• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Contoura

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Contourasee related information
Date Initiated by FirmMarch 17, 2014
Date PostedOctober 06, 2014
Recall Status1 Terminated 3 on February 14, 2018
Recall NumberZ-0019-2015
Recall Event ID 67029
Product Classification Bed, ac-powered adjustable hospital - Product Code FNL
ProductArjoHuntleigh Contoura C1000 series beds are designed to accommodate patients weighing up to 990 lbs. (450 kg). Model C1000 is the basic bed. Model C1080 includes a weighing system and a power outlet for an ArjoHuntleigh approved mattress pump. Product Usage: The Contoura series bariatric beds are electrically operating medical beds with a maximum patient weight of 450 kg. This range is intended for use in hospitals to facilitate management of obese patients. An optional integrated weighing system helps with assessment and management of the patient if required.
Code Information 1) Model Number: C1000; Serial Numbers (Foreign): 62265 , 399945 , 399946 , 399949 , 416912 , 416914 , 423215 , 475309 , 530882 , 554079 , 554082 , 566463 , 584837 , 586554 , 592203 , 592204 , 592205 , 592206 , 597027 , 597028 , 605859 , 646152 , 664632 , 680670 , 680699 , 693924 , 695671 , 699563 , 712231 , 714227 , 714228 , 714229 , 716809 , 720277 , 725622 , 731631 , 745781 , 745782 , 747354 , 747355 , 748161 , 748162 , 748635 , 761635 , 766430 , 766430 , 766431 , 766431 , 771194 , 771195 , 771197 , 771198 , 798571 , 816742 , 821563 , 821564 , 821704 , 821705 , 821706 , 821707 , 822714 , 823713 , 823713 , 823714 , 832524 , 832525 , 832526 , 832527 , 832528 , 840918 , 840919 , 840919 , 840920 , 840920 , 840922 , 840923 , 840924 , 840925 , 840947 , 840948 , 840949 , 840950 , 847439 , 853387 , 855259 , 855260 , 855261 , 855262 , 871837 , 871840 , 871841 , 871842 , 871843 , 872720 , 873306 , 873307 , 873307 , 881203 , 881204 , 881205 , 881206 , 881207 , 889938 , 889939 , 901232 , 901233 , 901234 , 901236 , 901237 , 901244 , 904448 , 904449 , 904449. , 913969 , 916105 , 916108 , 924589 , 925887 , 929764 , 929772 , 945149 , 945151 , 945154 , 945160 , 8401918 , 648568Z1894 , 648569Z1894 , C1000 , M/M-1 , M/M-10 , M/M-2 , M/M-3 , M/M-4 , M/M-5 , M/M-6 , M/M-7 , M/M-8 , M/M-9 , P0100097 , P0100172 , P0100337 , P0100523 , P0100655 , P0100723 , P0101916 , P0101927 , P0101928 , P0101929 , P0101930 , P0101931 , P0102404 , P0102894 , P0102895 , P0102896 , P0102897 , P0102898 , P0103581 , P0103582 , P0103583 , P0106046 , P0106705 , P0106707 , P0108275 , P0108276 , P0109784 , P0110620 , P0111754 , P0114163 , P0114164 , P0115132 , P0118357 , P0118358 , P0120283 , P0120855 , P0120862 , P0121025 , P0121307 , P0121308 , P0121605 , P0121857 , P0122187 , P0122459 , P0125170 , P0126006 , P0126089 , P0126317 , P0126318 , P0126326 , P0126865 , P0126866 , P0126867 , P0128593 , P0128594 , P0130033 , P0132561 , P0133271 , P0133392 , P0133482 , P013391 , P0134514 , P0137082¿¿¿ , P0137164 , P0137540 , P0138061 , P0138062 , P0138242 , P0138520 , P0138710 , P0138780 , P0138781 , P0139072 , P0143580¿¿¿ , P0143581¿¿¿ , P0143714 , P0144844¿¿¿ , P0144845¿¿¿ , P0145951 , P0146630 , P0146631 , P0146632 , P0147602 , P0147657 , P0147790 , P0147791 , P0147792 , P0148713 , P015878 , P015879 , P015880 , P015881 , P015882 , P015883 , P015884 , P015885 , P015886 , P015887 , P015888 , P015889 , P015890 , P015891 , P015892 , P015893 , P015894 , P015895 , P015896 , P015897 , P016101 , P016102 , P0162202 , P0162203 , P0162204 , P0162368 , P0162369 , P0163463 , P016386 , P016387 , P016388 , P016389 , P016392 , P016394 , P0164651¿¿¿ , P0164652¿¿¿ , P0164972 , P0167357 , P0167465 , P0167466 , P0168015 , P0170588 , P0170689 , P0170690 , P0170968¿¿¿ , P0170969¿¿¿ , P0170981 , P0170982 , P0171406 , P0171918 , P0175182 , P0175368 , P0175369 , P0175370 , P0175545 , P0175546 , P0176591 , P0177062 , P0177063 , P017742 , P0178144 , P0178145 , P0178146 , P0178147 , P0178884 , P0178994 , P0179154 , P0180855 , P0180856 , P0180857 , P0180858 , P0181259 , P0181260 , P0182476¿¿¿ , P0183725 , P0183726 , P0183727 , P0183728 , P0183729 , P0183730 , P0183731 , P0183732 , P0184221 , P0184222 , P0185060 , P0185078 , P0185233 , P0185307 , P0185896 , P0189314 , P0189315 , P0189316 , P019458 , P019459 , P019460 , P019461 , P021789 , P021791 , P021888 , P021894 , P022021 , P022022 , P022023 , P022759 , P022760 , P022761 , P023170 , P025354 , P025355 , P025968 , P025969 , P026033 , P026034 , P026125 , P026126 , P026302 , P026303 , P027324 , P027461 , P027954 , P027956 , P029384 , P031686 , P031774 , P032359 , P032387 , P034422 , P034423 , P034732 , P035210 , P035600 , P035705 , P036000 , P037150 , P038617 , P038752 , P038753 , P038894 , P039590 , P040635 , P040812 , P041044 , P041473 , P041617 , P042386 , P042430 , P042473 , P042474 , P042595 , P042596 , P042657 , P042727 , P042728 , P042842 , P043140 , P043232 , P043521 , P045035 , P045367 , P046255 , P046256 , P047337 , P047480 , P049467 , P049468 , P050432 , P050489 , P052233 , P056073 , P056861 , P056862 , P058743 , P060073 , P060179 , P060529 , P060857 , P060858 , P060918 , P060919 , P060920 , P062505 , P062609 , P062847 , P063301 , P063853 , P064395 , P064483 , P064500 , P064575¿¿¿¿ , P064719¿¿¿¿ , P064815 , P064816 , P065140 , P065141 , P065193 , P065314 , P065400 , P065401 , P066218 , P066570 , P066615 , P067019 , P067086 , P067087 , P067393 , P067441 , P067482 , P068002 , P068266 , P068700 , P068988 , P070081 , P070435 , P070443 , P070713 , P070770 , P071349 , P071350 , P071451 , P073778 , P073847 , P073934 , P075170 , P076716 , P076719 , P076849 , P076950 , P078419 , P079322 , P079323 , P080384 , P083938 , P084481 , P084594 , P085062 , P086507 , P086508 , P086830 , P086831 , P087584¿¿¿¿ , P087585 , P087853 , P088197 , P088210 , P088272 , P088570 , P088655 , P092355 , P093705 , P093834 , P094960 , P094961 , P096673 , P096674 , P096675 , P097132 , P097311 , P097500 , P099166 , P099167 , P099444 , P099445 , P099721 , P100052 , P100053 , PO106703 , PO106704 , PO106706 , PO29057 , PO29058 , PO29059 , PO84691 , PO84692 , PO84693 , PO84694 , PO84695 , PO84696 , PO84697 , PO84698 ;   1) Model Number C1080; Serial Numbers (US): 416930 , 766425 , 473239 , 773201 , 851902 , 416917 , 491904 , 545574 , 473246 , 416932 , 473245 , 473238 , 82730 , 783942 , 416919 , 473243 , 545569 , 473234 , 416920 , 416926 , 473236 , 830902 , 416927 , 491906 , 416931 , 416921 , 491895 , 745779 , 561704 , 561689 , 766423 , 416922 , 774737 , 416925 , 416928 , 491908 , 473242 , 491900 , 822726 , 836910 , 545567 , 473233 , 561700 , 491903 , 836902 , 82996 , 473240 , 745778 , 747442 , 561691 , 561695 , 561701 , 491893 , 545579 , 725635 , 304528 , 304527 , 747444 , 745780 , 473241 , 545573 , 816751 , 816748 , 812748 , 836904 , 304530 , 491896 , 545580 , 545570 , 545568 , 561693 , 774730 , 473232 , 491901 , 725633 , 561699;  Serial Numbers (Foreign): 18045 , 18151 , 18152 , 18153 , 18154 , 18155 , 18366 , 93080 , 114143 , 276987 , 366998 , 378830 , 378831 , 386998 , 398332 , 398333 , 442500 , 442501 , 442502 , 475308 , 479362 , 479363 , 479364 , 479365 , 479367 , 493548 , 493549 , 493550 , 493551 , 493552 , 493553 , 493554 , 493555 , 493556 , 493557 , 511571 , 511572 , 511573 , 566464 , 566465 , 602665 , 684642 , 695358 , 701859 , 701948 , 725623 , 725624 , 725625 , 725626 , 725627 , 725629 , 725631 , 745783 , 751632 , 765776 , 765777 , 778573 , 778575 , 778576 , 778577 , 778578 , 822721 , 822721 , 822730 , 822730 , 822731 , 822732 , 830904 , 830905 , 831024 , 831025 , 840921 , 844771 , 848196 , 848199 , 848203 , 848206 , 848207 , 848931 , 857830 , 857831 , 871844 , 873308 , 873309 , 873310 , 873311 , 895437 , 895438 , 901247 , 901248 , 904455 , 905470 , 925888 , 925889 , 925890 , 925901 , 929765 , 929766 , 929769 , 934938 , 934939 , 935002 , 945140 , 945142 , 945143 , 945145 , 945146 , 945147 , C1080 , CO10800103 , CO10800104 , CO10800105 , CO10800106 , CO10800107 , D778574 , G/493557 , P0032040 , P0032041 , P0100336 , P0100409 , P0101003 , P0101329 , P0101330 , P0101331 , P0101332 , P0101333 , P0101510 , P0101511 , P0101512 , P0101513 , P0101514 , P0101515 , P0101516 , P0106045 , P0106415 , P0106416 , P0106417¿¿¿ , P0107780 , P0108277 , P0108715 , P0108716 , P0108717 , P0109782 , P0109783 , P0109954 , P0110233 , P0111026 , P0111276 , P0111872 , P0111873 , P0112155 , P0112156 , P0112157 , P0112158 , P0112955 , P0113151 , P0113364 , P0115133 , P0115301 , P0115302 , P0116051 , P0116614 , P0116615 , P0117468 , P0117880 , P0117881 , P0122188 , P0122460 , P0125766 , P0125767 , P0127623 , P0127849 , P0128225 , P0128420 , P0128875 , P0129052 , P0129155 , P0129451 , P0129572 , P0130354 , P0132046 , P0132726 , P0133301 , P0133777 , P0133835 , P0133836 , P0133837 , P0134515 , P0135473 , P0138875 , P0149304 , P0149305 , P0160131 , P0163554 , P0164049 , P0164216 , P0164438 , P0164532 , P0165078 , P0165079 , P0165080 , P0165081 , P0166465 , P0166466 , P0166834 , P0167425 , P0167651 , P0167682 , P0169577 , P0170398 , P0170970¿¿¿ , P0172128 , P0172129 , P0172423 , P0172424 , P0172425 , P0172426 , P0172623 , P0172718 , P0175806 , P0175807 , P0177061 , P0178143 , P0181261 , P0183905 , P0186868 , P0186870 , P0188652 , P018869 , P0189663 , P0189664 , P019147 , P019148 , P019149 , P019150 , P019151 , P019275 , P019276 , P019277 , P019283 , P019322 , P019323 , P019326 , P019327 , P019801 , P019802 , P019803 , P019804 , P020452 , P020453 , P021332 , P021333 , P021792 , P021794 , P022018 , P022019 , P023898 , P023901 , P024623 , P025150 , P025151 , P026436 , P026437 , P026438 , P026439 , P026440 , P026619 , P026620 , P026621 , P026804 , P027081 , P027082 , P027321 , P027322 , P027323 , P027744 , P027953 , P028299 , P030415 , P030416 , P030863 , P030864 , P031143 , P031144 , P031530 , P034344 , P034345 , P034346 , P034424 , P034425 , P034542 , P034611 , P035841 , P035976 , P035977 , P036591 , P039475 , P039476 , P039591 , P039749 , P041222 , P041223 , P041615 , P041616 , P041932 , P041972 , P041973 , P041974 , P042167 , P042282 , P042284 , P042285 , P048169 , P048170 , P050022 , P050433 , P051327¿¿¿¿ , P051626 , P053704 , P054140 , P054376 , P054593 , P056636 , P056860 , P058001 , P058621 , P058622 , P059546¿¿¿¿ , P059634¿¿¿¿ , P059635¿¿¿¿ , P060675 , P060676 , P062509 , P062978 , P064317 , P065625 , P065806 , P066048 , P067245 , P067504 , P068235 , P068236 , P069079 , P069178 , P069334 , P069335 , P069615 , P070949 , P071725 , P073528 , P075036 , P075802 , P075946 , P076331 , P076526 , P079075 , P079321 , P081497 , P081583 , P081584 , P081854 , P082674 , P083243 , P083666 , P084374 , P086826 , P087950 , P088488 , P094207 , P094411 , P096286 , P097604 , P0M6050 , PO137838 , PO146396 , PO190805 , PO19275 , PO47568 , R61213
Recalling Firm/
Manufacturer
Arjo Hospital Equipment AB
Verkstadsvagen 5
Eslov Sweden
For Additional Information Contact
800-323-1245 Ext. 57985
Manufacturer Reason
for Recall
There is a low but existing entrapment risk for a patient being treated on the Contoura 1000 and Contoura 1080 bariatric beds if they have a BMI < 40.
FDA Determined
Cause 2
Labeling design
ActionArjoHuntleigh sent an URGENT FIELD SAFETY NOTICE - IMMEDIATE ATTENTION REQUIRED letters dated March 17, 2014 to all affected cusomers. The letters included instructions for customers to: 1) ensure that the Contoura Bed Field Safety Notice brought to the attention of all caregivers and clinical staff and they are aware that the Contoura 1000 or 1080 medical bed is intended to be used as a bariatric medical bed, designed for the patients with BMI greater than 40; 2) ensure that all users of the Contoura 1000/1080 bed, applicable staff and caregivers are trained to ensure awareness of the revised Instructions for Use Manual and new bed label; 3) discard all previous versions of Contoura 1000/1080 Bed Instructions for Use Manuals and ensure that a copy of this Contoura Bed Field Safety Notice is placed in the enclosed revised Contoura 1000/1080 bed Instructions for Use Manual (746-173-AM-3 - 10/2013); 4) where applicable, include this requirement in subsequent training of applicable staff and caregivers; and, 5) complete and sign the enclosed Customer Response Form and return this form to the ArjoHuntleigh Regulatory Department. An ArjoHuntleigh Service Technician will contact the designated facility representative listed on the Customer Response Form to schedule service to apply the two labels shown above, free of charge. For question contact ArjoHuntleigh at 1-800-323-1245, ext. 57985.
Quantity in Commerce1,695 beds (worldwide)
DistributionWorldwide Distribution - US: Nationwide in the states of (AZ, CA, CO, CT, GA, ID, IL, MA, MD, MN, NC, NE, OH, PA, RI, SD, TX, VA, WI); and the countries of Australia, Austria, Bahrain, Belgium, Canada, Denmark, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Lebanon, Mexico, Netherlands, New Zealand, Norway, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-