| Class 1 Device Recall INRatio2 PT/INR Professional Monitoring System | |
Date Initiated by Firm | July 11, 2016 |
Date Posted | August 19, 2016 |
Recall Status1 |
Terminated 3 on July 20, 2017 |
Recall Number | Z-2360-2016 |
Recall Event ID |
74665 |
510(K)Number | K072727 |
Product Classification |
Test, time, prothrombin - Product Code GJS
|
Product | Alere INRatio2 PT/INR Professional Monitoring System
Model Number: 55128A
Product Usage:
The Alere INRatio/INRatio2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes. |
Code Information |
All Lots All Serial Numbers |
Recalling Firm/ Manufacturer |
Alere San Diego, Inc. 9975 Summers Ridge Rd San Diego CA 92121-2997
|
For Additional Information Contact | Jackie Lustig 781-341-4009 |
Manufacturer Reason for Recall | Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory. |
FDA Determined Cause 2 | Other |
Action | Alere issued a press release on July 11, 2016. An Urgent Medical Device Letter dated July 26, 2016 was sent to affected customers on July 29, 2016. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to consult with their healthcare provider as soon as possible to transition to an alternate method of PT/INR testing and to contact the firm regarding return or disposal of unused product and obtaining device replacement from a different manufacturer. For questions contact the Alere INRatid1 Recall Hotline at 1-866-723-2535. |
Distribution | Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = GJS
|
|
|
|