| | Class 2 Device Recall Siemens Dimension Vista(R) System B12 Flex(R) Reagent Cartridge |  |
| Date Initiated by Firm | May 08, 2012 |
| Date Posted | June 27, 2012 |
| Recall Status1 |
Terminated 3 on April 03, 2014 |
| Recall Number | Z-1869-2012 |
| Recall Event ID |
61889 |
| 510(K)Number | K071224 |
| Product Classification |
Radioassay, vitamin b12 - Product Code CDD
|
| Product | Siemens Dimension Vista(R) System B12 Flex(R) Reagent Cartridge (K6442)
B12 Flex Reagent Cartridge used with the Dimension Vista System for the in vitro diagnostic test for the quantitative measurement of Vitamin B12 in human serum and plasma. |
| Code Information |
B12 Flex(R) reagent cartridge (K6442) - IFBA Interference All lots -- 07304AA, Exp 5/28/08; 08021AB, Exp 1/20/09; 08042AC, Exp 2/10/09; 08071AA, Exp 3/11/09; 08114AC, Exp 4/23/09; 09019AC, Exp 1/19/10; 09019AB, Exp 1/19/10; 09040AA, Exp 2/10/10; 09068AA, Exp 3/9/10; 09085AB, Exp 3/26/10; 09110AE, Exp 4/20/10; 09140AD, Exp 5/19/10; 09180AC, Exp 6/29/10; 09215AA, Exp 8/3/10; 09236AC, Exp 8/24/10; 09279AC, Exp 10/6/10; 09307AA, Exp 11/3/10; 10012AA, Exp 1/13/11; 10039AA, Exp 2/5/11; 10060AB, Exp 3/1/11; 10074AB, Exp 3/16/11; 10097AB, Exp 4/7/11; 10109AD, Exp 4/19/11; 10123AA, Exp 5/3/11; 10138AB, Exp 5/18/11; 10155AC, Exp 6/3/11; 10181AA, Exp 7/1/11; 10208AE, Exp 7/27/11; 10208AD, Exp 7/27/11; 10221AC, Exp 8/9/11; 10237AD, Exp 8/25/11; 10243AE, Exp 8/31/11; 10265AB, Exp 9/22/11; 10281AB, Exp 10/8/11; 10301AC, Exp 10/28/11; 10322AB, Exp 11/18/11; 10341AC, Exp 12/6/11; 10341AD, Exp 12/7/11; 11012AB, Exp 1/13/12; 11038AF, Exp 2/9/12; 11038AG, Exp 2/8/12; 11066AE, Exp 3/7/12; 11087AD, Exp 3/27/12; 11118AB, Exp 4/28/12; 11145AA, Exp 5/24/12; 11171AA, Exp 6/21/12; 11234AA, Exp 8/21/12; 11234AB, Exp 8/21/12; 11250AA, Exp 9/6/12; 11263AA, Exp 9/19/12; 11283AA, Exp 10/9/12; 11297AB, Exp 10/23/12; 11312AA, Exp 11/6/12; 11332AA, Exp 11/27/12; 11346AA, Exp 12/11/12; 11361AB, Exp 12/27/12; 12011AA, Exp 1/15/13; 12030AB, Exp 1/30/13; 12044AB, Exp 2/13/12; 12059AB, Exp 3/27/13; 12079AB, Exp 3/18/13; and 12093AB, Exp 4/1/13. |
| FEI Number |
2517506
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 GBC Drive, Mailstop 514 PO BOX 6101 Newark DE 19714-6101
|
| For Additional Information Contact | 302-631-6311 |
Manufacturer Reason for Recall | Siemens Healthcare Diagnostics has confirmed that the presence of Intrinsic Factor Blocking Antibody in some patients can result in a false and significant elevation in the reported concentration of Vitamin B12 on the Dimension Vista System. This issue affects all lots of B12 Flex reagent cartridges on the Dimension Vista System. |
FDA Determined Cause 2 | Device Design |
| Action | Siemens Healthcare Diagnostics Inc. sent a Urgent Device Recall letter to all affected customers dated May 2012. The letter identified the product the problem and the action needed to be taken by the customer.
The letter instructed customers to immediately discontinue use of the device, quarantine and dispose of all inventory, ensure that your Medical Director reviews this letter as well as the attached physician letter. Please complete the attached Field Correction form and fax to 302-631-8467.
Further questions please call 302-631-6311. |
| Quantity in Commerce | 59,735 |
| Distribution | Worldwide Distribution -- USA (nationwide) and the countries of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Great Britian, Italy, Netherlands, Norway, Portugal, Slovakia, South Africa, Spain, Switzerland, United Arab Emirates, Canada, New Zealand, and South Korea. *** Center Recommended Depth is Retail*** |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CDD
|
|
|
|