Date Initiated by FirmMay 20, 2014
Date PostedAugust 29, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall NumberZ-2523-2014
Recall Event ID 68536
Product Classification General surgery tray (kit) - Product Code LRO
ProductGeneral Cosmetic Pack FCSC, Code 900-3113, contains: (1) COVER MAYO STAND REINFORCED LIF (10) LAP SPONGE PREWASH 18" X 18" XRD LIF (2) GOWN LARGE STANDARD SMS AAMIIII (1) TABLE COVER REINFORCED 50" X 90" LIF (2) TOWEL ABSORBENT 15" X 20" LIF (1) YANKAUER SUCTION TUBE WITHOUT VENT LIF (1) SYRINGE BULB 60CC LIF (1) BLADE SURGICAL #15 CARBON STEEL (1) NEEDLE HYPODERMIC 18G X 1% LIF (6) TOWEL CLOTH HUCK BLUE LIF (4) DRAPE UTILITY WITH TAPE L/F (1) NEEDLE HYPODERMIC 25 X 1% L/F (1) SYRINGE 20CC WITHOUT NEEDLE LUER LOCK LIF (2) SYRINGE 10CC WITHOUT NNEDEL LUER LOCK LIF (2) NEEDLE SPINAL ANESHT 20G X 3%" LIF (1) SKIN MARKER WITH RULER LIF (1) COUNTER NEEDLE & BLADE 20C FOAM/MAG L/F (1) LITE GLOVE (1) DRAPE BREAST/CHEST SMS LIF (1) TUBE SUCTION CONNECT%" X 12' LIF (10) GAUZE SPONGE 4" X 4" 16PLY L/F (2) GLOVE MEDIUM FREETOOUCH VYNIL POWDER FREE (1) TRAY 3/COMPARTMENT (1) WRAPPER 24" X 24" L/F (2) PK. SURG. GLOVE# 7% ENCORE POWDER FREE LATEX Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Code Information 900-3113, 2 lots: 131110435 131210803 
FEI Number 2648727
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information ContactRosemari Melendez
787-622-5151 Ext. 7540
Manufacturer Reason
for Recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
FDA Determined
Cause 2
Package design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity in Commerce2 lots/ 18 units, multiple units per lot
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-