|
Class 1 Device Recall EminiC |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
June 11, 2014 |
Date Posted |
July 18, 2014 |
Recall Status1 |
Terminated 3 on March 29, 2016 |
Recall Number |
Z-2068-2014 |
Recall Event ID |
68344 |
510(K)Number |
K052582 K043276
|
Product Classification |
Analyzer, gas, carbon-dioxide, gaseous-phase - Product Code CCK
|
Product |
DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES.
DATEX-OHMEDA S/5TM FM WITH L-FICU04 ANDL-FICUO4A SOFTWARE AND N-FCREC MODULE **(OUS only).
E-miniC and accessories is indicated for monitoring C02 and respiration rate of all hospital patients. |
Code Information |
Serial # Model # 6837005 E-MINIC-00 6838210 E-MINIC-00 6845207 E-MINIC-00 6845218 E-MINIC-00 6845228 E-MINIC-00 6845248 E-MINIC-00 6860768 E-MINIC-00 6860858 E-MINIC-00 6876048 E-MINIC-00 6876050 E-MINIC-00 6876052 E-MINIC-00 6876054 E-MINIC-00 6878048 E-MINIC-00 6888716 E-MINIC-00 6888726 E-MINIC-00 6888755 E-MINIC-00 6889935 E-MINIC-00 6889950 E-MINIC-00 6891094 E-MINIC-00 6891114 E-MINIC-00 6891115 E-MINIC-00 6891116 E-MINIC-00 6891117 E-MINIC-00 6891121 E-MINIC-00 6891130 E-MINIC-00 6891137 E-MINIC-00 6891139 E-MINIC-00 6892616 E-MINIC-00 6892626 E-MINIC-00 6827924 E-MINIC-00 6829318 E-MINIC-00 6829321 E-MINIC-00 6832785 E-MINIC-00 6832787 E-MINIC-00 6832788 E-MINIC-00 6832789 E-MINIC-00 6832791 E-MINIC-00 6832795 E-MINIC-00 6832796 E-MINIC-00 6832797 E-MINIC-00 6832802 E-MINIC-00 6832803 E-MINIC-00 6832804 E-MINIC-00 6832805 E-MINIC-00 6832813 E-MINIC-00 6832815 E-MINIC-00 6832824 E-MINIC-00 6832825 E-MINIC-00 6832831 E-MINIC-00 6834874 E-MINIC-00 6834877 E-MINIC-00 6834880 E-MINIC-00 6834883 E-MINIC-00 6834884 E-MINIC-00 6834885 E-MINIC-00 6834886 E-MINIC-00 6834887 E-MINIC-00 6834888 E-MINIC-00 6834889 E-MINIC-00 6834890 E-MINIC-00 6834891 E-MINIC-00 6834892 E-MINIC-00 6834893 E-MINIC-00 6834894 E-MINIC-00 6834895 E-MINIC-00 6834896 E-MINIC-00 6834898 E-MINIC-00 6834899 E-MINIC-00 6834902 E-MINIC-00 6834903 E-MINIC-00 6834904 E-MINIC-00 6834907 E-MINIC-00 6834909 E-MINIC-00 6834910 E-MINIC-00 6834911 E-MINIC-00 6834913 E-MINIC-00 6834914 E-MINIC-00 6834915 E-MINIC-00 6834916 E-MINIC-00 6834917 E-MINIC-00 6834918 E-MINIC-00 6834919 E-MINIC-00 6834920 E-MINIC-00 6834921 E-MINIC-00 6834922 E-MINIC-00 6836999 E-MINIC-00 6837002 E-MINIC-00 6837003 E-MINIC-00 6837010 E-MINIC-00 6837011 E-MINIC-00 6837013 E-MINIC-00 6838187 E-MINIC-00 6838196 E-MINIC-00 6838197 E-MINIC-00 6838198 E-MINIC-00 6842192 E-MINIC-00 6842196 E-MINIC-00 6842198 E-MINIC-00 6842212 E-MINIC-00 6842221 E-MINIC-00 6848886 E-MINIC-00 6848887 E-MINIC-00 6848888 E-MINIC-00 6852650 E-MINIC-00 6852655 E-MINIC-00 6855674 E-MINIC-00 6860683 E-MINIC-00 6860685 E-MINIC-00 6860698 E-MINIC-00 6860750 E-MINIC-00 6873156 E-MINIC-00 6879423 E-MINIC-00 6883991 E-MINIC-00 6883994 E-MINIC-00 6883995 E-MINIC-00 6883997 E-MINIC-00 6884000 E-MINIC-00 6884007 E-MINIC-00 6884014 E-MINIC-00 6884016 E-MINIC-00 6884022 E-MINIC-00 6884023 E-MINIC-00 6887178 E-MINIC-00 6887182 E-MINIC-00 6887183 E-MINIC-00 6887185 E-MINIC-00 6887186 E-MINIC-00 6887187 E-MINIC-00 6887189 E-MINIC-00 6887190 E-MINIC-00 6887194 E-MINIC-00 6887195 E-MINIC-00 6887199 E-MINIC-00 6887201 E-MINIC-00 6887203 E-MINIC-00 6887206 E-MINIC-00 6887209 E-MINIC-00 6887210 E-MINIC-00 6887221 E-MINIC-00 6887222 E-MINIC-00 6888655 E-MINIC-00 6888670 E-MINIC-00 6888672 E-MINIC-00 6888681 E-MINIC-00 6888683 E-MINIC-00 6888684 E-MINIC-00 6888686 E-MINIC-00 6888688 E-MINIC-00 6888699 E-MINIC-00 6888703 E-MINIC-00 6888749 E-MINIC-00 6888750 E-MINIC-00 6888754 E-MINIC-00 6888771 E-MINIC-00 6888772 E-MINIC-00 6888773 E-MINIC-00 6889916 E-MINIC-00 6889939 E-MINIC-00 6889967 E-MINIC-00 6891144 E-MINIC-00 6891146 E-MINIC-00 6891148 E-MINIC-00 6891150 E-MINIC-00 6891152 E-MINIC-00 6891162 E-MINIC-00 6891180 E-MINIC-00 6891182 E-MINIC-00 6891184 E-MINIC-00 6896279 E-MINIC-00 6896303 E-MINIC-00 6838209 E-MINIC-00 6838211 E-MINIC-00 6838212 E-MINIC-00 6892655 E-MINIC-00 6892720 E-MINIC-00 6832823 E-MINIC-00 6838243 E-MINIC-00 6838252 E-MINIC-00 6838256 E-MINIC-00 6838262 E-MINIC-00 6842175 E-MINIC-00 6842181 E-MINIC-00 6842190 E-MINIC-00 6842204 E-MINIC-00 6842206 E-MINIC-00 6842217 E-MINIC-00 6842218 E-MINIC-00 6842219 E-MINIC-00 6842224 E-MINIC-00 6842225 E-MINIC-00 6842226 E-MINIC-00 6842227 E-MINIC-00 6842228 E-MINIC-00 6842230 E-MINIC-00 6842231 E-MINIC-00 6842233 E-MINIC-00 6842234 E-MINIC-00 6842236 E-MINIC-00 6842239 E-MINIC-00 6842242 E-MINIC-00 6842244 E-MINIC-00 6848877 E-MINIC-00 6848878 E-MINIC-00 6848881 E-MINIC-00 6848883 E-MINIC-00 6848884 E-MINIC-00 6848885 E-MINIC-00 6848889 E-MINIC-00 6848895 E-MINIC-00 6848897 E-MINIC-00 6852617 E-MINIC-00 6852641 E-MINIC-00 6852656 E-MINIC-00 6852662 E-MINIC-00 6852667 E-MINIC-00 6855627 E-MINIC-00 6855630 E-MINIC-00 6855719 E-MINIC-00 6860661 E-MINIC-00 6860720 E-MINIC-00 6860799 E-MINIC-00 6860821 E-MINIC-00 6860823 E-MINIC-00 6860825 E-MINIC-00 6860827 E-MINIC-00 6867789 E-MINIC-00 6870556 E-MINIC-00 6870574 E-MINIC-00 6870594 E-MINIC-00 6873165 E-MINIC-00 6873182 E-MINIC-00 6873190 E-MINIC-00 6876007 E-MINIC-00 6878050 E-MINIC-00 6878054 E-MINIC-00 6878056 E-MINIC-00 6883990 E-MINIC-00 6883998 E-MINIC-00 6884001 E-MINIC-00 6884006 E-MINIC-00 6884008 E-MINIC-00 6884010 E-MINIC-00 6884015 E-MINIC-00 6884020 E-MINIC-00 6887188 E-MINIC-00 6887198 E-MINIC-00 6887204 E-MINIC-00 6887211 E-MINIC-00 6887215 E-MINIC-00 6887216 E-MINIC-00 6887226 E-MINIC-00 6888661 E-MINIC-00 6888711 E-MINIC-00 6888719 E-MINIC-00 6888724 E-MINIC-00 6889920 E-MINIC-00 6889936 E-MINIC-00 6889942 E-MINIC-00 6889943 E-MINIC-00 6889948 E-MINIC-00 6889954 E-MINIC-00 6889957 E-MINIC-00 6889958 E-MINIC-00 6889959 E-MINIC-00 6889963 E-MINIC-00 6889966 E-MINIC-00 6891059 E-MINIC-00 6891069 E-MINIC-00 6891070 E-MINIC-00 6891080 E-MINIC-00 6891081 E-MINIC-00 6891090 E-MINIC-00 6891131 E-MINIC-00 6891143 E-MINIC-00 6892659 E-MINIC-00 6892665 E-MINIC-00 6892676 E-MINIC-00 6892680 E-MINIC-00 6892681 E-MINIC-00 6892694 E-MINIC-00 6892700 E-MINIC-00 6892701 E-MINIC-00 6895014 E-MINIC-00 6896275 E-MINIC-00 6896372 E-MINIC-00 6898678 E-MINIC-00 6898687 E-MINIC-00 6898696 E-MINIC-00 6838247 E-MINIC-00 6838251 E-MINIC-00 6838254 E-MINIC-00 6838259 E-MINIC-00 6842167 E-MINIC-00 6842182 E-MINIC-00 6868294 E-MINIC-00 6868320 E-MINIC-00 6870587 E-MINIC-00 6873125 E-MINIC-00 6873128 E-MINIC-00 6873130 E-MINIC-00 6873131 E-MINIC-00 6873139 E-MINIC-00 6873141 E-MINIC-00 6824774 E-MINIC-00 6824782 E-MINIC-00 6824783 E-MINIC-00 6860780 E-MINIC-00 6845224 E-MINIC-00 6845235 E-MINIC-00 6824790 E-MINIC-00 6824792 E-MINIC-00 6824797 E-MINIC-00 6824799 E-MINIC-00 6824800 E-MINIC-00 6824802 E-MINIC-00 6824804 E-MINIC-00 6824811 E-MINIC-00 6824812 E-MINIC-00 6824818 E-MINIC-00 6832801 E-MINIC-00 6832817 E-MINIC-00 6834901 E-MINIC-00 6836976 E-MINIC-00 6836980 E-MINIC-00 6829311 E-MINIC-00 6848912 E-MINIC-00 6860830 E-MINIC-00 6832821 E-MINIC-00 6842180 E-MINIC-00 6860849 E-MINIC-00 6889956 E-MINIC-00 6891173 E-MINIC-00 6892607 E-MINIC-00 6832790 E-MINIC-00 6845227 E-MINIC-00 6891055 E-MINIC-00 6892614 E-MINIC-00 6822945 E-MINIC-00 6822949 E-MINIC-00 6832794 E-MINIC-00 6832812 E-MINIC-00 6836997 E-MINIC-00 6836998 E-MINIC-00 6838214 E-MINIC-00 6838224 E-MINIC-00 6838228 E-MINIC-00 6838236 E-MINIC-00 6838237 E-MINIC-00 6838245 E-MINIC-00 6842177 E-MINIC-00 6842241 E-MINIC-00 6845206 E-MINIC-00 6848890 E-MINIC-00 6848898 E-MINIC-00 6848908 E-MINIC-00 6848909 E-MINIC-00 6848913 E-MINIC-00 6848915 E-MINIC-00 6848916 E-MINIC-00 6848942 E-MINIC-00 6852653 E-MINIC-00 6852661 E-MINIC-00 6852678 E-MINIC-00 6860438 E-MINIC-00 6860667 E-MINIC-00 6860777 E-MINIC-00 6870568 E-MINIC-00 6873123 E-MINIC-00 6873154 E-MINIC-00 6888695 E-MINIC-00 6888712 E-MINIC-00 6888738 E-MINIC-00 6891128 E-MINIC-00 6891168 E-MINIC-00 6891175 E-MINIC-00 6891176 E-MINIC-00 6891178 E-MINIC-00 6891181 E-MINIC-00 6892618 E-MINIC-00 6892625 E-MINIC-00 6892646 E-MINIC-00 6892693 E-MINIC-00 6892698 E-MINIC-00 6868321 E-MINIC-00 6842220 E-MINIC-00 6848914 E-MINIC-00 6848924 E-MINIC-00 6848936 E-MINIC-00 6852611 E-MINIC-00 6852612 E-MINIC-00 6852619 E-MINIC-00 6852622 E-MINIC-00 6852664 E-MINIC-00 6852666 E-MINIC-00 6852668 E-MINIC-00 6852669 E-MINIC-00 6852674 E-MINIC-00 6852675 E-MINIC-00 6852676 E-MINIC-00 6852677 E-MINIC-00 6852679 E-MINIC-00 6852680 E-MINIC-00 6852681 E-MINIC-00 6852682 E-MINIC-00 6855629 E-MINIC-00 6855639 E-MINIC-00 6855643 E-MINIC-00 6855647 E-MINIC-00 6855649 E-MINIC-00 6855651 E-MINIC-00 6855653 E-MINIC-00 6855660 E-MINIC-00 6855661 E-MINIC-00 6855664 E-MINIC-00 6855671 E-MINIC-00 6855673 E-MINIC-00 6855678 E-MINIC-00 6855680 E-MINIC-00 6855681 E-MINIC-00 6855683 E-MINIC-00 6855685 E-MINIC-00 6855686 E-MINIC-00 6855691 E-MINIC-00 6855695 E-MINIC-00 6855699 E-MINIC-00 6855701 E-MINIC-00 6855702 E-MINIC-00 6855704 E-MINIC-00 6855705 E-MINIC-00 6855709 E-MINIC-00 6855712 E-MINIC-00 6855713 E-MINIC-00 6855715 E-MINIC-00 6855717 E-MINIC-00 6855718 E-MINIC-00 6855722 E-MINIC-00 6855725 E-MINIC-00 6883984 E-MINIC-00 6884013 E-MINIC-00 6884019 E-MINIC-00 6895016 E-MINIC-00 6896270 E-MINIC-00 6896308 E-MINIC-00 6896310 E-MINIC-00 6896312 E-MINIC-00 6896334 E-MINIC-00 6896351 E-MINIC-00 6896356 E-MINIC-00 6896357 E-MINIC-00 6898635 E-MINIC-00 6898659 E-MINIC-00 6898679 E-MINIC-00 6898693 E-MINIC-00 6898697 E-MINIC-00 6898706 E-MINIC-00 6898707 E-MINIC-00 6898708 E-MINIC-00 6898715 E-MINIC-00 6898716 E-MINIC-00 6898720 E-MINIC-00 6898724 E-MINIC-00 6898726 E-MINIC-00 6898727 E-MINIC-00 6898729 E-MINIC-00 6898732 E-MINIC-00 6832827 E-MINIC-00 6842222 E-MINIC-00 6842238 E-MINIC-00 6855656 E-MINIC-00 6855677 E-MINIC-00 6855720 E-MINIC-00 6855728 E-MINIC-00 6855729 E-MINIC-00 6855732 E-MINIC-00 6860439 E-MINIC-00 6860444 E-MINIC-00 6860450 E-MINIC-00 6860670 E-MINIC-00 6860672 E-MINIC-00 6860761 E-MINIC-00 6860772 E-MINIC-00 6860773 E-MINIC-00 6860778 E-MINIC-00 6883980 E-MINIC-00 6883986 E-MINIC-00 6883988 E-MINIC-00 6884003 E-MINIC-00 6884004 E-MINIC-00 6884005 E-MINIC-00 6887205 E-MINIC-00 6888687 E-MINIC-00 6888693 E-MINIC-00 6889965 E-MINIC-00 6891151 E-MINIC-00 6891174 E-MINIC-00 6892656 E-MINIC-00 6892672 E-MINIC-00 6892697 E-MINIC-00 6894976 E-MINIC-00 6894998 E-MINIC-00 6895005 E-MINIC-00 6895021 E-MINIC-00 6895022 E-MINIC-00 6895023 E-MINIC-00 6895025 E-MINIC-00 6896262 E-MINIC-00 6896264 E-MINIC-00 6896266 E-MINIC-00 6896268 E-MINIC-00 6896269 E-MINIC-00 6896271 E-MINIC-00 6896272 E-MINIC-00 6896273 E-MINIC-00 6896274 E-MINIC-00 6896276 E-MINIC-00 6896280 E-MINIC-00 6896282 E-MINIC-00 6896283 E-MINIC-00 6896287 E-MINIC-00 6896290 E-MINIC-00 6896291 E-MINIC-00 6896292 E-MINIC-00 6896294 E-MINIC-00 6896296 E-MINIC-00 6896297 E-MINIC-00 6896298 E-MINIC-00 6896301 E-MINIC-00 6896307 E-MINIC-00 6896309 E-MINIC-00 6896311 E-MINIC-00 6896313 E-MINIC-00 6896315 E-MINIC-00 6896317 E-MINIC-00 6896322 E-MINIC-00 6896323 E-MINIC-00 6896332 E-MINIC-00 6896333 E-MINIC-00 6896335 E-MINIC-00 6896336 E-MINIC-00 6896337 E-MINIC-00 6896338 E-MINIC-00 6896339 E-MINIC-00 6896340 E-MINIC-00 6896341 E-MINIC-00 6896342 E-MINIC-00 6896343 E-MINIC-00 6896344 E-MINIC-00 6896345 E-MINIC-00 6896346 E-MINIC-00 6896347 E-MINIC-00 6896348 E-MINIC-00 6896349 E-MINIC-00 6896350 E-MINIC-00 6896352 E-MINIC-00 6896353 E-MINIC-00 6896354 E-MINIC-00 6896355 E-MINIC-00 6896358 E-MINIC-00 6896359 E-MINIC-00 6896360 E-MINIC-00 6896361 E-MINIC-00 6896362 E-MINIC-00 6896363 E-MINIC-00 6896364 E-MINIC-00 6896365 E-MINIC-00 6896366 E-MINIC-00 6896367 E-MINIC-00 6896368 E-MINIC-00 6896369 E-MINIC-00 6896370 E-MINIC-00 6896371 E-MINIC-00 6896373 E-MINIC-00 6896374 E-MINIC-00 6896375 E-MINIC-00 6896376 E-MINIC-00 6898627 E-MINIC-00 6898628 E-MINIC-00 6898629 E-MINIC-00 6898631 E-MINIC-00 6898632 E-MINIC-00 6898633 E-MINIC-00 6898634 E-MINIC-00 6898636 E-MINIC-00 6898638 E-MINIC-00 6898639 E-MINIC-00 6898640 E-MINIC-00 6898641 E-MINIC-00 6898642 E-MINIC-00 6898643 E-MINIC-00 6898644 E-MINIC-00 6898645 E-MINIC-00 6898646 E-MINIC-00 6898649 E-MINIC-00 6898650 E-MINIC-00 6898651 E-MINIC-00 6898653 E-MINIC-00 6898654 E-MINIC-00 6898655 E-MINIC-00 6898656 E-MINIC-00 6898657 E-MINIC-00 6898658 E-MINIC-00 6898660 E-MINIC-00 6898661 E-MINIC-00 6898662 E-MINIC-00 6898663 E-MINIC-00 6898664 E-MINIC-00 6898665 E-MINIC-00 6898666 E-MINIC-00 6898667 E-MINIC-00 6898668 E-MINIC-00 6898669 E-MINIC-00 6898670 E-MINIC-00 6898671 E-MINIC-00 6898672 E-MINIC-00 6898673 E-MINIC-00 6898674 E-MINIC-00 6898675 E-MINIC-00 6898676 E-MINIC-00 6898677 E-MINIC-00 6898680 E-MINIC-00 6898681 E-MINIC-00 6898682 E-MINIC-00 6898683 E-MINIC-00 6898684 E-MINIC-00 6898685 E-MINIC-00 6898686 E-MINIC-00 6898688 E-MINIC-00 6898691 E-MINIC-00 6898699 E-MINIC-00 6898700 E-MINIC-00 6898703 E-MINIC-00 6898705 E-MINIC-00 6898710 E-MINIC-00 6898713 E-MINIC-00 6898717 E-MINIC-00 6898718 E-MINIC-00 6898719 E-MINIC-00 6898722 E-MINIC-00 6898723 E-MINIC-00 6898725 E-MINIC-00 6898728 E-MINIC-00 6898735 E-MINIC-00 6898737 E-MINIC-00 6898742 E-MINIC-00 6898746 E-MINIC-00 6898747 E-MINIC-00 6898748 E-MINIC-00 6898758 E-MINIC-00 6898759 E-MINIC-00 6898764 E-MINIC-00 6898766 E-MINIC-00 6898767 E-MINIC-00 6898773 E-MINIC-00 6898775 E-MINIC-00 6898776 E-MINIC-00 6898777 E-MINIC-00 6824807 E-MINIC-00 6824814 E-MINIC-00 6824815 E-MINIC-00 6824816 E-MINIC-00 6824821 E-MINIC-00 6827897 E-MINIC-00 6827938 E-MINIC-00 6829322 E-MINIC-00 6832806 E-MINIC-00 6832807 E-MINIC-00 6832808 E-MINIC-00 6832816 E-MINIC-00 6891063 E-MINIC-00 6891078 E-MINIC-00 6891084 E-MINIC-00 6891110 E-MINIC-00 6891113 E-MINIC-00 6891120 E-MINIC-00 6845231 E-MINIC-00 6855657 E-MINIC-00 6855662 E-MINIC-00 6855687 E-MINIC-00 6855688 E-MINIC-00 6860731 E-MINIC-00 6887197 E-MINIC-00 6888696 E-MINIC-00 6888698 E-MINIC-00 6888702 E-MINIC-00 6888713 E-MINIC-00 6888732 E-MINIC-00 6888764 E-MINIC-00 6889955 E-MINIC-00 6889971 E-MINIC-00 6891041 E-MINIC-00 6891054 E-MINIC-00 6891064 E-MINIC-00 6891067 E-MINIC-00 6891082 E-MINIC-00 6891086 E-MINIC-00 6891088 E-MINIC-00 6891091 E-MINIC-00 6891092 E-MINIC-00 6891101 E-MINIC-00 6891102 E-MINIC-00 6891103 E-MINIC-00 6891105 E-MINIC-00 6891106 E-MINIC-00 6891111 E-MINIC-00 6892639 E-MINIC-00 6892651 E-MINIC-00 6892654 E-MINIC-00 6892678 E-MINIC-00 6892682 E-MINIC-00 6892721 E-MINIC-00 6892722 E-MINIC-00 6896261 E-MINIC-00 6852613 E-MINIC-00 6852621 E-MINIC-00 6870549 E-MINIC-00 6870564 E-MINIC-00 6892663 E-MINIC-00 6891153 E-MINIC-00 6891155 E-MINIC-00 6891156 E-MINIC-00 6832822 E-MINIC-00 6838235 E-MINIC-00 6838238 E-MINIC-00 6838240 E-MINIC-00 6892689 E-MINIC-00 6892699 E-MINIC-00 6822971 E-MINIC-00 6842243 E-MINIC-00 6848891 E-MINIC-00 6873143 E-MINIC-00 6883992 E-MINIC-00 6884011 E-MINIC-00 6884017 E-MINIC-00 6827922 E-MINIC-00 6827931 E-MINIC-00 6827932 E-MINIC-00 6827937 E-MINIC-00 6829313 E-MINIC-00 6848905 E-MINIC-00 6891166 E-MINIC-00 6852660 E-MINIC-00 6887213 E-MINIC-00 6888704 E-MINIC-00 6888737 E-MINIC-00 6888742 E-MINIC-00 6888743 E-MINIC-00 6832811 E-MINIC-00 6834923 E-MINIC-00 6834924 E-MINIC-00 6834927 E-MINIC-00 6834932 E-MINIC-00 6836970 E-MINIC-00 6836972 E-MINIC-00 6836978 E-MINIC-00 6837014 E-MINIC-00 6838190 E-MINIC-00 6896278 E-MINIC-00 6870593 E-MINIC-00 6873121 E-MINIC-00 6860684 E-MINIC-00 6860847 E-MINIC-00 6888678 E-MINIC-00 6891119 E-MINIC-00 6891154 E-MINIC-00 6892621 E-MINIC-00 6892627 E-MINIC-00 6892633 E-MINIC-00 6892692 E-MINIC-00 6894986 E-MINIC-00 6896305 E-MINIC-00 6834933 E-MINIC-00 6836975 E-MINIC-00 6836977 E-MINIC-00 6883967 E-MINIC-00 6887217 E-MINIC-00 6887219 E-MINIC-00 6888757 E-MINIC-00 6888765 E-MINIC-00 6888770 E-MINIC-00 6889915 E-MINIC-00 6889917 E-MINIC-00 6889919 E-MINIC-00 6889928 E-MINIC-00 6889932 E-MINIC-00 6891122 E-MINIC-00 6891123 E-MINIC-00 6891124 E-MINIC-00 6891125 E-MINIC-00 6891126 E-MINIC-00 6891132 E-MINIC-00 6891134 E-MINIC-00 6892719 E-MINIC-00 6845246 E-MINIC-00 6824787 E-MINIC-00 6832793 E-MINIC-00 6832798 E-MINIC-00 6832799 E-MINIC-00 6838258 E-MINIC-00 6838261 E-MINIC-00 6870581 E-MINIC-00 6870566 E-MINIC-00 6870572 E-MINIC-00 6870573 E-MINIC-00 6824788 E-MINIC-00 6824793 E-MINIC-00 6824813 E-MINIC-00 6827892 E-MINIC-00 6827904 E-MINIC-00 6827907 E-MINIC-00 6827912 E-MINIC-00 6827913 E-MINIC-00 6827914 E-MINIC-00 6827927 E-MINIC-00 6870582 E-MINIC-00 6873158 E-MINIC-00 6845232 E-MINIC-00 6827916 E-MINIC-00 6827935 E-MINIC-00 6836982 E-MINIC-00 6836995 E-MINIC-00 6837001 E-MINIC-00 6838186 E-MINIC-00 6838189 E-MINIC-00 6838193 E-MINIC-00 6838201 E-MINIC-00 6838204 E-MINIC-00 6838244 E-MINIC-00 6838246 E-MINIC-00 6855716 E-MINIC-00 6855721 E-MINIC-00 6860759 E-MINIC-00 6860791 E-MINIC-00 6883975 E-MINIC-00 6883976 E-MINIC-00 6883977 E-MINIC-00 6883978 E-MINIC-00 6883979 E-MINIC-00 6883983 E-MINIC-00 6888697 E-MINIC-00 6888700 E-MINIC-00 6888707 E-MINIC-00 6892643 E-MINIC-00 6892660 E-MINIC-00 6892675 E-MINIC-00 6892683 E-MINIC-00 6892695 E-MINIC-00 6892702 E-MINIC-00 6892703 E-MINIC-00 6894971 E-MINIC-00 6894972 E-MINIC-00 6894983 E-MINIC-00 6896327 E-MINIC-00 6898702 E-MINIC-00 6898714 E-MINIC-00 6898721 E-MINIC-00 6898730 E-MINIC-00 6898731 E-MINIC-00 6860718 E-MINIC-00 6873178 E-MINIC-00 6873185 E-MINIC-00 6873193 E-MINIC-00 6876009 E-MINIC-00 6876010 E-MINIC-00 6876011 E-MINIC-00 6876015 E-MINIC-00 6876017 E-MINIC-00 6876018 E-MINIC-00 6876026 E-MINIC-00 6876037 E-MINIC-00 6876041 E-MINIC-00 6876042 E-MINIC-00 6876051 E-MINIC-00 6878075 E-MINIC-00 6878076 E-MINIC-00 6878077 E-MINIC-00 6878078 E-MINIC-00 6878079 E-MINIC-00 6878082 E-MINIC-00 6878085 E-MINIC-00 6878089 E-MINIC-00 6879405 E-MINIC-00 6879406 E-MINIC-00 6879408 E-MINIC-00 6879410 E-MINIC-00 6879411 E-MINIC-00 6879412 E-MINIC-00 6879413 E-MINIC-00 6879414 E-MINIC-00 6879416 E-MINIC-00 6879419 E-MINIC-00 6879426 E-MINIC-00 6879432 E-MINIC-00 6879433 E-MINIC-00 6879440 E-MINIC-00 6879442 E-MINIC-00 6879443 E-MINIC-00 6879444 E-MINIC-00 6879446 E-MINIC-00 6879447 E-MINIC-00 6879448 E-MINIC-00 6879449 E-MINIC-00 6879451 E-MINIC-00 6879452 E-MINIC-00 6879453 E-MINIC-00 6879454 E-MINIC-00 6883965 E-MINIC-00 6887179 E-MINIC-00 6887212 E-MINIC-00 6860806 E-MINIC-00 6860832 E-MINIC-00 6860833 E-MINIC-00 6867788 E-MINIC-00 6867790 E-MINIC-00 6822940 E-MINIC-00 6822942 E-MINIC-00 6822946 E-MINIC-00 6822948 E-MINIC-00 6827890 E-MINIC-00 6827921 E-MINIC-00 6827928 E-MINIC-00 6827929 E-MINIC-00 6827936 E-MINIC-00 6829315 E-MINIC-00 6829316 E-MINIC-00 6834925 E-MINIC-00 6834926 E-MINIC-00 6834929 E-MINIC-00 6834930 E-MINIC-00 6834931 E-MINIC-00 6836969 E-MINIC-00 6836971 E-MINIC-00 6836973 E-MINIC-00 6836974 E-MINIC-00 6836979 E-MINIC-00 6836981 E-MINIC-00 6836983 E-MINIC-00 6836984 E-MINIC-00 6836985 E-MINIC-00 6836986 E-MINIC-00 6836987 E-MINIC-00 6836988 E-MINIC-00 6836989 E-MINIC-00 6836990 E-MINIC-00 6836991 E-MINIC-00 6836992 E-MINIC-00 6836993 E-MINIC-00 6837000 E-MINIC-00 6837007 E-MINIC-00 6837009 E-MINIC-00 6837012 E-MINIC-00 6837016 E-MINIC-00 6837018 E-MINIC-00 6838185 E-MINIC-00 6838188 E-MINIC-00 6838199 E-MINIC-00 6838200 E-MINIC-00 6838202 E-MINIC-00 6838203 E-MINIC-00 6838205 E-MINIC-00 6838217 E-MINIC-00 6838219 E-MINIC-00 6838222 E-MINIC-00 6838263 E-MINIC-00 6842191 E-MINIC-00 6842193 E-MINIC-00 6842195 E-MINIC-00 6842203 E-MINIC-00 6842205 E-MINIC-00 6842207 E-MINIC-00 6842215 E-MINIC-00 6845201 E-MINIC-00 6845205 E-MINIC-00 6845208 E-MINIC-00 6845226 E-MINIC-00 6845237 E-MINIC-00 6845242 E-MINIC-00 6848907 E-MINIC-00 6848934 E-MINIC-00 6848940 E-MINIC-00 6848949 E-MINIC-00 6852616 E-MINIC-00 6852618 E-MINIC-00 6852624 E-MINIC-00 6852633 E-MINIC-00 6852665 E-MINIC-00 6852670 E-MINIC-00 6852689 E-MINIC-00 6852691 E-MINIC-00 6852698 E-MINIC-00 6852701 E-MINIC-00 6855626 E-MINIC-00 6855631 E-MINIC-00 6855668 E-MINIC-00 6855692 E-MINIC-00 6855700 E-MINIC-00 6860790 E-MINIC-00 6860795 E-MINIC-00 6860796 E-MINIC-00 6860805 E-MINIC-00 6860828 E-MINIC-00 6860837 E-MINIC-00 6867794 E-MINIC-00 6867827 E-MINIC-00 6867832 E-MINIC-00 6868301 E-MINIC-00 6868305 E-MINIC-00 6868306 E-MINIC-00 6868307 E-MINIC-00 6868310 E-MINIC-00 6868311 E-MINIC-00 6868313 E-MINIC-00 6868314 E-MINIC-00 6868315 E-MINIC-00 6868322 E-MINIC-00 6868324 E-MINIC-00 6868327 E-MINIC-00 6868328 E-MINIC-00 6868330 E-MINIC-00 6868331 E-MINIC-00 6868332 E-MINIC-00 6868333 E-MINIC-00 6868337 E-MINIC-00 6868340 E-MINIC-00 6868343 E-MINIC-00 6870550 E-MINIC-00 6870552 E-MINIC-00 6870553 E-MINIC-00 6870555 E-MINIC-00 6870557 E-MINIC-00 6870558 E-MINIC-00 6870562 E-MINIC-00 6870565 E-MINIC-00 6870576 E-MINIC-00 6870583 E-MINIC-00 6870584 E-MINIC-00 6870585 E-MINIC-00 6870586 E-MINIC-00 6870588 E-MINIC-00 6870591 E-MINIC-00 6870595 E-MINIC-00 6873122 E-MINIC-00 6873126 E-MINIC-00 6873127 E-MINIC-00 6873129 E-MINIC-00 6873134 E-MINIC-00 6873135 E-MINIC-00 6873136 E-MINIC-00 6873145 E-MINIC-00 6873148 E-MINIC-00 6873161 E-MINIC-00 6876008 E-MINIC-00 6876012 E-MINIC-00 6878066 E-MINIC-00 6878083 E-MINIC-00 6887224 E-MINIC-00 6888694 E-MINIC-00 6888709 E-MINIC-00 6888715 E-MINIC-00 6888723 E-MINIC-00 6888740 E-MINIC-00 6889940 E-MINIC-00 6891142 E-MINIC-00 6891149 E-MINIC-00 6891185 E-MINIC-00 6891187 E-MINIC-00 6892608 E-MINIC-00 6892620 E-MINIC-00 6892684 E-MINIC-00 6892696 E-MINIC-00 6894975 E-MINIC-00 6894978 E-MINIC-00 6894980 E-MINIC-00 6896318 E-MINIC-00 6896320 E-MINIC-00 6898701 E-MINIC-00 6898711 E-MINIC-00 6892609 E-MINIC-00 6852648 E-MINIC-00 6832786 E-MINIC-00 6827889 E-MINIC-00 6837015 E-MINIC-00 6848899 E-MINIC-00 6848901 E-MINIC-00 6848902 E-MINIC-00 6848911 E-MINIC-00 6848925 E-MINIC-00 6848930 E-MINIC-00 6855645 E-MINIC-00 6860714 E-MINIC-00 6860723 E-MINIC-00 6860728 E-MINIC-00 6883982 E-MINIC-00 6888729 E-MINIC-00 6888748 E-MINIC-00 6891159 E-MINIC-00 6891161 E-MINIC-00 6892604 E-MINIC-00 6892705 E-MINIC-00 6892708 E-MINIC-00 6892709 E-MINIC-00 6894973 E-MINIC-00 6837006 E-MINIC-00 6838191 E-MINIC-00 6838194 E-MINIC-00 6838216 E-MINIC-00 6838223 E-MINIC-00 6838227 E-MINIC-00 6838232 E-MINIC-00 6838242 E-MINIC-00 6818561 E-MINIC-00 6824817 E-MINIC-00 6827893 E-MINIC-00 6827920 E-MINIC-00 6827926 E-MINIC-00 6827930 E-MINIC-00 6829310 E-MINIC-00 6829312 E-MINIC-00 6829314 E-MINIC-00 6829320 E-MINIC-00 6832809 E-MINIC-00 6832810 E-MINIC-00 6832819 E-MINIC-00 6834928 E-MINIC-00 6838213 E-MINIC-00 6838215 E-MINIC-00 6838218 E-MINIC-00 6838226 E-MINIC-00 6838233 E-MINIC-00 6838234 E-MINIC-00 6838241 E-MINIC-00 6838249 E-MINIC-00 6842172 E-MINIC-00 6845210 E-MINIC-00 6845222 E-MINIC-00 6845233 E-MINIC-00 6848896 E-MINIC-00 6848917 E-MINIC-00 6848921 E-MINIC-00 6848938 E-MINIC-00 6848946 E-MINIC-00 6848950 E-MINIC-00 6852602 E-MINIC-00 6852607 E-MINIC-00 6852620 E-MINIC-00 6852626 E-MINIC-00 6852627 E-MINIC-00 6852629 E-MINIC-00 6852635 E-MINIC-00 6852642 E-MINIC-00 6852644 E-MINIC-00 6852646 E-MINIC-00 6852647 E-MINIC-00 6852652 E-MINIC-00 6852654 E-MINIC-00 6852658 E-MINIC-00 6852663 E-MINIC-00 6852686 E-MINIC-00 6855625 E-MINIC-00 6855635 E-MINIC-00 6855675 E-MINIC-00 6855694 E-MINIC-00 6855706 E-MINIC-00 6855708 E-MINIC-00 6855710 E-MINIC-00 6855711 E-MINIC-00 6855714 E-MINIC-00 6855731 E-MINIC-00 6860430 E-MINIC-00 6860431 E-MINIC-00 6860433 E-MINIC-00 6860434 E-MINIC-00 6860445 E-MINIC-00 6860447 E-MINIC-00 6860448 E-MINIC-00 6860449 E-MINIC-00 6860452 E-MINIC-00 6860662 E-MINIC-00 6860664 E-MINIC-00 6860669 E-MINIC-00 6860689 E-MINIC-00 6860719 E-MINIC-00 6860729 E-MINIC-00 6860740 E-MINIC-00 6860752 E-MINIC-00 6860753 E-MINIC-00 6860755 E-MINIC-00 6860798 E-MINIC-00 6867831 E-MINIC-00 6868308 E-MINIC-00 6868316 E-MINIC-00 6868317 E-MINIC-00 6868318 E-MINIC-00 6868329 E-MINIC-00 6873176 E-MINIC-00 6873192 E-MINIC-00 6876032 E-MINIC-00 6876043 E-MINIC-00 6876044 E-MINIC-00 6884002 E-MINIC-00 6888665 E-MINIC-00 6827919 E-MINIC-00 6848920 E-MINIC-00 6855726 E-MINIC-00 6860826 E-MINIC-00 6860851 E-MINIC-00 6860852 E-MINIC-00 6867791 E-MINIC-00 6867792 E-MINIC-00 6867796 E-MINIC-00 6867798 E-MINIC-00 6867809 E-MINIC-00 6867816 E-MINIC-00 6867818 E-MINIC-00 6867822 E-MINIC-00 6867828 E-MINIC-00 6868304 E-MINIC-00 6870551 E-MINIC-00 6888736 E-MINIC-00 6888766 E-MINIC-00 6889937 E-MINIC-00 6889961 E-MINIC-00 6891138 E-MINIC-00 6891147 E-MINIC-00 6891157 E-MINIC-00 6891167 E-MINIC-00 6891169 E-MINIC-00 6891177 E-MINIC-00 6891179 E-MINIC-00 6892613 E-MINIC-00 6892631 E-MINIC-00 6892657 E-MINIC-00 6892662 E-MINIC-00 6892664 E-MINIC-00 6892707 E-MINIC-00 6892712 E-MINIC-00 6892713 E-MINIC-00 6892716 E-MINIC-00 6896277 E-MINIC-00 6896306 E-MINIC-00 6896314 E-MINIC-00 6896326 E-MINIC-00 6896331 E-MINIC-00 6896288 E-MINIC-00 6896293 E-MINIC-00 6852672 E-MINIC-00 6891186 E-MINIC-00 6860781 E-MINIC-00 6860786 E-MINIC-00 6887184 E-MINIC-00 6892690 E-MINIC-00 6818574 E-MINIC-00 6822938 E-MINIC-00 6822939 E-MINIC-00 6822943 E-MINIC-00 6822944 E-MINIC-00 6822947 E-MINIC-00 6822951 E-MINIC-00 6822952 E-MINIC-00 6822953 E-MINIC-00 6822954 E-MINIC-00 6822955 E-MINIC-00 6822956 E-MINIC-00 6822957 E-MINIC-00 6822958 E-MINIC-00 6822959 E-MINIC-00 6822960 E-MINIC-00 6822962 E-MINIC-00 6822964 E-MINIC-00 6822965 E-MINIC-00 6822967 E-MINIC-00 6824777 E-MINIC-00 6824779 E-MINIC-00 6824820 E-MINIC-00 6832783 E-MINIC-00 6832784 E-MINIC-00 6832792 E-MINIC-00 6832800 E-MINIC-00 6832818 E-MINIC-00 6832828 E-MINIC-00 6832830 E-MINIC-00 6834897 E-MINIC-00 6834900 E-MINIC-00 6836994 E-MINIC-00 6837004 E-MINIC-00 6837008 E-MINIC-00 6837017 E-MINIC-00 6838192 E-MINIC-00 6838195 E-MINIC-00 6838206 E-MINIC-00 6838207 E-MINIC-00 6838208 E-MINIC-00 6838239 E-MINIC-00 6838253 E-MINIC-00 6838257 E-MINIC-00 6842188 E-MINIC-00 6842197 E-MINIC-00 6842202 E-MINIC-00 6842208 E-MINIC-00 6842209 E-MINIC-00 6842216 E-MINIC-00 6842223 E-MINIC-00 6842229 E-MINIC-00 6842232 E-MINIC-00 6842235 E-MINIC-00 6842237 E-MINIC-00 6842240 E-MINIC-00 6845203 E-MINIC-00 6845204 E-MINIC-00 6845212 E-MINIC-00 6845214 E-MINIC-00 6845215 E-MINIC-00 6845216 E-MINIC-00 6845217 E-MINIC-00 6845219 E-MINIC-00 6845220 E-MINIC-00 6845221 E-MINIC-00 6845223 E-MINIC-00 6845225 E-MINIC-00 6845229 E-MINIC-00 6845230 E-MINIC-00 6845234 E-MINIC-00 6845236 E-MINIC-00 6845238 E-MINIC-00 6845239 E-MINIC-00 6845240 E-MINIC-00 6845244 E-MINIC-00 6845249 E-MINIC-00 6845250 E-MINIC-00 6848900 E-MINIC-00 6848922 E-MINIC-00 6848926 E-MINIC-00 6848927 E-MINIC-00 6848928 E-MINIC-00 6848932 E-MINIC-00 6848933 E-MINIC-00 6848937 E-MINIC-00 6848945 E-MINIC-00 6852609 E-MINIC-00 6852614 E-MINIC-00 6852615 E-MINIC-00 6852623 E-MINIC-00 6852625 E-MINIC-00 6852630 E-MINIC-00 6852631 E-MINIC-00 6852634 E-MINIC-00 6852636 E-MINIC-00 6852637 E-MINIC-00 6852639 E-MINIC-00 6852640 E-MINIC-00 6852657 E-MINIC-00 6852673 E-MINIC-00 6852685 E-MINIC-00 6852690 E-MINIC-00 6852692 E-MINIC-00 6852693 E-MINIC-00 6852694 E-MINIC-00 6852695 E-MINIC-00 6852696 E-MINIC-00 6852697 E-MINIC-00 6852700 E-MINIC-00 6855624 E-MINIC-00 6855628 E-MINIC-00 6855632 E-MINIC-00 6855633 E-MINIC-00 6855634 E-MINIC-00 6855636 E-MINIC-00 6855637 E-MINIC-00 6855655 E-MINIC-00 6855658 E-MINIC-00 6855659 E-MINIC-00 6855663 E-MINIC-00 6855666 E-MINIC-00 6855682 E-MINIC-00 6855697 E-MINIC-00 6860660 E-MINIC-00 6860680 E-MINIC-00 6860682 E-MINIC-00 6860686 E-MINIC-00 6860690 E-MINIC-00 6860713 E-MINIC-00 6860748 E-MINIC-00 6860756 E-MINIC-00 6860758 E-MINIC-00 6860787 E-MINIC-00 6860803 E-MINIC-00 6860810 E-MINIC-00 6860811 E-MINIC-00 6860812 E-MINIC-00 6860817 E-MINIC-00 6860822 E-MINIC-00 6860834 E-MINIC-00 6860838 E-MINIC-00 6860843 E-MINIC-00 6860850 E-MINIC-00 6860853 E-MINIC-00 6860857 E-MINIC-00 6867795 E-MINIC-00 6867799 E-MINIC-00 6867800 E-MINIC-00 6867801 E-MINIC-00 6867802 E-MINIC-00 6867819 E-MINIC-00 6867821 E-MINIC-00 6867824 E-MINIC-00 6867830 E-MINIC-00 6868303 E-MINIC-00 6868335 E-MINIC-00 6868336 E-MINIC-00 6868338 E-MINIC-00 6868339 E-MINIC-00 6868341 E-MINIC-00 6870554 E-MINIC-00 6870563 E-MINIC-00 6870569 E-MINIC-00 6870570 E-MINIC-00 6870575 E-MINIC-00 6870589 E-MINIC-00 6870590 E-MINIC-00 6870592 E-MINIC-00 6873117 E-MINIC-00 6873138 E-MINIC-00 6873140 E-MINIC-00 6873147 E-MINIC-00 6873155 E-MINIC-00 6873187 E-MINIC-00 6876016 E-MINIC-00 6888666 E-MINIC-00 6888689 E-MINIC-00 6888714 E-MINIC-00 6888717 E-MINIC-00 6888718 E-MINIC-00 6888727 E-MINIC-00 6888730 E-MINIC-00 6888753 E-MINIC-00 6888760 E-MINIC-00 6888762 E-MINIC-00 6888769 E-MINIC-00 6889947 E-MINIC-00 6889949 E-MINIC-00 6889953 E-MINIC-00 6889964 E-MINIC-00 6889969 E-MINIC-00 6889972 E-MINIC-00 6889973 E-MINIC-00 6891039 E-MINIC-00 6891040 E-MINIC-00 6891042 E-MINIC-00 6891046 E-MINIC-00 6891050 E-MINIC-00 6891053 E-MINIC-00 6891058 E-MINIC-00 6891060 E-MINIC-00 6891061 E-MINIC-00 6891062 E-MINIC-00 6891065 E-MINIC-00 6891066 E-MINIC-00 6891068 E-MINIC-00 6891071 E-MINIC-00 6891072 E-MINIC-00 6891073 E-MINIC-00 6891074 E-MINIC-00 6891075 E-MINIC-00 6891076 E-MINIC-00 6891085 E-MINIC-00 6891104 E-MINIC-00 6891127 E-MINIC-00 6891133 E-MINIC-00 6891135 E-MINIC-00 6891140 E-MINIC-00 6891141 E-MINIC-00 6891164 E-MINIC-00 6892603 E-MINIC-00 6892605 E-MINIC-00 6892606 E-MINIC-00 6892610 E-MINIC-00 6892611 E-MINIC-00 6892612 E-MINIC-00 6892615 E-MINIC-00 6892617 E-MINIC-00 6892619 E-MINIC-00 6892622 E-MINIC-00 6892623 E-MINIC-00 6892624 E-MINIC-00 6892628 E-MINIC-00 6892629 E-MINIC-00 6892630 E-MINIC-00 6892634 E-MINIC-00 6892638 E-MINIC-00 6892641 E-MINIC-00 6892645 E-MINIC-00 6892648 E-MINIC-00 6892652 E-MINIC-00 6892658 E-MINIC-00 6892668 E-MINIC-00 6892687 E-MINIC-00 6892704 E-MINIC-00 6892710 E-MINIC-00 6892715 E-MINIC-00 6894974 E-MINIC-00 6894979 E-MINIC-00 6895029 E-MINIC-00 6896285 E-MINIC-00 6896300 E-MINIC-00 6896304 E-MINIC-00 6898692 E-MINIC-00 6898752 E-MINIC-00 6898760 E-MINIC-00 6898765 E-MINIC-00 6822973 E-MINIC-00 6822974 E-MINIC-00 6824780 E-MINIC-00 6824784 E-MINIC-00 6827908 E-MINIC-00 6827909 E-MINIC-00 6827925 E-MINIC-00 6827933 E-MINIC-00 6827934 E-MINIC-00 6829317 E-MINIC-00 6889941 E-MINIC-00 6891057 E-MINIC-00 6898745 E-MINIC-00 6822968 E-MINIC-00 6822969 E-MINIC-00 6824773 E-MINIC-00 6860692 E-MINIC-00 6860746 E-MINIC-00 6860807 E-MINIC-00 6860808 E-MINIC-00 6860835 E-MINIC-00 6860855 E-MINIC-00 6860856 E-MINIC-00 6867797 E-MINIC-00 6868323 E-MINIC-00 6883981 E-MINIC-00 6883985 E-MINIC-00 6883987 E-MINIC-00 6883989 E-MINIC-00 6883999 E-MINIC-00 6884021 E-MINIC-00 6888677 E-MINIC-00 6888685 E-MINIC-00 6888705 E-MINIC-00 6888710 E-MINIC-00 6889921 E-MINIC-00 6889929 E-MINIC-00 6895024 E-MINIC-00 6896263 E-MINIC-00 6898698 E-MINIC-00 6848923 E-MINIC-00 6873120 E-MINIC-00 6873124 E-MINIC-00 6873160 E-MINIC-00 6873163 E-MINIC-00 6889945 E-MINIC-00 6889946 E-MINIC-00 6889962 E-MINIC-00 6891077 E-MINIC-00 6891079 E-MINIC-00 6891097 E-MINIC-00 6891112 E-MINIC-00 6891163 E-MINIC-00 6891170 E-MINIC-00 6891172 E-MINIC-00 6891183 E-MINIC-00 6892632 E-MINIC-00 6892635 E-MINIC-00 6892671 E-MINIC-00 6892691 E-MINIC-00 6892711 E-MINIC-00 6892714 E-MINIC-00 6892717 E-MINIC-00 6892718 E-MINIC-00 6894984 E-MINIC-00 6896267 E-MINIC-00 6896324 E-MINIC-00 6898709 E-MINIC-00 6836996 E-MINIC-00 6888758 E-MINIC-00 6848919 E-MINIC-00 6898694 E-MINIC-00 6898755 E-MINIC-00 6898756 E-MINIC-00 6829319 E-MINIC-00 6832814 E-MINIC-00 6832820 E-MINIC-00 6832829 E-MINIC-00 6832832 E-MINIC-00 6842189 E-MINIC-00 6852606 E-MINIC-00 6852684 E-MINIC-00 6855638 E-MINIC-00 6855640 E-MINIC-00 6855641 E-MINIC-00 6855642 E-MINIC-00 6855644 E-MINIC-00 6855646 E-MINIC-00 6855648 E-MINIC-00 6855652 E-MINIC-00 6855654 E-MINIC-00 6855727 E-MINIC-00 6855730 E-MINIC-00 6860741 E-MINIC-00 6860789 E-MINIC-00 6860792 E-MINIC-00 6860797 E-MINIC-00 6860802 E-MINIC-00 6860809 E-MINIC-00 6860824 E-MINIC-00 6860842 E-MINIC-00 6860846 E-MINIC-00 6860848 E-MINIC-00 6860854 E-MINIC-00 6867803 E-MINIC-00 6867804 E-MINIC-00 6867805 E-MINIC-00 6867806 E-MINIC-00 6867808 E-MINIC-00 6867810 E-MINIC-00 6867812 E-MINIC-00 6867813 E-MINIC-00 6867815 E-MINIC-00 6867817 E-MINIC-00 6867820 E-MINIC-00 6867837 E-MINIC-00 6883993 E-MINIC-00 6883996 E-MINIC-00 6887191 E-MINIC-00 6887193 E-MINIC-00 6887200 E-MINIC-00 6887202 E-MINIC-00 6887218 E-MINIC-00 6888656 E-MINIC-00 6888682 E-MINIC-00 6888691 E-MINIC-00 6896257 E-MINIC-00 6896258 E-MINIC-00 6884009 E-MINIC-00 6884012 E-MINIC-00 6884018 E-MINIC-00 6894985 E-MINIC-00 6822941 E-MINIC-00-EN 6822950 E-MINIC-00-EN 6822961 E-MINIC-00-EN 6822963 E-MINIC-00-EN 6822966 E-MINIC-00-EN 6822970 E-MINIC-00-EN 6822972 E-MINIC-00-EN 6824775 E-MINIC-00-EN 6824776 E-MINIC-00-EN 6824778 E-MINIC-00-EN 6824781 E-MINIC-00-EN 6824785 E-MINIC-00-EN 6824786 E-MINIC-00-EN 6824789 E-MINIC-00-EN 6824791 E-MINIC-00-EN 6824794 E-MINIC-00-EN 6824795 E-MINIC-00-EN 6824796 E-MINIC-00-EN 6824798 E-MINIC-00-EN 6824801 E-MINIC-00-EN 6824803 E-MINIC-00-EN 6824805 E-MINIC-00-EN 6824806 E-MINIC-00-EN 6824809 E-MINIC-00-EN 6824810 E-MINIC-00-EN 6824819 E-MINIC-00-EN 6824822 E-MINIC-00-EN 6827894 E-MINIC-00-EN 6827895 E-MINIC-00-EN 6827898 E-MINIC-00-EN 6827899 E-MINIC-00-EN 6827900 E-MINIC-00-EN 6827901 E-MINIC-00-EN 6827902 E-MINIC-00-EN 6827903 E-MINIC-00-EN 6827905 E-MINIC-00-EN 6827906 E-MINIC-00-EN 6827910 E-MINIC-00-EN 6827911 E-MINIC-00-EN 6827915 E-MINIC-00-EN 6827917 E-MINIC-00-EN 6827923 E-MINIC-00-EN 6834875 E-MINIC-00-EN 6834876 E-MINIC-00-EN 6834878 E-MINIC-00-EN 6834879 E-MINIC-00-EN 6834881 E-MINIC-00-EN 6834882 E-MINIC-00-EN 6834905 E-MINIC-00-EN 6834906 E-MINIC-00-EN 6834908 E-MINIC-00-EN 6834912 E-MINIC-00-EN 6838220 E-MINIC-00-EN 6838221 E-MINIC-00-EN 6838225 E-MINIC-00-EN 6838229 E-MINIC-00-EN 6838230 E-MINIC-00-EN 6838248 E-MINIC-00-EN 6838250 E-MINIC-00-EN 6838260 E-MINIC-00-EN 6838264 E-MINIC-00-EN 6842166 E-MINIC-00-EN 6842168 E-MINIC-00-EN 6842169 E-MINIC-00-EN 6842170 E-MINIC-00-EN 6842171 E-MINIC-00-EN 6842173 E-MINIC-00-EN 6842178 E-MINIC-00-EN 6842179 E-MINIC-00-EN 6842184 E-MINIC-00-EN 6842185 E-MINIC-00-EN 6842186 E-MINIC-00-EN 6842187 E-MINIC-00-EN 6842194 E-MINIC-00-EN 6842199 E-MINIC-00-EN 6842200 E-MINIC-00-EN 6842201 E-MINIC-00-EN 6842210 E-MINIC-00-EN 6842211 E-MINIC-00-EN 6842213 E-MINIC-00-EN 6842214 E-MINIC-00-EN 6845202 E-MINIC-00-EN 6845209 E-MINIC-00-EN 6845211 E-MINIC-00-EN 6845213 E-MINIC-00-EN 6845241 E-MINIC-00-EN 6845243 E-MINIC-00-EN 6845245 E-MINIC-00-EN 6845247 E-MINIC-00-EN 6848872 E-MINIC-00-EN 6848873 E-MINIC-00-EN 6848874 E-MINIC-00-EN 6848875 E-MINIC-00-EN 6848876 E-MINIC-00-EN 6848879 E-MINIC-00-EN 6848882 E-MINIC-00-EN 6848893 E-MINIC-00-EN 6848894 E-MINIC-00-EN 6848903 E-MINIC-00-EN 6848904 E-MINIC-00-EN 6848906 E-MINIC-00-EN 6848910 E-MINIC-00-EN 6848918 E-MINIC-00-EN 6848929 E-MINIC-00-EN 6848935 E-MINIC-00-EN 6848939 E-MINIC-00-EN 6848941 E-MINIC-00-EN 6848943 E-MINIC-00-EN 6848944 E-MINIC-00-EN 6848947 E-MINIC-00-EN 6848948 E-MINIC-00-EN 6848951 E-MINIC-00-EN 6852603 E-MINIC-00-EN 6852604 E-MINIC-00-EN 6852605 E-MINIC-00-EN 6852608 E-MINIC-00-EN 6852610 E-MINIC-00-EN 6852628 E-MINIC-00-EN 6852632 E-MINIC-00-EN 6852638 E-MINIC-00-EN 6852643 E-MINIC-00-EN 6852645 E-MINIC-00-EN 6852649 E-MINIC-00-EN 6852659 E-MINIC-00-EN 6852671 E-MINIC-00-EN 6852683 E-MINIC-00-EN 6852687 E-MINIC-00-EN 6852688 E-MINIC-00-EN 6852699 E-MINIC-00-EN 6855650 E-MINIC-00-EN 6855665 E-MINIC-00-EN 6855667 E-MINIC-00-EN 6855669 E-MINIC-00-EN 6855670 E-MINIC-00-EN 6855672 E-MINIC-00-EN 6855676 E-MINIC-00-EN 6855679 E-MINIC-00-EN 6855684 E-MINIC-00-EN 6855689 E-MINIC-00-EN 6855690 E-MINIC-00-EN 6855693 E-MINIC-00-EN 6855696 E-MINIC-00-EN 6855698 E-MINIC-00-EN 6855703 E-MINIC-00-EN 6855707 E-MINIC-00-EN 6855723 E-MINIC-00-EN 6855724 E-MINIC-00-EN 6855733 E-MINIC-00-EN 6860429 E-MINIC-00-EN 6860432 E-MINIC-00-EN 6860435 E-MINIC-00-EN 6860436 E-MINIC-00-EN 6860437 E-MINIC-00-EN 6860440 E-MINIC-00-EN 6860441 E-MINIC-00-EN 6860442 E-MINIC-00-EN 6860443 E-MINIC-00-EN 6860446 E-MINIC-00-EN 6860451 E-MINIC-00-EN 6860453 E-MINIC-00-EN 6860663 E-MINIC-00-EN 6860666 E-MINIC-00-EN 6860668 E-MINIC-00-EN 6860671 E-MINIC-00-EN 6860673 E-MINIC-00-EN 6860674 E-MINIC-00-EN 6860675 E-MINIC-00-EN 6860676 E-MINIC-00-EN 6860677 E-MINIC-00-EN 6860678 E-MINIC-00-EN 6860679 E-MINIC-00-EN 6860681 E-MINIC-00-EN 6860687 E-MINIC-00-EN 6860688 E-MINIC-00-EN 6860691 E-MINIC-00-EN 6860693 E-MINIC-00-EN 6860694 E-MINIC-00-EN 6860695 E-MINIC-00-EN 6860696 E-MINIC-00-EN 6860697 E-MINIC-00-EN 6860699 E-MINIC-00-EN 6860701 E-MINIC-00-EN 6860702 E-MINIC-00-EN 6860703 E-MINIC-00-EN 6860704 E-MINIC-00-EN 6860705 E-MINIC-00-EN 6860706 E-MINIC-00-EN 6860707 E-MINIC-00-EN 6860708 E-MINIC-00-EN 6860709 E-MINIC-00-EN 6860710 E-MINIC-00-EN 6860711 E-MINIC-00-EN 6860712 E-MINIC-00-EN 6860715 E-MINIC-00-EN 6860716 E-MINIC-00-EN 6860717 E-MINIC-00-EN 6860721 E-MINIC-00-EN 6860722 E-MINIC-00-EN 6860724 E-MINIC-00-EN 6860725 E-MINIC-00-EN 6860726 E-MINIC-00-EN 6860727 E-MINIC-00-EN 6860730 E-MINIC-00-EN 6860732 E-MINIC-00-EN 6860733 E-MINIC-00-EN 6860734 E-MINIC-00-EN 6860735 E-MINIC-00-EN 6860736 E-MINIC-00-EN 6860737 E-MINIC-00-EN 6860738 E-MINIC-00-EN 6860739 E-MINIC-00-EN 6860742 E-MINIC-00-EN 6860743 E-MINIC-00-EN 6860744 E-MINIC-00-EN 6860745 E-MINIC-00-EN 6860747 E-MINIC-00-EN 6860749 E-MINIC-00-EN 6860751 E-MINIC-00-EN 6860754 E-MINIC-00-EN 6860757 E-MINIC-00-EN 6860760 E-MINIC-00-EN 6860762 E-MINIC-00-EN 6860763 E-MINIC-00-EN 6860764 E-MINIC-00-EN 6860765 E-MINIC-00-EN 6860766 E-MINIC-00-EN 6860767 E-MINIC-00-EN 6860769 E-MINIC-00-EN 6860770 E-MINIC-00-EN 6860771 E-MINIC-00-EN 6860774 E-MINIC-00-EN 6860775 E-MINIC-00-EN 6860776 E-MINIC-00-EN 6860779 E-MINIC-00-EN 6860782 E-MINIC-00-EN 6860783 E-MINIC-00-EN 6860784 E-MINIC-00-EN 6860785 E-MINIC-00-EN 6860788 E-MINIC-00-EN 6860793 E-MINIC-00-EN 6860794 E-MINIC-00-EN 6860800 E-MINIC-00-EN 6860801 E-MINIC-00-EN 6860804 E-MINIC-00-EN 6860813 E-MINIC-00-EN 6860814 E-MINIC-00-EN 6860815 E-MINIC-00-EN 6860816 E-MINIC-00-EN 6860818 E-MINIC-00-EN 6860819 E-MINIC-00-EN 6860820 E-MINIC-00-EN 6860829 E-MINIC-00-EN 6860831 E-MINIC-00-EN 6860836 E-MINIC-00-EN 6860839 E-MINIC-00-EN 6860840 E-MINIC-00-EN 6860841 E-MINIC-00-EN 6860844 E-MINIC-00-EN 6860845 E-MINIC-00-EN 6860859 E-MINIC-00-EN 6867793 E-MINIC-00-EN 6867807 E-MINIC-00-EN 6867811 E-MINIC-00-EN 6867814 E-MINIC-00-EN 6867823 E-MINIC-00-EN 6867825 E-MINIC-00-EN 6867826 E-MINIC-00-EN 6867829 E-MINIC-00-EN 6867833 E-MINIC-00-EN 6867834 E-MINIC-00-EN 6867835 E-MINIC-00-EN 6867836 E-MINIC-00-EN 6868295 E-MINIC-00-EN 6868296 E-MINIC-00-EN 6868297 E-MINIC-00-EN 6868298 E-MINIC-00-EN 6868299 E-MINIC-00-EN 6868300 E-MINIC-00-EN 6868302 E-MINIC-00-EN 6868309 E-MINIC-00-EN 6868312 E-MINIC-00-EN 6868319 E-MINIC-00-EN 6868325 E-MINIC-00-EN 6868326 E-MINIC-00-EN 6868334 E-MINIC-00-EN 6868342 E-MINIC-00-EN 6870559 E-MINIC-00-EN 6870560 E-MINIC-00-EN 6870561 E-MINIC-00-EN 6870567 E-MINIC-00-EN 6870571 E-MINIC-00-EN 6870577 E-MINIC-00-EN 6870578 E-MINIC-00-EN 6870579 E-MINIC-00-EN 6870580 E-MINIC-00-EN 6870596 E-MINIC-00-EN 6870597 E-MINIC-00-EN 6870598 E-MINIC-00-EN 6873118 E-MINIC-00-EN 6873119 E-MINIC-00-EN 6873132 E-MINIC-00-EN 6873133 E-MINIC-00-EN 6873137 E-MINIC-00-EN 6873142 E-MINIC-00-EN 6873144 E-MINIC-00-EN 6873146 E-MINIC-00-EN 6873149 E-MINIC-00-EN 6873150 E-MINIC-00-EN 6873151 E-MINIC-00-EN 6873152 E-MINIC-00-EN 6873153 E-MINIC-00-EN 6873157 E-MINIC-00-EN 6873159 E-MINIC-00-EN 6873162 E-MINIC-00-EN 6873164 E-MINIC-00-EN 6873166 E-MINIC-00-EN 6873167 E-MINIC-00-EN 6873168 E-MINIC-00-EN 6873169 E-MINIC-00-EN 6873170 E-MINIC-00-EN 6873171 E-MINIC-00-EN 6873172 E-MINIC-00-EN 6873173 E-MINIC-00-EN 6873174 E-MINIC-00-EN 6873175 E-MINIC-00-EN 6873177 E-MINIC-00-EN 6873179 E-MINIC-00-EN 6873180 E-MINIC-00-EN 6873181 E-MINIC-00-EN 6873183 E-MINIC-00-EN 6873184 E-MINIC-00-EN 6873186 E-MINIC-00-EN 6873188 E-MINIC-00-EN 6873189 E-MINIC-00-EN 6873191 E-MINIC-00-EN 6873194 E-MINIC-00-EN 6873195 E-MINIC-00-EN 6873196 E-MINIC-00-EN 6876005 E-MINIC-00-EN 6876006 E-MINIC-00-EN 6876013 E-MINIC-00-EN 6876014 E-MINIC-00-EN 6876019 E-MINIC-00-EN 6876020 E-MINIC-00-EN 6876021 E-MINIC-00-EN 6876022 E-MINIC-00-EN 6876023 E-MINIC-00-EN 6876024 E-MINIC-00-EN 6876025 E-MINIC-00-EN 6876027 E-MINIC-00-EN 6876028 E-MINIC-00-EN 6876029 E-MINIC-00-EN 6876030 E-MINIC-00-EN 6876031 E-MINIC-00-EN 6876033 E-MINIC-00-EN 6876034 E-MINIC-00-EN 6876035 E-MINIC-00-EN 6876036 E-MINIC-00-EN 6876038 E-MINIC-00-EN 6876039 E-MINIC-00-EN 6876040 E-MINIC-00-EN 6876045 E-MINIC-00-EN 6876046 E-MINIC-00-EN 6876047 E-MINIC-00-EN 6876049 E-MINIC-00-EN 6876053 E-MINIC-00-EN 6878047 E-MINIC-00-EN 6878049 E-MINIC-00-EN 6878051 E-MINIC-00-EN 6878052 E-MINIC-00-EN 6878053 E-MINIC-00-EN 6878055 E-MINIC-00-EN 6878057 E-MINIC-00-EN 6878058 E-MINIC-00-EN 6878059 E-MINIC-00-EN 6878060 E-MINIC-00-EN 6878061 E-MINIC-00-EN 6878062 E-MINIC-00-EN 6878063 E-MINIC-00-EN 6878064 E-MINIC-00-EN 6878065 E-MINIC-00-EN 6878067 E-MINIC-00-EN 6878068 E-MINIC-00-EN 6878069 E-MINIC-00-EN 6878070 E-MINIC-00-EN 6878071 E-MINIC-00-EN 6878072 E-MINIC-00-EN 6878073 E-MINIC-00-EN 6878074 E-MINIC-00-EN 6878080 E-MINIC-00-EN 6878081 E-MINIC-00-EN 6878084 E-MINIC-00-EN 6878086 E-MINIC-00-EN 6878087 E-MINIC-00-EN 6878088 E-MINIC-00-EN 6878090 E-MINIC-00-EN 6878091 E-MINIC-00-EN 6878092 E-MINIC-00-EN 6878093 E-MINIC-00-EN 6878094 E-MINIC-00-EN 6878095 E-MINIC-00-EN 6878096 E-MINIC-00-EN 6879407 E-MINIC-00-EN 6879409 E-MINIC-00-EN 6879415 E-MINIC-00-EN 6879417 E-MINIC-00-EN 6879418 E-MINIC-00-EN 6879420 E-MINIC-00-EN 6879421 E-MINIC-00-EN 6879422 E-MINIC-00-EN 6879424 E-MINIC-00-EN 6879425 E-MINIC-00-EN 6879427 E-MINIC-00-EN 6879428 E-MINIC-00-EN 6879429 E-MINIC-00-EN 6879430 E-MINIC-00-EN 6879431 E-MINIC-00-EN 6879434 E-MINIC-00-EN 6879435 E-MINIC-00-EN 6879436 E-MINIC-00-EN 6879437 E-MINIC-00-EN 6879438 E-MINIC-00-EN 6879439 E-MINIC-00-EN 6879441 E-MINIC-00-EN 6879445 E-MINIC-00-EN 6879450 E-MINIC-00-EN 6883966 E-MINIC-00-EN 6883968 E-MINIC-00-EN 6883969 E-MINIC-00-EN 6883970 E-MINIC-00-EN 6883971 E-MINIC-00-EN 6883973 E-MINIC-00-EN 6883974 E-MINIC-00-EN 6884024 E-MINIC-00-EN 6887180 E-MINIC-00-EN 6887181 E-MINIC-00-EN 6887196 E-MINIC-00-EN 6887207 E-MINIC-00-EN 6887208 E-MINIC-00-EN 6887214 E-MINIC-00-EN 6887220 E-MINIC-00-EN 6887225 E-MINIC-00-EN 6887227 E-MINIC-00-EN 6888653 E-MINIC-00-EN 6888654 E-MINIC-00-EN 6888657 E-MINIC-00-EN 6888658 E-MINIC-00-EN 6888659 E-MINIC-00-EN 6888660 E-MINIC-00-EN 6888662 E-MINIC-00-EN 6888663 E-MINIC-00-EN 6888664 E-MINIC-00-EN 6888667 E-MINIC-00-EN 6888668 E-MINIC-00-EN 6888669 E-MINIC-00-EN 6888671 E-MINIC-00-EN 6888673 E-MINIC-00-EN 6888674 E-MINIC-00-EN 6888675 E-MINIC-00-EN 6888676 E-MINIC-00-EN 6888679 E-MINIC-00-EN 6888680 E-MINIC-00-EN 6888690 E-MINIC-00-EN 6888701 E-MINIC-00-EN 6888706 E-MINIC-00-EN 6888708 E-MINIC-00-EN 6888720 E-MINIC-00-EN 6888722 E-MINIC-00-EN 6888728 E-MINIC-00-EN 6888731 E-MINIC-00-EN 6888733 E-MINIC-00-EN 6888734 E-MINIC-00-EN 6888735 E-MINIC-00-EN 6888739 E-MINIC-00-EN 6888741 E-MINIC-00-EN 6888744 E-MINIC-00-EN 6888745 E-MINIC-00-EN 6888746 E-MINIC-00-EN 6888747 E-MINIC-00-EN 6888751 E-MINIC-00-EN 6888752 E-MINIC-00-EN 6888756 E-MINIC-00-EN 6888759 E-MINIC-00-EN 6888761 E-MINIC-00-EN 6888763 E-MINIC-00-EN 6888768 E-MINIC-00-EN 6889914 E-MINIC-00-EN 6889922 E-MINIC-00-EN 6889923 E-MINIC-00-EN 6889924 E-MINIC-00-EN 6889925 E-MINIC-00-EN 6889930 E-MINIC-00-EN 6889931 E-MINIC-00-EN 6889938 E-MINIC-00-EN 6889944 E-MINIC-00-EN 6889951 E-MINIC-00-EN 6889960 E-MINIC-00-EN 6889968 E-MINIC-00-EN 6889970 E-MINIC-00-EN 6891038 E-MINIC-00-EN 6891043 E-MINIC-00-EN 6891044 E-MINIC-00-EN 6891047 E-MINIC-00-EN 6891048 E-MINIC-00-EN 6891049 E-MINIC-00-EN 6891051 E-MINIC-00-EN 6891052 E-MINIC-00-EN 6891056 E-MINIC-00-EN 6891087 E-MINIC-00-EN 6891089 E-MINIC-00-EN 6891098 E-MINIC-00-EN 6891107 E-MINIC-00-EN 6891109 E-MINIC-00-EN 6891118 E-MINIC-00-EN 6891129 E-MINIC-00-EN 6891136 E-MINIC-00-EN 6891145 E-MINIC-00-EN 6891158 E-MINIC-00-EN 6891160 E-MINIC-00-EN 6891165 E-MINIC-00-EN 6891171 E-MINIC-00-EN 6892636 E-MINIC-00-EN 6892637 E-MINIC-00-EN 6892640 E-MINIC-00-EN 6892642 E-MINIC-00-EN 6892644 E-MINIC-00-EN 6892647 E-MINIC-00-EN 6892650 E-MINIC-00-EN 6892653 E-MINIC-00-EN 6892661 E-MINIC-00-EN 6892666 E-MINIC-00-EN 6892667 E-MINIC-00-EN 6892669 E-MINIC-00-EN 6892670 E-MINIC-00-EN 6892673 E-MINIC-00-EN 6892674 E-MINIC-00-EN 6892677 E-MINIC-00-EN 6892685 E-MINIC-00-EN 6892686 E-MINIC-00-EN 6892688 E-MINIC-00-EN 6892706 E-MINIC-00-EN 6894977 E-MINIC-00-EN 6894981 E-MINIC-00-EN 6894982 E-MINIC-00-EN 6894987 E-MINIC-00-EN 6894988 E-MINIC-00-EN 6894989 E-MINIC-00-EN 6894990 E-MINIC-00-EN 6894992 E-MINIC-00-EN 6894993 E-MINIC-00-EN 6894994 E-MINIC-00-EN 6894995 E-MINIC-00-EN 6894996 E-MINIC-00-EN 6894997 E-MINIC-00-EN 6894999 E-MINIC-00-EN 6895000 E-MINIC-00-EN 6895002 E-MINIC-00-EN 6895003 E-MINIC-00-EN 6895004 E-MINIC-00-EN 6895006 E-MINIC-00-EN 6895007 E-MINIC-00-EN 6895008 E-MINIC-00-EN 6895009 E-MINIC-00-EN 6895010 E-MINIC-00-EN 6895011 E-MINIC-00-EN 6895012 E-MINIC-00-EN 6895013 E-MINIC-00-EN 6895015 E-MINIC-00-EN 6895017 E-MINIC-00-EN 6895018 E-MINIC-00-EN 6895019 E-MINIC-00-EN 6895020 E-MINIC-00-EN 6895026 E-MINIC-00-EN 6895027 E-MINIC-00-EN 6895028 E-MINIC-00-EN 6895030 E-MINIC-00-EN 6896259 E-MINIC-00-EN 6896260 E-MINIC-00-EN 6896265 E-MINIC-00-EN 6896284 E-MINIC-00-EN 6896286 E-MINIC-00-EN 6896289 E-MINIC-00-EN 6896295 E-MINIC-00-EN 6896299 E-MINIC-00-EN 6896302 E-MINIC-00-EN 6896319 E-MINIC-00-EN 6896321 E-MINIC-00-EN 6896325 E-MINIC-00-EN 6896328 E-MINIC-00-EN 6896329 E-MINIC-00-EN 6896330 E-MINIC-00-EN 6898736 E-MINIC-00-EN 6898740 E-MINIC-00-EN 6898743 E-MINIC-00-EN 6898744 E-MINIC-00-EN 6898749 E-MINIC-00-EN 6898750 E-MINIC-00-EN 6898751 E-MINIC-00-EN 6898762 E-MINIC-00-EN 6898763 E-MINIC-00-EN 6898768 E-MINIC-00-EN 6898770 E-MINIC-00-EN 6898771 E-MINIC-00-EN 6898774 E-MINIC-00-EN 6838231 E-MINIC-00-EN 6883972 E-MINIC-00-EN 6827918 E-MINIC-00-EN 6860700 E-MINIC-00-EN 6888721 E-MINIC-00-EN 6891083 E-MINIC-00-EN 6891093 E-MINIC-00-EN 6891096 E-MINIC-00-EN 6891099 E-MINIC-00-EN 6891100 E-MINIC-00-EN 6848880 E-MINIC-00-EN 6848892 E-MINIC-00-EN 6848931 E-MINIC-00-EN 6887192 E-MINIC-00-EN 6888725 E-MINIC-00-EN 6888767 E-MINIC-00-EN 6889918 E-MINIC-00-EN 6889926 E-MINIC-00-EN 6889927 E-MINIC-00-EN 6889933 E-MINIC-00-EN 6889934 E-MINIC-00-EN 6838255 E-MINIC-00-EN 6842165 E-MINIC-00-EN 6842174 E-MINIC-00-EN 6842176 E-MINIC-00-EN 6842183 E-MINIC-00-EN 6887223 E-MINIC-00-EN 6888692 E-MINIC-00-EN 6891095 E-MINIC-00-EN 6891108 E-MINIC-00-EN 6892649 E-MINIC-00-EN 6889952 E-MINIC-00-EN 6891045 E-MINIC-00-EN 6860665 E-MINIC-00-EN 6867681 M1013204 6875967 M1013204 6875971 M1013204 6875996 M1013204 6873208 M1013204 6875993 M1013204 6875998 M1013204 6884034 M1013204 6884044 M1013204 6884047 M1013204 6884048 M1013204 6884049 M1013204 6885857 M1013204 6885868 M1013204 6885869 M1013204 6885871 M1013204 6885872 M1013204 6885873 M1013204 6885874 M1013204 6885881 M1013204 6873201 M1013204 6873204 M1013204 6875997 M1013204 6827939 M1013204 6827946 M1013204 6827966 M1013204 6875988 M1013204 6870545 M1013204 6875942 M1013204 6875985 M1013204 6878240 M1013204 6873203 M1013204 6873216 M1013204 6873207 M1013204 6873214 M1013204 6867682 M1013204 6873198 M1013204 6873199 M1013204 6875939 M1013204 6875941 M1013204 6875945 M1013204 6875984 M1013204 6875987 M1013204 6845272 M1013204 6845278 M1013204 6845283 M1013204 6845286 M1013204 6870540 M1013204 6870542 M1013204 6873197 M1013204 6873215 M1013204 6875906 M1013204 6875932 M1013204 6875940 M1013204 6875974 M1013204 6867648 M1013204 6873205 M1013204 6875990 M1013204 6878239 M1013204 6884046 M1013204 6867654 M1013204 6873200 M1013204 6873212 M1013204 6875995 M1013204 6885855 M1013204 6829271 N-FC-00 6839565 N-FC-00 6839571 N-FC-00 6839575 N-FC-00 6839576 N-FC-00 6799192 N-FC-00 6829276 N-FC-00 6829284 N-FC-00 6829292 N-FC-00 6831245 N-FC-00 6831253 N-FC-00 6831254 N-FC-00 6831255 N-FC-00 6831256 N-FC-00 6831261 N-FC-00 6831262 N-FC-00 6831263 N-FC-00 6831264 N-FC-00 6831265 N-FC-00 6831268 N-FC-00 6831274 N-FC-00 6839572 N-FC-00 6839584 N-FC-00 6839592 N-FC-00 6839596 N-FC-00 6849858 N-FC-00 6849884 N-FC-00 6849886 N-FC-00 6849895 N-FC-00 6849897 N-FC-00 6849903 N-FC-00 6849904 N-FC-00 6849905 N-FC-00 6863479 N-FC-00 6863481 N-FC-00 6881148 N-FC-00 6881378 N-FC-00 6884073 N-FC-00 6889541 N-FC-00 6889543 N-FC-00 6881405 N-FC-00 6831252 N-FC-00 6863497 N-FC-00 6863506 N-FC-00 6881404 N-FC-00 6799201 N-FC-00 6829270 N-FC-00 6829278 N-FC-00 6829281 N-FC-00 6829282 N-FC-00 6829288 N-FC-00 6831240 N-FC-00 6831246 N-FC-00 6831275 N-FC-00 6837808 N-FC-00 6837810 N-FC-00 6839555 N-FC-00 6839558 N-FC-00 6839561 N-FC-00 6839573 N-FC-00 6839583 N-FC-00 6839593 N-FC-00 6839594 N-FC-00 6839595 N-FC-00 6849863 N-FC-00 6849896 N-FC-00 6849874 N-FC-00 6849879 N-FC-00 6849881 N-FC-00 6849883 N-FC-00 6849885 N-FC-00 6849889 N-FC-00 6849892 N-FC-00 6849894 N-FC-00 6863478 N-FC-00 6869027 N-FC-00 6869028 N-FC-00 6869030 N-FC-00 6869032 N-FC-00 6870920 N-FC-00 6870921 N-FC-00 6870922 N-FC-00 6870923 N-FC-00 6870925 N-FC-00 6870928 N-FC-00 6870930 N-FC-00 6870931 N-FC-00 6870932 N-FC-00 6870935 N-FC-00 6870937 N-FC-00 6870938 N-FC-00 6870939 N-FC-00 6870940 N-FC-00 6870941 N-FC-00 6870942 N-FC-00 6870943 N-FC-00 6870944 N-FC-00 6870945 N-FC-00 6870946 N-FC-00 6881384 N-FC-00 6881388 N-FC-00 6881393 N-FC-00 6881395 N-FC-00 6881396 N-FC-00 6881398 N-FC-00 6881403 N-FC-00 6884066 N-FC-00 6884067 N-FC-00 6884075 N-FC-00 6884083 N-FC-00 6885565 N-FC-00 6885566 N-FC-00 6885568 N-FC-00 6885569 N-FC-00 6885570 N-FC-00 6885571 N-FC-00 6885573 N-FC-00 6885574 N-FC-00 6885982 N-FC-00 6889544 N-FC-00 6889557 N-FC-00 6889559 N-FC-00 6889563 N-FC-00 6849893 N-FC-00 6829274 N-FC-00 6831237 N-FC-00 6885572 N-FC-00 6839568 N-FC-00 6839574 N-FC-00 6839597 N-FC-00 6839611 N-FC-00 6870958 N-FC-00 6799191 N-FC-00 6799193 N-FC-00 6799194 N-FC-00 6829286 N-FC-00 6829290 N-FC-00 6831266 N-FC-00 6831267 N-FC-00 6839556 N-FC-00 6839598 N-FC-00 6839599 N-FC-00 6839600 N-FC-00 6839601 N-FC-00 6849859 N-FC-00 6849873 N-FC-00 6849878 N-FC-00 6849880 N-FC-00 6849887 N-FC-00 6863496 N-FC-00 6870947 N-FC-00 6870948 N-FC-00 6870949 N-FC-00 6870950 N-FC-00 6870951 N-FC-00 6870952 N-FC-00 6879585 N-FC-00 6879586 N-FC-00 6879587 N-FC-00 6879588 N-FC-00 6881387 N-FC-00 6881399 N-FC-00 6881400 N-FC-00 6881401 N-FC-00 6885975 N-FC-00 6885990 N-FC-00 6829265 N-FC-00 6839603 N-FC-00 6839604 N-FC-00 6839605 N-FC-00 6839607 N-FC-00 6839608 N-FC-00 6839609 N-FC-00 6839610 N-FC-00 6839612 N-FC-00 6849856 N-FC-00 6849857 N-FC-00 6863505 N-FC-00 6831286 N-FC-00 6829268 N-FC-00 6831250 N-FC-00 6839557 N-FC-00 6839559 N-FC-00 6839570 N-FC-00 6839585 N-FC-00 6839586 N-FC-00 6839590 N-FC-00 6849901 N-FC-00 6863494 N-FC-00 6863500 N-FC-00 6863507 N-FC-00 6881380 N-FC-00 6831238 N-FC-00 6831251 N-FC-00 6831257 N-FC-00 6849867 N-FC-00 6839577 N-FC-00 6839578 N-FC-00 6839579 N-FC-00 6839580 N-FC-00 6839581 N-FC-00 6839582 N-FC-00 6839602 N-FC-00 6849861 N-FC-00 6849864 N-FC-00 6849865 N-FC-00 6849868 N-FC-00 6849869 N-FC-00 6849870 N-FC-00 6849871 N-FC-00 6849877 N-FC-00 6849888 N-FC-00 6849890 N-FC-00 6881147 N-FC-00 6881383 N-FC-00 6881406 N-FC-00 6884069 N-FC-00 6884079 N-FC-00 6884080 N-FC-00 6885973 N-FC-00 6885991 N-FC-00 6849898 N-FC-00 6849900 N-FC-00 6849902 N-FC-00 6863487 N-FC-00 6863489 N-FC-00 6863490 N-FC-00 6863492 N-FC-00 6863493 N-FC-00 6863495 N-FC-00 6863498 N-FC-00 6863501 N-FC-00 6863502 N-FC-00 6863503 N-FC-00 6881397 N-FC-00 6885567 N-FC-00 6884068 N-FC-00 6885989 N-FC-00 6889536 N-FC-00 6849875 N-FC-00 6849876 N-FC-00 6831272 N-FC-00 6829277 N-FC-00 6829280 N-FC-00 6829283 N-FC-00 6831241 N-FC-00 6831242 N-FC-00 6831243 N-FC-00 6831247 N-FC-00 6831248 N-FC-00 6831249 N-FC-00 6831277 N-FC-00 6831279 N-FC-00 6831280 N-FC-00 6831284 N-FC-00 6837809 N-FC-00 6837811 N-FC-00 6837812 N-FC-00 6837814 N-FC-00 6839553 N-FC-00 6839554 N-FC-00 6839560 N-FC-00 6839562 N-FC-00 6839563 N-FC-00 6839564 N-FC-00 6839566 N-FC-00 6839569 N-FC-00 6839587 N-FC-00 6839588 N-FC-00 6839589 N-FC-00 6849872 N-FC-00 6849891 N-FC-00 6849899 N-FC-00 6863482 N-FC-00 6863483 N-FC-00 6863484 N-FC-00 6863485 N-FC-00 6863486 N-FC-00 6863488 N-FC-00 6863491 N-FC-00 6863499 N-FC-00 6863504 N-FC-00 6869025 N-FC-00 6869026 N-FC-00 6869029 N-FC-00 6869031 N-FC-00 6869033 N-FC-00 6869034 N-FC-00 6870924 N-FC-00 6870926 N-FC-00 6870927 N-FC-00 6870929 N-FC-00 6870933 N-FC-00 6870934 N-FC-00 6870936 N-FC-00 6881385 N-FC-00 6881386 N-FC-00 6881389 N-FC-00 6881390 N-FC-00 6881391 N-FC-00 6881392 N-FC-00 6881394 N-FC-00 6884065 N-FC-00 6884070 N-FC-00 6884071 N-FC-00 6884072 N-FC-00 6884074 N-FC-00 6884081 N-FC-00 6884082 N-FC-00 6884084 N-FC-00 6885972 N-FC-00 6885974 N-FC-00 6885987 N-FC-00 6885988 N-FC-00 6889538 N-FC-00 6889539 N-FC-00 6889540 N-FC-00 6889551 N-FC-00 6889556 N-FC-00 6829269 N-FC-00 6829279 N-FC-00 6831269 N-FC-00 6831270 N-FC-00 6831281 N-FC-00 6837813 N-FC-00 6839567 N-FC-00 6829275 N-FC-00 6831278 N-FC-00 6837807 N-FC-00 6881402 N-FC-00 6829287 N-FC-00 6870959 N-FC-00 6881146 N-FC-00 6881377 N-FC-00 6881379 N-FC-00 6881381 N-FC-00 6839591 N-FC-00-EN 6831244 N-FC-00-EN 6831258 N-FC-00-EN 6831259 N-FC-00-EN 6831260 N-FC-00-EN 6884078 N-FC-00-EN 6849934 N-FCREC-00 6834302 N-FCREC-00 6834308 N-FCREC-00 6834309 N-FCREC-00 6834310 N-FCREC-00 6834311 N-FCREC-00 6834312 N-FCREC-00 6834313 N-FCREC-00 6834314 N-FCREC-00 6834315 N-FCREC-00 6834316 N-FCREC-00 6834317 N-FCREC-00 6834318 N-FCREC-00 6834319 N-FCREC-00 6834320 N-FCREC-00 6839180 N-FCREC-00 6839182 N-FCREC-00 6839183 N-FCREC-00 6839184 N-FCREC-00 6839185 N-FCREC-00 6839186 N-FCREC-00 6839187 N-FCREC-00 6839188 N-FCREC-00 6839189 N-FCREC-00 6839190 N-FCREC-00 6839191 N-FCREC-00 6839192 N-FCREC-00 6839193 N-FCREC-00 6839194 N-FCREC-00 6839195 N-FCREC-00 6839196 N-FCREC-00 6839197 N-FCREC-00 6839198 N-FCREC-00 6839199 N-FCREC-00 6839200 N-FCREC-00 6839201 N-FCREC-00 6839202 N-FCREC-00 6839203 N-FCREC-00 6839204 N-FCREC-00 6839205 N-FCREC-00 6839206 N-FCREC-00 6839207 N-FCREC-00 6839208 N-FCREC-00 6839209 N-FCREC-00 6839210 N-FCREC-00 6839211 N-FCREC-00 6846254 N-FCREC-00 6846256 N-FCREC-00 6846258 N-FCREC-00 6846259 N-FCREC-00 6849935 N-FCREC-00 6849940 N-FCREC-00 6849942 N-FCREC-00 6849946 N-FCREC-00 6849949 N-FCREC-00 6849950 N-FCREC-00 6863457 N-FCREC-00 6863459 N-FCREC-00 6863460 N-FCREC-00 6863462 N-FCREC-00 6863470 N-FCREC-00 6863475 N-FCREC-00 6877795 N-FCREC-00 6877796 N-FCREC-00 6879675 N-FCREC-00 6879678 N-FCREC-00 6879680 N-FCREC-00 6881407 N-FCREC-00 6877797 N-FCREC-00 6879677 N-FCREC-00 6863476 N-FCREC-00 6863477 N-FCREC-00 6876315 N-FCREC-00 6876321 N-FCREC-00 6876322 N-FCREC-00 6876324 N-FCREC-00 6876325 N-FCREC-00 6876326 N-FCREC-00 6876327 N-FCREC-00 6876331 N-FCREC-00 6876333 N-FCREC-00 6876334 N-FCREC-00 6877787 N-FCREC-00 6877788 N-FCREC-00 6877789 N-FCREC-00 6877790 N-FCREC-00 6877791 N-FCREC-00 6877792 N-FCREC-00 6877798 N-FCREC-00 6879589 N-FCREC-00 6879590 N-FCREC-00 6879591 N-FCREC-00 6879592 N-FCREC-00 6879593 N-FCREC-00 6879594 N-FCREC-00 6879595 N-FCREC-00 6879596 N-FCREC-00 6879597 N-FCREC-00 6879598 N-FCREC-00 6839765 N-FCREC-00 6877785 N-FCREC-00 6876323 N-FCREC-00 6881423 N-FCREC-00 6881424 N-FCREC-00 6849939 N-FCREC-00 6839181 N-FCREC-00 6846255 N-FCREC-00 6876332 N-FCREC-00 6877793 N-FCREC-00 6877794 N-FCREC-00 6879599 N-FCREC-00 6879673 N-FCREC-00 6879676 N-FCREC-00 6879679 N-FCREC-00 6879681 N-FCREC-00 6879682 N-FCREC-00 6879683 N-FCREC-00 6881144 N-FCREC-00 6881145 N-FCREC-00 6881409 N-FCREC-00 6881410 N-FCREC-00 6881411 N-FCREC-00 6881412 N-FCREC-00 6881413 N-FCREC-00 6881414 N-FCREC-00 6881415 N-FCREC-00 6881416 N-FCREC-00 6881417 N-FCREC-00 6881419 N-FCREC-00 6892569 N-FCREC-00 6849936 N-FCREC-00 6849937 N-FCREC-00 6849938 N-FCREC-00 6849941 N-FCREC-00 6881408 N-FCREC-00 6881418 N-FCREC-00 6830983 N-FCREC-00-EN 6830984 N-FCREC-00-EN 6830985 N-FCREC-00-EN 6830986 N-FCREC-00-EN 6830987 N-FCREC-00-EN 6830992 N-FCREC-00-EN 6830993 N-FCREC-00-EN 6839212 N-FCREC-00-EN 6839761 N-FCREC-00-EN 6839762 N-FCREC-00-EN 6839763 N-FCREC-00-EN 6839764 N-FCREC-00-EN 6839766 N-FCREC-00-EN 6839767 N-FCREC-00-EN 6839768 N-FCREC-00-EN 6846241 N-FCREC-00-EN 6846242 N-FCREC-00-EN 6846243 N-FCREC-00-EN 6846244 N-FCREC-00-EN 6846245 N-FCREC-00-EN 6846246 N-FCREC-00-EN 6846247 N-FCREC-00-EN 6846248 N-FCREC-00-EN 6846249 N-FCREC-00-EN 6846250 N-FCREC-00-EN 6846251 N-FCREC-00-EN 6846252 N-FCREC-00-EN 6846253 N-FCREC-00-EN 6846257 N-FCREC-00-EN 6846260 N-FCREC-00-EN 6849908 N-FCREC-00-EN 6849909 N-FCREC-00-EN 6849910 N-FCREC-00-EN 6849911 N-FCREC-00-EN 6849912 N-FCREC-00-EN 6849913 N-FCREC-00-EN 6849914 N-FCREC-00-EN 6849915 N-FCREC-00-EN 6849916 N-FCREC-00-EN 6849917 N-FCREC-00-EN 6849918 N-FCREC-00-EN 6849919 N-FCREC-00-EN 6849920 N-FCREC-00-EN 6849921 N-FCREC-00-EN 6849922 N-FCREC-00-EN 6849923 N-FCREC-00-EN 6849924 N-FCREC-00-EN 6849925 N-FCREC-00-EN 6849926 N-FCREC-00-EN 6849927 N-FCREC-00-EN 6849928 N-FCREC-00-EN 6849930 N-FCREC-00-EN 6849931 N-FCREC-00-EN 6849932 N-FCREC-00-EN 6849933 N-FCREC-00-EN 6849943 N-FCREC-00-EN 6849944 N-FCREC-00-EN 6849945 N-FCREC-00-EN 6849947 N-FCREC-00-EN 6849948 N-FCREC-00-EN 6849951 N-FCREC-00-EN 6849952 N-FCREC-00-EN 6849953 N-FCREC-00-EN 6849954 N-FCREC-00-EN 6849955 N-FCREC-00-EN 6863448 N-FCREC-00-EN 6863449 N-FCREC-00-EN 6863450 N-FCREC-00-EN 6863451 N-FCREC-00-EN 6863452 N-FCREC-00-EN 6863453 N-FCREC-00-EN 6863454 N-FCREC-00-EN 6863455 N-FCREC-00-EN 6863456 N-FCREC-00-EN 6863458 N-FCREC-00-EN 6863461 N-FCREC-00-EN 6863463 N-FCREC-00-EN 6863464 N-FCREC-00-EN 6863465 N-FCREC-00-EN 6863466 N-FCREC-00-EN 6863467 N-FCREC-00-EN 6863469 N-FCREC-00-EN 6863471 N-FCREC-00-EN 6863472 N-FCREC-00-EN 6863473 N-FCREC-00-EN 6863474 N-FCREC-00-EN 6876305 N-FCREC-00-EN 6876306 N-FCREC-00-EN 6876307 N-FCREC-00-EN 6876308 N-FCREC-00-EN 6876309 N-FCREC-00-EN 6876310 N-FCREC-00-EN 6876311 N-FCREC-00-EN 6876312 N-FCREC-00-EN 6876313 N-FCREC-00-EN 6876314 N-FCREC-00-EN 6876316 N-FCREC-00-EN 6876317 N-FCREC-00-EN 6876318 N-FCREC-00-EN 6876319 N-FCREC-00-EN 6876320 N-FCREC-00-EN 6876328 N-FCREC-00-EN 6876329 N-FCREC-00-EN 6876330 N-FCREC-00-EN 6877784 N-FCREC-00-EN 6877786 N-FCREC-00-EN 6879600 N-FCREC-00-EN 6879672 N-FCREC-00-EN 6879684 N-FCREC-00-EN 6879685 N-FCREC-00-EN 6881422 N-FCREC-00-EN 6881425 N-FCREC-00-EN 6881426 N-FCREC-00-EN 6881427 N-FCREC-00-EN 6881428 N-FCREC-00-EN 6881429 N-FCREC-00-EN 6881432 N-FCREC-00-EN 6881433 N-FCREC-00-EN 6881434 N-FCREC-00-EN 6881435 N-FCREC-00-EN 6881436 N-FCREC-00-EN 6905190 N-FCREC-00-EN 6905206 N-FCREC-00-EN 6879671 N-FCREC-00-EN 6849929 N-FCREC-00-EN 6863468 N-FCREC-00-EN 6879674 N-FCREC-00-EN |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
Technical Support 262-513-4122
|
Manufacturer Reason for Recall |
Failure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value may impair clinical decision making for both mechanically and spontaneously ventilated patients because of incorrect low CO2 values.
|
FDA Determined Cause 2 |
Device Design |
Action |
GE Healthcare sent an Urgent Medical Device Correction letter dated June 11, 2014, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, and Product Correction and Contact Information. For questions customers were instructed to contact Technical Support at 1-800-558-7044 or their local Service Representative. |
Quantity in Commerce |
2,948 (761 units US, 2,187 units OUS) |
Distribution |
Worldwide Distribution - USA nationwide, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BRAZIL, CANADA, CHILE, CHINA, Croatia, CZECH REPUBLIC, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, IRAQ, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, MALAYSIA, NETHERLANDS, NEW ZEALAND, OMAN, PERU, POLAND, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, LEBANON, REPUBLIC OF MALDOVIA, COSTA RICA, GABON, INDONESIA, JAMAICA, NIGERIA, NORWAY, NEPAL, PAKISTAN, PANAMA, PHILIPPINES, SENEGAL, SOUTH AFRICA, THAILAND, TUNISIA, and CAMBODIA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = CCK and Original Applicant = DATEX-OHMEDA 510(K)s with Product Code = CCK and Original Applicant = GE HEALTHCARE
|
|
|
|