Date Initiated by Firm |
May 20, 2014 |
Date Posted |
August 29, 2014 |
Recall Status1 |
Terminated 3 on August 14, 2017 |
Recall Number |
Z-2428-2014 |
Recall Event ID |
68536 |
Product Classification |
General surgery tray (kit) - Product Code LRO
|
Product |
GENERAL PACK - CONTENTS: (4) ABSORBENT TOWEL 15" x 20" LIF (1) MAYO STAND COVER REINFORCED LIF ( 1) NEEDLE & BLADE COUNTER 20c FOAM/MAG LIF (2) UTILITY BOWL 16 oz. LIF (1) GLASINE BAG PLAIN ( 1) SUTURE BAG FLORAL (4) DRAPE UTILITY WITH TAPE LIF ( 1) TABLE COVER REINFORCED 50" X 90" LIF (1) SURGICAL BLADE# 10 STAINLESS STEEL ( 1) SURGICAL DURAPREP SOLUTION 26mL LIF ( 1) LAPAROTOMY T DRAPE STD SMS (2) DRAPE SHEET 41" X 58" SMS LIF (2) GOWN IMPERVIOUS REINFORCED X-LARGE SMS LIF (1 0) GAUZE 4"X 4" X 4 PLY ( 1) CAUTERY TIP POLISHER LIF (5) LAP SPONGES PRE-WASH XRD LIF ( 1) CAUTERY PENCIL ROCKER SWITCH LIF (1 0) GAUZE SPONGE 4" X 4" 16 PLY XRD LIF ( 1) TUBE SUCTION CONNECT. 3/16" X 12' LIF ( 1) SURGICAL BLADE# 15 CARBON STEEL ( 1) CONTAINER SPECIMEN 4 oz. WITH LID & LABEL
Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents
|
Code Information |
Product code 900-2003, 29 lots: 112030702 112041277 112062472 112093923 112103923 112114386 112114472 112125311 113015438 113025954 113026205 113036457 113047011 113047156 113057447 113057696 113057814 113068159 113078779 113089205 113099569 131110221 131110405 131110582 140211725 140211874 140312378 140312461 140412832 |
Recalling Firm/ Manufacturer |
Customed, Inc Calle Igualdad Final #7 Fajardo PR 00738
|
For Additional Information Contact |
Rosemari Melendez 787-622-5151 Ext. 7540
|
Manufacturer Reason for Recall |
Customed has determined that there is the possibility that packaging integrity may be compromised on differ catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
|
FDA Determined Cause 2 |
Package design/selection |
Action |
Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit.
Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540 |
Quantity in Commerce |
29 lots; 2144 units (multiple units per lot) |
Distribution |
Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|