| Date Initiated by Firm | January 25, 2017 |
| Create Date | March 09, 2017 |
| Recall Status1 |
Terminated 3 on October 06, 2017 |
| Recall Number | Z-1432-2017 |
| Recall Event ID |
76459 |
| 510(K)Number | K993425 |
| Product Classification |
Accelerator, linear, medical - Product Code IYE
|
| Product | PRIMUS HI and Mevatron M2/Primus Mid-Energy, Accelerator, Linear, Medical
The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer |
| Code Information |
Device Model Number 4504200 and 1940035 Serial numbers: 3009 2958 2863 2972 2646 3079 2516 2962 3055 1452 2702 2125 3066 2689 2228 2564 2507 1904 3075 2375 3013 3068 2596 3025 3049 2965 2045 3084 2855 2906 2292 2384 2067 2818 3089 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355-1418
|
| For Additional Information Contact | Marlynne Galloway 610-448-6478 |
Manufacturer Reason for Recall | Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Siemens sent a Customer Safety Advisory Notice (CSAN) dated February 7, 2017, to affected customers to inform them of a planned hardware update regarding the mechanics of the overhead suspension of their Digital Linear Accelerator. The notice also explained the actions Siemens is taking to perform the necessary fix at the affected sites. Customers were advised to notify anyone at their organization that should be aware of this information. Also, the safety notice should be forwarded to the new owner and Siemens also requested they be informed of the identity of the device's new owner where possible. Customers with questions were instructed to call 1-800-888-7436. For questions regarding this recall call 610-448-6478. |
| Quantity in Commerce | 35 systems |
| Distribution | Nationwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IYE
|