• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Carelink and Carelink Express Monitor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Carelink and Carelink Express Monitorsee related information
Date Initiated by FirmMarch 31, 2016
Create DateMay 04, 2016
Recall Status1 Terminated 3 on September 25, 2017
Recall NumberZ-1605-2016
Recall Event ID 73780
PMA NumberP890003S228 P890003S102 
Product Classification Pulse generator, permanent, implantable - Product Code NVZ
ProductMedtronic CareLink" Monitor (2490C) and Medtronic CareLink Express" Monitor (2020B) Product Usage: The 2490C CareLink Home Monitors and 2020B CareLink Express Monitors are remote monitoring systems that interrogate implanted devices and transmit the data to Medtronic s CareLink Network for viewing by the physician. The data is transmitted either through an analog telephone line, a cellular connection or an internet connection.
Code Information Model Serial Number 2020B LPH003245P 2020B LPH004235P 2020B LPH004270P 2020B LPH004448P 2020B LPH004578P 2020B LPH006282P 2020B LPH006598P 2020B LPH006892P 2490C IJX010143A 2490C IJX010218A 2490C IJX010291A 2490C IJX010304A 2490C IJX010320A 2490C IJX010414A 2490C IJX010416A 2490C IJX010441A 2490C IJX010474A 2490C IJX010489A 2490C IJX010501A 2490C IJX010521A 2490C IJX014234A 2490C IJX014380A 2490C IJX014382A 2490C IJX014387A 2490C IJX014436A 2490C IJX014473A 2490C IJX014502A 2490C IJX014591A 2490C IJX019468A 2490C IJX019469A 2490C IJX019550A 2490C IJX019568A 2490C IJX019647A 2490C IJX019681A 2490C IJX019716A 2490C IJX019729A 2490C IJX019731A 2490C IJX019752A 2490C IJX019753A 2490C IJX019756A 2490C IJX019782A 2490C IJX019824A 2490C IJX019845A 2490C IJX019905A 2490C IJX019933A 2490C IJX019944A 2490C IJX019947A 2490C IJX020167A 2490C IJX020175A 2490C IJX020197A 2490C IJX020240A 2490C IJX020247A 2490C IJX020371A 2490C IJX020403A 2490C IJX020409A 2490C IJX020419A 2490C IJX020450A 2490C IJX020467A 2490C IJX020566A 2490C IJX020620A 2490C IJX020715A 2490C IJX020774A 2490C IJX020781A 2490C IJX020803A 2490C IJX020806A 2490C IJX020811A 2490C IJX020846A 2490C IJX020861A 2490C IJX020862A 2490C IJX020914A 2490C IJX020920A 2490C IJX020987A 2490C IJX020988A 2490C IJX021117A 2490C IJX021158A 2490C IJX021172A 2490C IJX021239A 2490C IJX021335A 2490C IJX021355A 2490C IJX021422A 2490C IJX021477A 2490C IJX021512A 2490C IJX021545A 2490C IJX021551A 2490C IJX021582A 2490C IJX021599A 2490C IJX021610A 2490C IJX021621A 2490C IJX021631A 2490C IJX021646A 2490C IJX021684A 2490C IJX021777A 2490C IJX021783A 2490C IJX021804A 2490C IJX021809A 2490C IJX025346A 2490C IJX028261A 2490C IJX028274A 2490C IJX028283A 2490C IJX028314A 2490C IJX028322A 2490C IJX028333A 2490C IJX028351A 2490C IJX028406A 2490C IJX028432A 2490C IJX028434A 2490C IJX028495A 2490C IJX028503A 2490C IJX028515A 2490C IJX028537A 2490C IJX028543A 2490C IJX028550A 2490C IJX028551A 2490C IJX028568A 2490C IJX028600A 2490C IJX028602A 2490C IJX028635A 2490C IJX028647A 2490C IJX028695A 2490C IJX028698A 2490C IJX028700A 2490C IJX028731A 2490C IJX028745A 2490C IJX030967A 2490C IJX030979A 2490C IJX030993A 2490C IJX031030A 2490C IJX031052A 2490C IJX031112A 2490C IJX031183A 2490C IJX031222A 2490C IJX031262A 2490C IJX031269A 2490C IJX031276A 2490C IJX032098A 2490C IJX032167A 2490C IJX032242A 2490C IJX032243A 2490C IJX032308A 2490C IJX036530A 2490C IJX036532A 2490C IJX036541A 2490C IJX036542A 2490C IJX041980A 2490C IJX041999A 2490C IJX042002A 2490C IJX042033A 2490C IJX042076A 2490C IJX042117A 2490C IJX042136A 2490C IJX042152A 2490C IJX042198A 2490C IJX042210A 2490C IJX043527A 2490C IJX043553A 2490C IJX043590A 2490C IJX043592A 2490C IJX043596A 2490C IJX043609A 2490C IJX043716A 2490C IJX043740A 2490C IJX043741A 2490C IJX043783A 2490C IJX049201A 2490C IJX049221A 2490C IJX049223A 2490C IJX049250A 2490C IJX049273A 2490C IJX049283A 2490C IJX049353A 2490C IJX049363A 2490C IJX049527A 2490C IJX049536A 2490C IJX049833A 2490C IJX049834A 2490C IJX049840A 2490C IJX049843A 2490C IJX049866A 2490C IJX049899A 2490C IJX049900A 2490C IJX049905A 2490C IJX049916A 2490C IJX049937A 2490C IJX050066A 2490C IJX050075A 2490C IJX050157A 2490C IJX050166A 2490C IJX050181A 2490C IJX050198A 2490C IJX050201A 2490C IJX050236A 2490C IJX050237A 2490C IJX050242A 2490C IJX050244A 2490C IJX050271A 2490C IJX057326A 2490C IJX057347A 2490C IJX057374A 2490C IJX057376A 2490C IJX057393A 2490C IJX057404A 2490C IJX057420A 2490C IJX057497A 2490C IJX057603A 2490C IJX057608A 2490C IJX057610A 2490C IJX057622A 2490C IJX057623A 2490C IJX057648A 2490C IJX057671A 2490C IJX057683A 2490C IJX057725A 2490C IJX057781A 2490C IJX057808A 2490C IJX057817A 2490C IJX057821A 2490C IJX057825A 2490C IJX057834A 2490C IJX057838A 2490C IJX057907A 2490C IJX057967A 2490C IJX057978A 2490C IJX058293A 2490C IJX058297A 2490C IJX058300A 2490C IJX058333A 2490C IJX058352A 2490C IJX058354A 2490C IJX058378A 2490C IJX058396A 2490C IJX058436A 2490C IJX058443A 2490C IJX058453A 2490C IJX058470A 2490C IJX058519A 2490C IJX058550A 2490C IJX058556A 2490C IJX058565A 2490C IJX058570A 2490C IJX058586A 2490C IJX058611A 2490C IJX058669A 2490C IJX058704A 2490C IJX058766A 2490C IJX064936A 2490C IJX064975A 2490C IJX065043A 2490C IJX065061A 2490C IJX065074A 2490C IJX065109A 2490C IJX065122A 2490C IJX065140A 2490C IJX065147A 2490C IJX065148A 2490C IJX065160A 2490C IJX065166A 2490C IJX065182A 2490C IJX065206A 2490C IJX065305A 2490C IJX065310A 2490C IJX065321A 2490C IJX065325A 2490C IJX065359A 2490C IJX065362A 2490C IJX065364A 2490C IJX065400A 2490C IJX065410A 2490C IJX065416A 2490C IJX065449A 2490C IJX065468A 2490C IJX065470A 2490C IJX065473A 2490C IJX065474A 2490C IJX065490A 2490C IJX065491A 2490C IJX065496A 2490C IJX065502A 2490C IJX065507A 2490C IJX065531A 2490C IJX065532A 2490C IJX065549A 2490C IJX065554A 2490C IJX065582A 2490C IJX065587A 2490C IJX065588A 2490C IJX065625A 2490C IJX065695A 2490C IJX065783A 2490C IJX066483A 2490C IJX066504A 2490C IJX066511A 2490C IJX066526A 2490C IJX066549A 2490C IJX066559A 2490C IJX066563A 2490C IJX066566A 2490C IJX066584A 2490C IJX066589A 2490C IJX066594A 2490C IJX066601A 2490C IJX066613A 2490C IJX066642A 2490C IJX066700A 2490C IJX066713A 2490C IJX075368A 2490C IJX075370A 2490C IJX075392A 2490C IJX075393A 2490C IJX075401A 2490C IJX075428A 2490C IJX075484A 2490C IJX075486A 2490C IJX075527A 2490C IJX075556A 2490C IJX075576A 2490C IJX075656A 2490C IJX075692A 2490C IJX075706A 2490C IJX075771A 2490C IJX075794A 2490C IJX075832A 2490C IJX075837A 2490C IJX075838A 2490C IJX075849A 2490C IJX075988A 2490C IJX076011A 2490C IJX076029A 2490C IJX076032A 2490C IJX076592A 2490C IJX076617A 2490C IJX076722A 2490C IJX076730A 2490C IJX076755A 2490C IJX076756A 2490C IJX076780A 2490C IJX076846A 2490C IJX076885A 2490C IJX076908A 2490C IJX076912A 2490C IJX076917A 2490C IJX076952A 2490C IJX076962A 2490C IJX076983A 2490C IJX076988A 2490C IJX076990A 2490C IJX076996A 2490C IJX077087A 2490C IJX077136A 2490C IJX077144A 2490C IJX077194A 2490C IJX078017A 2490C IJX078035A 2490C IJX078043A 2490C IJX078084A 2490C IJX078130A 2490C IJX078138A 2490C IJX078141A 2490C IJX078178A 2490C IJX078179A 2490C IJX078189A 2490C IJX078191A 2490C IJX078201A 2490C IJX078256A 2490C IJX078263A 2490C IJX078278A 2490C IJX078282A 2490C IJX078289A 2490C IJX080778A 2490C IJX080790A 2490C IJX080792A 2490C IJX080794A 2490C IJX080840A 2490C IJX080873A 2490C IJX080883A 2490C IJX080885A 2490C IJX080976A 2490C IJX081005A 2490C IJX081021A 2490C IJX081068A 2490C IJX083316A 2490C IJX083326A 2490C IJX083422A 2490C IJX083435A 2490C IJX083449A 2490C IJX083705A 2490C IJX083706A 2490C IJX083728A 2490C IJX083741A 2490C IJX083802A 2490C IJX083817A 2490C IJX083833A 2490C IJX083855A 2490C IJX084030A 2490C IJX084046A 2490C IJX084051A 2490C IJX084055A 2490C IJX084159A 2490C IJX084173A 2490C IJX089855A 2490C IJX089963A 2490C IJX090000A 2490C IJX090028A 2490C IJX090037A 2490C IJX090183A 2490C IJX090228A 2490C IJX090312A 2490C IJX090317A 2490C IJX090328A 2490C IJX090389A 2490C IJX090401A 2490C IJX090402A 2490C IJX090464A 2490C IJX090508A 2490C IJX090514A 2490C IJX090565A 2490C IJX090585A 2490C IJX090590A 2490C IJX090612A 2490C IJX090622A 2490C IJX090637A 2490C IJX090680A 2490C IJX095795A 2490C IJX095809A 2490C IJX095815A 2490C IJX095827A 2490C IJX095835A 2490C IJX095837A 2490C IJX095911A 2490C IJX095940A 2490C IJX095975A 2490C IJX095987A 2490C IJX096014A 2490C IJX096025A 2490C IJX096079A 2490C IJX096104A 2490C IJX096113A 2490C IJX096173A 2490C IJX096248A 2490C IJX096263A 2490C IJX096264A 2490C IJX096302A 2490C IJX096325A 2490C IJX096326A 2490C IJX096342A 2490C IJX097406P 2490C IJX097538P 2490C IJX097571P 2490C IJX104171A 2490C IJX104177A 2490C IJX104196A 2490C IJX104203A 2490C IJX104226A 2490C IJX104269A 2490C IJX104279A 2490C IJX104405A 2490C IJX104431A 2490C IJX104437A 2490C IJX104464A 2490C IJX104471A 2490C IJX106151A 2490C IJX106480A 2490C IJX106526A 2490C IJX106527A 2490C IJX106554A 2490C IJX106604A 2490C IJX106612A 2490C IJX106754A 2490C IJX106792A 2490C IJX106795A 2490C IJX106839A 2490C IJX106846A 2490C IJX106981A 2490C IJX106985A 2490C IJX107000A 2490C IJX107006A 2490C IJX107086A 2490C IJX107139A 2490C IJX107208A 2490C IJX107247A 2490C IJX107258A 2490C IJX107276A 2490C IJX107428A 2490C IJX112502A 2490C IJX112517A 2490C IJX112540A 2490C IJX112547A 2490C IJX112581A 2490C IJX112584A 2490C IJX112638A 2490C IJX112652A 2490C IJX112659A 2490C IJX112662A 2490C IJX112664A 2490C IJX112690A 2490C IJX112731A 2490C IJX112770A 2490C IJX112806A 2490C IJX112849A 2490C IJX112851A 2490C IJX112852A 2490C IJX112911A 2490C IJX113034A 2490C IJX113104A 2490C IJX113131A 2490C IJX113191A 2490C IJX113209A 2490C IJX114594A 2490C IJX114634A 2490C IJX114696A 2490C IJX114707A 2490C IJX114767A 2490C IJX114769A 2490C IJX114780A 2490C IJX119334A 2490C IJX119394A 2490C IJX119419A 2490C IJX119428A 2490C IJX119496A 2490C IJX119519A 2490C IJX119627A 2490C IJX119630A 2490C IJX119648A 2490C IJX119729A 2490C IJX119801A 2490C IJX119808A 2490C IJX119820A 2490C IJX121094A 2490C IJX121148A 2490C IJX121155A 2490C IJX121181A 2490C IJX121253A 2490C IJX121290A 2490C IJX121293A 2490C IJX121298A 2490C IJX121306A 2490C IJX121388A 2490C IJX121395A 2490C IJX124582A 2490C IJX124587A 2490C IJX124591A 2490C IJX124611A 2490C IJX124635A 2490C IJX124639A 2490C IJX124658A 2490C IJX124676A 2490C IJX124679A 2490C IJX124687A 2490C IJX124700A 2490C IJX124715A 2490C IJX124717A 2490C IJX124728A 2490C IJX124732A 2490C IJX124734A 2490C IJX124753A 2490C IJX124764A 2490C IJX124769A 2490C IJX124772A 2490C IJX124782A 2490C IJX124818A 2490C IJX124831A 2490C IJX124832A 2490C IJX124860A 2490C IJX124889A 2490C IJX124890A 2490C IJX124933A 2490C IJX124940A 2490C IJX124953A 2490C IJX124964A 2490C IJX125761A 2490C IJX125764A 2490C IJX125913A 2490C IJX125925A 2490C IJX125980A 2490C IJX126200A 2490C IJX126255A 2490C IJX127441A 2490C IJX127456A 2490C IJX127458A 2490C IJX127471A 2490C IJX127496A 2490C IJX127498A 2490C IJX127521A 2490C IJX127539A 2490C IJX127552A 2490C IJX127555A 2490C IJX127579A 2490C IJX127581A 2490C IJX127592A 2490C IJX127598A 2490C IJX127600A 2490C IJX127605A 2490C IJX127610A 2490C IJX127616A 2490C IJX127637A 2490C IJX127641A 2490C IJX127676A 2490C IJX127691A 2490C IJX127723A 2490C IJX127765A 2490C IJX127775A 2490C IJX127915A 2490C IJX130885A 2490C IJX130888A 2490C IJX130905A 2490C IJX137179P 2490C IJX137212P 2490C IJX137236P 2490C IJX137252P 2490C IJX137272P 2490C IJX137291P 2490C IJX137299P 2490C IJX137327P 2490C IJX137335P 2490C IJX139699A 2490C IJX139719A 2490C IJX139745A 2490C IJX139777A 2490C IJX139805A 2490C IJX139887A 2490C IJX139908A 2490C IJX139916A 2490C IJX139962A 2490C IJX139964A 2490C IJX139971A 2490C IJX139987A 2490C IJX140031A 2490C IJX140032A 2490C IJX140083A 2490C IJX140090A 2490C IJX140107A 2490C IJX140111A 2490C IJX140116A 2490C IJX140143A 2490C IJX140144A 2490C IJX140176A 2490C IJX140195A 2490C IJX140212A 2490C IJX140256A 2490C IJX140273A 2490C IJX140305A 2490C IJX140317A 2490C IJX140423A 2490C IJX140468A 2490C IJX140491A 2490C IJX140544A 2490C IJX140566A 2490C IJX140575A 2490C IJX140636A 2490C IJX140659A 2490C IJX140714A 2490C IJX140718A 2490C IJX140793A 2490C IJX140827A 2490C IJX140847A 2490C IJX140869A 2490C IJX140877A 2490C IJX140907A 2490C IJX140953A 2490C IJX140954A 2490C IJX140968A 2490C IJX140974A 2490C IJX141044A 2490C IJX141057A 2490C IJX145888A 2490C IJX145893A 2490C IJX145917A 2490C IJX145924A 2490C IJX145947A 2490C IJX145993A 2490C IJX145998A 2490C IJX146000A 2490C IJX146021A 2490C IJX146042A 2490C IJX146046A 2490C IJX146056A 2490C IJX146057A 2490C IJX146059A 2490C IJX146075A 2490C IJX146081A 2490C IJX146086A 2490C IJX146107A 2490C IJX146129A 2490C IJX146137A 2490C IJX146141A 2490C IJX146146A 2490C IJX146157A 2490C IJX146161A 2490C IJX146162A 2490C IJX146165A 2490C IJX146179A 2490C IJX150029A 2490C IJX150073A 2490C IJX150148A 2490C IJX150158A 2490C IJX150200A 2490C IJX150233A 2490C IJX150247A 2490C IJX150249A 2490C IJX150258A 2490C IJX150282A 2490C IJX150319A 2490C IJX150323A 2490C IJX150332A 2490C IJX150335A 2490C IJX150339A 2490C IJX150367A 2490C IJX150403A 2490C IJX150412A 2490C IJX150427A 2490C IJX150473A 2490C IJX150566A 2490C IJX150631A 2490C IJX150717A 2490C IJX150725A 2490C IJX150728A 2490C IJX150732A 2490C IJX150738A 2490C IJX150744A 2490C IJX150759A 2490C IJX151044A 2490C IJX151118A 2490C IJX151136A 2490C IJX162824A 2490C IJX162825A 2490C IJX162844A 2490C IJX162851A 2490C IJX162854A 2490C IJX162856A 2490C IJX162857A 2490C IJX162874A 2490C IJX162917A 2490C IJX162925A 2490C IJX162944A 2490C IJX162947A 2490C IJX162973A 2490C IJX162995A 2490C IJX163006A 2490C IJX163041A 2490C IJX163044A 2490C IJX163115A 2490C IJX163172A 2490C IJX163174A 2490C IJX163207A 2490C IJX163231A 2490C IJX163236A 2490C IJX163241A 2490C IJX163282A 2490C IJX163287A 2490C IJX163306A 2490C IJX163335A 2490C IJX163368A 2490C IJX163372A 2490C IJX163385A 2490C IJX163474A 2490C IJX163479A 2490C IJX163481A 2490C IJX163487A 2490C IJX163501A 2490C IJX163514A 2490C IJX163518A 2490C IJX163559A 2490C IJX163566A 2490C IJX163570A 2490C IJX163590A 2490C IJX163592A 2490C IJX163599A 2490C IJX163611A 2490C IJX163636A 2490C IJX163638A 2490C IJX163646A 2490C IJX163672A 2490C IJX163680A 2490C IJX163699A 2490C IJX163717A 2490C IJX164705A 2490C IJX164711A 2490C IJX164713A 2490C IJX164717A 2490C IJX164718A 2490C IJX164721A 2490C IJX164735A 2490C IJX164746A 2490C IJX164760A 2490C IJX164766A 2490C IJX164779A 2490C IJX164830A 2490C IJX164890A 2490C IJX164910A 2490C IJX164911A 2490C IJX164969A 2490C IJX164995A 2490C IJX165042A 2490C IJX165054A 2490C IJX165085A 2490C IJX165090A 2490C IJX165095A 2490C IJX165148A 2490C IJX167164A 2490C IJX167192A 2490C IJX167273A 2490C IJX167274A 2490C IJX167290A 2490C IJX167303A 2490C IJX167342A 2490C IJX167347A 2490C IJX167404A 2490C IJX167472A 2490C IJX167480A 2490C IJX167486A 2490C IJX167489A 2490C IJX167497A 2490C IJX167499A 2490C IJX167524A 2490C IJX167548A 2490C IJX167604A 2490C IJX167617A 2490C IJX167625A 2490C IJX167676A 2490C IJX167682A 2490C IJX167708A 2490C IJX167719A 2490C IJX167728A 2490C IJX177710A 2490C IJX177728A 2490C IJX177736A 2490C IJX177737A 2490C IJX177738A 2490C IJX177741A 2490C IJX177743A 2490C IJX177744A 2490C IJX177749A 2490C IJX177750A 2490C IJX177758A 2490C IJX177760A 2490C IJX177761A 2490C IJX177768A 2490C IJX177784A 2490C IJX177791A 2490C IJX177806A 2490C IJX177815A 2490C IJX177816A 2490C IJX177818A 2490C IJX177831A 2490C IJX177837A 2490C IJX177851A 2490C IJX177852A 2490C IJX177876A 2490C IJX177878A 2490C IJX177897A 2490C IJX177917A 2490C IJX177926A 2490C IJX177932A 2490C IJX177933A 2490C IJX177935A 2490C IJX177957A 2490C IJX177977A 2490C IJX177979A 2490C IJX179386A 2490C IJX179434A 2490C IJX179482A 2490C IJX179502A 2490C IJX179505A 2490C IJX179547A 2490C IJX179549A 2490C IJX179554A 2490C IJX179570A 2490C IJX179597A 2490C IJX179616A 2490C IJX179620A 2490C IJX179629A 2490C IJX179646A 2490C IJX179781A 2490C IJX179822A 2490C IJX179840A 2490C IJX179871A 2490C IJX179894A 2490C IJX179898A 2490C IJX179937A 2490C IJX179947A 2490C IJX179981A 2490C IJX179985A 2490C IJX180008A 2490C IJX180024A 2490C IJX184596A 2490C IJX184608A 2490C IJX184643A 2490C IJX184646A 2490C IJX184753A 2490C IJX184754A 2490C IJX184777A 2490C IJX184787A 2490C IJX185629A 2490C IJX185631A 2490C IJX185639A 2490C IJX185647A 2490C IJX185648A 2490C IJX185667A 2490C IJX185698A 2490C IJX185778A 2490C IJX185779A 2490C IJX185791A 2490C IJX185801A 2490C IJX185809A 2490C IJX185826A 2490C IJX185877A 2490C IJX185886A 2490C IJX185889A 2490C IJX185896A 2490C IJX185915A 2490C IJX185945A 2490C IJX185948A 2490C IJX185950A 2490C IJX185952A 2490C IJX186019A 2490C IJX186039A 2490C IJX186049A 2490C IJX186052A 2490C IJX186059A 2490C IJX186110A 2490C IJX186111A 2490C IJX186128A 2490C IJX186171A 2490C IJX186177A 2490C IJX186218A 2490C IJX186269A 2490C IJX186275A 2490C IJX188572A 2490C IJX188608A 2490C IJX188620A 2490C IJX188677A 2490C IJX188678A 2490C IJX188689A 2490C IJX188695A 2490C IJX188723A 2490C IJX188730A 2490C IJX188735A 2490C IJX188736A 2490C IJX188745A 2490C IJX188746A 2490C IJX188752A 2490C IJX188754A 2490C IJX188757A 2490C IJX188772A 2490C IJX188809A 2490C IJX188847A 2490C IJX188898A 2490C IJX188954A 2490C IJX188991A 2490C IJX189011A 2490C IJX189043A 2490C IJX189051A 2490C IJX189081A 2490C IJX189118A 2490C IJX189193A 2490C IJX189196A 2490C IJX189209A 2490C IJX189224A 2490C IJX189255A 2490C IJX189280A 2490C IJX189292A 2490C IJX189293A 2490C IJX189324A 2490C IJX189325A 2490C IJX189326A 2490C IJX189337A 2490C IJX189346A 2490C IJX189350A 2490C IJX189470A 2490C IJX189472A 2490C IJX189493A 2490C IJX189494A 2490C IJX189536A 2490C IJX189546A 2490C IJX189549A 2490C IJX189596A 2490C IJX189608A 2490C IJX195465A 2490C IJX195470A 2490C IJX195519A 2490C IJX195549A 2490C IJX195580A 2490C IJX195594A 2490C IJX195641A 2490C IJX195681A 2490C IJX195707A 2490C IJX195767A 2490C IJX195781A 2490C IJX195794A 2490C IJX195811A 2490C IJX195837A 2490C IJX195843A 2490C IJX195878A 2490C IJX195899A 2490C IJX195960A 2490C IJX195969A 2490C IJX195979A 2490C IJX195986A 2490C IJX196000A 2490C IJX196022A 2490C IJX196031A 2490C IJX196034A 2490C IJX196104A 2490C IJX196157A 2490C IJX196184A 2490C IJX196192A 2490C IJX196215A 2490C IJX196280A 2490C IJX196291A 2490C IJX196405A 2490C IJX196431A 2490C IJX196433A 2490C IJX196437A 2490C IJX196442A 2490C IJX196452A 2490C IJX196455A 2490C IJX196571A 2490C IJX196585A 2490C IJX196608A 2490C IJX196631A 2490C IJX198990A 2490C IJX198998A 2490C IJX199009A 2490C IJX199011A 2490C IJX199013A 2490C IJX199032A 2490C IJX199069A 2490C IJX199075A 2490C IJX199104A 2490C IJX199121A 2490C IJX199169A 2490C IJX199173A 2490C IJX199176A 2490C IJX199184A 2490C IJX199197A 2490C IJX199210A 2490C IJX199252A 2490C IJX199255A 2490C IJX199272A 2490C IJX199291A 2490C IJX199320A 2490C IJX199325A 2490C IJX199331A 2490C IJX199338A 2490C IJX199340A 2490C IJX199371A 2490C IJX199391A 2490C IJX199419A 2490C IJX199421A 2490C IJX199429A 2490C IJX199458A 2490C IJX199461A 2490C IJX199525A 2490C IJX199546A 2490C IJX199560A 2490C IJX199566A 2490C IJX199607A 2490C IJX199613A 2490C IJX199623A 2490C IJX199660A 2490C IJX199665A 2490C IJX199682A 2490C IJX199687A 2490C IJX199692A 2490C IJX199711A 2490C IJX199737A 2490C IJX207528A 2490C IJX207529A 2490C IJX207555A 2490C IJX207587A 2490C IJX207682A 2490C IJX207766A 2490C IJX207770A 2490C IJX207773A 2490C IJX207774A 2490C IJX207823A 2490C IJX207824A 2490C IJX207856A 2490C IJX207866A 2490C IJX207892A 2490C IJX207895A 2490C IJX207898A 2490C IJX207902A 2490C IJX207908A 2490C IJX207983A 2490C IJX208008A 2490C IJX208017A 2490C IJX208023A 2490C IJX208032A 2490C IJX208033A 2490C IJX208053A 2490C IJX208126A 2490C IJX208217A 2490C IJX208233A 2490C IJX208241A 2490C IJX208338A 2490C IJX208339A 2490C IJX208360A 2490C IJX208361A 2490C IJX208369A 2490C IJX208397A 2490C IJX208460A 2490C IJX208487A 2490C IJX215227A 2490C IJX215263A 2490C IJX215266A 2490C IJX215273A 2490C IJX215294A 2490C IJX215335A 2490C IJX215358A 2490C IJX215416A 2490C IJX215472A 2490C IJX215475A 2490C IJX215514A 2490C IJX215520A 2490C IJX215550A 2490C IJX215593A 2490C IJX215612A 2490C IJX215624A 2490C IJX215694A 2490C IJX215722A 2490C IJX215751A 2490C IJX215756A 2490C IJX215757A 2490C IJX215759A 2490C IJX215766A 2490C IJX215771A 2490C IJX215781A 2490C IJX215783A 2490C IJX215795A 2490C IJX215824A 2490C IJX215837A 2490C IJX215844A 2490C IJX215885A 2490C IJX215908A 2490C IJX215927A 2490C IJX215929A 2490C IJX215938A 2490C IJX215960A 2490C IJX215961A 2490C IJX215962A 2490C IJX216026A 2490C IJX216033A 2490C IJX216044A 2490C IJX216072A 2490C IJX216079A 2490C IJX216105A 2490C IJX216143A 2490C IJX216153A 2490C IJX216189A 2490C IJX216208A 2490C IJX216210A 2490C IJX216949A 2490C IJX216978A 2490C IJX216981A 2490C IJX216989A 2490C IJX217037A 2490C IJX217058A 2490C IJX217091A 2490C IJX217138A 2490C IJX217146A 2490C IJX217157A 2490C IJX217180A 2490C IJX217187A 2490C IJX217192A 2490C IJX217196A 2490C IJX217207A 2490C IJX217210A 2490C IJX217278A 2490C IJX217323A 2490C IJX217325A 2490C IJX217334A 2490C IJX217336A 2490C IJX217349A 2490C IJX217360A 2490C IJX217387A 2490C IJX217392A 2490C IJX217410A 2490C IJX217417A 2490C IJX217438A 2490C IJX217475A 2490C IJX217547A 2490C IJX217580A 2490C IJX217616A 2490C IJX217623A 2490C IJX217645A 2490C IJX217665A 2490C IJX217984A 2490C IJX218020A 2490C IJX218030A 2490C IJX218038A 2490C IJX218098A 2490C IJX218108A 2490C IJX218113A 2490C IJX218116A 2490C IJX218145A 2490C IJX218219A 2490C IJX218226A 2490C IJX218296A 2490C IJX218325A 2490C IJX218339A 2490C IJX218359A 2490C IJX218373A 2490C IJX218393A 2490C IJX219993A 2490C IJX219997A 2490C IJX220063A 2490C IJX220149A 2490C IJX220156A 2490C IJX220171A 2490C IJX220231A 2490C IJX220235A 2490C IJX220241A 2490C IJX220718A 2490C IJX220721A 2490C IJX220763A 2490C IJX220876A 2490C IJX220929A 2490C IJX220950A 2490C IJX220981A 2490C IJX220985A 2490C IJX221008A 2490C IJX221039A 2490C IJX221063A 2490C IJX221077A 2490C IJX221078A 2490C IJX221092A 2490C IJX221145A 2490C IJX221152A 2490C IJX221155A 2490C IJX221203A 2490C IJX221212A 2490C IJX221221A 2490C IJX221330A 2490C IJX221401A 2490C IJX221472A 2490C IJX221494A 2490C IJX221503A 2490C IJX221511A 2490C IJX221517A 2490C IJX221588A 2490C IJX221594A 2490C IJX221610A 2490C IJX221611A 2490C IJX221628A 2490C IJX221632A 2490C IJX221652A 2490C IJX221655A 2490C IJX221694A 2490C IJX221735A 2490C IJX221790A 2490C IJX221826A 2490C IJX221837A 2490C IJX221901A 2490C IJX225284A 2490C IJX225293A 2490C IJX225309A 2490C IJX225321A 2490C IJX225329A 2490C IJX225337A 2490C IJX225360A 2490C IJX225361A 2490C IJX225363A 2490C IJX225410A 2490C IJX225421A 2490C IJX225446A 2490C IJX225448A 2490C IJX225449A 2490C IJX225459A 2490C IJX225470A 2490C IJX225482A 2490C IJX225500A 2490C IJX225503A 2490C IJX225512A 2490C IJX225522A 2490C IJX225523A 2490C IJX225531A 2490C IJX225534A 2490C IJX225538A 2490C IJX225547A 2490C IJX225568A 2490C IJX225605A 2490C IJX225608A 2490C IJX225610A 2490C IJX225615A 2490C IJX225620A 2490C IJX225625A 2490C IJX225627A 2490C IJX225631A 2490C IJX225645A 2490C IJX225646A 2490C IJX225650A 2490C IJX225653A 2490C IJX225678A 2490C IJX225698A 2490C IJX225699A 2490C IJX225708A 2490C IJX225719A 2490C IJX225727A 2490C IJX225729A 2490C IJX225732A 2490C IJX227047A 2490C IJX227048A 2490C IJX227053A 2490C IJX227057A 2490C IJX227059A 2490C IJX227101A 2490C IJX227116A 2490C IJX227128A 2490C IJX227141A 2490C IJX227142A 2490C IJX227150A 2490C IJX227156A 2490C IJX227167A 2490C IJX227176A 2490C IJX227190A 2490C IJX227193A 2490C IJX227196A 2490C IJX227210A 2490C IJX227247A 2490C IJX227290A 2490C IJX227311A 2490C IJX227339A 2490C IJX227361A 2490C IJX227365A 2490C IJX227383A 2490C IJX227386A 2490C IJX227392A 2490C IJX227403A 2490C IJX227432A 2490C IJX227445A 2490C IJX227482A 2490C IJX227485A 2490C IJX227487A 2490C IJX227537A 2490C IJX227543A 2490C IJX227562A 2490C IJX227586A 2490C IJX227651A 2490C IJX227694A 2490C IJX227731A 2490C IJX227737A 2490C IJX227781A 2490C IJX227782A 2490C IJX227810A 2490C IJX227818A 2490C IJX227823A 2490C IJX227868A 2490C IJX227904A 2490C IJX227958A 2490C IJX227960A 2490C IJX227963A 2490C IJX228023A 2490C IJX228034A 2490C IJX228039A 2490C IJX228087A 2490C IJX228110A 2490C IJX228114A 2490C IJX228136A 2490C IJX228138A 2490C IJX228146A 2490C IJX228151A 2490C IJX228160A 2490C IJX228208A 2490C IJX228224A 2490C IJX228262A 2490C IJX228282A 2490C IJX228283A 2490C IJX228284A 2490C IJX228299A 2490C IJX228332A 2490C IJX228369A 2490C IJX228370A 2490C IJX228372A 2490C IJX228387A 2490C IJX228389A 2490C IJX228396A 2490C IJX228398A 2490C IJX228434A 2490C IJX228449A 2490C IJX228463A 2490C IJX228469A 2490C IJX228477A 2490C IJX228479A 2490C IJX228515A 2490C IJX228518A 2490C IJX228526A 2490C IJX228530A 2490C IJX228541A 2490C IJX228558A 2490C IJX228565A 2490C IJX235788A 2490C IJX235792A 2490C IJX235793A 2490C IJX235803A 2490C IJX235806A 2490C IJX235817A 2490C IJX235864A 2490C IJX235872A 2490C IJX235988A 2490C IJX235999A 2490C IJX236013A 2490C IJX236021A 2490C IJX236037A 2490C IJX236057A 2490C IJX236080A 2490C IJX236092A 2490C IJX236123A 2490C IJX236136A 2490C IJX236141A 2490C IJX236142A 2490C IJX236143A 2490C IJX236178A 2490C IJX236205A 2490C IJX236215A 2490C IJX236241A 2490C IJX236269A 2490C IJX236280A 2490C IJX236281A 2490C IJX236304A 2490C IJX236313A 2490C IJX236334A 2490C IJX236358A 2490C IJX236372A 2490C IJX236397A 2490C IJX236409A 2490C IJX236414A 2490C IJX236464A 2490C IJX236487A 2490C IJX236508A 2490C IJX236509A 2490C IJX236512A 2490C IJX236549A 2490C IJX236583A 2490C IJX236629A 2490C IJX236681A 2490C IJX236684A 2490C IJX236689A 2490C IJX236714A 2490C IJX236747A 2490C IJX236748A 2490C IJX237572A 2490C IJX237593A 2490C IJX237609A 2490C IJX237615A 2490C IJX237655A 2490C IJX237680A 2490C IJX237681A 2490C IJX237691A 2490C IJX237702A 2490C IJX237734A 2490C IJX237755A 2490C IJX237759A 2490C IJX237763A 2490C IJX237784A 2490C IJX237878A 2490C IJX237925A 2490C IJX237932A 2490C IJX237939A 2490C IJX237948A 2490C IJX237967A 2490C IJX237979A 2490C IJX238012A 2490C IJX238022A 2490C IJX238030A 2490C IJX238045A 2490C IJX238067A 2490C IJX238085A 2490C IJX238091A 2490C IJX239422A 2490C IJX239440A 2490C IJX239452A 2490C IJX239473A 2490C IJX239494A 2490C IJX239505A 2490C IJX239506A 2490C IJX239533A 2490C IJX239534A 2490C IJX239648A 2490C IJX239697A 2490C IJX239705A 2490C IJX239719A 2490C IJX239729A 2490C IJX239764A 2490C IJX239815A 2490C IJX239869A 2490C IJX239914A 2490C IJX239933A 2490C IJX239993A 2490C IJX240010A 2490C IJX240045A 2490C IJX240046A 2490C IJX240051A 2490C IJX240057A 2490C IJX240060A 2490C IJX240063A 2490C IJX240067A 2490C IJX240077A 2490C IJX240094A 2490C IJX240099A 2490C IJX240102A 2490C IJX240106A 2490C IJX240151A 2490C IJX240202A 2490C IJX240235A 2490C IJX240239A 2490C IJX240241A 2490C IJX240242A 2490C IJX240278A 2490C IJX240307A 2490C IJX240313A 2490C IJX240345A 2490C IJX240348A 2490C IJX240353A 2490C IJX240358A 2490C IJX240368A 2490C IJX240369A 2490C IJX240387A 2490C IJX240389A 2490C IJX240607A 2490C IJX241436A 2490C IJX241443A 2490C IJX241444A 2490C IJX241519A 2490C IJX241556A 2490C IJX241626A 2490C IJX241643A 2490C IJX241648A 2490C IJX241687A 2490C IJX241691A 2490C IJX241695A 2490C IJX241726A 2490C IJX241734A 2490C IJX241749A 2490C IJX241754A 2490C IJX241758A 2490C IJX241768A 2490C IJX241769A 2490C IJX241775A 2490C IJX241779A 2490C IJX241825A 2490C IJX241839A 2490C IJX241844A 2490C IJX241846A 2490C IJX241847A 2490C IJX241848A 2490C IJX241972A 2490C IJX241973A 2490C IJX241974A 2490C IJX241987A 2490C IJX241988A 2490C IJX242002A 2490C IJX242007A 2490C IJX242010A 2490C IJX242116A 2490C IJX242119A 2490C IJX242144A 2490C IJX242146A 2490C IJX242168A 2490C IJX242170A 2490C IJX242183A 2490C IJX242188A 2490C IJX242190A 2490C IJX242196A 2490C IJX242203A 2490C IJX242205A 2490C IJX242215A 2490C IJX242259A 2490C IJX242278A 2490C IJX242291A 2490C IJX242302A 2490C IJX242303A 2490C IJX242315A 2490C IJX242318A 2490C IJX242319A 2490C IJX242353A 2490C IJX242358A 2490C IJX242400A 2490C IJX242403A 2490C IJX242431A 2490C IJX242444A 2490C IJX242449A 2490C IJX242474A 2490C IJX242497A 2490C IJX242500A 2490C IJX242538A 2490C IJX242551A 2490C IJX242563A 2490C IJX242588A 2490C IJX242704A 2490C IJX242726A 2490C IJX242748A 2490C IJX242872A 2490C IJX242896A 2490C IJX247605A 2490C IJX247613A 2490C IJX247619A 2490C IJX247620A 2490C IJX247625A 2490C IJX247667A 2490C IJX247678A 2490C IJX247680A 2490C IJX247688A 2490C IJX247697A 2490C IJX247726A 2490C IJX247727A 2490C IJX247730A 2490C IJX247731A 2490C IJX247746A 2490C IJX247766A 2490C IJX247796A 2490C IJX247805A 2490C IJX247843A 2490C IJX247846A 2490C IJX247854A 2490C IJX247886A 2490C IJX247890A 2490C IJX247891A 2490C IJX247907A 2490C IJX247919A 2490C IJX247925A 2490C IJX247926A 2490C IJX247941A 2490C IJX247948A 2490C IJX247953A 2490C IJX247954A 2490C IJX247966A 2490C IJX247968A 2490C IJX247979A 2490C IJX248006A 2490C IJX248012A 2490C IJX248021A 2490C IJX248026A 2490C IJX248032A 2490C IJX248049A 2490C IJX248062A 2490C IJX248066A 2490C IJX248071A 2490C IJX248080A 2490C IJX248085A 2490C IJX248120A 2490C IJX248140A 2490C IJX248156A 2490C IJX248157A 2490C IJX248166A 2490C IJX248196A 2490C IJX248202A 2490C IJX248204A 2490C IJX248218A 2490C IJX248219A 2490C IJX248236A 2490C IJX248279A 2490C IJX248370A 2490C IJX248401A 2490C IJX248504A 2490C IJX248532A 2490C IJX248540A 2490C IJX248643A 2490C IJX248693A 2490C IJX248717A 2490C IJX248777A 2490C IJX248814A 2490C IJX248825A 2490C IJX248828A 2490C IJX248858A 2490C IJX248895A 2490C IJX248899A 2490C IJX248924A 2490C IJX248942A 2490C IJX248967A 2490C IJX248972A 2490C IJX248989A 2490C IJX249023A 2490C IJX249029A 2490C IJX249055A 2490C IJX249084A 2490C IJX249090A 2490C IJX249104A 2490C IJX253628A 2490C IJX253630A 2490C IJX253638A 2490C IJX253642A 2490C IJX253644A 2490C IJX253652A 2490C IJX253656A 2490C IJX253657A 2490C IJX253670A 2490C IJX253681A 2490C IJX253689A 2490C IJX253694A 2490C IJX253805A 2490C IJX253837A 2490C IJX253860A 2490C IJX253870A 2490C IJX253882A 2490C IJX253908A 2490C IJX253950A 2490C IJX253958A 2490C IJX254040A 2490C IJX254042A 2490C IJX254097A 2490C IJX254144A 2490C IJX254179A 2490C IJX254202A 2490C IJX254222A 2490C IJX254228A 2490C IJX254235A 2490C IJX254237A 2490C IJX254255A 2490C IJX254264A 2490C IJX254267A 2490C IJX254293A 2490C IJX254300A 2490C IJX254310A 2490C IJX254324A 2490C IJX254337A 2490C IJX254352A 2490C IJX254379A 2490C IJX254388A 2490C IJX254399A 2490C IJX254403A 2490C IJX254408A 2490C IJX254411A 2490C IJX254418A 2490C IJX254419A 2490C IJX254444A 2490C IJX254449A 2490C IJX254469A 2490C IJX254487A 2490C IJX254537A 2490C IJX254549A 2490C IJX254572A 2490C IJX254574A 2490C IJX254601A 2490C IJX254624A 2490C IJX254640A 2490C IJX254667A 2490C IJX254686A 2490C IJX254697A 2490C IJX254716A 2490C IJX254723A 2490C IJX254763A 2490C IJX254826A 2490C IJX254827A 2490C IJX254828A 2490C IJX254866A 2490C IJX254867A 2490C IJX254868A 2490C IJX254982A 2490C IJX254984A 2490C IJX254992A 2490C IJX254996A 2490C IJX254997A 2490C IJX255026A 2490C IJX255047A 2490C IJX255078A 2490C IJX255099A 2490C IJX255102A 2490C IJX255104A 2490C IJX255177A 2490C IJX255269A 2490C IJX255295A 2490C IJX255350A 2490C IJX255362A 2490C IJX255364A 2490C IJX255382A 2490C IJX255383A 2490C IJX255431A 2490C IJX255437A 2490C IJX255448A 2490C IJX255453A 2490C IJX255454A 2490C IJX255462A 2490C IJX260626A 2490C IJX260631A 2490C IJX260638A 2490C IJX260642A 2490C IJX260647A 2490C IJX260651A 2490C IJX260675A 2490C IJX260676A 2490C IJX260678A 2490C IJX260718A 2490C IJX260726A 2490C IJX260748A 2490C IJX260769A 2490C IJX260771A 2490C IJX260798A 2490C IJX260858A 2490C IJX260872A 2490C IJX260881A 2490C IJX260905A 2490C IJX260906A 2490C IJX260913A 2490C IJX260914A 2490C IJX260924A 2490C IJX260943A 2490C IJX260972A 2490C IJX261045A 2490C IJX261070A 2490C IJX261079A 2490C IJX261081A 2490C IJX261098A 2490C IJX261100A 2490C IJX261101A 2490C IJX261147A 2490C IJX261162A 2490C IJX261170A 2490C IJX261225A 2490C IJX261252A 2490C IJX261271A 2490C IJX261278A 2490C IJX261280A 2490C IJX261299A 2490C IJX261302A 2490C IJX261338A 2490C IJX261361A 2490C IJX261362A 2490C IJX261370A 2490C IJX261428A 2490C IJX261445A 2490C IJX261454A 2490C IJX261456A 2490C IJX261468A 2490C IJX261470A 2490C IJX261499A 2490C IJX261500A 2490C IJX261509A 2490C IJX261518A 2490C IJX261519A 2490C IJX261522A 2490C IJX261526A 2490C IJX261532A 2490C IJX261552A 2490C IJX261561A 2490C IJX261613A 2490C IJX261625A 2490C IJX261631A 2490C IJX261643A 2490C IJX261646A 2490C IJX261651A 2490C IJX261672A 2490C IJX261674A 2490C IJX261680A 2490C IJX261702A 2490C IJX261719A 2490C IJX261742A 2490C IJX261771A 2490C IJX261773A 2490C IJX261791A 2490C IJX261795A 2490C IJX261814A 2490C IJX261815A 2490C IJX261816A 2490C IJX261830A 2490C IJX261865A 2490C IJX261869A 2490C IJX261890A 2490C IJX261936A 2490C IJX261946A 2490C IJX261951A 2490C IJX261958A 2490C IJX261968A 2490C IJX261969A 2490C IJX264663A 2490C IJX264672A 2490C IJX264680A 2490C IJX264696A 2490C IJX264702A 2490C IJX264709A 2490C IJX264734A 2490C IJX264738A 2490C IJX264745A 2490C IJX264751A 2490C IJX264761A 2490C IJX264771A 2490C IJX264782A 2490C IJX264786A 2490C IJX264789A 2490C IJX264818A 2490C IJX264823A 2490C IJX264828A 2490C IJX264837A 2490C IJX264840A 2490C IJX264842A 2490C IJX264853A 2490C IJX264855A 2490C IJX264857A 2490C IJX264861A 2490C IJX264886A 2490C IJX264887A 2490C IJX264897A 2490C IJX264904A 2490C IJX264916A 2490C IJX264919A 2490C IJX264924A 2490C IJX264935A 2490C IJX264956A 2490C IJX264957A 2490C IJX264962A 2490C IJX264964A 2490C IJX264965A 2490C IJX264970A 2490C IJX264972A 2490C IJX264975A 2490C IJX264985A 2490C IJX265002A 2490C IJX265013A 2490C IJX265020A 2490C IJX265021A 2490C IJX265054A 2490C IJX265074A 2490C IJX265085A 2490C IJX265087A 2490C IJX265090A 2490C IJX265091A 2490C IJX265125A 2490C IJX265127A 2490C IJX265134A 2490C IJX265182A 2490C IJX265201A 2490C IJX265205A 2490C IJX265227A 2490C IJX265240A 2490C IJX265258A 2490C IJX265260A 2490C IJX265270A 2490C IJX265280A 2490C IJX265314A 2490C IJX265324A 2490C IJX265331A 2490C IJX265336A 2490C IJX265363A 2490C IJX265381A 2490C IJX265382A 2490C IJX265389A 2490C IJX265398A 2490C IJX265428A 2490C IJX265437A 2490C IJX265447A 2490C IJX265477A 2490C IJX265487A 2490C IJX265519A 2490C IJX265521A 2490C IJX265533A 2490C IJX265541A 2490C IJX265542A 2490C IJX265549A 2490C IJX265595A 2490C IJX265598A 2490C IJX265599A 2490C IJX265623A 2490C IJX265644A 2490C IJX265651A 2490C IJX265653A 2490C IJX265662A 2490C IJX265665A 2490C IJX265687A 2490C IJX265707A 2490C IJX265714A 2490C IJX265730A 2490C IJX265735A 2490C IJX265742A 2490C IJX265746A 2490C IJX265757A 2490C IJX265787A 2490C IJX265816A 2490C IJX265835A 2490C IJX265842A 2490C IJX265921A 2490C IJX265927A 2490C IJX265979A 2490C IJX265987A 2490C IJX265995A 2490C IJX266010A 2490C IJX266030A 2490C IJX266032A 2490C IJX266033A 2490C IJX266043A 2490C IJX266081A 2490C IJX266084A 2490C IJX266110A 2490C IJX266126A 2490C IJX266133A 2490C IJX266136A 2490C IJX266203A 2490C IJX266218A 2490C IJX266265A 2490C IJX266274A 2490C IJX266280A 2490C IJX266286A 2490C IJX266313A 2490C IJX266327A 2490C IJX266355A 2490C IJX266357A 2490C IJX266359A 2490C IJX266377A 2490C IJX266378A 2490C IJX266395A 2490C IJX266401A 2490C IJX266412A 2490C IJX266447A 2490C IJX270507A 2490C IJX273296A 2490C IJX273305A 2490C IJX273324A 2490C IJX273378A 2490C IJX273383A 2490C IJX273489A 2490C IJX273504A 2490C IJX273519A 2490C IJX273538A 2490C IJX273561A 2490C IJX273573A 2490C IJX273591A 2490C IJX273607A 2490C IJX273611A 2490C IJX273657A 2490C IJX273658A 2490C IJX273669A 2490C IJX273685A 2490C IJX273691A 2490C IJX273721A 2490C IJX273727A 2490C IJX273736A 2490C IJX273752A 2490C IJX273754A 2490C IJX273770A 2490C IJX273872A 2490C IJX273910A 2490C IJX273963A 2490C IJX273976A 2490C IJX273988A 2490C IJX273996A 2490C IJX273997A 2490C IJX274019A 2490C IJX274067A 2490C IJX274069A 2490C IJX274073A 2490C IJX274082A 2490C IJX274099A 2490C IJX274127A 2490C IJX274128A 2490C IJX274168A 2490C IJX274169A 2490C IJX274191A 2490C IJX274194A 2490C IJX274201A 2490C IJX274255A 2490C IJX274287A 2490C IJX274305A 2490C IJX274306A 2490C IJX274316A 2490C IJX274324A 2490C IJX274361A 2490C IJX278482A 2490C IJX278486A 2490C IJX278488A 2490C IJX278490A 2490C IJX278492A 2490C IJX278493A 2490C IJX278505A 2490C IJX278510A 2490C IJX278527A 2490C IJX278537A 2490C IJX278566A 2490C IJX278575A 2490C IJX278576A 2490C IJX278587A 2490C IJX278611A 2490C IJX278620A 2490C IJX278621A 2490C IJX278640A 2490C IJX278669A 2490C IJX278671A 2490C IJX278696A 2490C IJX278715A 2490C IJX278716A 2490C IJX278735A 2490C IJX278752A 2490C IJX278773A 2490C IJX278792A 2490C IJX278794A 2490C IJX278841A 2490C IJX278848A 2490C IJX278866A 2490C IJX278908A 2490C IJX278910A 2490C IJX278919A 2490C IJX278928A 2490C IJX278936A 2490C IJX278984A 2490C IJX279004A 2490C IJX279012A 2490C IJX279016A 2490C IJX279020A 2490C IJX279023A 2490C IJX279039A 2490C IJX279126A 2490C IJX279147A 2490C IJX279161A 2490C IJX279632A 2490C IJX279646A 2490C IJX279650A 2490C IJX279691A 2490C IJX279714A 2490C IJX279726A 2490C IJX279730A 2490C IJX279731A 2490C IJX279778A 2490C IJX279792A 2490C IJX279797A 2490C IJX279802A 2490C IJX279804A 2490C IJX279817A 2490C IJX280483A 2490C IJX280492A 2490C IJX280502A 2490C IJX280505A 2490C IJX281885A 2490C IJX281919A 2490C IJX281962A 2490C IJX281973A 2490C IJX281995A 2490C IJX282012A 2490C IJX282026A 2490C IJX282031A 2490C IJX282041A 2490C IJX282047A 2490C IJX282052A 2490C IJX282059A 2490C IJX282094A 2490C IJX282117A 2490C IJX282118A 2490C IJX282121A 2490C IJX282141A 2490C IJX282147A 2490C IJX282154A 2490C IJX282160A 2490C IJX282166A 2490C IJX282173A 2490C IJX282174A 2490C IJX282197A 2490C IJX282202A 2490C IJX282232A 2490C IJX282245A 2490C IJX282246A 2490C IJX282261A 2490C IJX282287A 2490C IJX282291A 2490C IJX282303A 2490C IJX282306A 2490C IJX282318A 2490C IJX282330A 2490C IJX282333A 2490C IJX282343A 2490C IJX282350A 2490C IJX282368A 2490C IJX282373A 2490C IJX282403A 2490C IJX282479A 2490C IJX282501A 2490C IJX282504A 2490C IJX282518A 2490C IJX282525A 2490C IJX282527A 2490C IJX282529A 2490C IJX282532A 2490C IJX282580A 2490C IJX282599A 2490C IJX282610A 2490C IJX282787A 2490C IJX282812A 2490C IJX282821A 2490C IJX282845A 2490C IJX282847A 2490C IJX282851A 2490C IJX282858A 2490C IJX282864A 2490C IJX282868A 2490C IJX282883A 2490C IJX282884A 2490C IJX282885A 2490C IJX282891A 2490C IJX282893A 2490C IJX282909A 2490C IJX282911A 2490C IJX282914A 2490C IJX282927A 2490C IJX282929A 2490C IJX282931A 2490C IJX282936A 2490C IJX282938A 2490C IJX282939A 2490C IJX282948A 2490C IJX282949A 2490C IJX282954A 2490C IJX282957A 2490C IJX282973A 2490C IJX282975A 2490C IJX282986A 2490C IJX282994A 2490C IJX282997A 2490C IJX283008A 2490C IJX283011A 2490C IJX283013A 2490C IJX283016A 2490C IJX283042A 2490C IJX283043A 2490C IJX283054A 2490C IJX283060A 2490C IJX283076A 2490C IJX283101A 2490C IJX283111A 2490C IJX283112A 2490C IJX283116A 2490C IJX283154A 2490C IJX283159A 2490C IJX283168A 2490C IJX283220A 2490C IJX283239A 2490C IJX285813A 2490C IJX285819A 2490C IJX285828A 2490C IJX285836A 2490C IJX285896A 2490C IJX285902A 2490C IJX285905A 2490C IJX285908A 2490C IJX285938A 2490C IJX285954A 2490C IJX285955A 2490C IJX285971A 2490C IJX289737A 2490C IJX289739A 2490C IJX289767A 2490C IJX289769A 2490C IJX289794A 2490C IJX289803A 2490C IJX289845A 2490C IJX289874A 2490C IJX289906A 2490C IJX289907A 2490C IJX289912A 2490C IJX289927A 2490C IJX289955A 2490C IJX289957A 2490C IJX289970A 2490C IJX289990A 2490C IJX290007A 2490C IJX290061A 2490C IJX290062A 2490C IJX290075A 2490C IJX290089A 2490C IJX290100A 2490C IJX290103A 2490C IJX290127A 2490C IJX290153A 2490C IJX290160A 2490C IJX290218A 2490C IJX290730A 2490C IJX290765A 2490C IJX290801A 2490C IJX290832A 2490C IJX290844A 2490C IJX290846A 2490C IJX290861A 2490C IJX290866A 2490C IJX290874A 2490C IJX290880A 2490C IJX290893A 2490C IJX290898A 2490C IJX290934A 2490C IJX290964A 2490C IJX290977A 2490C IJX290986A 2490C IJX291070A 2490C IJX291071A 2490C IJX291095A 2490C IJX291129A 2490C IJX291139A 2490C IJX291145A 2490C IJX291152A 2490C IJX291176A 2490C IJX291230A 2490C IJX291252A 2490C IJX291278A 2490C IJX291337A 2490C IJX291392A 2490C IJX291393A 2490C IJX291427A 2490C IJX291428A 2490C IJX291432A 2490C IJX291439A 2490C IJX291448A 2490C IJX291462A 2490C IJX291514A 2490C IJX291541A 2490C IJX291555A 2490C IJX291566A 2490C IJX291569A 2490C IJX291576A 2490C IJX291603A 2490C IJX291647A 2490C IJX291663A 2490C IJX291682A 2490C IJX291686A 2490C IJX291701A 2490C IJX291703A 2490C IJX291716A 2490C IJX291725A 2490C IJX291752A 2490C IJX291762A 2490C IJX291763A 2490C IJX291805A 2490C IJX291806A 2490C IJX291808A 2490C IJX291810A 2490C IJX291813A 2490C IJX291826A 2490C IJX291849A 2490C IJX291857A 2490C IJX291859A 2490C IJX291880A 2490C IJX291902A 2490C IJX291912A 2490C IJX291928A 2490C IJX291969A 2490C IJX291971A 2490C IJX291975A 2490C IJX291982A 2490C IJX291996A 2490C IJX292003A 2490C IJX292010A 2490C IJX292021A 2490C IJX292050A 2490C IJX292084A 2490C IJX292088A 2490C IJX292197A 2490C IJX292203A 2490C IJX292214A 2490C IJX292232A 2490C IJX292247A 2490C IJX292257A 2490C IJX292276A 2490C IJX292304A 2490C IJX292313A 2490C IJX292325A 2490C IJX292374A 2490C IJX292384A 2490C IJX292393A 2490C IJX292394A 2490C IJX292423A 2490C IJX292426A 2490C IJX292455A 2490C IJX292462A 2490C IJX294520A 2490C IJX294528A 2490C IJX294549A 2490C IJX294555A 2490C IJX294567A 2490C IJX294575A 2490C IJX294613A 2490C IJX294615A 2490C IJX294632A 2490C IJX294639A 2490C IJX294640A 2490C IJX294642A 2490C IJX294665A 2490C IJX294669A 2490C IJX294733A 2490C IJX294756A 2490C IJX294775A 2490C IJX294777A 2490C IJX294779A 2490C IJX294796A 2490C IJX294811A 2490C IJX294818A 2490C IJX294830A 2490C IJX294846A 2490C IJX294910A 2490C IJX294941A 2490C IJX294966A 2490C IJX294984A 2490C IJX295004A 2490C IJX295006A 2490C IJX295030A 2490C IJX295086A 2490C IJX295119A 2490C IJX295126A 2490C IJX295162A 2490C IJX295166A 2490C IJX295182A 2490C IJX295183A 2490C IJX295184A 2490C IJX295198A 2490C IJX295206A 2490C IJX295253A 2490C IJX295259A 2490C IJX295261A 2490C IJX295262A 2490C IJX296807A 2490C IJX296809A 2490C IJX296862A 2490C IJX296864A 2490C IJX296879A 2490C IJX296944A 2490C IJX296961A 2490C IJX296963A 2490C IJX296968A 2490C IJX296975A 2490C IJX296995A 2490C IJX297040A 2490C IJX297048A 2490C IJX297053A 2490C IJX297057A 2490C IJX297088A 2490C IJX297094A 2490C IJX297109A 2490C IJX297110A 2490C IJX297115A 2490C IJX297144A 2490C IJX297155A 2490C IJX297156A 2490C IJX297168A 2490C IJX297236A 2490C IJX297250A 2490C IJX297741A 2490C IJX297756A 2490C IJX301632A 2490C IJX301635A 2490C IJX301652A 2490C IJX301671A 2490C IJX301687A 2490C IJX301728A 2490C IJX301744A 2490C IJX301745A 2490C IJX301752A 2490C IJX301757A 2490C IJX301763A 2490C IJX301778A 2490C IJX301821A 2490C IJX301831A 2490C IJX301849A 2490C IJX301877A 2490C IJX301881A 2490C IJX301884A 2490C IJX301887A 2490C IJX301943A 2490C IJX301954A 2490C IJX301955A 2490C IJX301962A 2490C IJX301969A 2490C IJX301971A 2490C IJX301979A 2490C IJX302031A 2490C IJX302035A 2490C IJX302038A 2490C IJX302046A 2490C IJX302057A 2490C IJX302132A 2490C IJX302600A 2490C IJX302602A 2490C IJX302612A 2490C IJX302613A 2490C IJX302624A 2490C IJX302632A 2490C IJX302639A 2490C IJX302648A 2490C IJX302651A 2490C IJX302656A 2490C IJX302660A 2490C IJX302685A 2490C IJX302694A 2490C IJX302706A 2490C IJX302708A 2490C IJX302732A 2490C IJX302739A 2490C IJX302741A 2490C IJX302742A 2490C IJX302745A 2490C IJX302747A 2490C IJX302755A 2490C IJX302790A 2490C IJX302831A 2490C IJX302852A 2490C IJX302867A 2490C IJX302892A 2490C IJX302895A 2490C IJX302904A 2490C IJX302916A 2490C IJX302941A 2490C IJX302953A 2490C IJX302961A 2490C IJX302975A 2490C IJX302998A 2490C IJX303003A 2490C IJX303005A 2490C IJX303031A 2490C IJX303046A 2490C IJX303050A 2490C IJX307786A 2490C IJX307796A 2490C IJX307841A 2490C IJX307863A 2490C IJX307910A 2490C IJX307933A 2490C IJX307945A 2490C IJX307954A 2490C IJX307970A 2490C IJX307975A 2490C IJX307976A 2490C IJX308008A 2490C IJX308026A 2490C IJX308032A 2490C IJX308146A 2490C IJX308216A 2490C IJX308219A 2490C IJX308265A 2490C IJX308269A 2490C IJX308297A 2490C IJX308310A 2490C IJX308421A 2490C IJX308424A 2490C IJX308436A 2490C IJX308485A 2490C IJX308501A 2490C IJX308529A 2490C IJX308538A 2490C IJX308553A 2490C IJX308559A 2490C IJX308561A 2490C IJX308612A 2490C IJX308664A 2490C IJX308672A 2490C IJX308678A 2490C IJX308680A 2490C IJX308695A 2490C IJX308698A 2490C IJX308699A 2490C IJX308743A 2490C IJX308749A 2490C IJX308752A 2490C IJX308763A 2490C IJX308797A 2490C IJX308804A 2490C IJX308807A 2490C IJX308812A 2490C IJX308822A 2490C IJX308826A 2490C IJX308834A 2490C IJX308837A 2490C IJX308858A 2490C IJX308859A 2490C IJX308888A 2490C IJX308890A 2490C IJX308917A 2490C IJX308937A 2490C IJX308944A 2490C IJX308945A 2490C IJX308952A 2490C IJX308963A 2490C IJX308968A 2490C IJX308978A 2490C IJX308984A 2490C IJX308987A 2490C IJX308990A 2490C IJX308991A 2490C IJX309001A 2490C IJX309004A 2490C IJX309007A 2490C IJX309014A 2490C IJX309019A 2490C IJX309023A 2490C IJX309027A 2490C IJX309030A 2490C IJX309055A 2490C IJX309069A 2490C IJX309072A 2490C IJX309086A 2490C IJX309087A 2490C IJX309094A 2490C IJX309103A 2490C IJX309107A 2490C IJX309117A 2490C IJX309120A 2490C IJX309122A 2490C IJX309127A 2490C IJX309134A 2490C IJX309140A 2490C IJX309150A 2490C IJX309164A 2490C IJX309173A 2490C IJX309179A 2490C IJX309180A 2490C IJX309181A 2490C IJX309182A 2490C IJX309186A 2490C IJX309191A 2490C IJX309196A 2490C IJX309201A 2490C IJX309213A 2490C IJX309219A 2490C IJX309220A 2490C IJX309221A 2490C IJX309225A 2490C IJX309227A 2490C IJX309229A 2490C IJX309243A 2490C IJX309248A 2490C IJX309251A 2490C IJX309260A 2490C IJX309267A 2490C IJX309274A 2490C IJX309275A 2490C IJX318230A 2490C IJX318236A 2490C IJX318237A 2490C IJX318255A 2490C IJX318276A 2490C IJX318295A 2490C IJX318297A 2490C IJX318312A 2490C IJX318317A 2490C IJX318326A 2490C IJX318398A 2490C IJX318399A 2490C IJX318439A 2490C IJX318441A 2490C IJX318455A 2490C IJX318463A 2490C IJX318468A 2490C IJX318477A 2490C IJX318479A 2490C IJX318485A 2490C IJX318493A 2490C IJX318499A 2490C IJX318525A 2490C IJX318532A 2490C IJX318534A 2490C IJX318540A 2490C IJX318548A 2490C IJX318550A 2490C IJX318562A 2490C IJX318582A 2490C IJX318592A 2490C IJX318598A 2490C IJX318605A 2490C IJX318637A 2490C IJX318654A 2490C IJX318665A 2490C IJX318680A 2490C IJX318708A 2490C IJX318709A 2490C IJX318723A 2490C IJX318748A 2490C IJX318751A 2490C IJX318760A 2490C IJX318786A 2490C IJX318798A 2490C IJX318827A 2490C IJX318833A 2490C IJX318850A 2490C IJX318973A 2490C IJX318977A 2490C IJX318982A 2490C IJX318983A 2490C IJX318996A 2490C IJX319022A 2490C IJX319038A 2490C IJX319058A 2490C IJX319103A 2490C IJX319107A 2490C IJX319110A 2490C IJX319124A 2490C IJX319128A 2490C IJX319142A 2490C IJX319147A 2490C IJX319153A 2490C IJX319154A 2490C IJX319182A 2490C IJX319202A 2490C IJX319212A 2490C IJX319232A 2490C IJX319236A 2490C IJX319273A 2490C IJX319329A 2490C IJX319345A 2490C IJX319347A 2490C IJX319369A 2490C IJX319374A 2490C IJX319384A 2490C IJX319621A 2490C IJX319626A 2490C IJX319628A 2490C IJX319660A 2490C IJX319662A 2490C IJX319681A 2490C IJX319709A 2490C IJX319711A 2490C IJX319714A 2490C IJX319718A 2490C IJX319728A 2490C IJX319742A 2490C IJX319745A 2490C IJX319749A 2490C IJX319751A 2490C IJX319762A 2490C IJX319764A 2490C IJX319773A 2490C IJX319776A 2490C IJX319779A 2490C IJX319791A 2490C IJX319799A 2490C IJX319810A 2490C IJX319814A 2490C IJX319819A 2490C IJX319834A 2490C IJX319836A 2490C IJX319846A 2490C IJX319849A 2490C IJX319862A 2490C IJX319866A 2490C IJX319879A 2490C IJX319903A 2490C IJX319948A 2490C IJX319951A 2490C IJX319982A 2490C IJX319987A 2490C IJX319989A 2490C IJX320022A 2490C IJX320033A 2490C IJX320060A 2490C IJX320068A 2490C IJX321917A 2490C IJX321922A 2490C IJX321934A 2490C IJX321956A 2490C IJX321969A 2490C IJX321974A 2490C IJX321975A 2490C IJX321981A 2490C IJX321983A 2490C IJX321997A 2490C IJX322003A 2490C IJX322034A 2490C IJX322041A 2490C IJX322048A 2490C IJX322091A 2490C IJX322095A 2490C IJX322097A 2490C IJX322108A 2490C IJX322117A 2490C IJX322132A 2490C IJX322141A 2490C IJX322147A 2490C IJX322215A 2490C IJX322218A 2490C IJX322232A 2490C IJX322236A 2490C IJX322238A 2490C IJX322239A 2490C IJX322244A 2490C IJX322247A 2490C IJX322248A 2490C IJX322259A 2490C IJX322275A 2490C IJX322280A 2490C IJX322312A 2490C IJX322324A 2490C IJX322340A 2490C IJX322355A 2490C IJX322361A 2490C IJX322366A 2490C IJX322373A 2490C IJX322410A 2490C IJX322411A 2490C IJX322412A 2490C IJX325477A 2490C IJX325490A 2490C IJX325504A 2490C IJX325542A 2490C IJX325543A 2490C IJX325557A 2490C IJX325571A 2490C IJX325577A 2490C IJX325578A 2490C IJX325580A 2490C IJX325583A 2490C IJX325586A 2490C IJX325600A 2490C IJX325620A 2490C IJX325627A 2490C IJX325642A 2490C IJX325644A 2490C IJX325707A 2490C IJX325725A 2490C IJX325734A 2490C IJX325771A 2490C IJX325785A 2490C IJX325788A 2490C IJX325809A 2490C IJX325842A 2490C IJX325844A 2490C IJX325845A 2490C IJX325884A 2490C IJX325894A 2490C IJX325909A 2490C IJX325926A 2490C IJX325927A 2490C IJX325963A 2490C IJX325983A 2490C IJX325985A 2490C IJX325986A 2490C IJX325987A 2490C IJX326010A 2490C IJX326036A 2490C IJX326083A 2490C IJX326119A 2490C IJX326136A 2490C IJX326145A 2490C IJX326150A 2490C IJX326173A 2490C IJX326175A 2490C IJX326193A 2490C IJX326273A 2490C IJX326286A 2490C IJX326296A 2490C IJX326308A 2490C IJX326320A 2490C IJX326322A 2490C IJX326327A 2490C IJX326331A 2490C IJX326333A 2490C IJX326334A 2490C IJX326346A 2490C IJX326359A 2490C IJX326360A 2490C IJX326361A 2490C IJX326362A 2490C IJX326365A 2490C IJX326369A 2490C IJX326391A 2490C IJX326395A 2490C IJX326399A 2490C IJX326410A 2490C IJX326441A 2490C IJX326501A 2490C IJX326532A 2490C IJX326541A 2490C IJX326574A 2490C IJX326607A 2490C IJX326610A 2490C IJX326616A 2490C IJX326623A 2490C IJX326627A 2490C IJX326637A 2490C IJX326643A 2490C IJX326663A 2490C IJX326674A 2490C IJX326680A 2490C IJX326687A 2490C IJX326717A 2490C IJX326727A 2490C IJX326730A 2490C IJX326750A 2490C IJX326772A 2490C IJX326862A 2490C IJX326895A 2490C IJX326900A 2490C IJX326904A 2490C IJX326919A 2490C IJX326936A 2490C IJX326948A 2490C IJX326995A 2490C IJX326997A 2490C IJX327005A 2490C IJX327006A 2490C IJX327034A 2490C IJX327044A 2490C IJX327063A 2490C IJX327081A 2490C IJX327086A 2490C IJX327104A 2490C IJX327129A 2490C IJX327140A 2490C IJX327145A 2490C IJX327146A 2490C IJX327176A 2490C IJX327183A 2490C IJX327188A 2490C IJX327192A 2490C IJX327195A 2490C IJX327197A 2490C IJX327202A 2490C IJX327206A 2490C IJX327214A 2490C IJX327218A 2490C IJX327222A 2490C IJX327234A 2490C IJX327236A 2490C IJX327239A 2490C IJX327241A 2490C IJX327251A 2490C IJX327283A 2490C IJX327294A 2490C IJX327309A 2490C IJX327324A 2490C IJX327441A 2490C IJX327443A 2490C IJX328491A 2490C IJX328497A 2490C IJX328503A 2490C IJX328533A 2490C IJX328535A 2490C IJX328536A 2490C IJX328541A 2490C IJX328544A 2490C IJX328561A 2490C IJX328565A 2490C IJX328571A 2490C IJX328588A 2490C IJX328589A 2490C IJX328596A 2490C IJX328602A 2490C IJX328606A 2490C IJX328612A 2490C IJX328625A 2490C IJX328627A 2490C IJX328635A 2490C IJX328636A 2490C IJX328641A 2490C IJX328648A 2490C IJX328650A 2490C IJX328657A 2490C IJX328671A 2490C IJX328678A 2490C IJX328681A 2490C IJX328691A 2490C IJX328692A 2490C IJX328698A 2490C IJX328699A 2490C IJX328701A 2490C IJX328714A 2490C IJX328715A 2490C IJX328728A 2490C IJX328738A 2490C IJX328741A 2490C IJX328742A 2490C IJX328750A 2490C IJX328751A 2490C IJX328770A 2490C IJX328774A 2490C IJX328775A 2490C IJX328778A 2490C IJX328782A 2490C IJX328789A 2490C IJX328800A 2490C IJX328803A 2490C IJX328823A 2490C IJX328835A 2490C IJX328842A 2490C IJX328843A 2490C IJX328844A 2490C IJX328845A 2490C IJX328850A 2490C IJX328851A 2490C IJX328869A 2490C IJX328874A 2490C IJX328876A 2490C IJX328881A 2490C IJX328885A 2490C IJX328892A 2490C IJX328894A 2490C IJX328898A 2490C IJX328906A 2490C IJX328908A 2490C IJX328922A 2490C IJX328931A 2490C IJX328935A 2490C IJX328945A 2490C IJX328959A 2490C IJX328961A 2490C IJX328974A 2490C IJX328975A 2490C IJX328977A 2490C IJX330799A 2490C IJX330828A 2490C IJX335261A 2490C IJX335263A 2490C IJX335283A 2490C IJX335304A 2490C IJX335312A 2490C IJX335313A 2490C IJX335317A 2490C IJX335319A 2490C IJX335330A 2490C IJX335336A 2490C IJX335346A 2490C IJX335350A 2490C IJX335360A 2490C IJX335381A 2490C IJX335387A 2490C IJX335395A 2490C IJX335409A 2490C IJX335436A 2490C IJX335478A 2490C IJX335549A 2490C IJX335552A 2490C IJX335553A 2490C IJX335563A 2490C IJX335577A 2490C IJX335579A 2490C IJX335596A 2490C IJX335667A 2490C IJX335680A 2490C IJX335685A 2490C IJX335701A 2490C IJX335704A 2490C IJX335707A 2490C IJX335710A 2490C IJX335712A 2490C IJX335722A 2490C IJX335723A 2490C IJX335737A 2490C IJX335740A 2490C IJX335743A 2490C IJX335749A 2490C IJX335754A 2490C IJX335756A 2490C IJX335758A 2490C IJX335760A 2490C IJX335797A 2490C IJX335830A 2490C IJX335919A 2490C IJX335934A 2490C IJX335944A 2490C IJX335958A 2490C IJX335971A 2490C IJX335978A 2490C IJX335992A 2490C IJX335994A 2490C IJX336004A 2490C IJX336007A 2490C IJX336033A 2490C IJX336035A 2490C IJX336070A 2490C IJX336083A 2490C IJX336104A 2490C IJX336158A 2490C IJX336173A 2490C IJX336176A 2490C IJX336180A 2490C IJX336193A 2490C IJX336195A 2490C IJX336197A 2490C IJX336204A 2490C IJX336212A 2490C IJX336236A 2490C IJX336241A 2490C IJX336244A 2490C IJX336251A 2490C IJX336273A 2490C IJX336288A 2490C IJX336306A 2490C IJX336308A 2490C IJX336320A 2490C IJX336346A 2490C IJX336378A 2490C IJX336431A 2490C IJX336433A 2490C IJX336452A 2490C IJX336504A 2490C IJX336517A 2490C IJX336518A 2490C IJX336533A 2490C IJX336539A 2490C IJX336540A 2490C IJX336544A 2490C IJX336564A 2490C IJX336566A 2490C IJX336610A 2490C IJX336622A 2490C IJX336624A 2490C IJX336650A 2490C IJX336657A 2490C IJX336658A 2490C IJX336667A 2490C IJX336675A 2490C IJX336706A 2490C IJX336730A 2490C IJX336773A 2490C IJX336779A 2490C IJX336784A 2490C IJX336803A 2490C IJX336815A 2490C IJX336820A 2490C IJX336833A 2490C IJX336863A 2490C IJX336873A 2490C IJX336899A 2490C IJX336903A 2490C IJX336914A 2490C IJX336917A 2490C IJX336943A 2490C IJX336955A 2490C IJX336960A 2490C IJX336964A 2490C IJX336967A 2490C IJX336968A 2490C IJX336971A 2490C IJX336972A 2490C IJX336976A 2490C IJX336978A 2490C IJX336992A 2490C IJX336995A 2490C IJX337028A 2490C IJX337033A 2490C IJX337058A 2490C IJX337064A 2490C IJX337068A 2490C IJX337070A 2490C IJX337085A 2490C IJX337088A 2490C IJX337123A 2490C IJX337128A 2490C IJX337137A 2490C IJX337151A 2490C IJX337182A 2490C IJX337215A 2490C IJX337227A 2490C IJX337229A 2490C IJX337233A 2490C IJX337234A 2490C IJX337250A 2490C IJX342764A 2490C IJX342771A 2490C IJX342800A 2490C IJX342815A 2490C IJX342817A 2490C IJX342832A 2490C IJX342857A 2490C IJX342887A 2490C IJX342901A 2490C IJX342903A 2490C IJX342976A 2490C IJX342987A 2490C IJX342996A 2490C IJX342998A 2490C IJX343001A 2490C IJX343005A 2490C IJX343026A 2490C IJX343036A 2490C IJX343045A 2490C IJX343066A 2490C IJX343067A 2490C IJX343071A 2490C IJX343078A 2490C IJX343091A 2490C IJX343105A 2490C IJX343137A 2490C IJX343141A 2490C IJX343144A 2490C IJX343160A 2490C IJX343161A 2490C IJX343162A 2490C IJX343172A 2490C IJX343242A 2490C IJX343304A 2490C IJX343357A 2490C IJX343388A 2490C IJX343397A 2490C IJX343426A 2490C IJX343437A 2490C IJX343443A 2490C IJX343447A 2490C IJX343495A 2490C IJX343536A 2490C IJX343564A 2490C IJX343597A 2490C IJX343630A 2490C IJX343689A 2490C IJX343732A 2490C IJX343750A 2490C IJX343769A 2490C IJX343772A 2490C IJX343778A 2490C IJX343791A 2490C IJX343792A 2490C IJX343793A 2490C IJX343860A 2490C IJX343863A 2490C IJX343887A 2490C IJX343889A 2490C IJX343915A 2490C IJX343918A 2490C IJX343953A 2490C IJX343964A 2490C IJX344005A 2490C IJX344008A 2490C IJX344009A 2490C IJX344010A 2490C IJX344014A 2490C IJX344027A 2490C IJX344031A 2490C IJX344061A 2490C IJX344074A 2490C IJX344078A 2490C IJX344091A 2490C IJX344143A 2490C IJX344176A 2490C IJX344184A 2490C IJX347882A 2490C IJX347897A 2490C IJX347902A 2490C IJX347904A 2490C IJX347910A 2490C IJX347918A 2490C IJX347924A 2490C IJX347977A 2490C IJX347978A 2490C IJX347986A 2490C IJX348016A 2490C IJX348017A 2490C IJX348020A 2490C IJX348022A 2490C IJX348027A 2490C IJX348034A 2490C IJX348039A 2490C IJX348040A 2490C IJX348049A 2490C IJX348076A 2490C IJX348102A 2490C IJX348107A 2490C IJX349889A 2490C IJX349893A 2490C IJX349976A 2490C IJX349992A 2490C IJX349994A 2490C IJX349995A 2490C IJX350005A 2490C IJX350010A 2490C IJX350012A 2490C IJX350051A 2490C IJX350076A 2490C IJX350137A 2490C IJX350139A 2490C IJX350150A 2490C IJX350175A 2490C IJX350192A 2490C IJX350260A 2490C IJX350282A 2490C IJX350297A 2490C IJX350314A 2490C IJX350316A 2490C IJX350327A 2490C IJX350333A 2490C IJX350337A 2490C IJX350339A 2490C IJX350340A 2490C IJX350352A 2490C IJX350706A 2490C IJX350707A 2490C IJX350733A 2490C IJX350777A 2490C IJX350821A 2490C IJX350827A 2490C IJX350828A 2490C IJX350852A 2490C IJX350886A 2490C IJX350891A 2490C IJX350919A 2490C IJX350924A 2490C IJX350926A 2490C IJX350967A 2490C IJX350979A 2490C IJX350982A 2490C IJX350999A 2490C IJX351006A 2490C IJX351014A 2490C IJX351016A 2490C IJX351017A 2490C IJX351037A 2490C IJX351053A 2490C IJX351062A 2490C IJX351105A 2490C IJX351135A 2490C IJX351137A 2490C IJX351149A 2490C IJX351203A 2490C IJX351251A 2490C IJX351261A 2490C IJX351269A 2490C IJX351270A 2490C IJX351376A 2490C IJX351396A 2490C IJX351400A 2490C IJX351415A 2490C IJX351419A 2490C IJX351466A 2490C IJX351482A 2490C IJX351496A 2490C IJX351501A 2490C IJX351505A 2490C IJX351507A 2490C IJX351518A 2490C IJX351519A 2490C IJX351522A 2490C IJX351574A 2490C IJX351613A 2490C IJX351629A 2490C IJX351649A 2490C IJX351655A 2490C IJX351683A 2490C IJX351703A 2490C IJX351722A 2490C IJX351734A 2490C IJX351738A 2490C IJX351739A 2490C IJX351740A 2490C IJX351743A 2490C IJX351752A 2490C IJX351753A 2490C IJX351760A 2490C IJX351779A 2490C IJX356897A 2490C IJX356903A 2490C IJX356919A 2490C IJX356929A 2490C IJX356972A 2490C IJX357024A 2490C IJX357077A 2490C IJX357098A 2490C IJX357104A 2490C IJX357114A 2490C IJX357142A 2490C IJX357152A 2490C IJX357173A 2490C IJX357201A 2490C IJX357216A 2490C IJX357237A 2490C IJX357258A 2490C IJX357263A 2490C IJX357274A 2490C IJX357285A 2490C IJX357291A 2490C IJX357319A 2490C IJX357320A 2490C IJX357333A 2490C IJX357342A 2490C IJX357353A 2490C IJX357367A 2490C IJX357369A 2490C IJX357374A 2490C IJX357375A 2490C IJX357381A 2490C IJX357390A 2490C IJX357424A 2490C IJX357438A 2490C IJX357442A 2490C IJX357449A 2490C IJX357456A 2490C IJX357463A 2490C IJX357465A 2490C IJX357469A 2490C IJX357531A 2490C IJX357539A 2490C IJX357542A 2490C IJX357601A 2490C IJX357605A 2490C IJX357609A 2490C IJX357640A 2490C IJX357651A 2490C IJX357675A 2490C IJX357679A 2490C IJX357681A 2490C IJX357682A 2490C IJX357687A 2490C IJX357689A 2490C IJX358500A 2490C IJX358510A 2490C IJX358511A 2490C IJX358512A 2490C IJX358513A 2490C IJX358516A 2490C IJX358519A 2490C IJX358523A 2490C IJX358534A 2490C IJX358535A 2490C IJX358543A 2490C IJX358548A 2490C IJX358556A 2490C IJX358564A 2490C IJX358565A 2490C IJX358568A 2490C IJX358572A 2490C IJX358582A 2490C IJX358586A 2490C IJX358587A 2490C IJX358588A 2490C IJX358600A 2490C IJX358646A 2490C IJX358650A 2490C IJX358651A 2490C IJX358652A 2490C IJX358655A 2490C IJX358656A 2490C IJX358657A 2490C IJX358658A 2490C IJX358660A 2490C IJX358717A 2490C IJX358723A 2490C IJX358732A 2490C IJX358736A 2490C IJX358741A 2490C IJX358749A 2490C IJX358774A 2490C IJX358785A 2490C IJX358792A 2490C IJX358797A 2490C IJX358828A 2490C IJX358843A 2490C IJX358909A 2490C IJX358911A 2490C IJX358916A 2490C IJX358924A 2490C IJX358928A 2490C IJX358953A 2490C IJX358989A 2490C IJX359005A 2490C IJX359009A 2490C IJX359019A 2490C IJX359021A 2490C IJX359026A 2490C IJX359031A 2490C IJX359042A 2490C IJX359050A 2490C IJX359079A 2490C IJX359085A 2490C IJX359091A 2490C IJX359101A 2490C IJX359108A 2490C IJX359110A 2490C IJX359112A 2490C IJX359186A 2490C IJX359206A 2490C IJX359233A 2490C IJX359234A 2490C IJX359262A 2490C IJX359269A 2490C IJX359285A 2490C IJX359307A 2490C IJX359340A 2490C IJX359415A 2490C IJX359426A 2490C IJX359435A 2490C IJX359451A 2490C IJX359481A 2490C IJX359483A 2490C IJX359508A 2490C IJX359517A 2490C IJX359518A 2490C IJX359527A 2490C IJX359543A 2490C IJX359578A 2490C IJX359587A 2490C IJX359597A 2490C IJX359607A 2490C IJX359609A 2490C IJX359615A 2490C IJX359673A 2490C IJX359702A 2490C IJX359708A 2490C IJX359711A 2490C IJX359732A 2490C IJX359946A 2490C IJX359954A 2490C IJX359956A 2490C IJX359959A 2490C IJX366327A 2490C IJX366354A 2490C IJX366364A 2490C IJX367026A 2490C IJX367027A 2490C IJX367042A 2490C IJX367045A 2490C IJX367046A 2490C IJX367073A 2490C IJX367081A 2490C IJX367098A 2490C IJX367100A 2490C IJX367101A 2490C IJX367110A 2490C IJX367125A 2490C IJX367126A 2490C IJX367151A 2490C IJX367152A 2490C IJX367154A 2490C IJX367160A 2490C IJX367170A 2490C IJX367185A 2490C IJX367203A 2490C IJX367205A 2490C IJX367208A 2490C IJX367210A 2490C IJX367217A 2490C IJX367243A 2490C IJX367248A 2490C IJX367253A 2490C IJX367257A 2490C IJX367268A 2490C IJX367277A 2490C IJX367279A 2490C IJX367294A 2490C IJX367301A 2490C IJX367307A 2490C IJX367311A 2490C IJX367312A 2490C IJX367313A 2490C IJX367334A 2490C IJX367337A 2490C IJX367354A 2490C IJX367380A 2490C IJX367391A 2490C IJX367397A 2490C IJX367398A 2490C IJX367401A 2490C IJX368848A 2490C IJX368852A 2490C IJX368876A 2490C IJX368885A 2490C IJX368915A 2490C IJX368974A 2490C IJX368988A 2490C IJX368995A 2490C IJX368999A 2490C IJX369001A 2490C IJX369003A 2490C IJX369034A 2490C IJX369035A 2490C IJX369053A 2490C IJX369056A 2490C IJX369071A 2490C IJX369082A 2490C IJX369103A 2490C IJX369111A 2490C IJX369136A 2490C IJX369151A 2490C IJX369175A 2490C IJX369183A 2490C IJX369184A 2490C IJX369187A 2490C IJX369196A 2490C IJX369198A 2490C IJX369211A 2490C IJX369219A 2490C IJX369241A 2490C IJX369251A 2490C IJX369289A 2490C IJX369316A 2490C IJX369318A 2490C IJX369323A 2490C IJX369327A 2490C IJX369349A 2490C IJX369415A 2490C IJX369421A 2490C IJX369463A 2490C IJX369484A 2490C IJX369495A 2490C IJX369513A 2490C IJX369515A 2490C IJX369536A 2490C IJX369537A 2490C IJX369542A 2490C IJX369556A 2490C IJX369558A 2490C IJX369579A 2490C IJX369580A 2490C IJX369596A 2490C IJX369616A 2490C IJX369626A 2490C IJX369637A 2490C IJX369640A 2490C IJX369678A 2490C IJX369682A 2490C IJX369694A 2490C IJX369710A 2490C IJX369726A 2490C IJX369727A 2490C IJX369738A 2490C IJX369751A 2490C IJX369769A 2490C IJX369772A 2490C IJX369779A 2490C IJX369785A 2490C IJX369792A 2490C IJX369800A 2490C IJX369802A 2490C IJX369808A 2490C IJX369814A 2490C IJX369815A 2490C IJX369818A 2490C IJX369832A 2490C IJX369838A 2490C IJX369839A 2490C IJX369873A 2490C IJX369912A 2490C IJX369937A 2490C IJX369954A 2490C IJX369955A 2490C IJX369956A 2490C IJX369963A 2490C IJX369967A 2490C IJX370027A 2490C IJX370029A 2490C IJX370037A 2490C IJX370070A 2490C IJX370094A 2490C IJX370115A 2490C IJX370129A 2490C IJX370130A 2490C IJX370137A 2490C IJX370140A 2490C IJX373397A 2490C IJX373399A 2490C IJX373447A 2490C IJX373472A 2490C IJX373485A 2490C IJX373488A 2490C IJX373494A 2490C IJX373495A 2490C IJX373496A 2490C IJX373500A 2490C IJX373503A 2490C IJX373526A 2490C IJX373549A 2490C IJX373552A 2490C IJX373553A 2490C IJX373571A 2490C IJX373578A 2490C IJX373604A 2490C IJX373623A 2490C IJX373630A 2490C IJX373634A 2490C IJX373637A 2490C IJX373655A 2490C IJX373659A 2490C IJX373662A 2490C IJX373670A 2490C IJX373671A 2490C IJX373687A 2490C IJX373694A 2490C IJX373702A 2490C IJX373725A 2490C IJX376161A 2490C IJX376173A 2490C IJX376180A 2490C IJX376188A 2490C IJX376190A 2490C IJX376194A 2490C IJX376195A 2490C IJX376199A 2490C IJX376201A 2490C IJX376203A 2490C IJX376213A 2490C IJX376219A 2490C IJX376228A 2490C IJX376242A 2490C IJX376250A 2490C IJX376273A 2490C IJX376289A 2490C IJX376294A 2490C IJX376297A 2490C IJX376301A 2490C IJX376303A 2490C IJX376307A 2490C IJX376308A 2490C IJX376310A 2490C IJX376324A 2490C IJX376328A 2490C IJX376332A 2490C IJX376338A 2490C IJX376345A 2490C IJX376346A 2490C IJX376352A 2490C IJX376353A 2490C IJX376399A 2490C IJX376413A 2490C IJX376418A 2490C IJX376423A 2490C IJX376426A 2490C IJX376437A 2490C IJX376443A 2490C IJX376446A 2490C IJX376453A 2490C IJX376456A 2490C IJX376464A 2490C IJX376466A 2490C IJX376494A 2490C IJX376512A 2490C IJX376519A 2490C IJX376530A 2490C IJX376533A 2490C IJX376557A 2490C IJX376562A 2490C IJX376565A 2490C IJX376571A 2490C IJX376575A 2490C IJX376577A 2490C IJX376630A 2490C IJX376634A 2490C IJX376636A 2490C IJX376641A 2490C IJX376647A 2490C IJX376650A 2490C IJX376652A 2490C IJX376657A 2490C IJX376658A 2490C IJX376681A 2490C IJX376682A 2490C IJX376691A 2490C IJX376722A 2490C IJX376727A 2490C IJX376764A 2490C IJX376765A 2490C IJX376786A 2490C IJX376840A 2490C IJX376847A 2490C IJX376856A 2490C IJX376881A 2490C IJX376884A 2490C IJX376890A 2490C IJX376914A 2490C IJX376915A 2490C IJX376927A 2490C IJX376934A 2490C IJX376965A 2490C IJX377036A 2490C IJX377038A 2490C IJX377046A 2490C IJX377057A 2490C IJX377069A 2490C IJX377074A 2490C IJX377078A 2490C IJX377079A 2490C IJX377086A 2490C IJX377124A 2490C IJX377130A 2490C IJX377132A 2490C IJX377134A 2490C IJX377135A 2490C IJX377188A 2490C IJX377190A 2490C IJX377222A 2490C IJX377228A 2490C IJX377262A 2490C IJX377266A 2490C IJX377291A 2490C IJX377305A 2490C IJX377325A 2490C IJX377326A 2490C IJX377343A 2490C IJX377350A 2490C IJX377356A 2490C IJX377364A 2490C IJX377374A 2490C IJX377411A 2490C IJX377424A 2490C IJX377436A 2490C IJX377442A 2490C IJX377482A 2490C IJX377497A 2490C IJX377507A 2490C IJX377520A 2490C IJX377536A 2490C IJX377539A 2490C IJX377568A 2490C IJX377577A 2490C IJX377588A 2490C IJX377601A 2490C IJX377609A 2490C IJX377612A 2490C IJX377615A 2490C IJX377631A 2490C IJX377633A 2490C IJX377681A 2490C IJX377683A 2490C IJX377693A 2490C IJX377694A 2490C IJX377696A 2490C IJX377707A 2490C IJX377715A 2490C IJX377718A 2490C IJX377732A 2490C IJX377740A 2490C IJX377777A 2490C IJX377796A 2490C IJX377799A 2490C IJX377822A 2490C IJX377830A 2490C IJX377847A 2490C IJX377848A 2490C IJX377850A 2490C IJX377861A 2490C IJX377888A 2490C IJX377890A 2490C IJX377916A 2490C IJX377920A 2490C IJX377924A 2490C IJX377942A 2490C IJX377944A 2490C IJX377945A 2490C IJX377998A 2490C IJX378033A 2490C IJX378051A 2490C IJX378055A 2490C IJX378062A 2490C IJX378063A 2490C IJX378087A 2490C IJX378089A 2490C IJX378091A 2490C IJX378112A 2490C IJX378120A 2490C IJX378144A 2490C IJX378152A 2490C IJX378155A 2490C IJX378159A 2490C IJX386073A 2490C IJX386076A 2490C IJX386089A 2490C IJX386090A 2490C IJX386097A 2490C IJX386098A 2490C IJX386106A 2490C IJX386118A 2490C IJX386120A 2490C IJX386125A 2490C IJX386134A 2490C IJX386138A 2490C IJX386140A 2490C IJX386167A 2490C IJX386169A 2490C IJX386172A 2490C IJX386178A 2490C IJX386199A 2490C IJX386225A 2490C IJX386230A 2490C IJX386240A 2490C IJX386241A 2490C IJX386245A 2490C IJX386247A 2490C IJX386259A 2490C IJX386279A 2490C IJX386314A 2490C IJX386323A 2490C IJX386341A 2490C IJX386349A 2490C IJX386355A 2490C IJX386375A 2490C IJX386403A 2490C IJX386405A 2490C IJX386435A 2490C IJX386465A 2490C IJX386469A 2490C IJX386477A 2490C IJX386484A 2490C IJX386490A 2490C IJX386502A 2490C IJX386507A 2490C IJX386511A 2490C IJX386540A 2490C IJX386543A 2490C IJX386551A 2490C IJX386567A 2490C IJX386580A 2490C IJX386581A 2490C IJX386591A 2490C IJX386593A 2490C IJX386599A 2490C IJX386602A 2490C IJX386616A 2490C IJX386620A 2490C IJX386636A 2490C IJX386655A 2490C IJX386663A 2490C IJX386705A 2490C IJX386742A 2490C IJX386759A 2490C IJX386762A 2490C IJX386771A 2490C IJX386773A 2490C IJX386783A 2490C IJX386794A 2490C IJX386869A 2490C IJX386907A 2490C IJX386912A 2490C IJX386920A 2490C IJX386925A 2490C IJX386970A 2490C IJX387030A 2490C IJX387049A 2490C IJX387057A 2490C IJX387089A 2490C IJX387092A 2490C IJX387135A 2490C IJX387148A 2490C IJX387163A 2490C IJX387171A 2490C IJX387172A 2490C IJX387173A 2490C IJX387188A 2490C IJX387203A 2490C IJX387214A 2490C IJX387243A 2490C IJX387254A 2490C IJX387269A 2490C IJX387276A 2490C IJX387289A 2490C IJX387295A 2490C IJX387298A 2490C IJX387319A 2490C IJX387328A 2490C IJX387333A 2490C IJX387343A 2490C IJX387349A 2490C IJX387360A 2490C IJX387375A 2490C IJX387379A 2490C IJX387387A 2490C IJX387406A 2490C IJX387418A 2490C IJX387431A 2490C IJX387502A 2490C IJX387539A 2490C IJX387571A 2490C IJX387573A 2490C IJX387578A 2490C IJX387585A 2490C IJX387587A 2490C IJX387595A 2490C IJX387620A 2490C IJX387686A 2490C IJX387698A 2490C IJX387699A 2490C IJX387730A 2490C IJX387740A 2490C IJX387742A 2490C IJX387781A 2490C IJX387790A 2490C IJX387821A 2490C IJX387919A 2490C IJX387921A 2490C IJX387927A 2490C IJX387931A 2490C IJX387980A 2490C IJX387991A 2490C IJX387992A 2490C IJX388003A 2490C IJX388010A 2490C IJX388014A 2490C IJX388044A 2490C IJX388062A 2490C IJX388075A 2490C IJX388084A 2490C IJX388124A 2490C IJX388139A 2490C IJX388147A 2490C IJX389703A 2490C IJX389704A 2490C IJX389706A 2490C IJX389709A 2490C IJX389715A 2490C IJX389717A 2490C IJX389726A 2490C IJX389728A 2490C IJX389737A 2490C IJX389750A 2490C IJX389754A 2490C IJX389760A 2490C IJX389764A 2490C IJX389776A 2490C IJX389779A 2490C IJX389783A 2490C IJX389788A 2490C IJX389791A 2490C IJX389812A 2490C IJX389815A 2490C IJX389816A 2490C IJX389817A 2490C IJX389818A 2490C IJX389823A 2490C IJX389829A 2490C IJX389835A 2490C IJX389836A 2490C IJX389844A 2490C IJX389856A 2490C IJX389868A 2490C IJX389882A 2490C IJX389888A 2490C IJX389910A 2490C IJX389932A 2490C IJX391562A 2490C IJX391575A 2490C IJX391584A 2490C IJX391585A 2490C IJX391589A 2490C IJX391603A 2490C IJX391610A 2490C IJX391619A 2490C IJX391636A 2490C IJX391651A 2490C IJX391661A 2490C IJX391665A 2490C IJX391674A 2490C IJX391702A 2490C IJX391708A 2490C IJX391718A 2490C IJX391747A 2490C IJX391763A 2490C IJX391773A 2490C IJX391783A 2490C IJX391802A 2490C IJX391804A 2490C IJX391825A 2490C IJX391843A 2490C IJX391844A 2490C IJX391846A 2490C IJX391852A 2490C IJX391876A 2490C IJX391889A 2490C IJX391908A 2490C IJX391918A 2490C IJX391919A 2490C IJX391925A 2490C IJX391938A 2490C IJX391942A 2490C IJX391960A 2490C IJX391970A 2490C IJX391990A 2490C IJX391991A 2490C IJX391995A 2490C IJX391996A 2490C IJX392001A 2490C IJX392003A 2490C IJX392052A 2490C IJX392053A 2490C IJX392058A 2490C IJX392111A 2490C IJX392137A 2490C IJX392175A 2490C IJX392176A 2490C IJX392191A 2490C IJX392209A 2490C IJX392254A 2490C IJX392257A 2490C IJX392258A 2490C IJX392279A 2490C IJX392309A 2490C IJX392320A 2490C IJX392328A 2490C IJX392340A 2490C IJX392358A 2490C IJX392365A 2490C IJX392371A 2490C IJX392421A 2490C IJX392431A 2490C IJX392450A 2490C IJX392461A 2490C IJX392480A 2490C IJX392481A 2490C IJX392484A 2490C IJX392499A 2490C IJX392537A 2490C IJX392544A 2490C IJX392547A 2490C IJX392550A 2490C IJX392571A 2490C IJX392573A 2490C IJX392576A 2490C IJX392579A 2490C IJX392580A 2490C IJX392590A 2490C IJX392596A 2490C IJX392603A 2490C IJX392606A 2490C IJX392611A 2490C IJX392621A 2490C IJX392631A 2490C IJX392640A 2490C IJX392648A 2490C IJX392655A 2490C IJX392661A 2490C IJX392663A 2490C IJX392678A 2490C IJX392722A 2490C IJX392735A 2490C IJX401638A 2490C IJX401663A 2490C IJX401669A 2490C IJX401691A 2490C IJX401706A 2490C IJX401709A 2490C IJX401715A 2490C IJX401768A 2490C IJX401769A 2490C IJX401776A 2490C IJX401780A 2490C IJX401784A 2490C IJX401786A 2490C IJX401794A 2490C IJX401809A 2490C IJX401872A 2490C IJX401888A 2490C IJX401931A 2490C IJX401942A 2490C IJX401947A 2490C IJX401952A 2490C IJX401955A 2490C IJX401963A 2490C IJX401972A 2490C IJX401979A 2490C IJX401985A 2490C IJX401995A 2490C IJX402051A 2490C IJX402081A 2490C IJX402097A 2490C IJX402115A 2490C IJX402132A 2490C IJX402134A 2490C IJX402139A 2490C IJX402156A 2490C IJX402164A 2490C IJX402167A 2490C IJX402169A 2490C IJX402188A 2490C IJX402190A 2490C IJX402193A 2490C IJX402199A 2490C IJX402201A 2490C IJX402202A 2490C IJX402207A 2490C IJX402219A 2490C IJX402220A 2490C IJX402231A 2490C IJX402236A 2490C IJX402238A 2490C IJX402242A 2490C IJX402251A 2490C IJX402255A 2490C IJX402268A 2490C IJX402276A 2490C IJX402279A 2490C IJX402284A 2490C IJX402287A 2490C IJX402288A 2490C IJX402293A 2490C IJX402304A 2490C IJX402310A 2490C IJX402318A 2490C IJX402319A 2490C IJX402321A 2490C IJX402327A 2490C IJX402335A 2490C IJX402338A 2490C IJX402341A 2490C IJX402344A 2490C IJX402355A 2490C IJX402361A 2490C IJX402363A 2490C IJX402370A 2490C IJX402677A 2490C IJX406510A 2490C IJX406513A 2490C IJX406525A 2490C IJX406550A 2490C IJX406559A 2490C IJX406570A 2490C IJX406572A 2490C IJX406578A 2490C IJX406594A 2490C IJX406601A 2490C IJX406604A 2490C IJX406631A 2490C IJX406668A 2490C IJX406676A 2490C IJX406679A 2490C IJX406680A 2490C IJX406683A 2490C IJX406687A 2490C IJX406690A 2490C IJX406695A 2490C IJX406698A 2490C IJX406707A 2490C IJX406723A 2490C IJX406728A 2490C IJX406783A 2490C IJX406787A 2490C IJX406790A 2490C IJX406793A 2490C IJX406797A 2490C IJX406828A 2490C IJX406830A 2490C IJX406855A 2490C IJX406871A 2490C IJX406874A 2490C IJX406882A 2490C IJX406884A 2490C IJX406889A 2490C IJX406920A 2490C IJX406930A 2490C IJX406934A 2490C IJX406942A 2490C IJX406947A 2490C IJX406951A 2490C IJX406982A 2490C IJX406983A 2490C IJX407012A 2490C IJX407015A 2490C IJX407016A 2490C IJX407018A 2490C IJX407019A 2490C IJX407024A 2490C IJX407027A 2490C IJX407050A 2490C IJX407061A 2490C IJX407066A 2490C IJX407111A 2490C IJX407136A 2490C IJX407172A 2490C IJX407194A 2490C IJX407199A 2490C IJX407206A 2490C IJX407207A 2490C IJX407227A 2490C IJX407278A 2490C IJX407286A 2490C IJX407317A 2490C IJX407374A 2490C IJX407391A 2490C IJX407396A 2490C IJX407397A 2490C IJX407400A 2490C IJX407401A 2490C IJX407409A 2490C IJX407412A 2490C IJX407435A 2490C IJX407471A 2490C IJX407478A 2490C IJX407480A 2490C IJX407483A 2490C IJX407544A 2490C IJX407554A 2490C IJX407591A 2490C IJX407600A 2490C IJX407639A 2490C IJX407675A 2490C IJX407680A 2490C IJX407689A 2490C IJX407694A 2490C IJX407696A 2490C IJX407699A 2490C IJX407709A 2490C IJX407711A 2490C IJX407724A 2490C IJX407727A 2490C IJX407729A 2490C IJX407732A 2490C IJX407755A 2490C IJX407759A 2490C IJX407785A 2490C IJX407798A 2490C IJX407876A 2490C IJX407889A 2490C IJX407912A 2490C IJX407921A 2490C IJX407940A 2490C IJX407957A 2490C IJX407973A 2490C IJX407981A 2490C IJX407995A 2490C IJX407996A 2490C IJX411545A 2490C IJX411552A 2490C IJX411556A 2490C IJX411561A 2490C IJX411568A 2490C IJX411571A 2490C IJX411572A 2490C IJX411590A 2490C IJX411620A 2490C IJX411621A 2490C IJX411627A 2490C IJX411628A 2490C IJX411640A 2490C IJX411648A 2490C IJX411657A 2490C IJX411661A 2490C IJX411666A 2490C IJX411670A 2490C IJX411680A 2490C IJX411682A 2490C IJX411683A 2490C IJX411688A 2490C IJX411689A 2490C IJX411698A 2490C IJX411703A 2490C IJX411707A 2490C IJX411713A 2490C IJX411714A 2490C IJX411721A 2490C IJX411722A 2490C IJX411730A 2490C IJX800020A 2490C IJX800047A 2490C IJX800059A 2490C IJX800074A 2490C IJX800077A 2490C IJX800080A 2490C IJX800118A 2490C IJX800125A 2490C IJX800169A 2490C IJX800173A 2490C IJX800175A 2490C IJX800195A 2490C IJX800216A 2490C IJX800218A 2490C IJX800221A 2490C IJX800229A 2490C IJX800238A 2490C IJX800268A 2490C IJX800304A 2490C IJX800310A 2490C IJX800316A 2490C IJX800325A 2490C IJX800335A 2490C IJX800341A 2490C IJX800354A 2490C IJX800377A 2490C IJX800400A 2490C IJX800402A 2490C IJX800404A 2490C IJX800414A 2490C IJX800432A 2490C IJX800478A 2490C IJX800479A 2490C IJX800487A 2490C IJX800514A 2490C IJX800524A 2490C IJX800527A 2490C IJX800530A 2490C IJX800532A 2490C IJX800545A 2490C IJX800553A 2490C IJX800566A 2490C IJX800575A 2490C IJX800587A 2490C IJX800593A 2490C IJX800597A 2490C IJX800636A 2490C IJX800637A 2490C IJX800663A 2490C IJX800672A 2490C IJX800677A 2490C IJX800690A 2490C IJX800703A 2490C IJX800710A 2490C IJX800714A 2490C IJX800720A 2490C IJX800751A 2490C IJX800790A 2490C IJX800815A 2490C IJX800819A 2490C IJX800819P 2490C IJX800822A 2490C IJX800825A 2490C IJX800871A 2490C IJX800890A 2490C IJX800891A 2490C IJX800921A 2490C IJX800937A 2490C IJX800988A 2490C IJX800990A 2490C IJX800999A 2490C IJX801013P 2490C IJX801015A 2490C IJX801016A 2490C IJX801019A 2490C IJX801026P 2490C IJX801028A 2490C IJX801033A 2490C IJX801053A 2490C IJX801054A 2490C IJX801059P 2490C IJX801066A 2490C IJX801103A 2490C IJX801107A 2490C IJX801125A 2490C IJX801142P 2490C IJX801152A 2490C IJX801175A 2490C IJX801180A 2490C IJX801183A 2490C IJX801189A 2490C IJX801195A 2490C IJX801196A 2490C IJX801207A 2490C IJX801230A 2490C IJX801235A 2490C IJX801251A 2490C IJX801253A 2490C IJX801263A 2490C IJX801268A 2490C IJX801271A 2490C IJX801272A 2490C IJX801281A 2490C IJX801304A 2490C IJX801311A 2490C IJX801315A 2490C IJX801334P 2490C IJX801340A 2490C IJX801353A 2490C IJX801356P 2490C IJX801360A 2490C IJX801372A 2490C IJX801404A 2490C IJX801415A 2490C IJX801416A 2490C IJX801422A 2490C IJX801426A 2490C IJX801429P 2490C IJX801433A 2490C IJX801434A 2490C IJX801436P 2490C IJX801440A 2490C IJX801442A 2490C IJX801462A 2490C IJX801469A 2490C IJX801473A 2490C IJX801488A 2490C IJX801499A 2490C IJX801521A 2490C IJX801529A 2490C IJX801583A 2490C IJX801589P 2490C IJX801597A 2490C IJX801600A 2490C IJX801619A 2490C IJX801620A 2490C IJX801629A 2490C IJX801640P 2490C IJX801649A 2490C IJX801650A 2490C IJX801653A 2490C IJX801661P 2490C IJX801665A 2490C IJX801677A 2490C IJX801689A 2490C IJX801695A 2490C IJX801699A 2490C IJX801704A 2490C IJX801705A 2490C IJX801708A 2490C IJX801735A 2490C IJX801737A 2490C IJX801739A 2490C IJX801743A 2490C IJX801751A 2490C IJX801754A 2490C IJX801768A 2490C IJX801768P 2490C IJX801785A 2490C IJX801787A 2490C IJX801789A 2490C IJX801793A 2490C IJX801798A 2490C IJX801799A 2490C IJX801816A 2490C IJX801828A 2490C IJX801830A 2490C IJX801832A 2490C IJX801833A 2490C IJX801839A 2490C IJX801842A 2490C IJX801852A 2490C IJX801863A 2490C IJX801866A 2490C IJX801872A 2490C IJX801876A 2490C IJX801905A 2490C IJX801921A 2490C IJX801923A 2490C IJX801930A 2490C IJX801935A 2490C IJX801938A 2490C IJX801939A 2490C IJX801964A 2490C IJX801977A 2490C IJX801978A 2490C IJX801994A 2490C IJX802010A 2490C IJX802031P 2490C IJX802046A 2490C IJX802063A 2490C IJX802068A 2490C IJX802098A 2490C IJX802136A 2490C IJX802139A 2490C IJX802142A 2490C IJX802161A 2490C IJX802163A 2490C IJX802170A 2490C IJX802180A 2490C IJX802181A 2490C IJX802211P 2490C IJX802217A 2490C IJX802239A 2490C IJX802253A 2490C IJX802260A 2490C IJX802275A 2490C IJX802281A 2490C IJX802297A 2490C IJX802319A 2490C IJX802328A 2490C IJX802354A 2490C IJX802361A 2490C IJX802364A 2490C IJX802379A 2490C IJX802413A 2490C IJX802414A 2490C IJX802445A 2490C IJX802494A 2490C IJX802503A 2490C IJX802539P 2490C IJX802549A 2490C IJX802554A 2490C IJX802569P 2490C IJX802592A 2490C IJX802594P 2490C IJX802609A 2490C IJX802615A 2490C IJX802626A 2490C IJX802642A 2490C IJX802644A 2490C IJX802674A 2490C IJX802677A 2490C IJX802682A 2490C IJX802714A 2490C IJX802723A 2490C IJX802729P 2490C IJX802780A 2490C IJX802807A 2490C IJX802808A 2490C IJX802846A 2490C IJX802850A 2490C IJX802851A 2490C IJX802854A 2490C IJX802877A 2490C IJX802879A 2490C IJX802896A 2490C IJX802899A 2490C IJX802922A 2490C IJX802937A 2490C IJX802944A 2490C IJX802945A 2490C IJX802952A 2490C IJX802954A 2490C IJX802959A 2490C IJX803002P 2490C IJX803018A 2490C IJX803019A 2490C IJX803025A 2490C IJX803052A 2490C IJX803057A 2490C IJX803058A 2490C IJX803074A 2490C IJX803082A 2490C IJX803091A 2490C IJX803107A 2490C IJX803111A 2490C IJX803117A 2490C IJX803124A 2490C IJX803147A 2490C IJX803182A 2490C IJX803184A 2490C IJX803214A 2490C IJX803223A 2490C IJX803256A 2490C IJX803259A 2490C IJX803266A 2490C IJX803284P 2490C IJX803304A 2490C IJX803310A 2490C IJX803317P 2490C IJX803320A 2490C IJX803323A 2490C IJX803324A 2490C IJX803331A 2490C IJX803342P 2490C IJX803364A 2490C IJX803367A 2490C IJX803375A 2490C IJX803385A 2490C IJX803389A 2490C IJX803399A 2490C IJX803401A 2490C IJX803407A 2490C IJX803444A 2490C IJX803457A 2490C IJX803470A 2490C IJX803471A 2490C IJX803483A 2490C IJX803485A 2490C IJX803493A 2490C IJX803494A 2490C IJX803498A 2490C IJX803512A 2490C IJX803516A 2490C IJX803529A 2490C IJX803540A 2490C IJX803544A 2490C IJX803558A 2490C IJX803596A 2490C IJX803596P 2490C IJX803626A 2490C IJX803639A 2490C IJX803669A 2490C IJX803692A 2490C IJX803716A 2490C IJX803745A 2490C IJX803752P 2490C IJX803824A 2490C IJX803835P 2490C IJX803850A 2490C IJX803851A 2490C IJX803852A 2490C IJX803869A 2490C IJX803879A 2490C IJX803883A 2490C IJX803891P 2490C IJX803892A 2490C IJX803919A 2490C IJX803944A 2490C IJX803949A 2490C IJX803949P 2490C IJX803952A 2490C IJX803958P 2490C IJX803970P 2490C IJX803994A 2490C IJX803999A 2490C IJX804022A 2490C IJX804023A 2490C IJX804030A 2490C IJX804031P 2490C IJX804037A 2490C IJX804058A 2490C IJX804101A 2490C IJX804110A 2490C IJX804134A 2490C IJX804139A 2490C IJX804141A 2490C IJX804177A 2490C IJX804184A 2490C IJX804196A 2490C IJX804199A 2490C IJX804205A 2490C IJX804207A 2490C IJX804217A 2490C IJX804228P 2490C IJX804231A 2490C IJX804235A 2490C IJX804238A 2490C IJX804282P 2490C IJX804316A 2490C IJX804328A 2490C IJX804340A 2490C IJX804349A 2490C IJX804356A 2490C IJX804358A 2490C IJX804368A 2490C IJX804379A 2490C IJX804386A 2490C IJX804402A 2490C IJX804402P 2490C IJX804419A 2490C IJX804420A 2490C IJX804430P 2490C IJX804433A 2490C IJX804436A 2490C IJX804439A 2490C IJX804449A 2490C IJX804462A 2490C IJX804475A 2490C IJX804513A 2490C IJX804519P 2490C IJX804524A 2490C IJX804528A 2490C IJX804530A 2490C IJX804530P 2490C IJX804531A 2490C IJX804536A 2490C IJX804537A 2490C IJX804582A 2490C IJX804583A 2490C IJX804590A 2490C IJX804604A 2490C IJX804609A 2490C IJX804624A 2490C IJX804672A 2490C IJX804683A 2490C IJX804688A 2490C IJX804690A 2490C IJX804720A 2490C IJX804721A 2490C IJX804733A 2490C IJX804742A 2490C IJX804775A 2490C IJX804792A 2490C IJX804801A 2490C IJX804804A 2490C IJX804810A 2490C IJX804815A 2490C IJX804842A 2490C IJX804846A 2490C IJX804848A 2490C IJX804857A 2490C IJX804859A 2490C IJX804860A 2490C IJX804873A 2490C IJX804874A 2490C IJX804879A 2490C IJX804880A 2490C IJX804886A 2490C IJX804892A 2490C IJX804907A 2490C IJX804910A 2490C IJX804912A 2490C IJX804913A 2490C IJX804916A 2490C IJX804922A 2490C IJX804929A 2490C IJX804948A 2490C IJX804967A 2490C IJX804968A 2490C IJX804977A 2490C IJX804978A 2490C IJX804983A 2490C IJX805004A 2490C IJX805005A 2490C IJX805027A 2490C IJX805037A 2490C IJX805040A 2490C IJX805054A 2490C IJX805061A 2490C IJX805067A 2490C IJX805078A 2490C IJX805079A 2490C IJX805081A 2490C IJX805086A 2490C IJX805093A 2490C IJX805116A 2490C IJX805119A 2490C IJX805129A 2490C IJX805132A 2490C IJX805141A 2490C IJX805144A 2490C IJX805155A 2490C IJX805157A 2490C IJX805160A 2490C IJX805161A 2490C IJX805172A 2490C IJX805174A 2490C IJX805187A 2490C IJX805194A 2490C IJX805200A 2490C IJX805210A 2490C IJX805218A 2490C IJX805219A 2490C IJX805222A 2490C IJX805225A 2490C IJX805226A 2490C IJX805248A 2490C IJX805249A 2490C IJX805255A 2490C IJX805278A 2490C IJX805285A 2490C IJX805288A 2490C IJX805346A 2490C IJX805373A 2490C IJX805377A 2490C IJX805383A 2490C IJX805385A 2490C IJX805404A 2490C IJX805417A 2490C IJX805450A 2490C IJX805479A 2490C IJX805481A 2490C IJX805483A 2490C IJX805498A 2490C IJX805499A 2490C IJX805503A 2490C IJX805504A 2490C IJX805521A 2490C IJX805526A 2490C IJX805559A 2490C IJX805561A 2490C IJX805563A 2490C IJX805568A 2490C IJX805569A 2490C IJX805602P 2490C IJX805606A 2490C IJX805610A 2490C IJX805622A 2490C IJX805628A 2490C IJX805673P 2490C IJX805684A 2490C IJX805708A 2490C IJX805721P 2490C IJX805759A 2490C IJX805771A 2490C IJX805792A 2490C IJX805804A 2490C IJX805930A 2490C IJX805956A 2490C IJX805968A 2490C IJX805976A 2490C IJX805980A 2490C IJX805984A 2490C IJX805993A 2490C IJX806001A 2490C IJX806006A 2490C IJX806084A 2490C IJX806091P 2490C IJX806102P 2490C IJX806127P 2490C IJX806144P 2490C IJX806214P 2490C IJX806221A 2490C IJX806241P 2490C IJX806358A 2490C IJX806363A 2490C IJX806425A 2490C IJX806448A 2490C IJX806464A 2490C IJX806465A 2490C IJX806472A 2490C IJX806476A 2490C IJX806479A 2490C IJX806490A 2490C IJX806495A 2490C IJX806509A 2490C IJX806510A 2490C IJX806517A 2490C IJX806530A 2490C IJX806535A 2490C IJX806546A 2490C IJX806551A 2490C IJX806557A 2490C IJX806566A 2490C IJX806575A 2490C IJX806599A 2490C IJX806636A 2490C IJX806650A 2490C IJX806657A 2490C IJX806659A 2490C IJX806660A 2490C IJX806662A 2490C IJX806664A 2490C IJX806669P 2490C IJX806686A 2490C IJX806712A 2490C IJX806754P 2490C IJX806763A 2490C IJX806776A 2490C IJX806780A 2490C IJX806794P 2490C IJX806835A 2490C IJX806847A 2490C IJX806871A 2490C IJX806872A 2490C IJX806886A 2490C IJX806889A 2490C IJX806899A 2490C IJX806901A 2490C IJX806961A 2490C IJX806966A 2490C IJX806973A 2490C IJX806998A 2490C IJX807000A 2490C IJX807011A 2490C IJX807021A 2490C IJX807053A 2490C IJX807159A 2490C IJX807161A 2490C IJX807188A 2490C IJX807200A 2490C IJX807208A 2490C IJX807212A 2490C IJX807398A 2490C IJX807412A 2490C IJX807415A 2490C IJX807419A 2490C IJX807443P 2490C IJX807465A 2490C IJX807504A 2490C IJX807571A 2490C IJX807575A 2490C IJX807584A 2490C IJX807585A 2490C IJX807592A 2490C IJX807600A 2490C IJX807605A 2490C IJX807607A 2490C IJX807625A 2490C IJX807631A 2490C IJX807634A 2490C IJX807648A 2490C IJX807674A 2490C IJX807675A 2490C IJX807703P 2490C IJX807717A 2490C IJX807725A 2490C IJX807734P 2490C IJX807735A 2490C IJX807750A 2490C IJX807764A 2490C IJX807765A 2490C IJX807773A 2490C IJX807823A 2490C IJX807826P 2490C IJX807834A 2490C IJX807860A 2490C IJX807862A 2490C IJX807868A 2490C IJX807882A 2490C IJX807899A 2490C IJX807908A 2490C IJX807908P 2490C IJX807929A 2490C IJX807931A 2490C IJX807938A 2490C IJX807941A 2490C IJX807944A 2490C IJX807945A 2490C IJX808003P 2490C IJX808016P 2490C IJX808022A 2490C IJX808156A 2490C IJX808157A 2490C IJX808159A 2490C IJX808186A 2490C IJX808242P 2490C IJX808269A 2490C IJX808275A 2490C IJX808276A 2490C IJX808294A 2490C IJX808297A 2490C IJX808301A 2490C IJX808305A 2490C IJX808306A 2490C IJX808308A 2490C IJX808319A 2490C IJX808325A 2490C IJX808340A 2490C IJX808345A 2490C IJX808348A 2490C IJX808357A 2490C IJX808383A 2490C IJX808399A 2490C IJX808405A 2490C IJX808405P 2490C IJX808421A 2490C IJX808424A 2490C IJX808539P 2490C IJX808573P 2490C IJX808582P 2490C IJX809032A 2490C IJX809546A 2490C IJX809555P 2490C IJX809591P 2490C IJX809621P 2490C IJX809622A 2490C IJX809640P 2490C IJX809697A 2490C IJX809723A 2490C IJX809732A 2490C IJX809738P 2490C IJX809772A 2490C IJX809793P 2490C IJX809821P 2490C IJX809824A 2490C IJX809852P 2490C IJX809892A 2490C IJX809897P 2490C IJX809924P 2490C IJX809936P 2490C IJX809976P 2490C IJX809979P 2490C IJX810022P 2490C IJX810092P 2490C IJX810159P 2490C IJX810163P 2490C IJX810226A 2490C IJX810261A 2490C IJX810269A 2490C IJX810328P 2490C IJX810365P 2490C IJX810387P 2490C IJX810396A 2490C IJX810519A 2490C IJX810603A 2490C IJX810604A 2490C IJX810607A 2490C IJX810622A 2490C IJX810794A 2490C IJX810863A 2490C IJX810876A 2490C IJX810908A 2490C IJX810971A 2490C IJX810988A 2490C IJX811232A 2490C IJX811234A 2490C IJX811242A 2490C IJX811248A 2490C IJX811255A 2490C IJX811257A 2490C IJX811261A 2490C IJX811279A 2490C IJX811280A 2490C IJX811287A 2490C IJX811300A 2490C IJX811308A 2490C IJX811310A 2490C IJX811311A 2490C IJX811318A 2490C IJX811319A 2490C IJX811322A 2490C IJX811323A 2490C IJX812009A 2490C IJX812272A   
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure
8200 Coral Sea St NE
Mounds View MN 55112-4391
For Additional Information ContactLocal Medtronic Representative
763-514-4000
Manufacturer Reason
for Recall
A recent firmware update developed by Medtronic for the 2490C CareLink Monitors and 2020B CareLink Express Monitors included incorrect data on the country analog modem dial-up configuration table used to dial into the CareLink Network. This incorrect data resulted in the removal of the 0 prefix necessary for select countries as well as an incorrect alignment of phone numbers for other select cou
FDA Determined
Cause 2
Software design (manufacturing process)
ActionMedtronic sent an "Urgent Field Safety Notice" dated April 2016to their customers. The letter was addressed to Healthcare Professional. The letter identified the affected product, problem and customer actions to be taken. For questions contact your local Medtronic Representative.
Quantity in Commerce5038
DistributionAUSTRIA, AUSTRALIA, BELGIUM, SWITZERLAND, GERMANY, DENMARK, FRANCE, GREAT BRITAIN, HUNGARY, JAPAN, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, SWEDEN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = NVZ
-
-