| Date Initiated by Firm | June 29, 2015 |
| Date Posted | August 05, 2015 |
| Recall Status1 |
Terminated 3 on December 07, 2016 |
| Recall Number | Z-2341-2015 |
| Recall Event ID |
71711 |
| 510(K)Number | K123529 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Artis Q systems;
dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. |
| Code Information |
Model Number of device: 10848282 10848281 10848355 10848354 serial numbers: 121006 121013 109016 121011 123008 123003 123011 121014 109011 121004 121018 121036 121012 109015 109009 121019 121010 121015 121016 109007 111000 121020 121009 121056 121057 109021 121008 109020 109027 109010 123010 121041 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc
51 Valley Stream Pkwy
Malvern
PA
19355-1406
|
| For Additional Information Contact | Customer Support 610-219-6300 |
Manufacturer Reason for Recall | The possibility exists for a large amount of liquid to seep through the top cover of the table and penetrate the components inside the table. If this occurs, table movements may no longer be possible. No injuries reported. |
FDA Determined Cause 2 | Device Design |
| Action | A customer safety notice, dated June 29, 2015, was sent to direct accounts to alert them about the issue and that Siemens' service organization will be contacting them to perform the corrective action. |
| Distribution | US Nationwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|