• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BD Phoenix PMIC107

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BD Phoenix PMIC107see related information
Date Initiated by FirmMay 11, 2017
Create DateNovember 25, 2017
Recall Status1 Terminated 3 on February 06, 2018
Recall NumberZ-0171-2018
Recall Event ID 77370
510(K)NumberK082538 
Product Classification System, test, automated, antimicrobial susceptibility, short incubation - Product Code LON
ProductBD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to test antibiotic susceptibility. The inducible macrolide-linosamide-steptogramin B (iMLSb) resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. --- Product is packaged 25 panels to 1 box.
Code Information Lot 7031818, Exp. Date 2/28/2018
Recalling Firm/
Manufacturer
Becton Dickinson & Co.
7 Loveton Cir
Sparks MD 21152-9212
For Additional Information ContactMs. Gail Griffiths
410-316-4054
Manufacturer Reason
for Recall
BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.
FDA Determined
Cause 2
Manufacturing material removal
ActionURGENT PRODUCT RECALL LETTER (dated May 2017) were sent to customers via a UPS mailing on May 11, 2017. Customers were instructed to discontinue use of the affected lot and discard any remaining product. BD will replace the discarded inventory. Please complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. Please complete and fax the form to: Attention: Regulatory Compliance, Fax: 410-316-4258 or e-mail to: BDRC2@bd.com. For further assistance, please contact BD (Customer/Technical) Support at 1-800-638-8663 in the USA.
Quantity in CommerceUS: 7750 units
DistributionCT, VA, PA, TX, FL, NC, WI, NM
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = LON
-
-