• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Synthes 5.0mm Variable Locking Screw

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Synthes 5.0mm Variable Locking Screwsee related information
Date Initiated by FirmFebruary 05, 2014
Date PostedMarch 04, 2014
Recall Status1 Terminated 3 on August 31, 2015
Recall NumberZ-1125-2014
Recall Event ID 67498
510(K)NumberK110354 
Product Classification condylar plate fixation Implant - Product Code JDP
ProductSynthes 5.0mm Variable Locking Screw This device is part of the Variable Angle Locking Compression Plate family and this technology included variable angle Combi holes in the plate shaft.The variable angle Combi hole provides the flexibility of axial compression and variable locking capacity throughout the length of the plate shaft.
Code Information part no. 02.231.240, lot no. 823111
FEI Number 3005180112
Recalling Firm/
Manufacturer
Synthes, Inc.
1302 Wrights Ln E
West Chester PA 19380-3417
For Additional Information ContactCustomer Support
610-719-5000
Manufacturer Reason
for Recall
Labeling correction: The package insert in one 5.0mm Variable Locking Screw package was a GP0006 (Anterior Cervical Vertebrae Plate System) insert, which should have been a GP2848 (4.5mm VA-LCP Curved Condylar Plate System insert).
FDA Determined
Cause 2
Employee error
ActionSynthes sent an Urgent Field Safety Notification/Medical Device Labeling Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Exchage the previously shipped package insert (GP0006) with the correct package insert (GP2848 provided with the notification). Review the package insert. Forward the Field Safety Notification to anyone in their facility that needs to be informed. If the package insert has been forwarded to another facility, contact that facility. Maintain awareness of this notice until all package inserts have been exchanged. Maintain a copy of the notice. Customers with questions were instructed to maintain a copy of the notice. For questions regarding this recall call 610-719-5000.
Quantity in Commerce45
DistributionNationwide Distribution including GA, CA, TX, MO, OH, PA, WA, AL, IN, AZ, MI, IA, TN, KY, NC, VA, NY, LA, IL, UT, and KS.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = JDP
-
-