• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Qiagen artus CMV RG PCR ASR (Analyte Specific Reagent)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Qiagen artus CMV RG PCR ASR (Analyte Specific Reagent)see related information
Date Initiated by FirmJanuary 30, 2013
Date PostedMay 14, 2013
Recall Status1 Terminated 3 on July 01, 2013
Recall NumberZ-1321-2013
Recall Event ID 64650
Product Classification Antibody igm,if, cytomegalovirus virus - Product Code LKQ
ProductQIAGEN artus CMV RG PCR ASR (24) (Catalog number 4503223) Product Usage: The artus CMV RG PCR ASR is designed to detect and to quantify nucleic acid specific to human cytomegalovirus (CMV) using real-time PCR technology (PCR) in the Rotor-Gene 2000/3000 (RG).
Code Information Catalog Number 4503223, Lot #142345647, Exp: 06/09/14
FEI Number 3003572099
Recalling Firm/
Manufacturer
QIAGEN Gaithersburg, Inc.
1201 Clopper Rd
Gaithersburg MD 20878-4000
For Additional Information ContactDonna Sowers
301-944-7284
Manufacturer Reason
for Recall
CMV RG PCR kits may contain a mixture of incorrect vials among the correct vials.
FDA Determined
Cause 2
Mixed-up of materials/components
ActionQiagen Gaithersburg, Inc. issued a PRODUCT FIELD ACTION letter dated February 4, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect the vials prior to use to ensure that the vials are labeled "CMV RG Master" Vials labeled "EBV TM Master" should NOT be used and count the vials and discard per internal laboratory procedures. Qiagen will replace free of charge or credit any kits discovered with incorrect master mix vials. Please complete page 2 of this notice as soon as possible and email or fax it back to us at: (661) 702-3854, techservice-na@qiagen.com. If you have additional questions or concerns polease do not hestiate to call QIAGEN Technical Services at 1-800-344-3631 option 2 or by e-mail techservice-na@qiagen.com.
Quantity in Commerce90 kits in total
DistributionUSA Nationwide Distribution including the states of: CA, DC, FL, GA, MA, MD, MI, and IL.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-