| | Class 2 Device Recall SenDx Medical Inc. ABL80FLEX |  |
| Date Initiated by Firm | January 30, 2012 |
| Date Posted | April 09, 2012 |
| Recall Status1 |
Terminated 3 on July 13, 2012 |
| Recall Number | Z-1384-2012 |
| Recall Event ID |
61178 |
| 510(K)Number | K051804 |
| Product Classification |
Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
|
| Product | SenDx Medical Inc
ABL80FLEX
Blood Gas, pH, Electrolyte, and Metabolite System
The ABL80 FLEX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, and hematocrit in whole blood. The ABL80 FLEX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. |
| Code Information |
Model number: 393-839 Serial number(s): 304157, 304160 and 304162 to 305414 for ABL80 FLEX analyzers |
Recalling Firm/ Manufacturer |
Sendx Medical Inc 1945 Palomar Oaks Way Carlsbad CA 92011-1300
|
| For Additional Information Contact | 760-603-6300 |
Manufacturer Reason for Recall | The product can cause the analyzer to report values outside the published performance specifications for pH, pCO2, cNa+, cK+, cCa2+ and cCl-. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | SenDx Medical sent Customers notification letters dated January 30th, 2012, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer.
We have recently become aware of a possible condition in a limited number of ABLBO FLEX analyzers configured with a specific Control PCe lerinted Circuit Board), which can cause the analyzer to report values outside the published performance specifications for pH, pCOz, cNa*, cK*, cCa2* and cCl-.
Customers were instructed to immediately implement Alternative 1 or 2 until the signal data PCB is replaced.
Alterative I:
Stop the use of all affected analyzers until corrective actions can be performed on these analyzers.
Alternative 2:
Verify proper function of the affected analyzers immediately following every
patient sample measurement by measuring a manual QC sample.
Customers were asked to complete the fax form acknowledging receipt of the notice and provided a contact for Technical Support at 1-800-736-0600 opt.4. |
| Quantity in Commerce | 819 analyzers and 44 signal data PCBs |
| Distribution | One US consignee in state of OH. Other consignees are international such as Australia, France, Netherlands, Denmark, China, United Kingdom, Turkey, Switzerland, Japan, India, Poland, Spain, and Germany. Distribution list submitted with supporting doc.s |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CHL
|
|
|
|