Class 2 Device Recall Brilliance CT 64, Ingenuity Core, and Ingenuity Core128 scanner

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Brilliance CT 64, Ingenuity Core, and Ingenuity Core128 scannersee related information
Date Initiated by FirmApril 15, 2013
Date PostedJuly 02, 2013
Recall Status1 Terminated 3 on January 12, 2015
Recall NumberZ-1646-2013
Recall Event ID 65139
510(K)NumberK033326 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductComputed Tomography X-ray systems Product Usage: The Brilliance CT 64, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Code Information The following system code numbers and associated serial numbers are affected by this recall:  System Code #728231, Serial #: 4001, 4003, 4016, 4020, 9009, 9029, 9030, 9033, 9034, 9037, 9038, 9040, 9041, 9043, 9044, 9045, 9048, 9049, 9051, 9052, 9054, 9056, 9058, 9059, 9062, 9063, 9064, 9065, 9066, 9067, 9068, 9069, 9070, 9071, 9072, 9073, 9074, 9075, 9077, 9080, 9081, 9083, 9086, 9087, 9088, 9089, 9092, 9093, 9094, 9096, 9098, 9099, 9104, 9105, 9106, 9107, 9108, 9112, 9118, 9120, 9126, 9127, 9131, 9135, 9137, 9138, 9140, 9142, 9143, 9148, 9151, 9156, 9159, 9160, 9162, 9163, 9166, 9171, 9172, 9173, 9175, 9177, 9180, 9187, 9190, 9195, 9196, 9197, 9201, 9202, 9205, 9206, 9208, 9213, 9214, 9215, 9216, 9217, 9218, 9235, 9243, 9259, 9501, 9502, 9503, 9504, 9505, 9506, 9507, 9508, 9509, 9510, 9511, 9512, 9513, 9514, 9515, 9516, 9517, 9518, 9519, 9520, 9521, 9522, 9523, 9524, 9525, 9526, 9527, 9528, 9529, 9530, 9531, 9532, 9533, 9534, 9535, 9536, 9537, 9538, 9539, 9540, 9541, 9542, 9543, 9545, 9546, 9547, 9548, 9549, 9550, 9551, 9552, 9553, 9554, 9555, 9556, 9557, 9558, 9559, 9560, 9561, 9562, 9563, 9564, 9565, 9566, 9567, 9568, 9569, 9570, 9571, 9572, 9573, 9574, 9575, 9576, 9577, 9578, 9579, 9580, 9581, 9582, 9583, 9584, 9585, 9586, 9587, 9588, 9589, 9590, 9591, 9592, 9593, 9594, 9596, 9597, 9598, 9599, 9600, 9601, 9602, 9603, 9604, 9605, 9606, 9607, 9608, 9610, 9611, 9612, 9613, 9614, 9615, 9616, 9617, 9618, 9619, 9620, 9621, 9622, 9623, 9624, 9625, 9626, 9627, 9628, 9629, 9630, 9631, 9632, 9633, 9634, 9635, 9636, 9637, 9638, 9639, 9640, 9641, 9642, 9643, 9644, 9645, 9646, 9647, 9648, 9649, 9650, 9651, 9652, 9653, 9654, 9655, 9656, 9657, 9658, 9659, 9660, 9661, 9662, 9663, 9664, 9665, 9666, 9667, 9668, 9670, 9671, 9672, 9673, 9674, 9675, 9676, 9677, 9678, 9679, 9680, 9681, 9682, 9683, 9684, 9685, 9686, 9687, 9688, 9689, 9690, 9691, 9692, 9693, 9694, 9695, 9697, 9698, 9699, 9700, 9701, 9702, 9703, 9704, 9705, 9706, 9707, 9708, 9709, 9710, 9711, 9712, 9713, 9714, 9715, 9716, 9717, 9718, 9719, 9720, 9721, 9722, 9723, 9724, 9725, 9726, 9727, 9728, 9729, 9730, 9731, 9732, 9734, 9735, 9736, 9737, 9738, 9739, 9740, 9741, 9742, 9743, 9744, 9745, 9746, 9747, 9748, 9749, 9750, 9751, 9752, 9753, 9754, 9755, 9756, 9757, 9758, 9759, 9760, 9761, 9762, 9763, 9764, 9765, 9766, 9767, 9768, 9769, 9770, 9771, 9772, 9773, 9774, 9775, 9776, 9777, 9778, 9779, 9780, 9781, 9782, 9783, 9784, 9785, 9786, 9787, 9788, 9789, 9790, 9791, 9792, 9793, 9794, 9795, 9796, 9797, 9798, 9799, 9800, 9801, 9802, 9803, 9804, 9805, 9806, 9807, 9808, 9809, 9810, 9811, 9812, 9813, 9814, 9815, 9816, 9817, 9818, 9819, 9820, 9821, 9822, 9823, 9824, 9825, 9826, 9827, 9829, 9830, 9831, 9832, 9833, 9834, 9835, 9836, 9837, 9838, 9839, 9840, 9841, 9842, 9843, 9844, 9845, 9846, 9847, 9848, 9849, 9850, 9852, 9853, 9854, 9855, 9856, 9857, 9858, 9859, 9860, 9861, 9862, 9863, 9864, 9865, 9866, 9867, 9868, 9869, 9870, 9871, 9872, 9873, 9874, 9875, 9876, 9877, 9878, 9879, 9880, 9881, 9882, 9883, 9884, 9885, 9886, 9887, 9888, 9889, 9890, 9891, 9892, 9893, 9894, 9895, 9896, 9898, 9899, 9900, 9901, 9902, 9903, 9904, 9905, 9906, 9907, 9908, 9909, 9910, 9911, 9912, 9913, 9914, 9915, 9916, 9917, 9918, 9919, 9920, 9921, 9922, 9923, 9924, 9925, 9926, 9927, 9929, 9930, 9931, 9933, 9936, 9937, 9938, 9940, 9941, 9942, 9946, 9947, 9948, 9950, 9951, 9952, 9953, 9954, 9956, 9957, 9958, 9959, 9960, 9961, 9963, 9964, 9965, 9966, 9968, 9969, 9970, 9971, 9972, 9974, 9975, 9976, 9978, 9980, 9981, 9982, 9983, 9984, 9985, 9986, 9987, 9989, 9991, 9993, 9994, 9995, 9996, 9997, 9998, 9999,  10000, 10001, 10002, 10003, 10004, 10005, 10006, 10007, 10008, 10009, 10011, 10012, 10014, 10015, 10016, 10017, 10018, 10019, 10020, 10021, 10022, 10023, 10024, 10025, 10026, 10027, 10028, 10030, 10031, 10032, 10033, 10034, 10035, 10036, 10037, 10039, 10040, 10041, 10042, 10044, 10045, 10046, 10047, 10048, 10049, 10050, 10051, 10052, 10053, 10054, 10056, 10057, 10058, 10059, 10060, 10061, 10064, 10065, 10066, 10067, 10068, 10069, 10070, 10071, 10072, 10073, 10075, 10077, 10078, 10079, 10080, 10081, 10083, 10084, 10085, 10086, 10087, 10088, 10089, 10090, 10092, 10094, 10095, 10096, 10097, 10098, 10099, 10100, 10101, 10102, 10103, 10104, 10105, 10106, 10107, 10108, 10109, 10110, 10111, 10112, 10113, 10114, 10115, 10116, 10117, 10118, 10119, 10120, 10121, 10122, 10123, 10124, 10125, 10126, 10127, 10128, 10129, 10130, 10131, 10132, 10134, 10135, 10136, 10137, 10138, 10139, 10140, 10141, 10142, 10143, 10144, 10145, 10146, 10147, 10148, 10149, 10150, 10151, 10152, 10153, 10154, 10155, 10156, 10157, 10158, 10160, 10161, 10162, 10164, 10165, 10166, 10167, 10168, 10169, 10170, 10171, 10172, 10174, 10176, 10177, 10178, 10179, 10180, 10181, 10183, 10184, 10185, 10186, 10187, 10188, 10190, 10191, 10192, 10193, 10194, 10195, 10196, 10197, 10198, 10200, 10201, 10202, 10203, 10204, 10205, 10206, 10207, 10208, 10209, 10210, 10211, 10212, 10213, 10214, 10215, 10216, 10217, 10218, 10219, 10220, 10221, 10222, 10223, 10224, 10225, 10226, 10227, 10228, 10229, 10230, 10231, 10234, 10235, 10237, 10238, 10239, 10240, 10241, 10242, 10243, 10244, 10245, 10246, 10247, 10248, 10250, 10251, 10252, 10253, 10254, 10255, 10256, 10257, 10258, 10259, 10260, 10261, 10262, 10263, 10264, 10265, 10266, 10267, 10268, 10269, 10270, 10271, 10272, 10273, 10274, 10275, 10276, 10277, 10278, 10279, 10280, 10281, 10282, 10283, 10284, 10285, 10286, 10287, 10288, 10291, 10292, 10293, 10294, 10295, 10296, 10297, 10298, 10299, 10300, 10301, 10302, 10303, 10304, 10305, 10307, 10308, 10309, 10310, 10311, 10312, 10313, 10314, 10315, 10316, 10317, 10318, 10319, 10320, 10321, 10322, 10323, 10324, 10325, 10327, 10330, 10331, 10332, 10333, 10334, 10335, 10336, 10337, 10338, 10339, 10341, 10342, 10343, 10344, 10345, 10346, 10347, 10348, 10349, 10350, 10351, 10352, 10353, 10354, 10355, 10357, 10358, 10359, 10360, 10361, 10362, 10363, 10364, 10365, 10366, 10367, 10368, 10369, 10370, 10371, 10372, 10373, 10374, 10375, 10376, 10378, 10379, 10380, 10381, 10383, 10384, 10385, 10386, 10388, 10389, 10390, 10391, 10392, 10393, 10394, 10395, 10396, 10397, 10398, 10399, 10400, 10401, 10402, 10403, 10404, 10405, 10406, 10407, 10408, 10409, 10410, 10411, 10412, 10413, 10414, 10415, 10416, 10417, 10418, 10419, 10420, 10421, 10422, 10424, 10425, 10426, 10427, 10428, 10429, 10430, 10431, 10432, 10434, 10435, 10436, 10437, 10438, 10439, 10440, 10441, 10442, 10443, 10444, 10445, 10446, 10447, 10448, 10449, 10450, 10451, 10452, 10453, 10454, 10455, 10456, 10457, 10458, 10459, 10460, 10461, 10462, 10463, 10464, 10465, 10466, 10467, 10468, 10469, 10470, 10471, 10472, 10473, 10474, 10475, 10476, 10477, 10478, 10479, 10480, 10481, 10482, 10483, 10484, 10485, 10486, 10487, 10488, 10489, 10490, 10491, 10492, 10494, 10495, 10496, 10497, 10498, 10499, 10500, 10502, 10503, 10504, 10505, 10506, 10507, 10508, 10510, 10511, 10512, 10513, 10514, 10515, 10516, 10517, 10518, 10519, 10520, 10521, 10522, 10523, 10525, 10526, 10527, 10528, 10530, 10531, 10532, 10533, 10534, 10535, 10536, 10537, 10538, 10539, 10540, 10541, 10542, 10543, 10545, 10546, 10547, 10548, 10549, 10550, 10551, 10552, 10553, 10554, 10555, 10556, 10557, 10558, 10559, 10560, 10561, 10562, 10563, 10564, 10565, 10566, 10567, 10568, 10569, 10570, 10571, 10572, 10573, 10574, 10575, 10576, 10577, 10578, 10579, 10580, 10581, 10582, 10583, 10584, 10585, 10586, 10587, 10588, 10589, 10590, 10591, 10592, 10593, 10594, 10595, 10596, 10597, 10598, 10599, 10600, 10601, 10602, 10603, 10604, 10605, 10606, 10607, 10608, 10609, 10610, 10611, 10612, 10613, 10614, 10615, 10616, 10617, 10618, 10619, 10621, 10622, 10623, 10624, 10625, 10626, 10627, 10628, 10629, 10630, 10631, 10632, 10633, 10634, 10635, 10636, 10637, 10638, 10639, 10640, 10641, 10642, 10643, 10644, 10645, 10646, 10647, 10648, 10649, 10650, 10651, 10652, 10653, 10654, 10655, 10656, 10657, 10658, 10659, 10660, 10661, 10662, 10663, 10664, 10665, 10666, 10667, 10668, 10669, 10670, 10671, 10672, 10673, 10674, 10675, 10676, 10677, 10678, 10679, 10680, 10681, 10682, 10683, 10684, 10685, 10686, 10687, 10688, 10689, 10690, 10691, 10692, 10693, 10694, 10695, 10696, 10697, 10698, 10699, 10700, 10701, 10702, 10703, 10704, 10705, 10706, 10707, 10708, 10710, 10712, 10713, 10714, 10715, 10717, 10718, 10720, 10721, 10722, 10723, 10724, 10725, 10726, 10727, 10728, 10730, 10731, 10733, 10734, 10735, 10736, 10737, 10738, 10739, 10740, 10741, 10742, 10743, 10744, 10745, 10747, 10748, 10749, 10750, 10751, 10752, 10753, 10755, 10756, 10757, 10758, 10759, 10760, 10761, 10762, 10763, 10764, 10765, 10766, 10767, 29002, 29003, 29005, 29007, 29014, 29016, 29018, 29022, 29024, 29025, 29026, 29027, 29034, 29035, 29037, 29038, 29044, 29045, 29049, 29054, 29056, 29057, 29062, 29070, 29086, 29092, 29097, 29107, 29111, 29113, 29124, 29138, 64015, 64040, 64051, 64117, 64135, 64151, 64152, 64153, 64158, 64163, 64167, 64168, 64173, 90029, 90030, 90031, 90033, 90036, 90037, 90041, 90042, 90045, 90046, 90047, 90050, 90057, 90058, 90060, 90062, 90063, 90066, 90070, 90072, 90073, 90077, 90078, 90079, 90080, 90081, 90082, 90084, 90086, 90087, 90089, 90091, 90092, 90094, 90096, 90097, 90098, 90099, 90100, 90102, 90105, 90106, 90107, 90117, 90118, 90120, 90123, 90124, 90129, 90131, 90132, 90136, 90137, 90139, 90140, 90141, 90143, 90147, 90150, 90151, 90152, 90153, 90154, 90156, 90157, 90159, 90160, 90161, 90162, 90163, 90167, 90168, 90169, 90175, 90177, 90180, 90182, 90185, 90187, 90205, 90208, 90213, 90218, 90221, 90223, 95003, 95008, 95009, 95011, 95012, 95013, 95014, 95015, 95016, 95017, 95018, 95019, 95020, 95021, 95022, 95023, 95024, 95025, 95029, 95030, 95031, 95032, 95033, 95034, 95035, 95036, 95037, 95038, 95039, 95040, 95041, 95042, 95043, 95044, 95045, 95047, 95048, 95049, 95050, 95051, 95052, 95053, 95054, 95055, 95056, 95057, 95059, 95060, 95061, 95062, 95063, 95064, 95065, 95066, 95067, 95068, 95069, 95071, 95072, 95073, 95074, 95075, 95076, 95077, 95078, 95079, 95080, 95081, 95082, 95083, 95084, 95085, 95086, 95087, 95088, 95089, 95090, 95091, 95092, 95093, 95094, 95095, 95096, 95097, 95098, 95099, 95100, 95101, 95102, 95103, 95104, 95105, 95106, 95107, 95108, 95109, 95110, 95111, 95112, 95113, 95114, 95116, 95117, 95118, 95119, 95120, 95121, 95122, 95123, 95124, 95125, 95126, 95127, 95128, 95129, 95130, 95131, 95132, 95133, 95134, 95135, 95136, 95137, 95138, 95139, 95140, 95141, 95142, 95143, 95144, 95145, 95146, 95147, 95149, 95150, 95151, 95152, 95153, 95154, 95155, 95156, 95157, 95158, 95159, 95161, 95162, 95163, 95164, 95165, 95167, 95168, 95169, 95170, 95171, 95172, 95173, 95175, 95176, 95177, 95178, 95179, 95180, 95182, 95183, 95184, 95185, 95186, 95187, 95188, 95189, 95190, 95191, 95192, 95193, 95194, 95195, 95196, 95197, 95198, 95199, 95200, 95201, 95202, 95203, 95204, 95205, 95206, 95207, 95208, 95210, 95212, 95213, 95214, 95215, 95216, 95217, 95218, 95219, 95220, 95221, 95222, 95223, 95224, 95225, 95226, 95227, 95228, 95229, 95231, 95232, 95233, 95234, 95235, 95236, 95237, 95238, 95239, 95240, 95241, 95242, 95243, 95244, 95245, 95246, 95247, 95249, 95250, 95251, 95252, 95253, 95254, 95255, 95256, 95257, 95258, 95259, 95260, 95261, 95262, 95263, 95264, 95265, 95266, 95267, 95268, 95269, 95270, 95271, 95272, 95273, 95274, 95275, 95276, 95277, 95278, 95279, 95280, 95281, 95282, 95283, 95284, 95285, 95286, 95287, 95288, 95289, 95290, 95291, 95292, 95293, 95294, 95296, 95297, 95298, 95299, 95300, 95301, 95302, 95303, 95304, 95305, 95308, 95309, 95310, 95311, 95312, 95313, 95314, 95315, 95316, 95317, 95318, 95319, 95320, 95321, 95322, 95323, 95324, 95325, 95326, 95327, 95328, 95329, 95330, 95331, 95332, 95333, 95334, 95335, 95336, 95337, 95338, 95339, 95340, 95342, 95343, 95344, 95345, 95346, 95347, 95348, 95349, 95350, 95351, 95352, 95353, 95354, 95355, 95356, 95357, 95358, 95359, 95360, 95361, 95362, 95363, 95364, 95365, 95366, 95369, 95370, 95371, 95374, 95375, 95376, 95377, 95378, 95379, 95380, 95381, 95382, 95383, 95385, 95386, 95387, 95388, 95389, 95390, 95391, 95393, 95394, 95395, 95396, 95397, 95398, 95399, 95400, 95401, 95402, 95403, 95404, 95405, 95406, 95407, 95408, 95409, 95410, 95411, 95412, 95413, 95414, 95415, 95416, 95417, 95418, 95419, 95420, 95421, 95422, 95424, 95425, 95426, 95427, 95428, 95429, 95430, 95431, 95432, 95433, 95434, 95435, 95436, 95437, 95438, 95439, 95440, 95441, 95442, 95443, 95444, 95445, 95446, 95447, 95448, 95449, 95450, 95451, 95452, 95453, 95454, 95455, 95456, 95457, 95458, 95459, 95461, 95462, 95463, 95464, 95465, 95466, 95467, 95468, 95469, 95470, 95471, 95472, 95473, 95474, 95475, 95476, 95477, 95479, 95480, 95481, 95482, 95483, 95484, 95485, 95486, 95487, 95488, 95489, 95490, 95491, 95492, 95493, 95494, 95495, 95496, 95497, 95498, 95499, 95500, 95501, 95502, 95503, 95504, 95505, 95506, 95507, 95508, 95509, 95510, 95511, 95512, 95513, 95514, 95515, 95516, 95517, 95518, 95519, 95520, 95521, 95522, 95523, 95524, 95525, 95526, 95527, 95528, 95529, 95530, 95531, 95532, 95533, 95534, 95535, 95536, 95537, 95538, 95540, 95541, 95542, 95543, 95544, 95545, 95546, 95547, 95548, 95549, 95550, 95551, 95552, 95554, 95555, 95556, 95557, 95558, 95559, 95560, 95561, 95563, 95564, 95565, 95566, 95567, 95568, 95569, 95570, 95571, 95572, 95573, 95574, 95575, 95577, 95578, 95579, 95580, 95581, 95582, 95583, 95584, 95585, 95586, 95587, 95588, 95589, 95590, 95591, 95592, 95595, 95596, 95597, 95598, 95599, 95600, 95601, 95602, 95603, 95604, 95605, 95606, 95607, 95608, 95609, 95610, 95611, 95612, 95613, 95614, 95615, 95616, 95617, 95618, 95620, 95621, 95622, 95623, 95624, 95625, 95626, 95627, 95628, 95629, 95630, 95631, 95632, 95633, 95634, 95635, 95636, 95637, 95638, 95639, 95640, 95641, 95642, 95643, 95644, 95645, 95646, 95647, 95648, 95649, 95650, 95651, 95652, 95653, 95654, 95655, 95656, 95657, 95658, 95659, 95660, 95661, 95662, 95664, 95665, 95666, 95667, 95668, 95669, 95670, 95671, 95672, 95673, 95674, 95675, 95676, 95677, 95678, 95679, 95680, 95681, 95682, 95683, 95684, 95685, 95686, 95687, 95688, 95689, 95690, 95691, 95692, 95693, 95694, 95695, 95696, 95697, 95698, 95699, 95700, 95702, 95703, 95704, 95705, 95706, 95707, 95708, 95709, 95710, 95711, 95712, 95713, 95714, 95715, 95716, 95718, 95719, 95720, 95721, 95722, 95723, 95724, 95725, 95726, 95727, 95729, 95730, 95731, 95732, 95733, 95734, 95735, 95736, 95737, 95738, 95739, 95740, 95741, 95742, 95743, 95744, 95745, 95746, 95747, 95748, 95749, 95750, 95751, 95752, 95753, 95754, 95755, 95756, 95757, 95758, 95759, 95760, 95761, 95764, 95765, 95767, 95768, 95769, 95770, 95772, 95773, 95774, 95775, 95776, 95777, 95778, 95779, 95780, 95781, 95782, 95783, 95784, 95785, 95786, 95787, 95788, 95789, 95790, 95791, 95792, 95793, 95794, 95795, 95796, 95797, 95798, 95799, 95800, 95801, 95802, 95803, 95804, 95805, 95806, 95807, 95808, 95809, 95810, 95811, 95812, 95814, 95815, 95816, 95817, 95818, 95819, 95820, 95821, 95822, 95823  System Code #728321, Serial #: 310001, 310002, 310003, 310004, 310005, 310006, 310007, 310010, 310011, 310012, 310013, 310014, 310015, 310016, 310017, 310018, 310019, 310020, 310021, 310022, 310023, 310024, 310025, 310026, 310027, 310028, 310029, 310030, 310031, 310032, 310033, 310034, 310035, 310036, 310037, 310038, 310040, 310041, 310043, 310044, 310045, 310046, 310047, 310048, 310049, 310050, 310051, 310052, 310053, 310054, 310055, 310056, 310057, 310058, 310059, 310060, 310061, 310062, 310063, 310065, 310066, 310067, 310068, 310069, 310071, 310072, 310073, 310074, 310075, 310076, 310077, 310078, 310079, 310080, 310082, 310083, 310084, 310085, 310086, 310087, 310088, 310089, 310090, 310091, 310092, 310093, 310094, 310095, 310096, 333001, 333007 System Code #728323, Serial #: 320002, 320003, 320004, 320005, 320006, 320007, 320008, 320009, 320010, 320011, 320013, 320014, 320015, 320017, 320018, 320019, 320020, 320022, 320023, 320024, 320025, 320026, 320027, 320028, 320029, 320031, 320032, 320034, 320035, 320036, 320037, 320038, 320039, 320040, 320041, 320043, 320044, 320045, 320046, 320047.320048, 320049, 320050, 320051, 320052, 320053, 320054, 320055, 320056, 320057, 320059, 320060, 320062, 320063, 320064, 320065, 320066, 320067, 320068, 320069, 320070, 320071, 320072, 320074, 320075, 320078, 320079
FEI Number 1525965
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland OH 44143-2131
For Additional Information Contact
800-722-9377
Manufacturer Reason
for Recall
Philips Healthcare received reports from the field that certain Brilliance CT 64, Ingenuity Core, and Ingenuity Core128 systems running software versions 2.6.1 or 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images, which may impair the diagnostic quality of affected images.
FDA Determined
Cause 2
Software design
ActionPhilips Healthcare sent an Urgent Medical device Correction letter dated April 15, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter states that Philips Healthcare is implementing software update versions 2.6.2 or 3.5.5 to resolve the issue and a Philips Field Service Engineer will contact each site to schedule the software installation. For questions contact your local Philips representative or local Philips Healthcare office.
Quantity in Commerce2348 total units: 2182 Brilliance CT 64 units; 95 Ingenuity Core units; and 71 Ingenuity Core128 units.
DistributionUSA Nationwide Distribution - including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Total Product Life CycleTPLC Device Report