• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall Chromovision High Definition (HD) Wireless Camera

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall Chromovision High Definition (HD) Wireless Camerasee related information
Date Initiated by FirmJuly 10, 2012
Date PostedAugust 06, 2012
Recall Status1 Terminated 3 on June 06, 2013
Recall NumberZ-2148-2012
Recall Event ID 62683
Product Classification Camera, cine, surgical, without audio - Product Code FWH
ProductChromovision High Definition (HD) Wireless Camera, Pt # CC-9715900. Visual recording of surgical procedures.
Code Information Serial numbers: 350111-S10017, 350111-S10018, 350111-S10019, 350111-S10020, 350111-S10021, 350111-S10022, 350111-S10023, 350111-S10024, 350111-S10025, 350111-S10026, 350111-S10027, 350111-S10028, 350111-S10030, 350111-S10031, 350111-S10032, 350111-S10033, 350111-S10034, 350111-S10035, 350111-S10036, 350111-S10037, 350111-S10038, 350111-S10039, 350111-S10040, 350111-S10041, 350111-S10043, 350111-S10044, 350111-S10045, 350111-S10046, 350111-S10047, 350111-S10048, 350111-S10049, 350111-S10050, 350111-S10051, 350111-S10052, 350111-S10053, 350111-S10054, 350111-S10055, 350111-S10056, 350111-S10060, 350111-S10061, 350111-S10062, 350111-S10063, 350111-S10064, 350111-S10065, 350111-S10066, 350111-S10067, 350111-S10068, 350111-S10069, 350111-S10070, 350111-S10071, 350111-S10072, 350111-S10073, 350111-S10074, 350111-S10075, 350111-S10076, 350111-S10077, 350111-S10078, 350111-S10079, 350111-S10080, 350111-S10081, 350111-S10082, 350111-S10083, 350111-S10084, 350111-S10085, 350111-S10086, 350111-S10087, 350111-S10088, 350111-S10089, 350111-S10090, 350111-S10091, 350111-S10092, 350111-S10093, 350111-S10094, 350111-S10095, 350111-S10096, 350111-S10097, 350111-S10098, 350111-S10099, 350111-S10100, 350111-S10101, 350111-S10102, 350111-S10103, 350111-S10104, 350111-S10105, 350111-S10106, 350111-S10107, 350111-S10108, 350111-S10109, 350111-S10110, 350111-S10111, 350111-S10112, 350111-S10113, 350111-S10114, 350111-S10115, 350111-S10116, 350111-S10117, 350111-S10118, 350111-S10119, 350111-S10120, 350111-S10121, 350111-S10122, 350111-S10123, 350111-S10124, 350111-S10125, 350111-S10126, 350111-S10127, 350111-S10128, 350111-S10129, 350111-S10130, 350111-S10131, 350111-S10132, 350111-S10133, 350111-S10134, 350111-S10135, 350111-S10136, 350111-S10137, 350111-S10138, 350111-S10139, 350111-S10140, 350111-S10141, 350111-S10142, 350111-S10143, 350111-S10144, 350111-S10145, 350111-S10146, 350111-S10147, 350111-S10148, 350111-S10149, 350111-S10150, 350111-S10151, 350111-S10152, 350111-S10153, 350111-S10154, 350111-S10155, 350111-S10156, 350111-S10157, 350111-S10158, 350111-S10159, 350111-S10160, 350111-S10161, 350111-S10162, 350111-S10164, 350111-S10165, 350111-S10166, 350111-S10167, 350111-S10168, 350111-S10169, 350111-S10170, 350111-S10171, 350111-S10172, 350111-S10173, 350111-S10176, 350111-S10177, 350111-S10178, 350111-S10179, 350111-S10180, 350111-S10181, 350111-S10192, 350111-S10193, 350111-S10194, 350111-S10195, 350111-S10196, 350111-S10197, 350111-S10198, 350111-S10199, 350111-S10200, 350111-S10201, 350111-S10202, 350111-S10203, 350111-S10204, 350111-S10205, 350111-S10206, 350111-S10207, 350111-S10208, 350111-S10209, 350111-S10210, 350111-S10211, 350111-S10212, 350111-S10213, 350111-S10215, 350111-S10216, 350111-S10218, 350111-S10219, 350111-S10220, 350111-S10221, 350111-S10222, 350111-S10223, 350111-S10224, 350111-S10225, 350111-S10226, 350111-S10227, 350111-S10228, 350111-S10229, 350111-S10230, 350111-S10231, 350111-S10232, 350111-S10233, 350111-S10234, 350111-S10235, 350111-S10236, 350111-S10237, 350111-S10238, 350111-S10239, 350111-S10248, 350111-S10249, 350111-S10250, 350111-S10251, 350111-S10252, 350111-S10253, 350111-S10254, 350111-S10255, 350111-S10256, 350111-S10257, 350111-S10258, 350111-S10259, 350111-S10260, 350111-S10261, 350111-S10262, 350111-S10263, 350111-S10265, 350111-S10266, 350111-S10267, 350111-S10268, 350111-S10269, 350111-S10270, 350111-S10273, 350111-S10274, 350111-S10276, 350111-S10277, 350111-S10278, 350111-S10279, 350111-S10280, 350111-S10281, 350111-S10282, 350111-S10283, 350111-S10284, 350111-S10285, 350111-S10286, 350111-S10287, 350111-S10288, 350111-S10289, 350111-S10290, 350111-S10291, 350111-S10292, 350111-S10293, 350111-S10295, 350111-S10296, 350111-S10297, 350111-S10298, 350111-S10300, 350111-S10301, 350111-S10302, 350111-S10303, 350111-S10305, 350111-S10306, 350111-S10307, 350111-S10309 and 350111-S10310.
FEI Number 1220685
Recalling Firm/
Manufacturer
Berchtold Corp.
1950 Hanahan Rd
N Charleston SC 29406-4878
For Additional Information ContactChris Mannarino
843-569-6100
Manufacturer Reason
for Recall
Defective camera: There is a possibility that the connection of the wires to the connector may become loose, causing image distortion or loss.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionBERCHTOLD Corporation sent a letter dated July 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. BERCHTOLD will replace the customer's cameras at no cost to the customer. Customers with questions should call 843-569-6100, ext 322. For questions regarding this recall call 843-569-6100.
Quantity in Commerce17 cameras
DistributionNationwide Distribution including FL, MI, MS, and NH.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-