Date Initiated by Firm |
March 22, 2017 |
Create Date |
April 14, 2017 |
Recall Status1 |
Terminated 3 on May 17, 2021 |
Recall Number |
Z-1820-2017 |
Recall Event ID |
76810 |
510(K)Number |
K031333 K133292 K141979
|
Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
Product |
Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029; Allura Xper FD10 OR Table 722033; Allura Xper FD10/10 OR Table 722034; Allura Xper FD20/10 OR Table 722036; Allure Xper FD20/20 722038; Allura Xper FD20/20 biplane OR Table 722039 |
Code Information |
Devices with Laird Chiller. |
Recalling Firm/ Manufacturer |
Philips Electronics North America Corporation 3000 Minuteman Rd Andover MA 01810-1032
|
For Additional Information Contact |
800-722-9377
|
Manufacturer Reason for Recall |
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Consignees were notified of the recall via letter sent on 3/22/2017. The letter explains the issue, actions that should be taken by the customer, and actions to be taken by Philips. |
Quantity in Commerce |
6992 (total for all devices) |
Distribution |
Nationwide, worldwide including Canada |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = OWB and Original Applicant = PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 510(K)s with Product Code = OWB and Original Applicant = PHILIPS MEDICAL SYSTEMS NETHERLAND, B.V. 510(K)s with Product Code = OWB and Original Applicant = PHILIPS MEDICAL SYSTEMS NORTH AMERICA, INC.
|