| | Class 2 Device Recall bioMerieux PREVI Isola System |  |
| Date Initiated by Firm | February 10, 2014 |
| Date Posted | March 12, 2014 |
| Recall Status1 |
Terminated 3 on March 16, 2017 |
| Recall Number | Z-1177-2014 |
| Recall Event ID |
67463 |
| Product Classification |
General Purpose, microbiology, diagnostic Medical Device - Product Code LIB
|
| Product | bioMerieux PREVI Isola System (General purpose,microbiology, diagnostic Medical Device), (REF 29500/29500R), bioMerieux, Hazelwood, MO. |
| Code Information |
Serial number range: AS 180-00001 to AS 180-00476 |
| FEI Number |
1950204
|
Recalling Firm/ Manufacturer |
Biomerieux Inc 595 Anglum Rd Hazelwood MO 63042-2320
|
| For Additional Information Contact | Ryan LeMelle 314-731-8526 |
Manufacturer Reason for Recall | The firm has determined the product may fail to dispense the sample to the agar plate resulting in a "failure to dispense" in conjunction with urine specimens. This could lead to a "false" negative growth result. |
FDA Determined Cause 2 | Software design |
| Action | The firm sent a recall notification letter to all of their consignees on 2/10/2014. |
| Quantity in Commerce | 471 instruments |
| Distribution | Distribution was made to AK, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, MA, MO, NC, NJ, NY, OH, OR, PA, RI, SC, TN, TX, VA and WI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|