| | Class 2 Device Recall Diana |  |
| Date Initiated by Firm | March 08, 2016 |
| Create Date | April 11, 2016 |
| Recall Status1 |
Terminated 3 on August 17, 2016 |
| Recall Number | Z-1364-2016 |
| Recall Event ID |
73528 |
| Product Classification |
System/device, pharmacy compounding - Product Code NEP
|
| Product | Diana Automated Compounding System, Diana Onco Plus, Item No. ASN220
The Diana is an automated compounding device for the accurate, safe and efficient preparation of solutions by a pharmacist prior to delivery for infusion into patient. |
| Code Information |
All Serial Numbers |
| FEI Number |
2025816
|
Recalling Firm/ Manufacturer |
ICU Medical, Inc.
951 Calle Amanecer
San Clemente
CA
92673-6212
|
| For Additional Information Contact | 949-366-2183 |
Manufacturer Reason for Recall | ICU Medical Inc. has received a report of compounded volumes not meeting the accuracy specification by the Diana Compounding System when mixing volumes under 5ml. |
FDA Determined Cause 2 | Use error |
| Action | ICU Medical, Inc. sent customer notification letter dated March 8, 2016, to customers to inform them that ICU Medical Inc. has received a report of compounded volumes not meeting the accuracy specification by the Diana Compounding System when mixing volumes under 5ml. The letter informs the customers that ICU Medical will provide an updated Operator's Manual to provide clarity on the difference between Operating Range and Accuracy Range as soon as they are available. Customers are informed that the updated specifications are provided with the customer communication and to be used until the Operators Manual is provided. Customers are instructed to return the completed Medical Device Safety Communication Response form to ICU Medical via Fax (801) 264-1755 or by email at recall@icumed.com. Customers with any questions or require assistance relating to the customer notification letter are instructed to contact ICU Customer Service Monday through Friday between the hours of 8:30 AM and 4:00 PM Pacific time:
(866)8299025 and select option 8, or email the following address:
productreturns@icumed.com. |
| Quantity in Commerce | 36 units |
| Distribution | Nationwide Distribution to Nationwide in AL, AZ, CA, CO, CT, DC, GA, IL, MD, MI, MN, MO, NC, NH, NJ, NY, OH, OK, PA, TN, TX, VA..
Worldwide: AU, BE, BR, CA, CH, DE, DZ, EE, EG, ES, FI, FR, HK, HR, IL, IT, JO, KR, MX, NL, PL, SA, SE, SI, TR. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|