| | Class 2 Device Recall PKG, ROTATING HANDLE, W/RATCHET, INSULATED SHAFT |  |
| Date Initiated by Firm | November 17, 2014 |
| Date Posted | December 12, 2014 |
| Recall Status1 |
Terminated 3 on July 17, 2015 |
| Recall Number | Z-0621-2015 |
| Recall Event ID |
69838 |
| Product Classification |
Endoscope and/or accessories - Product Code KOG
|
| Product | PKG, ROTATING HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080230.
Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. |
| Code Information |
40727 41812 42276 42321 43396 140312 143267 245172 246071 246923 247037 340652 340653 440289 441514 441892 443352 443353 444984 444985 51299 0543138D 0543898D 0640771D 0641429D 0644144D 0646013D 0647316D 0647549D 0740185D 0744066D 0744695D 0746332D 0840071D 0840907D 0842105D 0843936D 0844813D 0846982D 0846986D 0847650D 0847661D 0847671D 92509 0941061D 0941604D 0941726D 0944836D 0945121D 0945472D 0945725D 0945884D 0945983D 1040247D 1040397D 1040426D 1040985D 1042094D 1042466D 1042503D 1042504D 1042940D 1042942D 1044541D 1045186D 1045983D 1046856D 1140701D 1141744D 1142005D 1143438D 1143439D 1144077D 1144676D 1144677D 1145548D 1145549D 1146121D 1240550D 1242235D 1243162 1243162D 1243163 1243596 1340609 1340610 1340610D 1341875 1341876 1342868 1342868D 1342869 1345090 1345090D 1345799 1440293 181299 511298 521298 |
Recalling Firm/ Manufacturer |
Stryker Endoscopy 5900 Optical Ct San Jose CA 95138-1400
|
| For Additional Information Contact | 408-754-2000 |
Manufacturer Reason for Recall | The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods |
FDA Determined Cause 2 | Error in labeling |
| Action | Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below:
www.stryker.com/IFU1000401070
Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314.
Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com.
For questions regarding this recall call 408-754-2000. |
| Quantity in Commerce | 180,573 units total |
| Distribution | Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|