 Class 2 Device Recall Regardsee related information
Date Initiated by FirmJanuary 09, 2017
Create DateMarch 24, 2017
Recall Status1 Terminated 3 on September 28, 2018
Recall NumberZ-1608-2017
Recall Event ID 76255
Product Classification General surgery tray - Product Code LRO
ProductRegard, Item Number: 800550001, Sterile, OR0639A - Hip Shoulder - Medical Center Santa Rosa
Code Information 027181A Exp Date 2017-03-30; 030175A Exp Date 2017-03-30; 030570A Exp Date 2017-11-30; 031809A Exp Date 2017-12-30; 032419A Exp Date 2018-03-25; 033639A Exp Date 2018-03-30;
FEI Number 3004513970
Recalling Firm/
Manufacturer
Resource Optimization & Innovation Llc
2909 N Neergard Ave
Springfield MO 65803-6317
For Additional Information ContactDebbie Seilling-Ferguson
417-820-2793
Manufacturer Reason
for Recall
Custom procedure trays contain a component, a light glove, that has been recalled by the manufacturer because it can split during application to a light handle breaking the sterile field.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionROi sent an Urgent Recall Notice dated January 9, 2017, to all affected consignees. The recalling firm instructed consignees to quarantine affected product and complete the response form. The recalling firm stated that they will issue over lableing for the affected packs. Sterile replacement light covers will be sent directly to the end user if available. Consignees were instructed to inform their subaccounts of the recall. For questions regarding this recall call 417-820-2793.
Quantity in Commerce209 kits
DistributionUS Distribution to LA and TX.
Total Product Life CycleTPLC Device Report