• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Irrigation Catheters

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Irrigation Catheterssee related information
Date Initiated by FirmAugust 09, 2012
Date PostedSeptember 25, 2012
Recall Status1 Terminated 3 on March 22, 2013
Recall NumberZ-2448-2012
Recall Event ID 63100
510(K)NumberK902256 
Product Classification Catheter, percutaneous - Product Code DQY
ProductIrrigation Catheters (Models IRR-XXX-XX): Model Numbers & Description: IRR-023-10F,CATHETER IRRIGATION 10F 23CM; IRR-080-4F, CATHETER, IRRIGATION 4F 80CM; IRR-080-6F, CATHETER, IRRIGATION 6F 80CM. Irrigation Catheters are indicated for irrigation in the peripheral vascular system and biliary tree.
Code Information Model Numbers: Lot Numbers IRR-023-10F: 1158230. IRR-080-4F: 1156082, 1170185, 1164648. IRR-080-6F: 1158960.
FEI Number 2027111
Recalling Firm/
Manufacturer
Applied Medical Resources Corp
22872 Avenida Empresa # 3
Rancho Santa Margarita CA 92688-2650
For Additional Information Contact
949-713-8000
Manufacturer Reason
for Recall
Applied Medical is recalling specific lot numbers of its Vascular Catheters because they may have a potential for packaging particulate matter to reside on the product.
FDA Determined
Cause 2
Material/Component Contamination
ActionApplied Medical sent an "URGENT: MEDICAL DEVICE RECALL" letter dated August 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to complete the attached Recall Notification Confirmation Form and to return it to the firm via e-mail to recall60476587@appliedmedical.com or fax to (949) 713-8832. For product return questions, customers were instructed to contact Sales Operations Administrator at (949) 713-8652.For questions regarding this recall contact the firm at (949) 713-8041.
Quantity in Commerce40,238 units for all products in Recall Event
DistributionWorldwide Distribution-USA (nationwide) and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Germany, Ecuador, Estonia, Spain, France, United Kingdom, Hong Kong, Ireland, Israel, Italy, Republic of Korea, Libyan Arab Jamahiriya, Malaysia, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Serbia, Saudi Arabia, Taiwan, and South Africa.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = DQY
-
-