• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Alere Triage BNP PN 98000XR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Alere Triage BNP PN 98000XRsee related information
Date Initiated by FirmMay 22, 2012
Date PostedJuly 06, 2012
Recall Status1 Terminated 3 on October 18, 2013
Recall NumberZ-1931-2012
Recall Event ID 61927
510(K)NumberK051781 
Product Classification Test,natriuretic peptide - Product Code NBC
ProductAlere Triage BNP PN 98000XR Intended to be used as an aid in the diagnosis and assessment of severity of congestive heart failure (also referred to as heart failure).
Code Information Phase 1:  K50671B, W49486B, W49487B, W49489B, W49498B, W49500B, W49873B, W49890B, W49895B, W49896B, W49934B, W49947B, W49948B, W49950B, W49962B, W49968B, W50257B, W50278B, W50279B, W50282B, W50287B, W50294B, and W50313B.  Phase 2:  K50665B, W49855B, W49862B, W49865B, W49887B, W49892B, W49893B, W49894B, W49900B, W49933B, W49939B, W49953B, W49958B, W49960B, W49961B, W50252B, W50268B, W50273B, W50274B, W50290B, W50292B, W50295B, W50297B, W50298B, W50304B, W50355B, W50357B, and W50375B.
FEI Number 1000125596
Recalling Firm/
Manufacturer
Alere San Diego, Inc.
9975 Summers Ridge Rd
San Diego CA 92121-2997
For Additional Information Contact
858-805-2000 Ext. 3015
Manufacturer Reason
for Recall
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
FDA Determined
Cause 2
Process design
ActionPhase 1 of the recall included a recall communication being initiated on May 22, 2012 with Alere forwarding a recall letter to all their customers who purchased the Triage CardioProfiler Panel, Triage Cardiac Panel, Triage Profiler SOB Panel, Triage BNP, and Triage D-dimer. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to discontinue use of the affected product, and to complete and fax the enclosed verification form within 10 days. Phase 2 of the recall included a recall communication letter being forwarded to customers on 6/11/12. The second letter included additional lots of Triage products that go beyond the lots included in an earlier recall notification letter. The letter informed the customers that the additional lots have significantly decreased precision relative to the package insert which could result in an increased frequency of false positive or false negative results. A third letter dated June 12, 2012 included 3 more lots for the Triage Cardio Profiler Panel PN 97100CP and Alere Triage Profiler SOB Panel PN 97300. The letter informed the customers that the additional lots have an increased frequency of Troponin I results >0.05 ng/mL for samples which are found to be below 0.05 ng/mL upon additional testing. Customers with questions about the information contained in the notifications, were instructed to contact Alere San Diego at (877) 308-8287 or at 9975 Summers Ridge Road, San Diego, CA 92121.
Quantity in Commerce30,083 kits
DistributionNationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = NBC
-
-