|
Class 2 Device Recall RUSCH ENDOBRONCHIAL TUBE SETS |
|
Date Initiated by Firm |
November 21, 2014 |
Date Posted |
August 18, 2015 |
Recall Status1 |
Terminated 3 on August 15, 2017 |
Recall Number |
Z-2408-2015 |
Recall Event ID |
71727 |
510(K)Number |
K951091
|
Product Classification |
Tube, bronchial (w/wo connector) - Product Code BTS
|
Product |
Bronchial Double Lumen Tube Set (Right), Sterile |
Code Information |
Product Code 162000260 Batch Number 14CT11; Product Code 116200280 Batch Number 13JT05, 13KT18, 14CT11, 14FT05, 14FT10; Product Code 116200350 Batch Number 13FT12R, 13GT11, 13HT07, 13KT23, 14AT08; Product Code 116200370 Batch Number 13KT23, 13LT14, 14DT25, 14ET23; Product Code 116200390 Batch Number 13HT12, 13JT08, 13KT18, 14ET10; Product Code 116200410 Batch Number 14ET05, 14ET22; Product Code 116201350 Batch Number 13FT12R, 13GT05; Product Code 116201410 Batch Number 14ET04; Product Code 116200-000260 Batch Number 13FT12, 13GE29J, 13GT29, 13HT12, 13HT17, 13IT10, 13JT05R, 13JT21, 13LT04, 14DT03, 14ET04, 14ET10, 14ET22, 14FT04, 14GT11, 14GT18, 14GT21, 14HT09, 14HT17; Product Code 116200-000260 - Batch Number 13FT12, 13GE29J, 13GT29, 13HT12, 13HT17, 13IT10, 13JT05R, 13JT21, 13LT04, 14DT03, 14ET04, 14ET10, 14ET22, 14FT04, 14GT11, 14GT18, 14GT21, 14HT09, 14HT17; Product Code 116200-000280 Batch Number 13FT06, 13FT12, 13GE29J, 13HT12, 13HT17, 13IT10, 13JT11, 13KT05, 13KT18, 13LT11, 14CT06, 14CT21, 14DT10, 14DT24, 14ET10, 14ET22, 14FT04, 14FT10, 14GT03, 14GT15, 14GT21, 14HT09, 14IT05, 14IT20; Product Code 116200-000350 Batch Number 13GE28J, 13GT05, 13GT11, 13GT23, 13GT29, 13HT07, 13HT12, 13HT17, 13IT10, 13IT13, 13JT08, 13JT20, 13JT21, 13KT05, 13KT12, 13KT23, 13LT04, 13LT08, 13LT11, 14AT04, 14AT08, 14BT21, 14CT06, 14CT16, 14CT20, 14CT21, 14DT06, 14DT24, 14ET04, 14ET05, 14FT22, 14FT04, 14FT10, 14FT13, 14FT20, 14GT05, 14GT15, 14HT09, 14HT20, 14HT23, 14IT05, 14IT08, 14JT09, 14JT12, 14JT24; Product Code 116200-000370 Batch Number 13FT17, 13GE28J, 13GT05, 13GT16, 13HT12, 13IT05, 13IT10, 13IT16, 13IT20R, 13JT08, 13JT15, 13KT05, 13KT12, 13KT23, 13LT08, 13LT14, 14AT08, 14AT12, 14BT21, 14CT06, 14CT16, 14DT03, 14DT13, 14DT19, 14DT20, 14DT24, 14DT25, 14ET05, 14ET23, 14ET28, 14FT04, 14FT13, 14FT20, 14GT03, 14GT18, 14HT03, 14HT09, 14HT12, 14HT17, 14IT08, 14IT17, 14JT21, 14JT24; Product Code 116200-000390 Batch Number 13FT12, 13FT18, 13GE28J, 13GT16, 13GT29, 13HT12, 13IT05, 13IT10, 13IT13, 13IT20, 13IT20R, 13JT05R, 13JT08, 13JT15, 13JT20R, 13KT05, 13KT18, 13LT08, 14BT15, 14BT16, 14BT21, 14CT11, 14CT16, 14DT13, 14DT15, 14ET04, 14ET10, 14FT04, 14FT16, 14GT03, 14GT15, 14GT21, 14HT03, 14HT17, 14HT23, 14IT05, 14IT08, 14IT17, 14IT20, 14JT12, 14JT24; Product Code 116200-000410 Batch Number 13FT12, 13GE30J, 13GT11, 13GT24, 13GT29, 13HT12, 13HT17, 13IT10, 13IT13, 13IT20R, 13JT05, 13JT05R, 13JT20, 13KT05, 13KT12, 13LT04, 13LT08, 14AT08, 14BT21, 14CT11, 14DT13, 14DT15, 14DT19, 14DT24, 14ET04, 14ET05, 14ET22, 14FT13, 14GT15, 14GT21, 14HT03, 14HT09, 14HT17, 14HT20, 14HT23, 14IT08 |
Recalling Firm/ Manufacturer |
Teleflex, Inc. 550 E Swedesford Rd Ste 400 Wayne PA 19087-1601
|
For Additional Information Contact |
610-378-0131
|
Manufacturer Reason for Recall |
Customer complaints reporting that the cobb connector detached from the main connector prior to use.
|
FDA Determined Cause 2 |
Other |
Action |
Recall letters were sent on 11/21/2014 to consignees requesting them to discontinue use and quarantine the products. The letter includes a reply form to be returned. The letter to the distributors requested a sub-recall, and included a recall letter to be sent to their accounts. |
Distribution |
AL, AK, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MD, MA, MI, MS, MO, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WV. WI, DC, Puerto Rico, Canada, Bolivia, Ecuador, Hong Kong, Malaysia, Philippines, Taiwan, Trinidad & Tobago |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = BTS and Original Applicant = RUSCH INTL.
|
|
|
|