| | Class 2 Device Recall 2008 MeDS Pump, Venofer(R) Pump |  |
| Date Initiated by Firm | April 01, 2013 |
| Date Posted | August 26, 2013 |
| Recall Status1 |
Terminated 3 on March 02, 2015 |
| Recall Number | Z-2082-2013 |
| Recall Event ID |
65080 |
| 510(K)Number | K121497 |
| Product Classification |
Dialyzer, high permeability with or without sealed dialysate system - Product Code KDI
|
| Product | 2008 MeDS Pump, Venofer(R) Pump
Product Usage: The 2008 MeDs pump is an optional accessory for use with the Fresenius 2008 Series Hemodialysis Machines and is intended to administer Venofer to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer, in conjunction with hemodialysis, is prescribed by a physician. |
| Code Information |
Serial numbers - TM-001065, TM-001283, TM-001284, TM-001325, TM-001411, TM-001426, TM-001465, TM-001476, TM-001529, TM-001967, TM-001968, TM-001975, TM-001985, TM-002015, TM-002016, TM-002017, TM-002018, TM-002021, TM-002022, TM-002095, TM-000015, TM-000028, TM-000031, TM-000041, TM-0048, TM-000050, TM-000051, TM-000053, TM-000055, TM-000059, TM-000064, TM-000065, TM-000068, TM-0069, TM-000213, TM-000215, TM-000218, TM-000219, TM-000070, TM-000072, TM-000074, TM-000075, TM-000078, TM-000081, TM-000082, TM-000083, TM-000085, TM-000086, TM-000087, TM-000088, TM-000089, TM-000092, TM-000040, TM-000077, TM-000098, TM-000101, TM-000102, TM-000103, TM-000104, TM-000105, TM-000106, TM-000107, TM-000108, TM-000109, TM-000112, TM-000113, TM-000114, TM-000115, TM-000116, TM-000122, TM-000123, TM-000124, TM-000125, TM-000126, TM-000128, TM-000145, TM-000002, TM-000003, TM-000004, TM-000005, TM-000006, TM-000007, TM-000008, TM-000009, TM-000010, TM-000012, TM-000013, TM-000014, TM-000017, TM-000018, TM-000019, TM-000020, TM-000023, TM-000024, TM-000025, TM-000033, TM-000036, TM-A595, TM-A610, TM-A627, TM-000220, TM-000227, TM-007364, TM-007371, TM-006848, TM-006860, TM-006875, TM-006879, TM-006880, TM-006881, TM-006882, TM-006883, TM-006884, TM-006885, TM-006886, TM-006887, TM-006890, TM-006891, TM-006892, TM-006893, TM-006897, TM-006898, TM-006911, TM-006912, TM-006913, TM-006914, TM-006915, TM-006916, TM-006917, TM-006918, TM-006919, and TM-006920. |
| FEI Number |
3005778453
|
Recalling Firm/ Manufacturer |
Fresenius Medical Care Holdings, Inc. dba Renal Solutions 770 Commonwealth Dr Ste 101 Warrendale PA 15086-7521
|
| For Additional Information Contact | Renal Solutions, Inc. Customer Service 724-772-6900 Ext. 140 |
Manufacturer Reason for Recall | Product is being recalled because the combination of the current 50 mg vial, the 2008 MeDS pump, and the 2008 MeDS pump tubing Set for CombiSet Bloodlines can lead to a situation where Venofer(R) delivery does not meet dosage accuracy requirements. This may lead to a slight over or under delivery of Venofer(R) that does not pose an acceptable risk to the patient.
Please note that this recall |
FDA Determined Cause 2 | Device Design |
| Action | Renal Solutions sent an Urgent Notification letter dated April 1, 2013 via certified mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to cease further use of the device. The letter informed customers that a Fresenius Medical Care Technical Service/Customer Service representative will contact customers to arrange for removal of the device. For questions contact Fresenius Medical Care Customer Service at 800-227-2572. |
| Quantity in Commerce | 132 |
| Distribution | USA Nationwide Distribution in the states of CA, MA, and NY. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|