• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall SoundBite Microphone Tube

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall SoundBite Microphone Tubesee related information
Date Initiated by FirmFebruary 02, 2012
Date PostedMarch 07, 2012
Recall Status1 Terminated 3 on March 23, 2012
Recall NumberZ-1183-2012
Recall Event ID 61114
510(K)NumberK110831 
Product Classification Hearing aid, bone conduction - Product Code LXB
ProductSoundBite Hearing System; Microphone Tube Size 4. For use in patients with moderately severe, severe or profound sensorineural hearing loss in one ear and normal hearing in the other or patients with conductive hearing loss where the pure tone average bond-conduction hearing threshold is better than or equal to 25 dB HL. Sonitus Medical, San Mateo, CA.
Code Information Model number P02019 - Right, lot M1102 and P020020 - Left, lot M1105.
FEI Number 3009200826
Recalling Firm/
Manufacturer
Sonitus Medical Inc.
1825 S Grant St Ste 350
San Mateo CA 94402-7033
For Additional Information ContactViorica Filimon
650-251-4185
Manufacturer Reason
for Recall
Sonitas received reports of stress cracking on the SoundBite Microphone Tube.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionLetters were sent to all direct SoundBite customers on February 9, 2012 via certified mail. Customers were asked to remove the old microphone tubes and replace with the new enclosed microphone tubes. Return mail instructions were provided for the old microphone tubes. Customer questions were directed to 650-251-4185.
Quantity in Commerce676 microphones, all sizes.
DistributionClass III Recall - Nationwide Distribution, including the states of CA, AZ, PA, MI, UT, NC, NY, WI, TX, FL and DC.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = LXB
-
-