• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall MicroVue SC5b9 Plus EIA (RUO)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall MicroVue SC5b9 Plus EIA (RUO)see related information
Date Initiated by FirmNovember 03, 2015
Create DateApril 24, 2018
Recall Status1 Terminated 3 on August 28, 2020
Recall NumberZ-1498-2018
Recall Event ID 79642
Product Classification Research use only/immunology devices - Product Code OTR
ProductMicroVue SC5b-9 Plus EIA (RUO), Model A020. Research use only (RUO) product. An enzyme immunoassay for the quantitation of SC5b-9 complex present in human plasma or serum.
Code Information Lot 052418
FEI Number 1000122536
Recalling Firm/
Manufacturer
Diagnostic Hybrids, Inc.
2005 E State St Ste 100
Athens OH 45701-2125
For Additional Information ContactRonald H. Lollar
740-589-3300
Manufacturer Reason
for Recall
A component of the kit was found to contain bacterial contamination.
FDA Determined
Cause 2
Material/Component Contamination
ActionCustomers were notified via letter dated November 3, 2015, to review their inventory for the specified lots, discard the kits, and contact the recalling firm for replacement. Certificates of destruction were also provided to customers. For further questions, please call (740) 589-3300.
Quantity in Commerce107 kits
DistributionWorldwide Distribution - USA (nationwide) to the states of : CO, CT, IA, IL, MN, MO, NC, NY, OH TX, and UT., and to the countries of : Switzerland, India, Hong Kong, and Japan.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-