|
Class 2 Device Recall Giraffe Spot PT |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
November 18, 2016 |
Create Date |
March 20, 2017 |
Recall Status1 |
Terminated 3 on July 13, 2018 |
Recall Number |
Z-1503-2017 |
Recall Event ID |
76283 |
510(K)Number |
K011549
|
Product Classification |
Unit, neonatal phototherapy - Product Code LBI
|
Product |
Giraffe Spot PT Ohmeda Medical Spot PT Lite Phototherapy System- The Spot PT Lite Phototherapy System provides light therapy for the treatment of hyperbilirubinemia, commonly known as neonatal jaundice, during the newborn period in the hospital |
Code Information |
See Attached Spreadsheat |
Recalling Firm/ Manufacturer |
Ohmeda Medical 8880 Gorman Rd Laurel MD 20723-5800
|
For Additional Information Contact |
410-888-5376
|
Manufacturer Reason for Recall |
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
According to the firm, this issue will be resolved by sending an Urgent Medical Device Correction customer letter to customers with affected Power Cords. This letter will inform the customer of the issue and provide safety instructions for immediate mitigation of the issue. The long-term solution is to re-place all affected Power Cords at no cost to the customer.
According to the firm, the plan for public notification is to contact all customers with affected Power Cords by means of an Urgent Medical Device Correction letter. In the US this will be sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX). This letter will be sent to the following titles within the affected accounts; Nurse Managers, Labor & Delivery/NICU, Bio-Medical Engineering Department Managers and Risk Management Directors
We will track all service requests for the field action to completion. These service requests will record the replacement of the affected Power Cords. Exceptions to completion of the service request will be recorded as well. Records of field action completion will be retained per our processes and procedures. |
Quantity in Commerce |
5,993 (5,080 US; 913 OUS) |
Distribution |
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela, |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LBI and Original Applicant = OHMEDA MEDICAL
|
|
|
|