|
Class 2 Device Recall MAYFIELD Composite Series Skull Clamp |
|
Date Initiated by Firm |
August 29, 2013 |
Date Posted |
November 13, 2013 |
Recall Status1 |
Terminated 3 on January 14, 2014 |
Recall Number |
Z-0210-2014 |
Recall Event ID |
66449 |
510(K)Number |
K120633
|
Product Classification |
Holder, head, neurosurgical (skull clamp) - Product Code HBL
|
Product |
Composite Series Skull Clamp, one clamp per tote.
The MAYFIELD¿ Skull Clamp (A3059) is a cranial stabilization device, designed to provide rigid skeletal fixation. The Skull Clamp is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary |
Code Information |
The following serial numbers are affected by this recall: SS120076, SS120082, SS120083, SS120084, SS120088, SS120089, SS120090, SS120092, SS120093, SS120094, SS120095, SS120111, SS120112, SS120113, SS120114, SS120115, SS120116, SS120117, SS120118, SS120119, SS120120, SS120121, SS120122, SS120123, SS120124, SS120125, SS120126, SS120127, SS120128, SS120130, SS120131, SS120133, SS120134, SS120135, SS120171, SS120252, SS120256, SS120257 |
Recalling Firm/ Manufacturer |
Integra LifeSciences Corp. 311 Enterprise Dr Plainsboro NJ 08536-3344
|
For Additional Information Contact |
Integra Customer Service 855-532-1723
|
Manufacturer Reason for Recall |
An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in a report of patient injury, adverse health consequence, or a delay / prolongation of the surgical procedure.
|
FDA Determined Cause 2 |
Device Design |
Action |
Integra sent an Urgent: Voluntary Medical Device Recall Notification dated August 29, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Your local Integra NeuroSpecialist will set up a time with you to discuss replacement options but in the interim, Integra is asking that you do the following:
1. Review your inventory and determine if you have any MAYFIELD¿ Composite Series Skull Clamps (A3059). If so, stop using them immediately and remove them from service to prevent any inadvertent use of the Composite Series Skull Clamp.
Either while the Integra NeuroSpecialist is on-site or independently;
2. Complete the attached form. If you do not have affected product, check the box: I do not have Composite Series Skull Clamps.
3. If you do have affected product, check the box: I do have Composite Series Skull Clamps. Record the total quantity and Serial No.(s) of all Skull Clamps in your inventory. The Serial No. is located on the clamp upright comprised of two letters (SS) followed by six numerals.
4. Complete the other information and return by email or fax as indicated on the form, or give it to the Integra NeuroSpecialist who will return it for you.
For assistance returning and exchanging these products or any other questions that you may have, please contact Integra Customer Service
at 1-855-532-1723.
We sincerely apologize for any inconvenience this may have caused and thank you for your cooperation. |
Quantity in Commerce |
37 units |
Distribution |
US Distribution including the states of AL, AR, CA, CT, FL, GA, IA, IL, MS, NY, OH, SC, TN, and TX. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HBL and Original Applicant = INTEGRA LIFESCIENCES CORPORATION
|
|
|
|