| | Class 2 Device Recall Spirit Select Bed |  |
| Date Initiated by Firm | June 28, 2016 |
| Date Posted | August 03, 2016 |
| Recall Status1 |
Terminated 3 on June 08, 2017 |
| Recall Number | Z-2332-2016 |
| Recall Event ID |
74560 |
| Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
| Product | Spirit Select Bed, A-C Powered Hospital Bed |
| Code Information |
Model Numbers: 5700, Spirit Select Bed (manufactured between November 1, 2014 January 13, 2016); Serial Numbers:029481 029482 029483 029484 029485 029486 029487 029488 029489 029490 029491 029492 029493 029494 029495 029496 029497 029498 029499 029500 029501 029502 029503 029504 029505 029506 029507 029508 029509 029510 029511 029512 029513 029514 029515 029516 029517 029518 029519 029520 029521 029522 029523 029524 029525 029526 029527 029528 029529 029530 029531 029532 029543 029544 029545 029546 029547 029548 029549 029550 029551 029552 029553 029554 029555 029556 029557 029565 029566 029567 029568 029569 029570 029571 029572 029573 029574 029575 029576 029577 029578 029579 029580 029581 029582 029583 029584 029585 029586 029587 029588 029589 029590 029591 029592 029593 029594 029595 029596 029597 029598 029599 029600 029601 029602 029603 029604 029605 029606 029607 029608 029609 029610 029611 029612 029613 029614 029615 029616 029617 029618 029619 029620 029621 029622 029623 029624 029625 029626 029627 029628 029629 029630 029631 029632 029633 029634 029635 029636 029637 029638 029660 029661 029662 029663 029664 029665 029666 029667 029668 029669 029670 029671 029672 029673 029674 029675 029676 029677 029678 029679 029680 029681 029682 029683 029684 029685 029686 029687 029688 029689 029690 029691 029692 029693 029694 029695 029696 029697 029698 029699 029700 029701 029702 029703 029704 029705 029706 029709 029710 029711 029712 029713 029714 029715 029716 029717 029718 029719 029720 029721 029722 029723 029724 029725 029726 029727 029728 029729 029730 029731 029732 029733 029734 029735 029736 029737 029738 029739 029740 029741 029742 029743 029744 029745 029746 029747 029748 029749 029750 029751 029752 029753 029754 029755 029756 029757 029758 029759 029760 029812 029813 029814 029815 029816 029817 029818 029819 029820 029821 029822 029823 029824 029825 029826 029827 029828 029829 029830 029831 029832 029833 029834 029835 029836 029837 029838 029839 029840 029841 029842 029843 029844 029845 029846 029847 029848 029849 029850 029851 029852 029853 029854 029855 029856 029857 029858 029859 029860 029861 029862 029863 029864 029865 029866 029867 029868 029869 029870 029871 029872 029873 029874 029875 029876 029877 029879 029880 029881 029882 029883 029884 029885 029886 029887 029888 029889 029890 029891 029892 029893 029894 029895 029896 029897 029898 029899 029900 029901 029916 029917 029918 029919 029920 029921 029922 029923 029924 029925 029926 029927 029928 029930 029931 029932 029933 029934 029935 029936 029937 029938 029939 029940 029941 029942 029943 029944 029945 029946 029947 029948 029949 029951 029952 029953 029954 029955 029956 029957 029958 029959 029960 029961 029962 029963 029964 029965 029966 029967 029968 029969 029970 029971 029972 029973 029974 029975 029976 029977 029978 029979 029980 029981 029982 029983 029984 029985 029986 029987 029988 029989 029990 029991 029992 029993 029994 029995 029996 029997 029998 029999 030000 030001 030002 030003 030004 030005 030006 030007 030008 030009 030010 030011 030012 030013 030014 030015 030016 030017 030018 030019 030020 030021 030022 030023 030024 030025 030026 030027 030028 030029 030030 030031 030032 030033 030034 030035 030103 030104 030105 030106 030107 030108 030109 030110 030111 030112 030113 030114 030115 030116 030117 030118 030119 030120 030121 030122 030123 030124 030125 030126 030127 030128 030129 030130 030131 030132 030133 030134 030135 030136 030137 030138 030139 030140 030141 030142 030143 030144 030145 030146 030147 030148 030162 030163 030164 030165 030166 030167 030168 030169 030170 030171 030172 030173 030174 030175 030176 030177 030178 030179 030180 030181 030182 030183 030184 030185 030186 030187 030188 030189 030190 030191 030192 030193 030194 030195 030196 030197 030198 030199 030200 030201 030202 030203 030222 030233 030234 030235 030239 030240 030250 030286 030287 030288 030289 030290 030291 030292 030293 030294 030295 030296 030297 030298 030299 030300 030301 030302 030303 030304 030305 030306 030307 030308 030309 030310 030311 030312 030313 030314 030315 030316 030317 030318 030319 030320 030321 030322 030323 030324 030325 030326 030327 030328 030329 030330 030331 030332 030333 030334 030335 030336 030337 030338 030339 030340 030341 030342 030343 030344 030345 030346 030347 030348 030349 030350 030351 030352 030353 030398 030399 030400 030401 030402 030403 030404 030405 030406 030412 030413 030414 030415 030416 030417 030419 030420 030421 030422 030423 030424 030425 030426 030427 030428 030429 030430 030431 030432
030433 030434 030435 030436 030437 030438 030476 030504 030505 030506 030507 030520 030521 030522 030524 030526 030528 030529 030530 030531 030532 030533 030534 030535 030536 030537 030538 030539 030540 030541 030542 030543 030544 030545 030549 030552 030553 030554 030555 030556 030580 030643 030644 030645 030766 030767 030926 030927 030928 030929 030930 030931 030932 030933 030934 030935 030936 030937 030938 030939 030940 030941 030942 030943 030944 030945 030946 030947 030948 030949 030950 030951 030952 030953 030954 030955 030956 030960 030970 030971 030972 030973 030974 030975 030976 030977 030978 030979 030980 030981 030982 030983 030984 030985 030986 030987 030988 030989 030990 030991 030992 030993 030994 030995 030996 030997 030998 030999 031000 031001 031002 031003 031004 031005 031006 031007 031008 031009 031010 031011 031012 031013 031014 031033 031034 031035 031036 031037 031038 031039 031040 031041 031042 031043 031044 031045 031046 031047 031048 031049 031050 031051 031052 031053 031054 031055 031056 031057 031058 031059 031060 031061 031062 031063 031064 031065 031066 031067 031068 031069 031070 031071 031072 031073 031074 031075 031076 031077 031078 031079 031080 031081 031082 031083 031084 031085 031086 031089 031090 031091 031092 031093 031094 031095 031096 031097 031098 031099 031100 031101 031102 031103 031104 031105 031106 031107 031108 031109 031110 031111 031112 031113 031114 031115 031116 031117 031118 031119 031120 031121 031122 031123 031124 031125 031126 031127 031128 031129 031130 031131 031132 031133 031134 031135 031136 031137 031138 031139 031140 031141 031142 031143 031144 031145 031146 031147 031148 031149 031150 031151 031152 031153 031154 031155 031156 031157 031158 031159 031160 031161 031162 031163 031164 031165 031166 031167 031168 031169 031170 031171 031172 031173 031174 031175 031176 031177 031178 031179 031180 031181 031182 031183 031184 031185 031186 031187 031188 031189 031190 031191 031192 031193 031194 031195 031196 031197 031198 031199 031200 031201 031202 031203 031204 031205 031206 031207 031208
031218 031219 031220 031221 031222 031223 031224 031225 031226 031227 031228 031229 031230 031231 031232 031233 031234 031235 031236 031237 031238 031239 031240 031241 031242 031243 031244 031245 031246 031247 031248 031249 031250 031251 031252 031253 031254 031255 031256 031264 031265 031266 031267 031268 031269 031270 031271 031272 031273 031274 031275 031276 031277 031278 031279 031280 031281 031282 031283 031284 031285 031286 031287 031288 031289 031290 031291 031292 031295 031296 031297 031298 031299 031300 031301 031302 031303 031304 031305 031306 031307 031308 031309 031310 031311 031312 031313 031314 031315 031316 031317 031318 031319 031320 031321 031322 031323 031324 031325 031327 031328 031329 031330 031331 031332 031333 031334 031335 031336 031337 031338 031339 031340 031341 031342 031343 031344 031345 031346 031347 031348 031349 031350 031351 031352 031353 031354 031355 031356 031357 031358 031359 031360 031361 031362 031363 031364 031365 031366 031367 031368 031369 031370 031371 031372 031373 031374 031375 031376 031377 031378 031379 031380 031381 031382 031383 031384 031385 031386 031387 031388 031389 031390 031391 031392 031393 031394 031395 031396 031397 031398 031399 031400 031401 031402 031417 031419 031420 031421 031422 031423 031424 031431 031463 031464 031465 031466 031467 031468 031469 031470 031471 031472 031473 031474 031475 031476 031477 031478 031479 031480 031481 031482 031483 031484 031485 031486 031487 031488 031489 031491 031492 031493 031494 031495 031496 031497 031498 031499 031500 031501 031502 031503 031504 031505 031506 031507 031508 031509 031547 031548 031549 031550 031551 031552 031553 031554 031555 031556 031557 031558 031559 031560 031561 031562 031563 031564 031565 031566 031567 031568 031569 031570 031571 031572 031573 031574 031575 031576 031577 031578 031579 031580 031581 031582 031583 031584 031585 031586 031591 031592 031593 031594 031595 031596 031597 031598 031599 031600 031601 031602 031603 031604 031605 031606 031607 031608 031609 031610 031611 031612 031613 031614 031615 031642 031643 031644
031645 031646 031647 031648 031649 031650 031651 031652 031653 031654 031655 031656 031657 031658 031665 031666 031667 031668 031669 031670 031671 031672 031673 031674 031675 031676 031677 031678 031679 031680 031682 031683 031684 031685 031686 031687 031688 031689 031690 031691 031692 031693 031694 031695 031696 031697 031698 031699 031700 031701 031702 031703 031704 031705 031706 031707 031708 031709 031710 031711 031712 031713 031714 031715 031716 031717 031718 031719 031720 031722 031723 031724 031725 031726 031727 031728 031729 031730 031731 031732 031733 031734 031735 031736 031737 031738 031739 031740 031741 031742 031743 031744 031745 031746 031747 031748 031749 031750 031751 031752 031753 031754 031755 031756 031757 031758 031759 031760 031761 031762 031763 031764 031765 031766 031767 031768 031769 031770 031771 031772 031773 031774 031775 031776 031796 031797 031798 031799 031800 031801 031802 031803 031804 031805 031806 031807 031808 031809 031810 031811 031812 031813 031814 031815 031816 031817 031818 031819 031820 031821 031822 031823 031824 031825 031826 031827 031828 031829 031830 031831 031832 031840 031843 031844 031845 031846 031847 031848 031849 031850 031851 031852 031853 031854 031855 031856 031857 031858 031859 031860 031861 031862 031863 031864 031865 031866 031867 031868 031869 031870 031871 031872 031873 031874 031875 031876 031877 031878 031879 031880 031881 031882 031883 031884 031885 031886 031887 031888 031889 031890 031891 031892 031893 031894 031895 031896 031897 031898 031899 031900 031901 031902 031903 031904 031905 031906 031907 031908 031909 031910 031911 031912 031913 031914 031915 031916 031917 031918 031919 031920 031921 031922 031923 031924 031925 031926 031927 031928 031929 031930 031931 031932 031933 031934 031935 031936 031937 031938 031939 031940 031941 031942 031943 031944 032020 032021 032022 032023 032024 032025 032026 032027 032028 032029 032030 032031 032055 032056 032057 032058 032059 032060 032061 032062 032063 032064 032065 032066 032067 032068 032069 032070 032071 032072 032073 032074 032075 032076 032077 032078 032079 032080 032081 032082 032083 032084 032085 032086 032087 032088 032089 032090 032091 032092 032093 032094 032095 032096 032097 032098 032099 032100 032101 032102 032103 032104 032105 032106 032107 032108 032109 032110 032111 032112 032113 032114 032115 032116 032117 032118 032119 032120 032121 032122 032123 032124 032125 032126 032127 032128 032129 032131 032132 032133 032134 032135 032136 032137 032138 032139 032140 032141 032142 032143 032144 032145 032146 032147 032148 032149 032150 032151 032152 032153 032190 032191 032192 032193 032194 032195 032196 032197 032198 032199 032200 032201 032202 032203 032204 032205 032206 032207 032208 032209 032210 032211 032212 032213 032214 032215 032216 032217 032218 032219 032220 032221 032222 032223 032224 032225 032226 032227 032228 032229 032230 032231 032232 032233 032234 032235 032236 032237 032238 032239 032240 032241 032242 032243 032244 032245 032246 032247 032248 032249 032250 032251 032252 032253 032254 032255 032256 032257 032258 032259 032260 032261 032262 032263 032264 032265 032266 032267 032268 032269 032270 032271 032272 032273 032274 032275 032276 032277 032278 032279 032280 032281 032282 032283 032284 032285 032286 032287 032288 032289 032290 032291 032292 032293 032294 032295 032296 032297 032298 032299 032300 032301 032302 032303 032304 032305 032306 032307 032308 032309 032310 032311 032312 032313 032314 032315 032316 032317 032318 032319 032320 032321 032322 032323 032324 032325 032326 032327 032328 032329 032330 032331 032332 032333 032334 032335 032336 032337 032338 032339 032340 032341 032342 032343 032344 032357 032358 032359 032360 032361 032362 032363 032364 032365 032366 032367 032368 032369 032370 032371 032372 032373 032374 032375 032376 032377 032378 032379 032380 032381 032382 032383 032384 032386 032387 032388 032389 032390 032391 032392 032393 032394 032395 032396 032397 032398 032399 032400 032401 032402 032403 032404 032405 032406 032407 032408 032409 032410 032411 032412 032413 032414 032415 032416 032417 032418 032419 032420 032421 032422 032423 032424 032425 032426 032427 032428 032429 032430 032431 032432 032433 032434 032435 032436 032437 032438 032439 032440 032441 032442 032443 032444 032445 032446 032447 032448 032449 032450 032451 032452 032453 032454 032455 032456 032457 032458 032459 032460 032461 032462 032463 032464 032465 032466 032467 032468 032469 032472 032473 032474 032475 032476 032477 032478 032479 032480 032481 032482 032483 032484 032485 032486 032487 032488 032489 032490 032491 032492 032493 032494 032495 032496 032497 032498 032499 032500 032501 032502 032503 032504 032505 032506 032507 032508 032509 032510 032511 032512 032513 032514 032515 032516 032517 032518 032519 032520 032521 032522 032523 032524 032535 032536 032537 032538 032539 032540 032541 032542 032543 032544 032545 032546 032547 032548 032549 032550 032551 032552 032553 032554 032555 032556 032557 032558 032559 032560 032561 032562 032563 032564 032565 032566 032567 032568 032569 032570 032571 032572 032573 032574 032575 032576 032577 032578 032579 032580 032581 032594 032595 032596 032597 032598 032599 032600 032601 032602 032603 032604 032605 032606 032607 032608 032609 032610 032612 032613 032614 032615 032616 032617 032618 032619 032620 032621 032622 032623 032624 032625 032626 032627 032628 032629 032630 032631 032632 032633 032634 032635 032636 032637 032638 032639 032640 032641 032642 032643 032644 032645 032646 032647 032648 032649 032650 032651 032652 032653 032654 032655 032656 032657 032658 032659 032660 032661 032662 032663 032664 032671 032672 032673 032674 032675 032676 032677 032678 032679 032680 032681 032682 032683 032684 032685 032686 032687 032688 032689 032690 032691 032692 032693 032694 032695 032696 032697 032698 032699 032700 032701 032702 032703 032704 032705 032706 032707 032708 032709 032710 032711 032712 032713 032714 032715 032716 032717 032718 032719 032720 032721 032722 032723 032724 032725 032726 032727 032728 032729 032730 032731 032732 032733 032734 032738 032739 032740 032741 032742 032743 032744 032745 032746 032760 032761 032762 032763 032764 032765 032766 032767 032768 032769 032770 032771 032772 032773 032774 032775 032776 032777 032778 032779 032780 032781 032782 032783 032784 032785 032786 032787 032788 032789 032790 032791 032792 032793 032794 032795 032796 032797 032825 032826 032827 032828 032829 032830 032831 032832 032833 032834 032835 032836 032837 032838 032839 032840 032841 032843 032844 032845 032846 032847 032848 032849 032850 032851 032852 032853 032854 032855 032856 032857 032858 032859 032860 032861 032862 032863 032864 032865 032866 032867 032868 032869 032870 032871 032872 032873 032874 032875 032876 032877 032878 032879 032880 032881 032882 032883 032884 032885 032886 032887 032888 032889 032890 032891 032892 032893 032894 032895 032896 032897 032898 032899 032900 032901 032902 032903 032904 032906 032907 032909 032910 032911 032912 032913 032914 032915 032916 032917 032918 032919 032920 032921 032922 032923 032924 032925 032926 032927 032928 032929 032930 032931 032932 032933 032934 032935 032936 032937 032938 032939 032940 032941 032942 032943 032944 032945 032946 032947 032948 032949 032950 032951 032952 032953 032954 032955 032956 032957 032958 032959 032960 032962 032963 032964 032965 032966 032967 032968 032969 032970 032971 032972 032973 032974 032975 032976 032977 032978 032979 032980 032982 032983 032984 032985 032986 032987 032988 032989 032990 032991 032992 032993 032994 032995 032996 032997 032998 032999 033000 033001 033002 033003 033004 033005 033006 033007 033013 033014 033015 033016 033017 033018 033019 033020 033021 033022 033023 033024 033025 033026 033027 033028 033029 033030 033031 033032 033033 033034 033035 033036 033037 033038 033039 033040 033041 033042 033043 033044 033045 033046 033047 033048 033049 033050 033051 033052 033053 033054 033055 033056 033057 033058 033059 033060 033061 033062 033063 033064 033065 033066 033067 033068 033069 033070 033071 033072 033073 033074 033075 033076 033077 033078 033079 033080 033081 033082 033083 033084 033085 033086 033087 033088 033089 033090 033091 033092 033093 033094 033095 033096 033097 033098 033099 033100 033101 033102 033103 033104 033105 033106 033107 033108 033109 033110 033111 033112 033113 033114 033115 033116 033117 033118 033119 033120 033121 033122 033123 033124 033125 033126 033127 033128 033129 033130 033131 033132 033133 033134 033135 033136 033137 033138 033139 033140 033141 033142 033143 033144 033145 033146 033147 033148 033149 033150 033151 033152 033153 033154 033155 033156 033157 033158 033159 033160 033161 033162 033163 033164 033165 033166 033167 033168 033169 033170 033171 033172 033173 033229 033230 033231 033232 033233 033234 033235 033236 033237 033238 033239 033240 033241 033242 033243 033244 033245 033246 033247 033249 033250 033251 033252 033253 033254 033255 033256 033257 033258 033259 033260 033261 033262 033263 033264 033265 033266 033267 033268 033269 033270 033271 033272 033273 033274 033275 033276 033277 033278 033287 033288 033289 033290 033291 033292 033293 033294 033295 033296 033297 033298 033299 033300 033301 033302 033303 033304 033305 033306 033307 033308 033309 033310 033311 033312 033313 033314 033315 033316 033317 033318 033319 033320 033321 033322 033323 033324 033325 033326 033327 033328 033329 033330 033331 033332 033333 033334 033335 033336 033337 033338 033339 033340 033341 033342 033343 033344 033345 033346 033347 033348 033349 033350 033351 033352 033353 033354 033355 033356 033357 033358 033359 033360 033361 033362 033363 033364 033365 033366 033367 033368 033369 033370 033371 033372 033373 033374 033375 033376 033377 033378 033379 033380 033381 033382 033383 033384 033385 033386 033387 033388 033389 033390 033391 033392 033393 033394 033395 033396 033397 033398 033399 033400 033401 033402 033403 033404 033405 033406 033407 033408 033409 033410 033411 033412 033413 033414 033415 033416 033417 033418 033419 033420 033423 033424 033425 033426 033427 033428 033429 033430 033431 033432 033433 033434 033435 033436 033437 033438 033439 033440 033441 033442 033443 033444 033445 033446 033447 033448 033449 033450 033451 033452 033453 033454 033455 033456 033457 033458 033459 033460 033461 033462 033463 033464 033465 033466 033467 033468 033469 033470 033471 033472 033473 033474 033475 033476 033477 033478 033479 033480 033481 033482 033483 033484 033485 033486 033487 033488 033489 033490 033491 033492 033493 033494 033495 033496 033497 033498 033499 033500 033501 033502 033503 033504 033505 033506 033507 033508 033509 033510 033511 033512 033513 033514 033515 033516 033517 033518 033519 033520 033521 033522 033523 033524 033525 033526 033527 033528 033529 033530 033531 033532 033533 033534 033535 033536 033537 033538 033539 033540 033541 033542 033543 033544 033545 033546 033547 033548 033549 033550 033551 033552 033553 033554 033555 033556 033557 033558 033559 033560 033561 033562 033563 033564 033565 033566 033567 033568 033569 033570 033571 033572 033573 033574 033575 033576 033577 033578 033579 033580 033581 033582 033583 033584 033585 033586 033587 033588 033589 033590 033591 033592 033593 033594 033595 033596 033597 033598 033599 033600 033601 033602 033603 033604 033605 033606 033607 033608 033609 033610 033611 033612 033613 033614 033615 033616 033617 033618 033619 033620 033621 033622 033623 033624 033625 033626 033627 033628 033629 033630 033631 033632 033633 033634 033635 033636 033637 033638 033639 033640 033641 033642 033643 033644 033645 033646 033647 033648 033649 033650 033651 033652 033653 033654 033655 033656 033657 033658 033659 033660 033761 033762 033763 033764 033885 033886 033887 033888 033889 033890 033891 033892 033893 033894 033895 033896 033897 033898 033899 033900 033901 033902 033903 033904 033906 033907 033908 033909 033910 033911 033912 033913 033914 033915 033916 033917 033945 033946 033947 033948 033949 033950 033951 033952 033953 033954 033955 033956 033957 033958 033959 033960 033961 033962 033963 033964 033965 033966 033967 033968 033969 033970 033971 033972 033973 033974 033975 033976 033977 033981 033982 033983 033984 033985 033986 033987 033988 033989 033990 033991 033992 033993 033994 033995 033996 033997 033998 033999 034000 034001 034002 034003 034004 034005 034006 034007 034008 034009 034010 034011 034012 034013 034014 034015 034016 034017 034018 034019 034020 034021 034022 034023 034024 034025 034026 034027 034028 034029 034030 034031 034032 034038 034039 034040 034041 034042 034043 034044 034045 034046 034047 034048 034049 034050 034051 034052 034053 034054 034055 034056 034057 034058 034059 034060 034061 034062 034063 034064 034065 034066 034067 034068 034069 034070 034071 034072 034073 034074 034075 034076 034077 034078 034079 034080 034081 034082 034083 034084 034085 034086 034087 034100 034101 034102 034103 034104 034105 034106 034107 034108 034109 034110 034111 034112 034113 034114 034115 034116 034117 034118 034119 034120 034121 034122 034123 034124 034125 034126 034127 034128 034129 034130 034131 034132 034133 034134 034135 034136 034137 034138 034139 034140 034141 034142 034143 034144 034145 034146 034147 034148 034149 034150 034151 034152 034153 034154 034155 034156 034157 034158 034159 034160 034161 034162 034163 034164 034165 034166 034167 034168 034169 034170 034171 034172 034173 034174 034175 034176 034177 034178 034179 034180 034181 034182 034211 034212 034213 034214 034215 034216 034217 034218 034251 034252 034253 034254 034255 034256 034257 034258 034259 034260 034266 034271 034272 034311 034312 034313 034314 034315 034316 034317 034318 034319 034320 034321 034322 034323 034324 034325 034326 034327 034328 034329 034330 034331 034332 034333 034334 034335 034336 034337 034338 034339 034340 034341 034342 034343 034344 034345 034346 034347 034348 034362 034363 034364 034365 034366 034367 034368 034369 034370 034371 034372 034373 034374 034375 034376 034377 034378 034379 034380 034381 034382 034383 034384 034385 034386 034387 034388 034389 034390 034391 034392 034393 034394 034395 034396 034397 034398 034399 034400 034401 034402 034403 034404 034405 034406 034407 034408 034409 034410 034411 034412 034413 034414 034415 034416 034427 034428 034429 034430 034431 034432 034433 034434 034435 034436 034437 034438 034439 034440 034441 034442 034443 034444 034450 034451 034452 034453 034454 034455 034456 034457 034458 034459 034460 034461 034462 034463 034464 034465 034466 034467 034468 034469 034470 034471 034472 034473 034474 034475 034476 034477 034478 034479 034480 034481 034482 034483 034484 034485 034486 034487 034488 034489 034490 034491 034492 034493 034494 034495 034496 034497 034498 034499 034500 034501 034502 034503 034504 034505 034506 034507 034508 034509 034510 034511 034512 034513 034514 034515 034516 035060 035061 035062 035063 035064 035065 035066 035067 035068 035069 035070 035071 035072 035073 035074 035075 035076 035077 035078 035079 035080 035081 035082 035083 035084 035085 035086 035087 035088 035089 035090 035091 035092 035093 035094 035095 035096 035097 035098 035099 035100 035101 035102 035103 035104 035105 035106 035107 035108 035109 035110 035111 035112 035113 035114 035115 035116 035117 035118 035119 035120 035121 035122 035123 035124 035125 035126 035127 035128 035129 035130 035131 035132 035151 035152 035153 035154 035155 035156 035157 035158 035245 035246 035247 035248 035249 035250 035251 035252 035253 035254 035255 035256 035257 035258 035259 035260 035261 035262 035263 035264 035265 035266 035267 035268 035269 035270 035271 035272 035273 035274 035275 035276 035277 035278 035279 035280 035281 035282 035283 035284 035285 035286 035287 035288 035289 035290 035291 035292 035293 035294 035295 035296 035297 035298 035299 035300 035301 035302 035303 035304 035305 035306 035307 035308 035309 035310 035311 035312 035313 035314 035315 035316 035317 035318 035319 035320 035321 035322 035323 035324 035325 035326 035327 035328 035329 035330 035332 035333 035334 035335 035336 035337 035338 035339 035340 035341 035342 035343 035345 035346 035347 035348 035349 035350 035351 035352 035353 035354 035355 035356 035357 035358 035359 035360 035361 035362 035363 035364 035365 035366 035367 035368 035369 035370 035371 035372 035373 035374 035375 035376 035377 035378 035379 035380 035381 035382 035383 035384 035385 035386 035387 035388 035390 035391 035392 035393 035394 035395 035396 035397 035398 035399 035400 035401 035402 035403 035404 035405 035406 035407 035408 035409 035410 035411 035412 035413 035414 035415 035416 035417 035418 035419 035420 035421 035422 035423 035424 035425 035432 035433 035434 035435 035436 035437 035438 035439 035440 035441 035442 035443 035444 035445 035446 035447 035448 035449 035450 035451 035452 035453 035454 035455 035456 035457 035458 035459 035460 035461 035462 035463 035464 035465 035466 035467 035468 035469 035470 035471 035472 035473 035474 035475 035476 035477 035478 035479 035480 035481 035482 035483 035484 035485 035486 035487 035488 035489 035490 035491 035492 035493 035494 035495 035496 035497 035498 035499 035500 035501 035502 035503 035504 035505 035506 035507 035508 035509 035510 035511 035512 035513 035514 035515 035516 035517 035518 035519 035520 035521 035522 035523 035524 035525 035526 035527 035528 035529 035530 035531 035532 035533 035534 035535 035536 035537 035538 035539 035540 035541 035542 035543 035544 035545 035546 035547 035548 035549 035550 035551 035552 035553 035554 035555 035556 035557 035558 035559 035560 035561 035562 035563 035564 035565 035566 035567 035568 035569 035570 035571 035572 035573 035574 035575 035576 035577 035578 035579 035580 035581 035582 035583 035584 035585 035586 035587 035588 035589 035590 035591 035592 035593 035594 035595 035596 035597 035598 035599 035600 035601 035602 035603 035604 035605 035606 035607 035608 035609 035610 035611 035612 035613 035614 035615 035616 035617 035618 035619 035620 035621 035622 035623 035624 035625 035626 035627 035628 035629 035630 035631 035632 035633 035634 035635 035636 035637 035638 035639 035640 035641 035642 035643 035644 035645 035646 035647 035648 035649 035650 035651 035652 035653 035654 035655 035656 035657 035658 035660 035661 035662 035663 035664 035665 035666 035667 035668 035669 035670 035671 035672 035673 035674 035675 035676 035677 035678 035679 035680 035681 035682 035683 035684 035685 035686 035687 035688 035689 035690 035691 035692 035693 035694 035695 035696 035697 035698 035699 035700 035701 035702 035703 035704 035705 035706 035707 035708 035709 035710 035711 035712 035713 035714 035715 035716 035717 035718 035719 035720 035721 035722 035723 035724 035725 035726 035727 035728 035729 035730 035731 035732 035733 035734 035735 035736 035737 035738 035739 035740 035741 035742 035743 035744 035745 035746 035747 035748 035749 035750 035751 035752 035753 035754 035755 035756 035757 035758 035759 035760 035761 035762 035763 035764 035765 035766 035767 035768 035769 035770 035771 035772 035773 035774 035775 035797 035798 035799 035800 035801 035802 035803 035804 035805 035806 035807 035808 035809 035810 035811 035812 035813 035814 035815 035816 035817 035818 035819 035820 035821 035822 035823 035824 035825 035826 035827 035828 035829 035830 035831 035832 035833 035834 035835 035836 035837 035838 035839 035840 035841 035842 035843 035844 035845 035846 035847 035848 035849 035850 035851 035852 035853 035854 035855 035856 035857 035858 035859 035860 035861 035862 035863 035864 035865 035866 035867 035868 035869 035870 035871 035872 035873 035874 035875 035876 035877 035878 035879 035880 035881 035882 035883 035884 035885 035886 035887 035888 035889 035890 035891 035892 035893 035894 035895 035896 035897 035898 035899 035900 035901 035902 035903 035904 035905 035906 035907 035908 035909 035910 035911 035912 035913 035914 035915 035916 035917 035918 035919 035920 035921 035922 035923 035924 035925 035926 035927 035928 035929 035930 035931 035932 035933 035934 035935 035936 035937 035938 035939 035940 035941 035942 035943 035944 035945 035946 035947 035948 035949 035950 035951 035952 035953 035954 035955 035956 035958 035959 035960 035961 035962 035963 035964 035965 035966 035967 035968 035969 035970 035971 035972 035973 035974 035975 035976 035977 035978 035979 035980 035981 035982 035983 035984 035985 035986 035987 035988 035989 035990 035991 035992 035993 035994 035995 035996 035997 035998 035999 036001 036002 036003 036004 036005 036006 036007 036008 036009 036010 036011 036012 036013 036014 036015 036016 036017 036018 036019 036020 036021 036022 036023 036024 036025 036026 036027 036028 036029 036030 036031 036032 036033 036034 036035 036036 036037 036038 036039 036040 036041 036042 036043 036044 036045 036046 036047 036048 036049 036050 036051 036052 036053 036054 036055 036056 036057 036059 036060 036061 036062 036063 036064 036065 036066 036067 036068 036075 036076 036077 036078 036079 036080 036081 036082 036083 036084 036085 036086 036087 036088 036089 036090 036091 036092 036093 036094 036095 036096 036097 036103 036104 036105 036106 036107 036108 036109 036110 036111 036112 036113 036114 036115 036116 036117 036118 036119 036120 036121 036122 036123 036124 036125 036126 036127 036128 036129 036130 036131 036132 036137 036138 036139 036140 036141 036142 036143 036144 036145 036146 036147 036148 036149 036150 036151 036152 |
Recalling Firm/ Manufacturer |
CHG Hospital Beds Inc 153 Towerline Place London Canada
|
| For Additional Information Contact | Renata Sila 800-327-0770 |
Manufacturer Reason for Recall | Stryker Medical is initiating a voluntary recall of the Spirit Select and Spirit Plus A-C powered hospital beds due to reports of hi-lo actuators broken at the mount ends, which could cause the bed to unexpectedly lower resulting in patient injury. |
FDA Determined Cause 2 | Component design/selection |
| Action | The firm, Stryker Medical, issued an "URGENT MEDICAL DEVICE RECALL" letter dated 7/8/2016 by FedEx to its customers and included a revised preventive maintenance checklist. The letter described the product, problem and actions to be taken. The customers were instructed to: locate the units listed on the attached business reply form; remove these units from service, if not possible to remove units from service, place units in the height position according to instructions; file the revised preventative maintenance checklist; return the Business Reply Form to confirm receipt of the notification by fax (269)488-8691 or email productfieldaction@stryker.com. If customers have loaned or sold any of the beds listed in this letter, please forward a copy of the recall notice to the new users and advise us of their new location in the space provided on the business reply form. Your Stryker Field service Representative will contact your facility to add support brackets to your beds.
Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
" Online: www.fda.gov/medwatch/report.htm
" Regular Mail: use postage-paid FDA form 3500 available at: www.fda.gov/MedWatch/getforms.htm. Mail to MedWatch , P.O. Box 3002, Rockville, MD 20847-3002
" Fax: 1-800-FDA-0178
If you have any questions or concerns, please contact Stryker Customer Service (1-800-327-0770). Our normal business hours are Monday-Friday 8 a.m-6 p.m. (EST). |
| Quantity in Commerce | 4,308 units total |
| Distribution | Worldwide Distribution: US (nationwide) including states of: AK , AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, and WI; and country of: Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|