Class 2 Device Recall EZPass Suture Retriever Nitinol Single Pack

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall EZPass Suture Retriever Nitinol Single Packsee related information
Date Initiated by FirmApril 15, 2014
Date PostedJune 16, 2014
Recall Status1 Terminated 3 on July 07, 2015
Recall NumberZ-1801-2014
Recall Event ID 68216
Product Classification Orthopedic manual surgical instrument - Product Code LXH
ProductEZPass Suture Retriever, Nitinol, Single Pack, Part number 10007380. Orthopedic Manual Surgical Instrument, Sterile, single use arthroscopic instrument utilized in conjunction with a nitinol wire to aid in passing suture through soft tissue.
Code Information Part numbers 110007380 Lot 223570 and 803860
FEI Number 1825034
Recalling Firm/
Manufacturer
Biomet, Inc.
56 E Bell Dr
Warsaw IN 46582-6989
For Additional Information Contact
574-267-6639 Ext. 1676
Manufacturer Reason
for Recall
The nitinol wire or monofilament may buckle within the EZPass Suture Passer and become kinked making the wire/monofilament difficult to use.
FDA Determined
Cause 2
Device Design
ActionOn 2/28/14, customers received a letter stating these devices have been recalled following an investigation which identified that the nitinol wire or monofilament may buckle within the EzPass Suture Passer and become kinked making the wire/monofilament difficult to use. The letter also stated, there is no adverse health outcome expected for the patient and this action requires the immediate location and discontinued use of the product and its return to Biomet.
Quantity in Commerce100 units
DistributionNationwide Distribution.
Total Product Life CycleTPLC Device Report