• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Minicat CT Scanner

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Minicat CT Scannersee related information
Date Initiated by FirmJuly 27, 2010
Create DateJune 26, 2015
Recall Status1 Terminated 3 on September 11, 2014
Recall NumberZ-0038-2014
Recall Event ID 66338
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductXoran Minicat00202
Code Information minicat
FEI Number 3004198450
Recalling Firm/
Manufacturer
Xoran Technologies, Inc.
5210 S State Rd
Ann Arbor MI 48108-7936
For Additional Information Contact
734-663-7194
Manufacturer Reason
for Recall
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha
FDA Determined
Cause 2
Radiation Control for Health and Safety Act
ActionAll of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component. Xoran's Research & Development group is evaluating the current specification to see if the field specifications are appropriate and to ensure that proper testing protocols are communicated and are in place for use by 3 rd party testing organizations. In addition, Xoran has initiated CAPA # 18 0 that is looking at the source block with regard to evaluating the stability of the dose rating.
Quantity in Commerce320
DistributionWorldwide Distribution - US nationwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-