 Class 2 Device Recall MAQUETsee related information
Date Initiated by FirmJanuary 31, 2014
Date PostedMarch 10, 2014
Recall Status1 Terminated 3 on April 24, 2015
Recall NumberZ-1165-2014
Recall Event ID 67515
510(K)NumberK090555 
Product Classification Heat-exchanger, cardiopulmonary bypass - Product Code DTR
ProductMAQUET PLEGIOX Cardioplegia Heat Exchanger The PLEGIOX Cardioplegia Heat Exchanger is used as part of a cardioplegia system to set and maintain temperature for given flow rates and within the given temperature range of blood cardioplegic and crystalloid cardioplegic solutions during extracorporeal circulation. The product is designed for single use only, for an application period of no longer than 6 hours.
Code Information Order Number Model Number and Description Lot number 70105.0647 BO-TOP 15200 OR Pack - HMO and RF 15674-05 70105.0647 BO-TOP 15200 OR Pack - HMO and RF 15929-05 70105.0647 BO-TOP 15200 OR Pack - HMO and RF 16115-06 70105.0647 BO-TOP 15200 OR Pack - HMO and RF 16346-06 70105.0967 BO-T 1260 Adult Cardioplegia Pack 15760-07 70105.0975 BO-TOP 26300 Custom Pack 15796-13 70105.1736 BEQ-TOP 4903 Open Heart Pack 15921-07 70105.1736 BEQ-TOP 4903 Open Heart Pack 16073-07 70105.1736 BEQ-TOP 4903 Open Heart Pack 16153-07 70105.1736 BEQ-TOP 4903 Open Heart Pack 16321-07 70105.1861 BO-TOP 14601 Open Heart Tubing Pack 15880-02 70105.2463 BO-T 25102 Custom Pack 15755-07 70105.2463 BO-T 25102 Custom Pack 15889-07 70105.2463 BO-T 25102 Custom Pack 16046-07 70105.2463 BO-T 25102 Custom Pack 16155-07 70105.2463 BO-T 25102 Custom Pack 16267-07 70105.2650 BO-TOP 23102 Scott &White OR CPB Pack 15818-10 70105.2659 BO-TOP 4903 Adult Open Heart Pack 15780-03 70105.2659 BO-TOP 4903 Adult Open Heart Pack 16047-03 70105.3832 BO-TOP 32400 Adult CPB Pack 15814-06 70105.3832 BO-TOP 32400 Adult CPB Pack 16063-06 70105.3832 BO-TOP 32400 Adult CPB Pack 16449-06 70105.5015 BO-TOP 36500 Cardiac Surgery 15658-13 70105.5015 BO-TOP 36500 Cardiac Surgery 15772-13 70105.5015 BO-TOP 36500 Cardiac Surgery 15883-16 70105.5015 BO-TOP 36500 Cardiac Surgery 16078-16 70105.5015 BO-TOP 36500 Cardiac Surgery 16250-16 70105.5015 BO-TOP 36500 Cardiac Surgery 16410-17 70105.6041 BEQ-T 4906 Pediatric CPB Pack w/o VKMO 15819-02 70105.6238 BO-TOP 38800 Open Heart Pack 15635-05 70105.6238 BO-TOP 38800 Open Heart Pack 16013-05 70105.6867 BO-TOP 4112 1/4x3/16 Neonatal Pack 16039-09 70105.6867 BO-TOP 4112 1/4x3/16 Neonatal Pack 16238-10 70105.6869 BO-TOP 4113 Pediatric 1/4x3/8 15574-03 70106.2237 T 24260 St Louis Children¿s Cardioplegia 15659-05 70106.2237 T 24260 St Louis Children¿s Cardioplegia 15967-05 70106.2707 BEQ-TOP 40000 Open Heart Pack 15892-05 70106.2707 BEQ-TOP 40000 Open Heart Pack 16122-07 70106.2707 BEQ-TOP 40000 Open Heart Pack 16224-07 70106.3071 BEQ-TOP 30901 Custom CPB Pack 15783-05 70106.3071 BEQ-TOP 30901 Custom CPB Pack 15923-05 70106.3071 BEQ-TOP 30901 Custom CPB Pack 15998-05 70106.3071 BEQ-TOP 30901 Custom CPB Pack 16239-05 70106.3071 BEQ-TOP 30901 Custom CPB Pack 16387-05 70106.3364 BO-T 41760 Robert Packer Plegiox Set 15845-03 70106.3364 BO-T 41760 Robert Packer Plegiox Set 15976-03 70106.3364 BO-T 41760 Robert Packer Plegiox Set 16120-03 70106.3623 BO-T 31261 Plegiox Pack 15800-04 70106.3623 BO-T 31261 Plegiox Pack 15924-04 70106.3623 BO-T 31261 Plegiox Pack 16112-04 70106.3623 BO-T 31261 Plegiox Pack 16145-04 70106.3623 BO-T 31261 Plegiox Pack 16184-04 70106.4327 BO-TOP 48400 OR Pack 16169-01 70106.4327 BO-TOP 48400 OR Pack 16310-01 70106.5315 BEQ-T 16403 Liver TX 16114-06 70106.6206 BEQ-TOP 38307 Adult Pack 15931-03 70106.6398 BEQ-TOP 48000 Open Heart WITH Oxygenator 15932-02 70106.6400 BEQ-TOP 48001 Open Heart WITH OXY 15933-02 70106.6461 BEQ-TOP 38310 Small Adult Pack 15934-02 70106.6664 BEQ-T 4903 Open Heart Pack with NO VKMO 16040-01 70106.6712 BO-T 14601 Open Heart WITHOUT VKMO 16067-01 70106.6713 BEQ-T 40000 Open Heart WITHOUT VKMO 16056-01 70106.6738 BO-T 26300 Transitional Cardiac Surgery Pk 16146-01 70106.6914 BEQ-T 4961 Cardioplegia Pack 4:1 16240-01
FEI Number 3008355164
Recalling Firm/
Manufacturer
Maquet Cardiovascular Us Sales, Llc
45 Barbour Pond Drive
Wayne NJ 07470
For Additional Information Contact
888-627-8383
Manufacturer Reason
for Recall
The MAQUET post-marketing vigilance program has identified a discrepancy that the desired cardioplegia temperature may not be obtained when the PLEGIOX Cardioplegia Heat Exchanger (PLEGIOX) is used in conjunction with a heater/cooler unit.
FDA Determined
Cause 2
Device Design
ActionMaquet Inc.sent an" Field Safety Notice" dated January 31, 2014, to all affected customers via Fed Ex. The letter identified the product the problem and the action needed to be taken by the customer. Any questions or additional information, can be obtained from the MAQUET representative, or MAQUET Customer Service at 1-888-627-8383 (press option 2, followed by option 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST. Please complete and return the enclosed confirmation form, acknowledging your receipt and understanding of this communication, by emailing a scanned copy to Fieldactions@maquet.com or via fax at (973) 807-9290.
Quantity in Commerce1,424 units
DistributionWorldwide Distribution - USA (nationwide) and the countries of Brazil, Germany and Turkey
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DTR
-
-