| | Class 2 Device Recall Symbia S Series |  |
| Date Initiated by Firm | August 12, 2014 |
| Date Posted | January 06, 2015 |
| Recall Status1 |
Terminated 3 on December 23, 2016 |
| Recall Number | Z-0911-2015 |
| Recall Event ID |
69114 |
| 510(K)Number | K131634 |
| Product Classification |
System, tomography, computed, emission - Product Code KPS
|
| Product | The Siemens Symbia S series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques: planar imaging, whole body imaging, and tomographic imaging for isotopes with energies up to 588 keV.
Software: The syngo MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT, and other imaging modalities. |
| Code Information |
1) Symbia S; Product Number: 8717741; Serial Numbers: 1001 ; 1002 ; 1003 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1027 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1034 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1215 ; 1216 ; 1217 ; 1218 ; 1219 ; 1220 ; 1221 ; 1222 ; 1223 ; 1224 ; 1225 ; 1226 ; 1227 ; 1228 ; 1229 ; 1230 ; 1231 ; 1232 ; 1233 ; 1234 ; 1235 ; 1236 ; 1237 ; 1238 ; 1239 ; 1240 ; 1241 ; 1242 ; 1243 ; 1244 ; 1245 ; 1246 ; 1247 ; 1248 ; 1249 ; 1250 ; 1251 ; 1252 ; 1253 ; 1254 ; 1255 ; 1256 ; 1257 ; 1258 ; 1259 ; 1260 ; 1261 ; 1262 ; 1263 ; 1264 ; 1265 ; 1266 ; 1267 ; 1268 ; 1269 ; 1270 ; 1271 ; 1272 ; 1273 ; 1274 ; 1275 ; 1276 ; 1277 ; 1278 ; 1279 ; 1280 ; 1281 ; 1282 ; 1283 ; 1284 ; 1285 ; 1286 ; 1287 ; 1288 ; 1289 ; 1290 ; 1291 ; 1292 ; 1293 ; 1294 ; 1295 ; 1296 ; 1297 ; 1298 ; 1299 ; 1300 ; 1301 ; 1302 ; 1303 ; 1304 ; 1305 ; 1306 ; 1307 ; 1308 ; 1310 ; 1311 ; 1312 ; 1313 ; 1314 ; 1315 ; 1317 ; 1318 ; 1319 ; 1320 ; 1321 ; 1322 ; 1323 ; 1324 ; 1325 ; 1326 ; 1327 ; 1328 ; 1329 ; 1330 ; 1331 ; 1332 ; 1333 ; 1334 ; 1335 ; 1336 ; 1337 ; 1338 ; 1339 ; 1341 ; 1342 ; 1343 ; 1344 ; 1345 ; 1346 ; 1347 ; 1348 ; 1349 ; 1350 ; 1351 ; 1352 ; 1353 ; 1354 ; 1355 ; 1356 ; 1357 ; 1358 ; 1359 ; 1360 ; 1363 ; 1364 ; 1365 ; 1366 ; 1367 ; 1368 ; 1369 ; 1370 ; 1371 ; 1372 ; 1373 ; 1374 ; 1375 ; 1376 ; 1377 ; 1378 ; 1379 ; 1380 ; 1381 ; 1382 ; 1383 ; 1384 ; 1385 ; 1386 ; 1387 ; 1388 ; 1389 ; 1390 ; 1391 ; 1392 ; 1393 ; 1394 ; 1395 ; 1396 ; 1397 ; 1398 ; 1399 ; 1400 ; 1401 ; 1402 ; 1403 ; 1404 ; 1405 ; 1406 ; 1407 ; 1408 ; 1409 ; 1410 ; 1411 ; 1412 ; 1413 ; 1414 ; 1415 ; 1416 ; 1417 ; 1418 ; 1419 ; 1420 ; 1421 ; 1422 ; 1423 ; 1424 ; 1425 ; 1426 ; 1427 ; 1428 ; 1429 ; 1430 ; 1431 ; 1432 ; 1433 ; 1434 ; 1435 ; 1436 ; 1437 ; 1438 ; 1439 ; 1440 ; 1441 ; 1442 ; 1443 ; 1444 ; 1445 ; 1446 ; 1447 ; 1448 ; 1449 ; 1450 ; 1451 ; 1452 ; 1453 ; 1454 ; 1455 ; 1456 ; 1457 ; 1458 ; 1459 ; 1460 ; 1461 ; 1462 ; 1463 ; 1464 ; 1465 ; 1466 ; 1467 ; 1468 ; 1469 ; 1470 ; 1471 ; 1472 ; 1473 ; 1474 ; 1475 ; 1476 ; 1477 ; 1478 ; 1479 ; 1480 ; 1481 ; 1482 ; 1483 ; 1484 ; 1485 ; 1486 ; 1487 ; 1488 ; 1489 ; 1490 ; 1491 ; 1492 ; 1493 ; 1494 ; 1495 ; 1496 ; 1497 ; 1498 ; 1499 ; 1500 ; 1501 ; 1502 ; 1503 ; 1504 ; 1505 ; 1506 ; 1507 ; 1508 ; 1509 ; 1510 ; 1511 ; 1512 ; 1513 ; 1514 ; 1515 ; 1516 ; 1517 ; 1518 ; 1519 ; 1520 ; 1521 ; 1522 ; 1523 ; 1524 ; 1525 ; 1526 ; 1527 ; 1528 ; 1529 ; 1530 ; 1531 ; 1532 ; 1533 ; 1534 ; 1535 ; 1536 ; 1537 ; 1538 ; 1539 ; 1540 ; 1541 ; 1542 ; 1543 ; 1544 ; 1545 ; 1546 ; 1547 ; 1548 ; 1549 ; 1550 ; 1551 ; 1552 ; 1553 ; 1554 ; 1555 ; 1556 ; 1557 ; 1558 ; 1559 ; 1560 ; 1561 ; 1562 ; 1563 ; 1564 ; 1565 ; 1566 ; 1567 ; 1568 ; 1569 ; 1570 ; 1571 ; 1572 ; 1573 ; 1574 ; 1575 ; 1576 ; 1577 ; 1578 ; 1579 ; 1580 ; 1581 ; 1582 ; 1583 ; 1584 ; 1585 ; 1586 ; 1588 ; 1589 ; 1590 ; 1591 ; 1592 ; 1593 ; 1594 ; 1595 ; 1596 ; 1597 ; 1598 ; 1599 ; 1600 ; 1601 ; 1602 ; 1603 ; 1604 ; 1605 ; 1606 ; 1607 ; 1608 ; 1609 ; 1610 ; 1611 ; 1612 ; 1613 ; 1614 ; 1615 ; 1616 ; 1617 ; 1618 ; 1619 ; 1620 ; 1621 ; 1622 ; 1623 ; 1624 ; 1625 ; 1626 ; 1627 ; 1628 ; 1629 ; 1630 ; 1631 ; 1632 ; 1633 ; 1634 ; 1635 ; 1636 ; 1637 ; 1638 ; 1639 ; 1640 ; 1641 ; 1642 ; 1643 ; 1644 ; 1645 ; 1646 ; 1647 ; 1648 ; 1649 ; 1650 ; 1651 ; 1652 ; 1653 ; 1654 ; 1655 ; 1656 ; 1657 ; 1658 ; 1659 ; 1660 ; 1661 ; 1662 ; 1663 ; 1664 ; 1665 ; 1666 ; 1667 ; 1668 ; 1669 ; 1670 ; 1671 ; 1672 ; 1673 ; 1674 ; 1675 ; 1676 ; 1677 ; 1678 ; 1679 ; 1680 ; 1681 ; 1682 ; 1683 ; 1684 ; 1685 ; 1686 ; 1687 ; 1688 ; 1689 ; 1690 ; 1691 ; 1692 ; 1693 ; 1694 ; 1695 ; 1696 ; 1697 ; 1698 ; 1699 ; 1700 ; 1701 ; 1702 ; 1703 ; 1704 ; 1705 ; 1706 ; 1707 ; 1708 ; 1709 ; 1710 ; 1711 ; 1712 ; 1713 ; 1714 ; 1715 ; 1716 ; 1717 ; 1718 ; 1719 ; 1720 ; 1721 ; 1722 ; 1723 ; 1724 ; 1725 ; 1726 ; 1727 ; 1728 ; 1729 ; 1730 ; 1731 ; 1732 ; 1733 ; 1734 ; 1735 ; 1736 ; 1737 ; 1738 ; 1739 ; 1740 ; 1741 ; 1742 ; 1743 ; 1744 ; 1745 ; 1746 ; 1747 ; 1748 ; 1749 ; 1750 ; 1751 ; 1752 ; 1753 ; 1754 ; 1755 ; 1756 ; 1757 ; 1758 ; 1759 ; 1760 ; 1761 ; 1762 ; 1763 ; 1764 ; 1765 ; 1766 ; 1767 ; 1768 ; 1769 ; 1770 ; 1771 ; 1772 ; 1773 ; 1774 ; 1775 ; 1776 ; 1777 ; 1778 ; 1779 ; 1780 ; 1781 ; 1782 ; 1783 ; 1784 ; 1785 ; 1786 ; 1787 ; 1788 ; 1789 ; 1790 ; 1791 ; 1792 ; 1793 ; 1794 ; 1795 ; 1796 ; 1797 ; 1798 ; 1799 ; 1800 ; 1801 ; 1802 ; 1803 ; 1804 ; 1805 ; 1806 ; 1807 ; 1808 ; 1809 ; 1810 ; 1811 ; 1812 ; 1813 ; 1814 ; 1815 ; 1816 ; 1817 ; 1818 ; 1819 ; 1820 ; 1821 ; 1822 ; 1823 ; 1824 ; 1825 ; 1826 ; 1827 ; 1828 ; 1829 ; 1830 ; 1831 ; 1832 ; 1833 ; 1834 ; 1835 ; 1836 ; 1837 ; 1838 ; 1839 ; 1840 ; 1841 ; 1842 ; 1843 ; 1844 ; 1845 ; 1846 ; 1847 ; 1848 ; 1849 ; 1850 ; 1851 ; 1852 ; 1853 ; 1854 ; 1855 ; 1856 ; 1857 ; 1858 ; 1859 ; 1860 ; 1861 ; 1862 ; 1863 ; 1864 ; 1865 ; 1866 ; 1867 ; 1868 ; 1869 ; 1870 ; 1871 ; 1872 ; 1873 ; 1874 ; 1875 ; 1876 ; 1877 ; 1878 ; 1879 ; 1880 ; 1881 ; 1882 ; 1883 ; 1884 ; 1885 ; 202703 ; 701158475 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc. 2501 Barrington Rd Hoffman Estates IL 60195-2061
|
| For Additional Information Contact | 800-888-7436 |
Manufacturer Reason for Recall | Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries. |
FDA Determined Cause 2 | Device Design |
| Action | Siemens sent Customer Safety Advisory Notice letters to all customers of record. The letters instructed customers to: 1) continue to use their system(s); 2) place the safety advisory in the system's instructions for use; and, 3) to anticipate that their local service engineer will contact them to perform an inspection and schedule any necessary repairs by December 2014. Customers were also instructed to share the letter with their support staff and, if the equipment was sold or transferred, to forward the safety notice to the new owner of the equipment.
Customers with questions can contact their local service representative or Siemens Medical Solutions at 800-888-7436. |
| Quantity in Commerce | 877 systems |
| Distribution | Worldwide Distribution - USA (naitonwide) and Internationally to Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|