 Class 2 Device Recall TruLight ALC, Trulight 5520 or 5320 lighting system.see related information
Date Initiated by FirmJanuary 09, 2012
Date PostedFebruary 13, 2012
Recall Status1 Terminated 3 on October 09, 2012
Recall NumberZ-1010-2012
Recall Event ID 61004
510(K)NumberK102758 
Product Classification Light, surgical, ceiling mounted - Product Code FSY
ProductSurgical lighting systems iLED and TruLight in use with the ALC+ function. The TRUMPF surgical lights are intended to locally illuminate an operating or examination area of the patients body with a high intensity light.
Code Information Material number 1583966
FEI Number 3003184737
Recalling Firm/
Manufacturer
Trumpf Medical Systems, Inc.
415 Jessen Ln
Charleston SC 29492
For Additional Information ContactLindsey Ronnenberg
843-534-0606
Manufacturer Reason
for Recall
The ALC+ function may not be properly locked during use resulting in the possibility of the handle falling off into the surgical field causing injury to the patient.
FDA Determined
Cause 2
Device Design
ActionThe firm, TRUMPF Medical Systems, Inc., sent an "Urgent Safety Information" notice/package dated January 9, 2012 to its customers. (Note: Some were delivered by TRUMPF representatives in person and some via mail with Urgent Safety identification on the envelope). The notice describes the product, problem and actions to be taken. TRUMPF will replace all sterilizable handles of this type in the field with new handles, in addition, TRUMPF will send new handles and verify disposal of the old version at the customers location. Distribution of new handles will start January 2012. The customers were instructed to ensure that all users and other relevent persons are informed of the Urgent Safety Information; store the information until the measure has been completed and complete and return the Confirmation of User Receipt form via mail to TRUMPF Medical Systems, Inc. 1046 LeGrand Blvd., Charleston, SC 29063 or email: Lindsey.ronnenberg@us.trumpf-med.com. If you have any questions, call 843-822-6939.
Quantity in Commerce8,733 units
DistributionWorldwide distribution: USA (nationwide) including states of: AL, CA, CO, CT, DE, FL, GA, IL, IN, MA, MD, MO, NC, NM, NJ, NY, OH, PA, TX, VA, VT, WA, and WI; and countries of: Australia; Austria, Belguim, China, Cyprus, Denmark, Estonia, Finland, France, Germany, Great Britain, Hong Kong, Indonesia, India, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea, Latvia, Luxembourg, Mexico, Netherlands, Norway, Pakistan, Peru, Philippines, Poland, Romania, Russian Federation, Saudi Arabia, Seychelles, Singapore, South Africa (Zuid Afirka), Spain, Sweden, Switzerland (Confederation of Helvetia), Taiwan, Thailand, Tunisia, Turkey, Uruguay, Venezuela, and Viet Nam.
Total Product Life CycleTPLC Device Report