 Class 2 Device Recall Panoramic Xray Model PC1000, Laser1000see related information
Date Initiated by FirmJuly 14, 2017
Date PostedJanuary 09, 2018
Recall Status1 Terminated 3 on July 23, 2018
Recall NumberZ-0517-2018
Recall Event ID 78105
510(K)NumberK870236 K882436 
Product Classification Unit, x-ray, extraoral with timer - Product Code EHD
ProductPC-1000, PC-1000/Laser1000, Model Numbers 800724-2 & 800724-3
Code Information Serial Numbers: 6001 to 19340
FEI Number 1832462
Recalling Firm/
Manufacturer
Panoramic Rental Corp.
4321 Goshen Rd
Fort Wayne IN 46818-1201
Manufacturer Reason
for Recall
The lift motor used to raise and lower the overhead chassis will wear if not properly maintained according to the maintenance schedule.
FDA Determined
Cause 2
Device Design
ActionOn July 14, 2017, Panoramic distributed Urgent Medical Device Correction notices to customers who own and have not taken action as required for the PC-1000 X-ray system machines. This means that your device does not have a critical safety feature that is intended to protect your patients and staff from harm. Customers are encouraged to immediately discontinue use of their machine and follow the instructions in the notice. Customers are encouraged to not resume use of the machine until a certified Panoramic service technician repairs the lift motor. Customers are advised to contact Panoramic to ensure any safety risks associated with the lift motor are identified and addressed. Action Required: (A) If your PC1000 machine is still in service at this location, you must immediately discontinue use and choose one of the following by returning the Acknowledgement Form or contacting the Panoramic Service Team by calling 888-381-2577 Monday through Friday 8 am - 6 pm. 1. An upgraded lift motor assembly at no cost to you; 2. A digital Panoramic rental unit to replace your PC1000, contingent on a signed one year rental contract (subject to availability); or 3. A new digital Panoramic or PanoramicCephalometric unit to replace your PC1000 at a reduced price (starting at $17,995 depending on machine type and subject to availability) (B) If your PC1000 machine is no longer in service or under your control: 4. If the machine is no longer in service, Panoramic will provide free removal of your machine.
Quantity in Commerce13,340 units
DistributionUS Nationwide distribution including Puerto Rico
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = EHD
-
-