| Date Initiated by Firm | August 03, 2017 |
| Create Date | October 24, 2017 |
| Recall Status1 |
Terminated 3 on September 27, 2023 |
| Recall Number | Z-0046-2018 |
| Recall Event ID |
77982 |
| 510(K)Number | K110413 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Power Processor 1K Stockyard.
The Power Processor performs all pre-analytical sample tube
preparation. |
| Code Information |
Software version PLC V26R10 |
| FEI Number |
2050012
|
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
Manufacturer Reason for Recall | Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard
can initiate a retrieval of a sample tube during the rack loading process, which should not
occur. This issue and associated complaint were discovered and filed internally. |
FDA Determined Cause 2 | Software design |
| Action | BEC will initiate a SW update- to fix the SW defect. This will be
executed via MOD (field modification) performed by Field Service. |
| Quantity in Commerce | 27 units |
| Distribution | China
France
Italy
Kuwait
Spain
Turkey
United States |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|