| | Class 2 Device Recall LUMEX A GrahamField Brand, Bed End Sert |  |
| Date Initiated by Firm | April 17, 2012 |
| Date Posted | May 11, 2012 |
| Recall Status1 |
Terminated 3 on July 15, 2013 |
| Recall Number | Z-1536-2012 |
| Recall Event ID |
61527 |
| Product Classification |
Bed, therapeutic, ac-powered, adjustable home-use - Product Code LLI
|
| Product | LUMEX A Graham-Field Brand, Bed End Sert. Low/Full-electric, Plastic Panels, Item 690-7604-000 Manufactured for : GF Heath Products, Inc., Atlanta, Georgia 30360 Made in China.
LUMEX A Graham-Field Brand, Bed End Sert. Low/Full-electric, Walnut Panels, Item 690-7004-000 Manufactured for : GF Heath Products, Inc., Atlanta, Georgia 30360 Made in China.
Therapeutic bed intended for home use. |
| Code Information |
Lot #'s: A0309, A0501, A0609, B0309, B0501, B0609, C0309,C0501 , D0309,E0309, F0309 Lot/serial number: A0309: 6970400900001 6970400900002 6970400900003 6970400900004 6970400900005 6970400900006 6970400900007 6970400900008 6970400900009 6970400900010 6970400900011 6970400900012 6970400900013 6970400900014 6970400900015 6970400900016 6970400900017 6970400900018 6970400900019 6970400900020 6970400900021 6970400900022 6970400900023 6970400900024 6970400900025 6970400900026 6970400900027 6970400900028 6970400900029 6970400900030 6970400900031 6970400900032 6970400900033 6970400900034 6970400900035 6970400900036 6970400900037 6970400900038 6970400900039 6970400900040 6970400900041 6970400900042 6970400900043 6970400900044 6970400900045 6970400900046 6970400900047 6970400900048 6970400900049 6970400900050 6970400900051 6970400900052 6970400900053 6970400900054 6970400900055 6970400900056 6970400900057 6970400900058 6970400900059 6970400900060 6970400900061 6970400900062 6970400900063 6970400900064 6970400900065 6970400900066 6970400900067 6970400900069 6970400900070 6970400900071 6970400900072 6970400900073 6970400900074 6970400900075 6970400900076 6970400900077 6970400900078 6970400900079 6970400900080 6970400900081 6970400900082 6970400900084 6970400900085 6970400900086 6970400900087 6970400900088 6970400900089 6970400900090 6970400900091 6970400900092 6970400900093 6970400900094 6970400900095 6970400900096 6970400900097 6970400900098 6970400900099 6970400900100 6970400900101 6970400900102 6970400900103 6970400900104 6970400900105 6970400900106 6970400900107 6970400900108 6970400900109 6970400900110 6970400900111 6970400900112 6970400900113 6970400900114 6970400900115 6970400900116 6970400900118 6970400900119 6970400900121 6970400900122 6970400900123 6970400900124 6970400900124 6970400900126 6970400900126 6970400900127 6970400900128 6970400900129 6970400900130 6970400900131 6970400900132 6970400900133 6970400900134 6970400900135 6970400900136 6970400900137 6970400900138 6970400900138 6970400900139 6970400900139 6970400900141 6970400900143 6970400900144 6970400900145 6970400900147 6970400900148 6970400900149 6970400900150 6976040900003 6976040900004 6976040900005 6976040900006 6976040900007 6976040900008 6976040900009 6976040900010 6976040900011 6976040900012 6976040900013 6976040900014 6976040900016 6976040900017 6976040900018 6976040900019 6976040900020 6976040900021 6976040900022 6976040900023 6976040900024 6976040900025 6976040900026 6976040900027 6976040900028 6976040900029 6976040900030 6976040900031 6976040900032 6976040900033 6976040900034 6976040900035 6976040900036 6976040900037 6976040900038 6976040900039 6976040900040 6976040900041 6976040900042 6976040900043 6976040900044 6976040900045 6976040900046 6976040900047 6976040900048 6976040900049 6976040900050 6976040900051 6976040900052 6976040900053 6976040900054 6976040900055 6976040900056 6976040900057 6976040900058 6976040900059 6976040900060 6976040900061 6976040900062 6976040900063 6976040900064 6976040900065 6976040900066 6976040900067 6976040900068 6976040900069 6976040900070 6976040900071 6976040900072 6976040900073 6976040900074 6976040900075 6976040900076 6976040900077 6976040900078 6976040900079 6976040900080 6976040900081 6976040900082 6976040900083 6976040900084 6976040900085 6976040900086 6976040900087 6976040900088 6976040900089 6976040900090 6976040900091 6976040900092 6976040900093 6976040900094 6976040900095 6976040900096 6976040900097 6976040900098 6976040900099 6976040900100 6976040900101 6976040900102 6976040900103 6976040900104 6976040900105 6976040900106 6976040900107 6976040900108 6976040900109 6976040900110 6976040900111 6976040900112 6976040900113 6976040900114 6976040900115 6976040900116 6976040900117 6976040900118 6976040900119 6976040900120 6976040900121 6976040900122 6976040900123 6976040900123 6976040900124 6976040900125 6976040900126 6976040900127 6976040900128 6976040900129 6976040900130 6976040900131 6976040900132 6976040900133 6976040900134 6976040900135 6976040900136 6976040900137 6976040900138 6976040900139 6976040900140 6976040900141 6976040900142 6976040900143 6976040900144 6976040900145 6976040900146 6976040900147 6976040900148 6976040900149 6976040900150 A0501: 6970400901021 6970400901022 6970400901023 6970400901024 6970400901025 6970400901026 6970400901027 6970400901029 6970400901030 6970400901031 6970400901032 6970400901033 6970400901034 6970400901035 6970400901036 6970400901037 6970400901038 6970400901039 6970400901040 6970400901041 6970400901042 6970400901043 6970400901044 6970400901045 6970400901047 6970400901048 6970400901049 6970400901050 6970400901051 6970400901052 6970400901054 6970400901055 6970400901056 6970400901057 6970400901058 6970400901059 6970400901060 6970400901061 6970400901063 6970400901064 6970400901065 6970400901066 6970400901067 6970400901069 6970400901070 6970400901071 6970400901073 6970400901074 6970400901075 6970400901076 6970400901077 6970400901078 6970400901079 6970400901080 6970400901081 6970400901082 6970400901083 6970400901084 6970400901085 6970400901086 6970400901087 6970400901088 6970400901089 6970400901090 6970400901092 6970400901094 6970400901095 6970400901096 6970400901101 6970400901102 6970400901103 6970400901104 6970400901105 6970400901106 6970400901107 6970400901108 6970400901110 6970400901111 6970400901113 6970400901114 6970400901115 6970400901117 6970400901118 6970400901119 6970400901120 6970400901121 6970400901122 6970400901123 6970400901125 6970400901126 6970400901127 6970400901128 6970400901129 6970400901130 6970400901131 6970400901132 6970400901133 6970400901136 6970400901137 6970400901139 6970400901140 6970400901141 6970400901142 6970400901143 6970400901144 6970400901145 6970400901146 6970400901149 6970400901150 6970400901151 6970400901152 6970400901153 6970400901154 6970400901155 6970400901156 6970400901157 6970400901158 6970400901159 6970400901161 6970400901163 6970400901164 6970400901165 6970400901166 6970400901167 6970400901168 6970400901169 6970400901170 6976040900992 6976040900993 6976040900994 6976040900995 6976040900996 6976040900997 6976040900998 6976040900999 6976040901000 6976040901001 6976040901003 6976040901004 6976040901007 6976040901008 6976040901010 6976040901012 6976040901013 6976040901014 6976040901018 6976040901020 6976040901021 6976040901022 6976040901023 6976040901024 6976040901025 6976040901026 6976040901027 6976040901028 6976040901029 6976040901030 6976040901031 6976040901032 6976040901033 6976040901034 6976040901035 6976040901036 6976040901037 6976040901038 6976040901039 6976040901040 6976040901041 6976040901046 6976040901047 6976040901048 6976040901050 6976040901051 6976040901052 6976040901053 6976040901054 6976040901055 6976040901056 6976040901058 6976040901059 6976040901060 6976040901061 6976040901062 6976040901064 6976040901065 6976040901067 6976040901068 6976040901070 6976040901074 6976040901075 6976040901076 6976040901077 6976040901078 6976040901080 6976040901082 6976040901083 6976040901085 6976040901087 6976040901088 6976040901090 6976040901092 6976040901093 6976040901095 6976040901096 6976040901097 6976040901098 6976040901099 6976040901100 6976040901101 6976040901102 6976040901103 6976040901104 6976040901105 6976040901106 6976040901107 6976040901108 6976040901109 6976040901110 6976040901111 6976040901122 6976040901123 6976040901124 6976040901125 6976040901126 6976040901127 6976040901128 6976040901129 6976040901130 6976040901131 6976040901132 6976040901133 6976040901134 6976040901135 6976040901136 6976040901137 6976040901138 6976040901139 6976040901140 6976040901141 A0609: 6970400901446 6970400901447 6970400901448 6970400901449 6970400901450 6970400901451 6970400901452 6970400901453 6970400901454 6970400901455 6970400901456 6970400901457 6970400901458 6970400901459 6970400901461 6970400901462 6970400901464 6970400901465 6970400901466 6970400901467 6970400901468 6970400901469 6970400901470 6970400901471 6970400901472 6970400901473 6970400901474 6970400901475 6970400901476 6970400901478 6970400901479 6970400901480 6970400901481 6970400901482 6970400901483 6970400901484 6970400901486 6970400901487 6970400901488 6970400901490 6970400901491 6970400901492 6970400901493 6970400901494 6970400901495 6970400901496 6970400901497 6970400901499 6970400901500 6970400901501 6970400901502 6970400901503 6970400901505 6970400901506 6970400901507 6970400901508 6970400901509 6970400901510 6970400901511 6970400901512 6970400901513 6970400901514 6970400901516 6970400901517 6970400901520 6970400901521 6970400901522 6970400901523 6970400901524 6970400901525 6970400901526 6970400901527 6970400901528 6970400901530 6970400901532 6970400901533 6970400901534 6970400901535 6970400901536 6970400901537 6970400901538 6970400901539 6970400901540 6970400901541 6970400901542 6970400901543 6970400901544 6970400901545 6976040901419 6976040901420 6976040901421 6976040901422 6976040901423 6976040901424 6976040901425 6976040901426 6976040901444 6976040901448 6976040901449 B0309: 6970400900151 6970400900152 6970400900153 6970400900154 6970400900155 6970400900156 6970400900157 6970400900158 6970400900159 6970400900160 6970400900161 6970400900162 6970400900163 6970400900164 6970400900165 6970400900168 6970400900169 6970400900170 6970400900171 6970400900172 6970400900173 6970400900174 6970400900175 6970400900176 6970400900177 6970400900178 6970400900179 6970400900180 6970400900182 6970400900183 6970400900184 6970400900185 6970400900186 6970400900187 6970400900188 6970400900189 6970400900190 6970400900192 6970400900193 6970400900194 6970400900195 6970400900196 6970400900197 6970400900198 6970400900199 6970400900200 6970400900201 6970400900202 6970400900205 6970400900206 6970400900208 6970400900210 6970400900211 6970400900212 6970400900213 6970400900214 6970400900215 6970400900216 6970400900217 6970400900218 6970400900219 6970400900220 6970400900221 6970400900222 6970400900225 6970400900226 6970400900228 6970400900229 6970400900230 6970400900231 6970400900232 6970400900233 6970400900234 6970400900236 6970400900237 6970400900238 6970400900239 6970400900240 6970400900241 6970400900242 6970400900243 6970400900244 6970400900245 6970400900246 6970400900247 6970400900248 6970400900249 6970400900250 6970400900252 6970400900253 6970400900254 6970400900255 6970400900256 6970400900257 6970400900258 6970400900259 6970400900260 6970400900261 6970400900262 6970400900263 6970400900265 6970400900266 6970400900267 6970400900268 6970400900269 6970400900270 6970400900271 6970400900272 6970400900273 6970400900274 6970400900275 6970400900276 6970400900277 6970400900278 6970400900279 6970400900280 6970400900281 6970400900282 6970400900283 6970400900284 6970400900285 6970400900286 6970400900288 6970400900289 6970400900290 6970400900291 6970400900292 6970400900293 6970400900294 6970400900295 6970400900296 6970400900297 6970400900299 6970400900300 6970400900301 6970400900302 6970400900303 6970400900304 6970400900305 6970400900306 6970400900307 6970400900308 6970400900309 6970400900311 6970400900312 6970400900313 6970400900314 6970400900315 6970400900316 6970400900317 6970400900318 6970400900319 6970400900323 6970400900324 6970400900325 6970400900326 6970400900327 6970400900328 6970400900329 6970400900330 6970400900331 6970400900332 6970400900333 6970400900334 6970400900335 6970400900336 6970400900338 6970400900339 6970400900340 6970400900341 6970400900342 6970400900343 6970400900344 6970400900345 6970400900347 6970400900348 6970400900349 6976040900151 6976040900152 6976040900153 6976040900154 6976040900155 6976040900156 6976040900157 6976040900158 6976040900159 6976040900161 6976040900162 6976040900163 6976040900165 6976040900166 6976040900167 6976040900168 6976040900169 6976040900171 6976040900172 6976040900174 6976040900180 6976040900181 6976040900182 6976040900183 6976040900185 6976040900189 6976040900190 6976040900191 6976040900192 6976040900193 6976040900194 6976040900195 6976040900196 6976040900197 6976040900198 6976040900199 6976040900200 6976040900201 6976040900202 6976040900203 6976040900204 6976040900208 6976040900209 6976040900210 6976040900211 6976040900213 6976040900214 6976040900216 6976040900217 6976040900218 6976040900220 6976040900221 6976040900222 6976040900224 6976040900225 6976040900226 6976040900228 6976040900229 6976040900230 6976040900231 6976040900232 6976040900233 6976040900234 6976040900235 6976040900236 6976040900237 6976040900238 6976040900239 6976040900240 6976040900241 6976040900242 6976040900243 6976040900244 6976040900245 6976040900246 6976040900247 6976040900249 6976040900250 6976040900251 6976040900252 6976040900253 6976040900254 6976040900255 6976040900256 6976040900257 6976040900258 6976040900259 6976040900260 6976040900261 6976040900262 6976040900263 6976040900264 6976040900265 6976040900266 6976040900267 6976040900269 6976040900270 6976040900271 6976040900272 6976040900273 6976040900274 6976040900275 6976040900276 6976040900277 6976040900278 6976040900279 6976040900280 6976040900281 6976040900284 6976040900285 6976040900287 6976040900288 6976040900289 6976040900291 6976040900293 6976040900294 6976040900295 6976040900296 6976040900297 6976040900298 6976040900299 6976040900300 6976040900301 6976040900303 6976040900304 6976040900305 6976040900308 6976040900309 6976040900310 6976040900312 6976040900313 6976040900314 6976040900315 6976040900316 6976040900318 6976040900319 6976040900320 6976040900326 6976040900327 6976040900329 6976040900330 6976040900331 6976040900332 6976040900334 6976040900336 6976040900337 6976040900338 6976040900341 6976040900342 6976040900343 6976040900344 6976040900345 6976040900346 6976040900347 6976040900348 6976040900349 6976040900350 B0501: 6970400901171 6970400901172 6970400901173 6970400901174 6970400901175 6970400901176 6970400901177 6970400901178 6970400901179 6970400901180 6970400901181 6970400901182 6970400901184 6970400901186 6970400901187 6970400901188 6970400901189 6970400901190 6970400901192 6970400901193 6970400901194 6970400901195 6970400901196 6970400901197 6970400901198 6970400901200 6970400901201 6970400901202 6970400901203 6970400901204 6970400901205 6970400901207 6970400901208 6970400901209 6970400901210 6970400901212 6970400901213 6970400901214 6970400901215 6970400901216 6970400901217 6970400901218 6970400901220 6970400901221 6970400901223 6970400901224 6970400901225 6970400901226 6970400901227 6970400901228 6970400901229 6970400901230 6970400901231 6970400901232 6970400901233 6970400901234 6970400901235 6970400901236 6970400901238 6970400901239 6970400901240 6970400901241 6970400901242 6970400901243 6970400901244 6970400901247 6970400901249 6970400901250 6970400901251 6970400901255 6970400901256 6970400901257 6970400901258 6970400901259 6970400901260 6970400901261 6970400901262 6970400901263 6970400901264 6970400901266 6970400901267 6970400901269 6970400901271 6970400901274 6970400901275 6970400901276 6970400901277 6970400901278 6970400901280 6970400901281 6970400901282 6970400901283 6970400901284 6970400901285 6970400901286 6970400901289 6970400901290 6970400901291 6970400901294 6970400901295 6976040901175 6976040901178 6976040901180 6976040901181 6976040901185 6976040901186 6976040901216 6976040901217 6976040901218 6976040901219 6976040901220 6976040901222 6976040901223 6976040901224 6976040901225 6976040901226 6976040901227 6976040901228 6976040901229 6976040901230 6976040901231 6976040901232 6976040901233 6976040901262 B0609: 6976040901477 6976040901478 6976040901479 6976040901480 6976040901481 6976040901482 6976040901483 6976040901485 6976040901490 6976040901493 6976040901494 6976040901496 6976040901497 6976040901498 6976040901499 6976040901500 6976040901501 6976040901502 6976040901503 6976040901504 6976040901505 6976040901506 6976040901510 6976040901513 6976040901514 6976040901515 6976040901516 6976040901517 6976040901518 6976040901519 6976040901520 6976040901521 6976040901522 6976040901523 6976040901524 6976040901525 6976040901526 6976040901527 6976040901530 6976040901531 6976040901532 6976040901533 6976040901534 6976040901535 6976040901539 6976040901542 6976040901543 6976040901544 6976040901545 6976040901546 6976040901547 6976040901548 6976040901549 6976040901550 6976040901551 6976040901552 6976040901553 6976040901554 6976040901555 6976040901556 6976040901558 6976040901559 6976040901560 6976040901561 6976040901562 6976040901564 6976040901565 6976040901566 6976040901567 6976040901568 6976040901569 6976040901570 6976040901571 6976040901572 6976040901573 6976040901574 6976040901575 6976040901576 C0309: 6970400900351 6970400900352 6970400900353 6970400900354 6970400900355 6970400900356 6970400900357 6970400900358 6970400900359 6970400900360 6970400900361 6970400900363 6970400900364 6970400900365 6970400900366 6970400900367 6970400900368 6970400900369 6970400900370 6970400900371 6970400900372 6970400900373 6970400900374 6970400900376 6970400900377 6970400900378 6970400900379 6970400900380 6970400900382 6970400900383 6970400900384 6970400900385 6970400900386 6970400900387 6970400900388 6970400900389 6970400900391 6970400900392 6970400900393 6970400900394 6970400900395 6970400900396 6970400900397 6970400900398 6970400900400 6970400900401 6970400900402 6970400900403 6970400900404 6970400900405 6970400900406 6970400900407 6970400900408 6970400900409 6970400900410 6970400900411 6970400900412 6970400900413 6970400900414 6970400900415 6970400900416 6970400900417 6970400900418 6970400900419 6970400900420 6970400900421 6970400900422 6970400900423 6970400900425 6970400900426 6970400900427 6970400900428 6970400900429 6970400900431 6970400900432 6970400900433 6970400900434 6970400900435 6970400900436 6970400900437 6970400900438 6970400900439 6970400900440 6970400900441 6970400900442 6970400900443 6970400900444 6970400900445 6970400900446 6970400900447 6970400900448 6970400900449 6970400900450 6970400900451 6970400900452 6970400900453 6970400900454 6970400900456 6970400900457 6970400900458 6970400900459 6970400900460 6970400900462 6970400900463 6970400900464 6970400900465 6970400900466 6970400900467 6970400900468 6970400900469 6970400900470 6970400900472 6970400900473 6970400900474 6970400900475 6970400900476 6970400900477 6970400900478 6970400900479 6970400900480 6970400900481 6970400900482 6970400900483 6970400900485 6970400900486 6970400900487 6970400900488 6970400900489 6970400900491 6970400900492 6970400900493 6970400900494 6970400900495 6970400900496 6970400900497 6970400900498 6970400900499 6970400900500 6970400900501 6970400900502 6970400900503 6970400900504 6970400900505 6970400900506 6970400900507 6970400900508 6970400900509 6970400900510 6970400900511 6970400900512 6970400900513 6970400900514 6970400900515 6970400900516 6970400900517 6970400900518 6970400900519 6970400900520 6970400900521 6970400900522 6970400900523 6970400900524 6970400900525 6970400900526 6970400900527 6970400900528 6970400900529 6970400900530 6970400900531 6970400900532 6970400900533 6970400900534 6970400900535 6970400900536 6970400900537 6970400900538 6970400900539 6970400900540 6970400900541 6970400900542 6970400900543 6970400900544 6970400900545 6970400900546 6970400900547 6970400900548 6970400900549 6970400900550 6976040900352 6976040900355 6976040900356 6976040900358 6976040900359 6976040900360 6976040900362 6976040900363 6976040900364 6976040900365 6976040900366 6976040900367 6976040900368 6976040900369 6976040900373 6976040900374 6976040900375 6976040900376 6976040900378 6976040900381 6976040900384 6976040900385 6976040900386 6976040900387 6976040900388 6976040900399 6976040900400 6976040900401 6976040900402 6976040900403 6976040900404 6976040900405 6976040900406 6976040900407 6976040900420 6976040900431 6976040900438 6976040900441 6976040900442 6976040900443 6976040900444 6976040900445 6976040900446 6976040900447 6976040900448 6976040900449 6976040900450 6976040900451 6976040900452 6976040900453 6976040900454 6976040900455 6976040900456 6976040900457 6976040900458 6976040900459 6976040900460 6976040900466 6976040900467 6976040900468 6976040900470 6976040900475 6976040900477 6976040900478 6976040900479 6976040900485 6976040900491 6976040900512 6976040900513 6976040900518 6976040900519 6976040900520 6976040900521 6976040900522 6976040900525 6976040900529 6976040900530 6976040900531 6976040900532 6976040900536 6976040900537 6976040900540 6976040900541 6976040900542 6976040900543 6976040900545 6976040900546 6976040900547 6976040900548 6976040900549 6976040900550 C0501: 6970400901296 6970400901297 6970400901298 6970400901301 6970400901302 6970400901304 6970400901305 6970400901306 6970400901307 6970400901308 6970400901309 6970400901310 6970400901312 6970400901313 6970400901314 6970400901315 6970400901316 6970400901317 6970400901318 6970400901319 6970400901320 6970400901321 6970400901322 6970400901323 6970400901324 6970400901325 6970400901326 6970400901327 6970400901328 6970400901329 6970400901330 6970400901331 6970400901332 6970400901333 6970400901334 6970400901335 6970400901336 6970400901337 6970400901338 6970400901339 6970400901340 6970400901341 6970400901342 6970400901343 6970400901344 6970400901345 6970400901346 6970400901347 6970400901348 6970400901349 6970400901350 6970400901351 6970400901352 6970400901353 6970400901355 6970400901356 6970400901357 6970400901358 6970400901359 6970400901360 6970400901361 6970400901362 6970400901363 6970400901364 6970400901365 6970400901366 6970400901367 6970400901368 6970400901369 6970400901370 6970400901371 6970400901372 6970400901373 6970400901374 6970400901375 6970400901376 6970400901377 6970400901378 6970400901379 6970400901380 6970400901381 6970400901382 6970400901383 6970400901384 6970400901385 6970400901387 6970400901388 6970400901389 6970400901390 6970400901391 6970400901393 6970400901395 6970400901396 6970400901397 6970400901398 6970400901399 6970400901400 6970400901401 6970400901402 6970400901403 6970400901404 6970400901405 6970400901406 6970400901407 6970400901408 6970400901409 6970400901410 6970400901411 6970400901412 6970400901413 6970400901414 6970400901415 6970400901416 6970400901417 6970400901418 6970400901419 6970400901421 6970400901422 6970400901426 6970400901428 6970400901429 6970400901430 6970400901431 6970400901432 6970400901433 6970400901434 6970400901435 6970400901436 6970400901437 6970400901438 6970400901439 6970400901440 6970400901441 6970400901442 6970400901445 6976040901300 6976040901308 6976040901310 6976040901322 6976040901324 6976040901327 6976040901328 6976040901329 6976040901330 6976040901331 6976040901332 6976040901333 6976040901334 6976040901335 6976040901336 6976040901361 6976040901362 6976040901369 6976040901377 6976040901378 6976040901379 6976040901380 6976040901381 6976040901382 6976040901383 6976040901386 6976040901396 6976040901397 6976040901398 6976040901399 6976040901400 6976040901401 6976040901402 6976040901403 6976040901404 6976040901405 6976040901406 6976040901407 6976040901408 6976040901409 6976040901410 6976040901411 6976040901412 6976040901413 6976040901414 6976040901415 6976040901416 D0309: 6970400900551 6970400900552 6970400900553 6970400900554 6970400900555 6970400900556 6970400900557 6970400900558 6970400900559 6970400900560 6970400900561 6970400900562 6970400900563 6970400900564 6970400900565 6970400900567 6970400900568 6970400900570 6970400900571 6970400900572 6970400900573 6970400900574 6970400900576 6970400900577 6970400900578 6970400900579 6970400900580 6970400900582 6970400900583 6970400900585 6970400900586 6970400900587 6970400900588 6970400900589 6970400900590 6970400900591 6970400900592 6970400900593 6970400900594 6970400900595 6970400900596 6970400900597 6970400900598 6970400900599 6970400900600 6970400900601 6970400900602 6970400900604 6970400900606 6970400900607 6970400900608 6970400900609 6970400900610 6970400900611 6970400900612 6970400900613 6970400900614 6970400900615 6970400900616 6970400900617 6970400900618 6970400900620 6970400900621 6970400900622 6970400900623 6970400900624 6970400900625 6970400900626 6970400900627 6970400900629 6970400900630 6970400900631 6970400900632 6970400900633 6970400900634 6970400900635 6970400900636 6970400900637 6970400900638 6970400900639 6970400900640 6970400900641 6970400900642 6970400900643 6970400900644 6970400900645 6970400900646 6970400900648 6970400900649 6970400900650 6970400900651 6970400900652 6970400900653 6970400900654 6970400900655 6970400900656 6970400900657 6970400900658 6970400900659 6970400900660 6970400900661 6970400900662 6970400900663 6970400900664 6970400900665 6970400900666 6970400900667 6970400900668 6970400900669 6970400900670 6970400900671 6970400900672 6970400900673 6970400900674 6970400900675 6970400900676 6970400900677 6970400900678 6970400900679 6970400900680 6970400900681 6970400900682 6970400900683 6970400900684 6970400900685 6970400900686 6970400900687 6970400900688 6970400900689 6970400900690 6970400900691 6970400900692 6970400900693 6970400900694 6970400900695 6970400900696 6970400900697 6970400900698 6970400900700 6970400900701 6970400900702 6970400900703 6970400900704 6970400900705 6970400900706 6970400900707 6970400900708 6970400900709 6970400900711 6970400900712 6970400900713 6970400900714 6970400900715 6970400900716 6970400900717 6970400900718 6970400900719 6970400900720 6970400900721 6970400900722 6970400900723 6970400900724 6970400900725 6970400900726 6970400900727 6970400900728 6970400900729 6970400900730 6970400900732 6970400900733 6970400900734 6970400900735 6970400900737 6970400900738 6970400900739 6970400900740 6970400900741 6970400900743 6970400900744 6970400900746 6970400900747 6970400900748 6970400900749 6976040900551 6976040900553 6976040900554 6976040900555 6976040900556 6976040900557 6976040900559 6976040900560 6976040900561 6976040900562 6976040900563 6976040900564 6976040900565 6976040900566 6976040900567 6976040900568 6976040900570 6976040900591 6976040900592 6976040900593 6976040900594 6976040900595 6976040900596 6976040900598 6976040900599 6976040900600 6976040900601 6976040900602 6976040900603 6976040900604 6976040900605 6976040900606 6976040900607 6976040900608 6976040900609 6976040900610 6976040900633 6976040900634 6976040900635 6976040900636 6976040900637 6976040900638 6976040900639 6976040900640 6976040900641 6976040900642 6976040900643 6976040900644 6976040900645 6976040900646 6976040900647 6976040900648 6976040900649 6976040900650 6976040900652 6976040900653 6976040900654 6976040900655 6976040900656 6976040900657 6976040900658 6976040900659 6976040900660 6976040900682 6976040900688 6976040900690 6976040900703 6976040900704 6976040900705 6976040900706 6976040900709 6976040900710 6976040900711 6976040900712 6976040900713 6976040900714 6976040900715 6976040900716 6976040900717 6976040900718 6976040900719 6976040900720 6976040900721 6976040900722 6976040900723 6976040900724 6976040900725 6976040900726 6976040900727 6976040900728 6976040900728 6976040900729 6976040900730 6976040900748 6976040900749 6976040900750 E0309: 6970400900751 6970400900752 6970400900753 6970400900756 6970400900757 6970400900758 6970400900759 6970400900760 6970400900761 6970400900762 6970400900764 6970400900765 6970400900767 6970400900769 6970400900770 6970400900773 6970400900774 6970400900775 6970400900776 6970400900778 6970400900779 6970400900780 6970400900785 6970400900786 6970400900787 6970400900788 6970400900789 6970400900791 6970400900792 6970400900793 6970400900794 6970400900796 6970400900797 6970400900801 6970400900802 6970400900803 6970400900806 6970400900807 6970400900808 6970400900809 6970400900810 6970400900811 6970400900812 6970400900814 6970400900815 6970400900816 6970400900818 6970400900819 6970400900820 6970400900821 6970400900822 6970400900823 6970400900824 6970400900826 6970400900827 6970400900828 6970400900829 6970400900832 6970400900833 6970400900834 6970400900835 6970400900836 6970400900837 6970400900838 6970400900839 6970400900840 6970400900841 6970400900842 6970400900843 6970400900844 6970400900845 6970400900846 6970400900847 6970400900848 6970400900849 6970400900850 6970400900851 6970400900852 6970400900853 6970400900855 6970400900856 6970400900857 6970400900858 6970400900859 6970400900860 6970400900861 6970400900862 6970400900863 6970400900864 6970400900865 6970400900866 6970400900869 6970400900870 6970400900871 6970400900872 6970400900873 6970400900876 6970400900877 6970400900878 6970400900879 6970400900880 6970400900881 6970400900882 6970400900883 6970400900884 6970400900885 6970400900886 6970400900887 6970400900888 6970400900889 6970400900890 6970400900891 6970400900892 6970400900893 6970400900894 6970400900895 6970400900896 6970400900897 6970400900898 6970400900899 6970400900900 6970400900901 6970400900904 6970400900905 6970400900906 6970400900907 6970400900908 6970400900909 6970400900910 6970400900912 6970400900913 6970400900915 6970400900916 6970400900917 6970400900919 6970400900920 6970400900923 6970400900924 6970400900925 6970400900927 6970400900928 6970400900929 6970400900930 6970400900933 6970400900934 6970400900935 6970400900936 6970400900937 6970400900938 6970400900939 6970400900940 6970400900945 6970400900948 6970400900949 6976040900781 6976040900786 6976040900788 6976040900790 6976040900791 6976040900792 6976040900793 6976040900794 6976040900795 6976040900796 6976040900797 6976040900798 6976040900801 6976040900802 6976040900803 6976040900804 6976040900805 6976040900806 6976040900807 6976040900808 6976040900809 6976040900810 6976040900811 6976040900812 6976040900813 6976040900814 6976040900815 6976040900818 6976040900819 6976040900820 6976040900821 6976040900822 6976040900823 6976040900826 6976040900827 6976040900828 6976040900829 6976040900830 6976040900831 6976040900832 6976040900833 6976040900834 6976040900835 6976040900836 6976040900837 6976040900838 6976040900839 6976040900840 6976040900841 6976040900842 6976040900843 6976040900844 6976040900845 6976040900846 6976040900847 6976040900848 6976040900849 6976040900850 6976040900851 6976040900852 6976040900853 6976040900855 6976040900856 6976040900857 6976040900859 6976040900860 6976040900861 6976040900862 6976040900863 6976040900867 6976040900869 6976040900870 6976040900873 6976040900874 6976040900880 6976040900883 6976040900884 6976040900885 6976040900887 6976040900889 6976040900890 6976040900891 6976040900892 6976040900893 6976040900894 6976040900895 6976040900897 6976040900900 6976040900901 6976040900902 6976040900904 6976040900908 6976040900909 6976040900911 6976040900912 6976040900914 6976040900915 6976040900916 6976040900917 6976040900918 6976040900919 6976040900920 6976040900924 6976040900925 6976040900928 6976040900929 6976040900930 6976040900931 6976040900933 6976040900935 6976040900938 6976040900941 6976040900942 6976040900943 6976040900945 6976040900948 F0309: 6970400900951 6970400900952 6970400900953 6970400900954 6970400900955 6970400900956 6970400900957 6970400900958 6970400900959 6970400900960 6970400900961 6970400900962 6970400900963 6970400900964 6970400900965 6970400900966 6970400900967 6970400900968 6970400900969 6970400900970 6970400900971 6970400900972 6970400900973 6970400900974 6970400900975 6970400900976 6970400900977 6970400900978 6970400900979 6970400900980 6970400900981 6970400900982 6970400900983 6970400900984 6970400900985 6970400900986 6970400900987 6970400900988 6970400900989 6970400900990 6970400900991 6970400900992 6970400900993 6970400900994 6970400900995 6970400900996 6970400900997 6970400900998 6970400900999 6970400901000 6970400901001 6970400901002 6970400901003 6970400901004 6970400901005 6970400901006 6970400901007 6970400901008 6970400901009 6970400901010 6970400901011 6970400901012 6970400901013 6970400901014 6970400901015 6970400901016 6970400901017 6970400901018 6970400901019 6970400901020 6976040900951 6976040900952 6976040900953 6976040900954 6976040900955 6976040900956 6976040900957 6976040900958 6976040900959 6976040900960 6976040900961 6976040900962 6976040900963 6976040900964 6976040900965 6976040900966 6976040900967 6976040900968 6976040900969 6976040900970 6976040900971 6976040900972 6976040900973 6976040900974 6976040900975 6976040900976 6976040900978 6976040900979 6976040900980 6976040900981 6976040900982 6976040900983 6976040900984 6976040900985 6976040900986 6976040900987 6976040900988 6976040900989 6976040900990 6976040900991 Serial Number: 6970400900184 6970400900595 6970400900764 6970400900934 6970400901036 6976040900150 6976040900211 6976040900298 6976040900368 6976040900832 6976040901037 6976040901062 6976040901097 6976040901191 6976040901400 6976040901536 6976040900001 6976040900008 |
| FEI Number |
2183191
|
Recalling Firm/ Manufacturer |
GF Health Products, Inc./d.b.a.Basic American Metal Products 336 Trowbridge Dr Fond Du Lac WI 54937-9103
|
| For Additional Information Contact | 920-929-8200 |
Manufacturer Reason for Recall | GF Health Products, Inc., is conducting a recall on the Lumex Patriot Full Electric Homecare "Bed Ends" manufactured 3/27/09, 5/9/09, 5/30/09 and 6/20/09. The supplier deviated from the print specifications of a machined pin to a sheared pin. This deviation has the potential to provide room for the pin to move during use. If failure were to occur, the bed could descend rapidly causing non-typic |
FDA Determined Cause 2 | Component change control |
| Action | GF Health Products, Inc. sent an "URGENT CORRECTION AND REMOVAL" letter dated April 17, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to immediately cease use of the affected product, secure the product, and to prevent the product from further distribution to customers and end users. Contact the firm at 1-800-556-6508 for questions regarding this notice. |
| Quantity in Commerce | 3095 pairs of bed ends |
| Distribution | Worldwide Distribution-USA (nationwide) including the states of AL, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, and WI and the countries of EL SALVADOR, CHILE, GUATEMALA, TRINIDAD, TOBAGO, and BAHAMAS. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|