U.S. Department of Health and Human Services

Class 2 Device Recall Elements Obturation Unit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Elements Obturation Unitsee related information
Date Initiated by FirmDecember 27, 2017
Create DateJanuary 26, 2018
Recall Status1 Terminated 3 on April 29, 2024
Recall NumberZ-0413-2018
Recall Event ID 78861
510(K)NumberK031664 
Product Classification Plugger, root canal, endodontic - Product Code EKR
ProductElements Obturation Unit (EOU) used with respective Buchanan Heat Pluggers Device is an endodontic root canal plugger. It is intended to be used in Dentistry to provide continuous heat at the tip of a dental instrument to test tooth response to thermal stimulus, for tissue cauterization and coagulation, to backfill and downpack gutta percha during Endodontic root canal treatment, and when in Extruder mode it is used only to backfill gutta percha during root canal obturation.
Code Information Part Number for Elements Obturation Unit (EOU): 973-0322; and Part Numbers for the Buchanan Heat Pluggers: 952-0004, 952-0003, 952-0005, 952-0006, 952-0007, 952-0031, 973-0219 (hot pulp test tip). All serial numbers of the EOU and respective Buchanan Heat Pluggers are included in the recall.
FEI Number 2016150
Recalling Firm/
Manufacturer
Ormco/Sybronendo
1332 S Lone Hill Ave
Glendora CA 91740-5339
Manufacturer Reason
for Recall
Buchanan Heat Pluggers will possibly overheat while used with the Elements Obturation Unit (EOU).
FDA Determined
Cause 2
Under Investigation by firm
ActionSybron Endo sent an Urgent: Medical Device Recall letter dated December 27, 2017. Customers were informed that Sybron Endo was still evaluating the root cause and would provide additional instructions once the root cause is determined. Distributors were asked to cease distributing any Elements Obturation Units (EOU), confirm if any are in inventory, return any affected devices, provide the notification to customers if further distributed, return any affected devices received from customers, and complete and return the acknowledgement form. Customers were instructed to examine current inventory for affected product, check the heat plugger to see if the NiTi icon appears and/or the tip turns glowing red, if it does do not use the heat plugger and return it to Sybron Endo, and complete and return the acknowledgement form. For further questions, please call the Customer Care department at (800) 537-7123.
Quantity in Commerce25,765 total
DistributionUSA (nationwide) distributed to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to Armenia, Australia, Canada, Chile, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Malaysia, Mexico, Netherlands, Pakistan, Panama, Philippines, Singapore, South Korea, Switzerland, Thailand, United Kingdom, Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = EKR
-
-