H294448 ,H294512 ,H294581 ,H294644 ,H294707 ,H294772 ,H294836 ,H294901 ,H294969 ,H295034 ,H295102 ,H295174 ,H295281  ,H294449 ,H294513 ,H294582 ,H294645 ,H294708 ,H294773 ,H294837 ,H294902 ,H294970 ,H295035 ,H295103 ,H295176 ,H295282  ,H294450 ,H294514 ,H294583 ,H294646 ,H294709 ,H294774 ,H294838 ,H294903 ,H294971 ,H295036 ,H295105 ,H295177 ,H295284  ,H294451 ,H294515 ,H294584 ,H294647 ,H294710 ,H294775 ,H294839 ,H294904 ,H294972 ,H295037 ,H295106 ,H295178 ,H295287  ,H294452 ,H294516 ,H294585 ,H294648 ,H294711 ,H294776 ,H294840 ,H294905 ,H294974 ,H295038 ,H295107 ,H295179 ,H295290  ,H294454 ,H294517 ,H294586 ,H294649 ,H294712 ,H294777 ,H294841 ,H294906 ,H294975 ,H295039 ,H295108 ,H295180 ,H295291  ,H294455 ,H294518 ,H294587 ,H294650 ,H294713 ,H294778 ,H294842 ,H294907 ,H294976 ,H295040 ,H295109 ,H295181 ,H295299  ,H294457 ,H294519 ,H294588 ,H294651 ,H294714 ,H294779 ,H294843 ,H294908 ,H294977 ,H295041 ,H295110 ,H295182 ,H295300  ,H294458 ,H294520 ,H294589 ,H294652 ,H294715 ,H294780 ,H294844 ,H294910 ,H294978 ,H295042 ,H295111 ,H295183 ,H295301  ,H294459 ,H294521 ,H294590 ,H294653 ,H294716 ,H294781 ,H294845 ,H294911 ,H294979 ,H295043 ,H295112 ,H295184  ,H294460 ,H294522 ,H294591 ,H294654 ,H294717 ,H294782 ,H294846 ,H294912 ,H294980 ,H295044 ,H295114 ,H295185  ,H294461 ,H294523 ,H294592 ,H294655 ,H294718 ,H294783 ,H294847 ,H294914 ,H294981 ,H295045 ,H295116 ,H295186  ,H294462 ,H294524 ,H294593 ,H294656 ,H294719 ,H294784 ,H294848 ,H294915 ,H294982 ,H295046 ,H295117 ,H295188  ,H294463 ,H294525 ,H294594 ,H294657 ,H294720 ,H294785 ,H294850 ,H294916 ,H294983 ,H295047 ,H295118 ,H295193  ,H294464 ,H294526 ,H294595 ,H294658 ,H294723 ,H294786 ,H294851 ,H294917 ,H294984 ,H295048 ,H295119 ,H295195  ,H294465 ,H294527 ,H294596 ,H294659 ,H294724 ,H294787 ,H294853 ,H294918 ,H294986 ,H295049 ,H295120 ,H295196  ,H294466 ,H294528 ,H294597 ,H294660 ,H294726 ,H294788 ,H294854 ,H294919 ,H294987 ,H295050 ,H295121 ,H295199  ,H294467 ,H294529 ,H294598 ,H294661 ,H294727 ,H294789 ,H294855 ,H294920 ,H294988 ,H295051 ,H295122 ,H295201  ,H294468 ,H294530 ,H294599 ,H294662 ,H294728 ,H294790 ,H294856 ,H294921 ,H294989 ,H295053 ,H295123 ,H295202  ,H294469 ,H294531 ,H294600 ,H294663 ,H294729 ,H294791 ,H294857 ,H294922 ,H294990 ,H295054 ,H295124 ,H295203  ,H294470 ,H294532 ,H294601 ,H294664 ,H294730 ,H294792 ,H294858 ,H294923 ,H294991 ,H295056 ,H295125 ,H295206  ,H294471 ,H294533 ,H294602 ,H294665 ,H294731 ,H294793 ,H294859 ,H294924 ,H294992 ,H295057 ,H295126 ,H295207  ,H294472 ,H294534 ,H294603 ,H294666 ,H294732 ,H294794 ,H294860 ,H294925 ,H294993 ,H295058 ,H295127 ,H295208  ,H294473 ,H294535 ,H294604 ,H294667 ,H294733 ,H294795 ,H294861 ,H294926 ,H294995 ,H295059 ,H295128 ,H295209  ,H294474 ,H294536 ,H294605 ,H294668 ,H294734 ,H294796 ,H294862 ,H294927 ,H294996 ,H295060 ,H295129 ,H295210  ,H294475 ,H294537 ,H294606 ,H294669 ,H294735 ,H294797 ,H294863 ,H294928 ,H294997 ,H295061 ,H295130 ,H295211  ,H294476 ,H294538 ,H294607 ,H294670 ,H294736 ,H294798 ,H294864 ,H294929 ,H294998 ,H295062 ,H295131 ,H295212  ,H294477 ,H294539 ,H294608 ,H294671 ,H294737 ,H294799 ,H294865 ,H294930 ,H294999 ,H295063 ,H295132 ,H295213  ,H294478 ,H294540 ,H294610 ,H294672 ,H294738 ,H294800 ,H294866 ,H294931 ,H295000 ,H295064 ,H295133 ,H295214  ,H294479 ,H294541 ,H294611 ,H294673 ,H294739 ,H294801 ,H294867 ,H294932 ,H295001 ,H295065 ,H295134 ,H295216  ,H294480 ,H294542 ,H294612 ,H294674 ,H294740 ,H294802 ,H294868 ,H294933 ,H295002 ,H295066 ,H295135 ,H295217  ,H294481 ,H294543 ,H294613 ,H294675 ,H294741 ,H294803 ,H294869 ,H294934 ,H295003 ,H295067 ,H295136 ,H295221  ,H294482 ,H294544 ,H294614 ,H294676 ,H294742 ,H294804 ,H294870 ,H294935 ,H295004 ,H295068 ,H295138 ,H295223  ,H294483 ,H294545 ,H294615 ,H294677 ,H294743 ,H294806 ,H294871 ,H294936 ,H295005 ,H295069 ,H295139 ,H295224  ,H294484 ,H294546 ,H294616 ,H294678 ,H294744 ,H294807 ,H294872 ,H294937 ,H295006 ,H295070 ,H295140 ,H295225  ,H294485 ,H294547 ,H294617 ,H294679 ,H294745 ,H294808 ,H294874 ,H294938 ,H295007 ,H295071 ,H295141 ,H295226  ,H294486 ,H294548 ,H294618 ,H294680 ,H294746 ,H294809 ,H294875 ,H294940 ,H295008 ,H295072 ,H295142 ,H295227  ,H294487 ,H294549 ,H294619 ,H294681 ,H294747 ,H294810 ,H294876 ,H294941 ,H295009 ,H295073 ,H295143 ,H295228  ,H294488 ,H294550 ,H294620 ,H294682 ,H294748 ,H294811 ,H294877 ,H294942 ,H295010 ,H295074 ,H295144 ,H295229  ,H294489 ,H294551 ,H294621 ,H294683 ,H294749 ,H294812 ,H294878 ,H294943 ,H295011 ,H295075 ,H295145 ,H295231  ,H294490 ,H294552 ,H294622 ,H294684 ,H294750 ,H294813 ,H294879 ,H294945 ,H295012 ,H295076 ,H295146 ,H295235  ,H294491 ,H294553 ,H294623 ,H294685 ,H294751 ,H294814 ,H294880 ,H294946 ,H295013 ,H295077 ,H295147 ,H295236  ,H294492 ,H294554 ,H294624 ,H294686 ,H294752 ,H294815 ,H294881 ,H294947 ,H295014 ,H295078 ,H295148 ,H295238  ,H294493 ,H294555 ,H294625 ,H294687 ,H294753 ,H294816 ,H294882 ,H294948 ,H295015 ,H295080 ,H295149 ,H295239  ,H294494 ,H294556 ,H294626 ,H294688 ,H294754 ,H294817 ,H294883 ,H294949 ,H295016 ,H295081 ,H295150 ,H295240  ,H294495 ,H294557 ,H294627 ,H294689 ,H294755 ,H294818 ,H294884 ,H294950 ,H295017 ,H295082 ,H295151 ,H295241  ,H294496 ,H294564 ,H294628 ,H294690 ,H294756 ,H294820 ,H294885 ,H294951 ,H295018 ,H295083 ,H295152 ,H295242  ,H294497 ,H294565 ,H294629 ,H294691 ,H294757 ,H294821 ,H294886 ,H294952 ,H295019 ,H295084 ,H295154 ,H295244  ,H294498 ,H294567 ,H294630 ,H294692 ,H294758 ,H294822 ,H294887 ,H294953 ,H295020 ,H295085 ,H295155 ,H295245  ,H294499 ,H294568 ,H294631 ,H294693 ,H294759 ,H294823 ,H294888 ,H294954 ,H295021 ,H295086 ,H295156 ,H295246  ,H294500 ,H294569 ,H294632 ,H294694 ,H294760 ,H294824 ,H294889 ,H294955 ,H295022 ,H295087 ,H295157 ,H295247  ,H294501 ,H294570 ,H294633 ,H294695 ,H294761 ,H294825 ,H294890 ,H294956 ,H295023 ,H295089 ,H295158 ,H295248  ,H294502 ,H294571 ,H294634 ,H294696 ,H294762 ,H294826 ,H294891 ,H294957 ,H295024 ,H295090 ,H295159 ,H295249      ,H294503 ,H294572 ,H294635 ,H294697 ,H294763 ,H294827 ,H294892 ,H294958 ,H295025 ,H295091 ,H295160 ,H295250  ,H294504 ,H294573 ,H294636 ,H294698 ,H294764 ,H294828 ,H294893 ,H294959 ,H295026 ,H295093 ,H295161 ,H295252  ,H294505 ,H294574 ,H294637 ,H294699 ,H294765 ,H294829 ,H294894 ,H294960 ,H295027 ,H295094 ,H295162 ,H295254  ,H294506 ,H294575 ,H294638 ,H294700 ,H294766 ,H294830 ,H294895 ,H294961 ,H295028 ,H295095 ,H295163 ,H295255  ,H294507 ,H294576 ,H294639 ,H294701 ,H294767 ,H294831 ,H294896 ,H294962 ,H295029 ,H295096 ,H295169 ,H295265   
Recalling Firm/
Manufacturer
Varian Medical Systems, Inc. Oncology Systems
911 Hansen Way
Palo Alto CA 94304-1028
For Additional Information Contact Peter J. Coronado
650-424-5731
Manufacturer Reason
for Recall
An event has been reported to Varian which entails excessive connector resistance, which caused the actual jaw positions to differ from the intended jaw positions without warning the operator.
FDA Determined
Cause 2
Device Design
Action Varian Medical Systems sent an Urgent Medical Device Correction - Urgent Field Safety Notice dated November 28, 2011 ***Expanded to include product #2 on December 16, 2011***, to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to establish a comprehensive quality assurance program before beginning treatments for the day, quality assurance tests must be performed. Additionally, any discrepancies in imaging field sizes vs planned field size should be brought to the attention of the medical physicist for investigation before treatment continues. Customers were asked to complete the attached Proof of Notification or Receipt Verification Card once they have read the document and return it to Varian Medical Systems. For question regarding this recall: USA and Canada 1-888-827-4265 Europe +41 41 749 8844
Quantity in Commerce 6158 units total
Distribution Worldwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = IYE and Original Applicant = VARIAN MEDICAL SYSTEMS, ONCOLOGY SYSTEMS
-
-