| | Class 2 Device Recall IMMULITE 2000/IMMULITE 2000XPi Systems CMM CMV IgM Assay |  |
| Date Initiated by Firm | August 30, 2016 |
| Create Date | October 27, 2016 |
| Recall Status1 |
Terminated 3 on June 07, 2018 |
| Recall Number | Z-0320-2017 |
| Recall Event ID |
75090 |
| 510(K)Number | K100433 |
| Product Classification |
Antibody igm,if, cytomegalovirus virus - Product Code LKQ
|
| Product | IMMULITE 2000/IMMULITE 2000XPi Systems CMM CMV IgM Assay, REF/Catalog Number L2KCM2(D) - 200 tests (US), SMN 10370302 (US), UDI 00630414985725 (US); and REF/Catalog Number L2KCM2 - 200 tests (OUS), SMN 10371320 (OUS), UDI 00630414961163 (OUS); IVD --- Siemens Healthcare Diagnostics Products Ltd., Llanberis, Gwynedd, LL55 4EL UK
For in vitro diagnostic use with IMMULITE 2000 analyzers for the qualitative detection of IgM antibodies to cytomegalovirus (CMV) in human serum or plasma (EDTA or heparinized), as an aid in the diagnosis of current and recent CMV infection in individuals with signs and symptoms of CMV infection or clinical suspicion of CMV infection. This assay is not FDA cleared or approved for use in testing (screening) blood or plasma donors, neonatal screening, or for use at point-of-care facilities. |
| Code Information |
UDI 00630414985725/Catalog # L2KCM2(D): Kit lots D255, D256, D257, D258, D259, D260, D261, D262, D263, D264, D266; UDI 00630414961163/Catalog # L2KCM2: Kit lots 255, 256, 257, 258, 259, 260, 261, 262, 263, 264, 266, 267 |
| FEI Number |
2432235
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
Manufacturer Reason for Recall | Multiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots 259 and above. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | The firm, Siemens, issued an "Urgent Medical Device Recall" Letter (IMC16-22.A.US, dated 8/30/2016) in the US and an "Urgent Field Safety Notice" (IMC16-22.A.OUS, dated August 2016) Outside the US (OUS) starting on 8/30/2016 The notices instructed customers to discontinue use of and discard the affected kit lots listed in the customer notice; review your inventory of the products to determine your laboratory's replacement needs; review the UMDR and or the UFSN with their Medical Director, In the US, the method of distribution is FedEx and Outside the US the distribution is determined at the country level; complete and return the Field Correction Effectiveness Check Form via Fax to: Customer Care Center at (312) 275-7795 within 30 days, and retain this letter with your laboratory records, and forward this letter to those who may have received this product.
If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative. |
| Quantity in Commerce | US: 326 kits; Foreign: 9633 kits |
| Distribution | Worldwide distribution- US (Nationwide) and countries of: United Arab Emirates, Afghanistan, Angola, Argentina, Australia, Bangladesh, Belgium, Bulgaria, Bahrain, Bolivia, Brazil, Belarus, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Dominican Republic, Algeria, Ecuador, Egypt, Spain, France, Great Britain, Greece, Guatemala, Honduras, Hungary, India, Iraq, Italy, Jordan, Canary Islands, Kyrgyzstan, Rep. Korea (S), Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macedonia, Mexico, Malaysia, Nicaragua, Netherlands, Norway, Nepal, Peru, Pakistan, Poland, Portugal, Paraguay, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, El Salvador, Syria, Turkmenistan, Tunisia, Turkey, Uruguay, Uzbekistan, Vietnam, Serbia,and South Africa. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = LKQ
|
|
|
|