• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall AngioJet SOLENT Omni Thrombectomy Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall AngioJet SOLENT Omni Thrombectomy Setsee related information
Date Initiated by FirmJuly 31, 2015
Date PostedAugust 13, 2015
Recall Status1 Terminated 3 on February 05, 2016
Recall NumberZ-2378-2015
Recall Event ID 71849
510(K)NumberK111182 
Product Classification Catheter, embolectomy - Product Code DXE
ProductAngioJet SOLENT Omni Thrombectomy Set: UPN 109681-004 & 109681-001, Catalog No 109681, Sterilized. For breaking apart and removing thrombus from vascular system, used with the AngioJet Ultra System in breaking apart and removing thrombus from A-V access conduits.
Code Information 162185 162318 
Recalling Firm/
Manufacturer
Boston Scientific Corporation
1 Scimed Pl
Maple Grove MN 55311-1565
For Additional Information ContactNicole Pshon
763-494-1556
Manufacturer Reason
for Recall
AngioJet products were distributed with an incorrect use-by (expiry) date. The correct date is Nov-2015, but the units are labeled with a use-by date of Dec-2015.
FDA Determined
Cause 2
Process control
ActionBoston Scientific sent an Urgent Medical Device Recall Removal letter dated July 31, 2015, to all affected consignees. The letter described the problem and the affected product. Consignees were advised to segregate the affected product and return it to Boston Scientific. They were also requested consignees to complete and return the Account Reply Verification Tracking Form. Contact your Sales Representative for any questions. For questions regarding this recall call 763-494-1556.
Quantity in Commercetotal of 9 units
DistributionWorldwide Distribution - US including AZ, MO, OH and Internationally to GERMANY and SPAIN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.